Cosinus CMP60116-1/2 User manual

MULTI-CHANNEL LOW NOISE AMPLIFICATION SYSTEM
CMP60116-1/2
INSTRUCTION MANUAL

MULTI-CHANNEL LOW NOISE AMPLIFICATION SYSTEM
CMP60116-1/2
INSTRUCTION MANUAL
DA00073361-001

i CMP60116-1/2
―――Preface ―――
Thank you for purchasing the “MULTI-CHANNEL LOW NOISE AMPLIFICATION SYSTEM
CMP60116-1/2”.
For safe and correct use of this product, please read the “Safety Precautions” section that
follows before attempting to use the instrument.
●Marks and symbols
For safe operation by the use and to prevent damage to the instrument, prelase give
attention to the following marks and symbols that are used in this manual.
This mark indicates information for preventing the possibiliy of death or serious
personal injury from electrical shock or other hazards or damage to the instrument in
the use or handling of this instrument.
This mark indicates information for preventing the possibility of injury to the use or
damage to the instrument and the use for handling of this instrument.
●This manual consists of the following chapters.
If using this product for the first time, start from “1. Outline”.
1. Outline
This chapter gives an overview and describes the features and applications of this
product and the simple operating principle of the product.
2. Preparation before Use
This chapter describes important preparations to be made before installation and
operation.
3. Panel Features and Basic Operations
This chapter explains the basic operations of the panels.
4. Maintenance
This chapter describes a method for simply inspecting operation.
5. Specifications
This chapter gives specifications (functions and perfoemance).
CAUTION
WARNING

ii CMP60116-1/2
Safety Precautions
To ensure safe use, be sure to observe the following safety precautions.
NF Corporation shall not be held liable for damages that arise from a failure to observe
these safety precautions or warnings or cautions in the instruction manual.
●Be sure to observe the information of the instcutions manual.
The instruction manual contains information for the safe operation of the product.
Be sure to read the information first before using the product.
All the warnings in the instruction manual are intended for preventing risks that
may lead to serious accidents. Ensure to obey them.
●Check the power supply voltage.
This product operates on the power supply voltage indicated in the instruction
manual. Prior to connecting the power supply, check that the voltage of the power
supply matches the rated power supply of the product.
●If you notice anything strange.
If this product emits smoke, an unusual smell or strange sound, immediately power it
off and stop using it.
If such an anomaly occurs, prevent anyone from using this product until it has been
repaired, and immediately report the problem to NF Corporation or one of out
representatives.
●Do not operate in an explosive atmosphere.
An explosion or other such hazard may result.
●Do not remove the cover.
Never remove the cover.
Even when the inside of this producti needs to be inspected, do not touch the inside.
All such inspections are to be performed by service technicians designated by NF
Corporation.
●Do not modify the product.
Never modify the product. Modification to the product may pose a new risk. We may
refuse the repair of a modified product.
●Ensure that water does not get into this product.
Using the product in wet condition may cause electric shock and fire. When water etc.
get into the product, immediately power it off, and contact NF Corporation or one of our
representatives.
● If lightning occures, power off and disconnect this product.
A lightning may cause an electric shockm a fire and a failure.

iii CMP60116-1/2
Safety Precautions
WARNING
●Safety symbols
Caution, refer to instruction manual.
This notifies the user of potential hazards and indicates that he/she
must refer to the instruction manual.
Caution, possibility of electric shock.
This indicates that an electric shock may occur under specific
conditions.
Warning
This contains information to avoid risks in equipment handling
that could result in loss life or bodily injury.
Caution
This contains information to avoid risks equipment handling that
could result in minor or moderate injury to person or damage to
property or the equipment.
●Other symbol
This indicates that the terminal or the outer conductor of the connector is
connected to the signal ground.
● Disposal of this product
a) Use the servies of an industrial waste disposal contractor for disposal of the entire
product.
b) This product does not include batteries.
c) This product does not include mercury.
CAUTION

