30 Washington Ave Ste D Haddonfield, NJ 08033
tel (800)707-5751 ●fax (800)721-2377 ●service@covita.net ●www.covita.net
1. Ensure you are performing a calibration in a fresh air environment.
2. Turn Micro+ ON by holding the on/off button down for three seconds.
3. Main Screen
a. On the main screen touch the gears icon.
4. Gears Screen
a. Touch the calibration icon to enter the calibration screen.
5. Calibration Mode
a. Touch the zero icon to zero the monitor. The monitor will
display an hour glass and then a check mark. Touch the
check mark to return back to the calibration screen.
b. Touch the calibration icon. (Note: the canister is no longer
crossed out)
c. Immediately turn on the gas by turning the valve. No fine tuning is
required because this regulator is fixed at 1 lpm.
d. The monitor’s reading will slowly climb from zero and a spinning
hour glass will be shown on the screen as well.
e. When you receive a check mark, turn OFF the gas.
Touch the check mark to lock in calibration & to return
previous menu.
f. Touch the home icon or return arrow to return to the main
screen.
g. Calibration Complete Please contact coVita with any questions.
**coVita offers annual maintenance packages.**