iv CMP60116-1/2
Contents
Page
Preface..................................................................................................................................i
Safety Precautions................................................................................................................ii
1. Outline................................................................................................................... 1-1
1.1 Overview................................................................................................................ 1-2
1.2 Features................................................................................................................. 1-2
1.3 Applications...........................................................................................................1-3
1.4 Circuit functions ................................................................................................... 1-3
2. Preparation before Use......................................................................................... 2-1
2.1 Checking before use .............................................................................................. 2-2
2.2 Conditions for installation location .....................................................................2-3
2.3 Power supply ......................................................................................................... 2-4
2.3.1 Connecting to low noise DC power supply LP5391...................................... 2-5
2.3.2 Connecting to CMP60116-3 ...........................................................................2-6
3. Panel Features and Basic Operations ................................................................. 3-1
3.1 Panel component names and functions ............................................................... 3-2
3.2 Input connection and installation........................................................................ 3-5
3.3 Output connection................................................................................................. 3-6
3.4 Turning on power and warm-up time ..................................................................3-6
4. Maintenance.......................................................................................................... 4-1
4.1 Before maintenance .............................................................................................. 4-2
4.2 Dairy maintenance ............................................................................................... 4-2
4.3 Storage, repacking and transportation ...............................................................4-2
4.4 Function test ......................................................................................................... 4-3
4.4.1 Consumption Current (with No Signal) .......................................................4-3
4.4.2 A (Non-inverting) input operation ................................................................ 4-4
4.4.3 −B (Inverting) input operation...................................................................... 4-5
5. Specifications ........................................................................................................ 5-1
5.1 Absolute maximum ratings .................................................................................. 5-2
5.2 CMP60116-1 .......................................................................................................... 5-2
5.2.1 Input ...............................................................................................................5-2
5.2.2 Output ............................................................................................................5-3
5.2.3 Amplifier ........................................................................................................ 5-3
5.2.4 Low-pass filter ............................................................................................... 5-4
5.2.5 Power supply..................................................................................................5-4
5.3 CMP60116-2 .......................................................................................................... 5-4
5.3.1 Input ...............................................................................................................5-4
5.3.2 Output ............................................................................................................5-5
5.3.3 Amplifier ........................................................................................................ 5-5
5.3.4 Low-pass filter ............................................................................................... 5-6

v CMP60116-1/2
5.3.5 Power supply..................................................................................................5-6
5.4 Power Connector Pin Assignment........................................................................ 5-6
5.5 General .................................................................................................................. 5-7
5.6 Notes......................................................................................................................5-7
5.7 External view ........................................................................................................5-8

vi CMP60116-1/2
Figures and Tables
Page
Figure 1-1 Block diagram............................................................................................. 1-3
Figure 2-1 Connecting to LP5391. ...............................................................................2-5
Figure 2-2 Connecting to CMP60116-3. ......................................................................2-6
Figure 2-3 Connection when driving 8 products with CMP60116-3.......................... 2-7
Figure 3-1 Front and rear panel views........................................................................3-4
Figure 4-1 Connection diagram for checking non-inverting input operation. .......... 4-4
Figure 4-2 Non-inverting input / output voltage waveform. ...................................... 4-4
Figure 4-3 Connection diagram for checking inverting input operation...................4-5
Figure 4-4 Inverting input / output voltage waveform...............................................4-5
Figure 5-1 External view.............................................................................................. 5-8
Table 2-1 List of contents ............................................................................................. 2-2
Table 4-1 Setting when checking current consumption ............................................. 4-3
Table 5-1 Power Connector Pin Assignment ............................................................... 5-6

CMP60116-1/2
1-1
1. Outline
1.1 Overview ................................................................................................................1-2
1.2 Features.................................................................................................................1-2
1.3 Applications............................................................................................................1-3
1.4 Circuit functions......................................................................................................1-3

1.1 Overview
1-2 CMP60116-1/2
1.1 Overview
This product is a multi-channel amplification system that incorporates four low noise
differential amplifiers. This product has the following performance.
CMP60116-1
・Input resistance 100 k
・Voltage gain 40 dB
・Frequency characteristic DC to 1 MHz
CMP60116-2
・Input resistance 1 M
・Voltage gain 40 dB
・Frequency characteristic DC to 20 MHz
1.2 Features
(1) Easy multi-channeling
(2) Multi-functional
Input coupling:
DC / AC
Input mode:
Single-ended input / Differential input / GND
LPF selection:
THRU / LPF ( fc = 1 MHz )
Offset adjustment:
Adjustable to 0 V
Selection of power supply GND potential:
FLOAT / EXTERNAL(Connection with the earth of the power supply)
(3) Connector with excellent shielding characteristics
Input and output connectors:
SMA connectors

1.3 Applications
1-3 CMP60116-1/2
1.3 Applications
The CMP60116-1/2 is high gain, wide-bandwidth, and low noise. It is widely used for
amplification of small signals.
(1) Preamplifier for the magnetic flux measurement by SQUID sensor, etc.
(2) Preamplifier for small voltage after voltage conversion of SEM, etc.
(3) Preamplifier for the sensor such as inertial, pressure, sonic, etc.
(4) Preamplifier for biological and chemical sensor.
(5) Improvement of noise characteristics for a lock-in amplifier.
1.4 Circuit functions
This product is a voltage amplifier that amplifies the voltage difference between the
Aand the −Bterminal. The gain is 40 dB. The input resistance of the CMP60116-1 is
100 k, and the input resistance of the CMP60116-2 is 1 M. The output resistance is
50 for both models.
Setting of input coupling mode (DC / AC), setting of input mode (differential input,
single-ended input, GND), setting of output low-pass filter (THRU / LPF), setting of
power supply GND (FLOAT / EXTERNAL) can be changed by each switch.
A
-B
+15V
-15V
+15V
-15V
+
-
A1
40dB OUTPUT
Cc
1MΩ
COUPLING
DC
AC
INPUT SEL
A
GND
INPUT SEL
-B
GND
1MΩ
Rin
+15V
-15V
OFFSET
LPF
fc=1MHz
OUTPUT
LPF
THRU
×4
+15V
-15V
POWER
INPUT
+15V 1
-15V 3
GND 2
GND 4
EXTERNAL 7
EXTERNAL 8
NC 5
NC 6
NC 9
1 +15V
3 -15V
2 GND
4 GND
7 EXTERNAL
8 EXTERNAL
5 NC
6 NC
9 NC
POWER
OUTPUT
1MΩGND
FLOAT
EXTERNAL
INPUT
FLOAT EXTERNAL
Rin
Cc
CMP60116-1 CMP60116-2
Cc 1 μF 0.1 μF
Rin 100 kΩ1 MΩ
Figure 1-1 Block diagram.

CMP60116-1/2
2-1
2. Preparation before Use
2.1 Checking before use ..............................................................................................2-2
2.2 Conditions for installation location .........................................................................2-3
2.3 Power supply..........................................................................................................2-4
2.3.1 Connecting to low noise DC power supply LP5391 ........................................2-5
2.3.2 Connecting to CMP60116-3 ............................................................................2-6

2.1 Checking before use
CMP60116-1/2
2-2
2.1 Checking before use
■ Safety check
Before using this product, read the following section of this manual.
●[Safety Precautions] at the beginning of this instruction manual.
● [2.3 Power supply]
■ Check apperance and accessories
First check for any damage that may have occurred during transportation.
After unpacking the product, refer to “Table 2-1 List of contents”and confirm that all
items are present in the quantities listed.
Table 2-1 List of contents
●Optional Item
Optional items below are separately available. If required, please contact NF or one
of our agents.
●Main unit ·····························································································1
●Instruction manual·················································································1

2.2 Conditions for installation location
CMP60116-1/2
2-3
2.2 Conditions for installation location
●The installation location shall meet the following temperature and humidity
conditions.
Specifications guaranteed: 18 °C to 28 °C, 5 %RH to 85 %RH
Operating: 0 °C to 40 °C,5 %RH to 85 %RH
Storage: −10 °C to 50 °C,5 %RH to 95 %RH
However, do not use the product if condencation is present.
●Use the product indoors at an altitude of up to 2000 m.
●Do not install the product in the following locations:
• Locations where flammable gases may be present.
There is a risk of explosion. Never install or use the product in such
locations.
• Outdoors, locations exposed to direct sunlight, near fire or heat sources.
The performance may not be satisfied, or a failure may occur.
•Locations where corrosive gases, water vapor dust, or too humid.
Malfunction or a failure may occur.
• Near an electromagnetic filed source, high-voltage product, or power line.
Noise may increase.
• Where is a lot of vibration.
Noise may increase, or a malfunction may result.

2.3 Power supply
CMP60116-1/2
2-4
2.3 Power supply
This product operates under the following power supply conditions.
・Stabilized DC power supply with ±15 V ±0.5 V, ± 300 mA or higher,
ripple noise1 mVrms or lower.
We provide the excellent stability and low noise performance DC power supply LP5391.
For information on those products, please contact the NF corporation or one of our agents.
Do not connect this product to an AC outlet, because doing so is dangerous.
・Incorrect voltage polarity of the power supply will damage this product.
・Supplying a voltage greater than ± 16.5 V will damage this product.
・This product does not operate by single power supply (+30 V / GND). Be sure to use a
dual power supply (+15 V / GND / −15 V).
Attention
WARNING

2.3 Power supply
CMP60116-1/2
2-5
2.3.1 Connecting to low noise DC power supply LP5391
The DC POWER CABLE (CMP60116-4) is available to connect this product to LP5391. If
you require the cable, please contact NF or one of our agents.
The following figure shows the connection using the CMP60116-4. The output of LP5391
is set as ±15 V.
Figure 2-1 Connecting to LP5391.
・Turn off the output switch of the power supply unit before connecting this product to the
power supply unit.
・Do not plug / unplug the DC power cable while the power supply is turned on. This
product may be damaged.
Attention
CMP60116-4

2.3 Power supply
CMP60116-1/2
2-6
2.3.2 Connecting to CMP60116-3
The CMP60116-3 attached DC power cable is available to connect this product to
CMP60116-3.
The following figure shows the connection using the CMP60116-3 attached DC power
cable. The output of CMP60116-3 is set as ±15 V.
Figure 2-2 Connecting to CMP60116-3.
・Turn off the output switch of the power supply unit before connecting this product to the
power supply unit.
・Do not plug / unplug the DC power cable while the power supply is turned on. This
product may be damaged.
CMP60116-3 attached DC power cable
Attention

2.3 Power supply
CMP60116-1/2
2-7
The CMP60116-3 can drive up to 8 units (32 CH) of this product. A dedicated DC cable
(0.25 m) is available for multiple connection, please contact NF or one of our agents.
The following figure shows the connection for driving 8 units with CMP60116-3. The
output of CMP60116-3 is set as ±15 V.
Figure 2-3 Connection when driving 8 products with CMP60116-3.
DC cable
DC cable
DC cable
DC cable
DC cable
DC cable
CMP60116-3 attached DC power cable

CMP60116-1/2
3-1
3. Panel Features and Basic Operations
3.1 Panel component names and functions .................................................................3-2
3.2 Input connection and installation............................................................................3-5
3.3 Output connection..................................................................................................3-6
3.4 Turning on power and warm-up time......................................................................3-6

3.1 Panel component names and functions
3-2 CMP60116-1/2
3.1 Panel component names and functions
See Figure 3-1, Front and rear panel views.
"Front panel"
①A
②−B
The Ais a non-inverting input connector of this product. The −Bis inverting input
connector. Both connectors are SMA connectors.
Differential signal iput voltage range is ±0.1 V. Common-mode input voltage range
is ±5 V.
③COUPLING
The COUPLING is a switch to select the input coupling method.
DC input coupling can be selected with "DC" and AC input coupling can be selected
with "AC".
The lower cutoff frequency is approximately 0.25 Hz when AC coupling is selected.
④INPUT SEL
This is a switch to select the input mode. The following input modes can be
selected by combining two switches.
1) "A" and "-B" : Differential input mode
2) "A" and "GND" : single-ended input mode (Non-inverting amplifier)
3) "GND" and "-B " : single-ended input mode (Inverting amplifier)
⑤OUTPUT (THRU / LPF)
Selection switch for output stage LPF.
The output of the amplifier is output as it is with "THRU".
It is output through LPF of cutoff frequency 1 MHz with "LPF".
The output resistance is 50 Ω.
⑥POWER LED
Lights up when the power supply voltage is applied.
⑦OFFSET
This is a potentiometer to adjust the offset voltage.
Using a screwdriver whose tip does not match or applying excessive force may
cause damage. The recommended tip size of screwdriver is 1.8 to 2.0 mm of width
and 0.4 to 0.5 mm of thickness. The adjustment torque is 15 mN·m maximum.
Table of contents
Other Cosinus Test Equipment manuals
Popular Test Equipment manuals by other brands

Sonotec
Sonotec Airborne Sound Sensor BS10 User documentation

Cleerline
Cleerline SSF-TKITP-500 manual

R&S
R&S RTB2K-COM4 Getting started

BS BATTERY
BS BATTERY BST 1000 instruction manual

American DJ
American DJ MEGA STROBE/DMX instruction manual

Harvest TEC
Harvest TEC 453 Installation and operation instruction manual