Curlin Medical 4000 Series User manual

Important Notification
A lack of compliance with proper positioning of the tubing could
result in infusion inaccuracies.
Please center the tubing in the pumping area when installing the
administration set.
CURLIN MEDICAL
4000 Series ™ and 2000 Series™*
Ambulatory Infusion Pumps
Clinician’s Guide
INTERMITTENT THERAPY
REFER TO USER’S MANUAL FOR DETAILED INSTRUCTIONS
KEYPAD FUNCTION
ON/OFF: Turns pump on and off. When the pump is running, the
PAUSE key must be used to stop infusion first. Then press the OFF
key to turn pump off.
RUN/PAUSE: Starts or pauses the pump. When the pump is paused,
pressing this key will resume pumping.
UP ARROW/ DOWN ARROW: Used to scroll through the display
screens.
NO/CHANGE: Rejects highlighted data.
YES/ ENTER: Accepts highlighted data.
HELP/OPTIONS: Used to request and exit help screen when pump is
paused, or go to options screen when pump is running.
./SILENCE: Used to silence alarm for 1 minute intervals, and as a
decimal point.
PRIME/BOLUS: Used to prime fluid through the tubing to clear it of
air.
LED INDICATORS
Green- pump is running
Yellow- standby
Red- pump in alarm status
SETTING UP THE PUMP
1Install 2 new “C” size alkaline batteries or connect Battery Pack, or
AC adapter to bottom of pump.
2Prepare IV bag and attach a new administration set.
3Eliminate all air from the bag and the administration set (see
PRIMING THE SET).
4Install set and close pump door.
5Attach set to venous access device using agency protocol.
Note: If a Check Valve Adapter is provided with your administration
set, remove it from the package and connect it to the distal end of the
set.
INSTALLING THE ADMINISTRATION SET
1Fully open the door of the pump.
2Close the slide-clamp on the administration set.
3Activate the Integral “Flow-Stop” by twisting and removing the
breakaway tab.
4Insert the blue tubing guide into the receptacle on right side of
pump in the direction of the blue arrow, positioning the tubing in
front of the door hinge.
5Center the tubing in the middle of the pumping fingers, covering
the yellow dot.
6Holding the yellow Integral “Flow-Stop” by the handle, insert it on an
angle into the receptacle on the left side of the pump in the direction of
the yellow arrow. CAUTION: Do not press down on the top of the
Integral “Flow-Stop” when inserting it into the receptacle. This
action could inadvertently open the Integral “Flow-Stop” and allow
fluid to flow through the tubing.
7Open the slide-clamp on the administration set
8Push the tubing slightly into the black Air Detector slot.
9Fully close the door of the pump and latch it securely.
PRIMING THE SET
CAUTION: Disconnect set from patient’s access site before priming.
By Gravity:
1Twist and remove breakaway tab from “Flow- Stop”.
2Squeeze “Flow-Stop” between thumb and forefinger to allow fluid
to flow through tubing.
Electronically:
Note: Set must be installed into the pump.
1Release any clamps on the administration set.
2Press and release PRIME key from “Run Options” menu
3Press and hold the PRIME key until set is free of air; release key to
stop (each prime cycle allows up to 6 ml of fluid).
4When prime is complete, press YES/ENTER to exit the prime
screen.
Flow-Stop handle
A
ir Detector
Data Interface/
Bolus Cord Port

PROGRAMMING FOR INTERMITTENT THERAPY
1Turn the pump on.
2Observe the self-test, the LED lights, and the power source screen.
3Select PROGRAM from the first selection menu.
4If multiple therapies are available, select INTERMITTENT from the
therapy selection menu.
5Complete the pre-prescription menu:
6Complete the prescription menu:
7Make selection from run option menu:
8Press NO to review prescription, OPTIONS to view options menu,
PRIME to prime set, or press RUN key to start infusion.
9While INTERMITTENT infusion is running, the display will alternate
between various information screens.
10 To interrupt or pause an infusion, press RUN/PAUSE key.
11 Make selection from the pause menu:
12 If REPEAT Rx is selected, a screen will appear to confirm the
dosing schedule:
13 When the infusion is complete, the following screen displays:
14 To stop a therapy, press PAUSE then turn pump off.
OPTIONS
1LOCK: Press NO to change Lock Setting (may need access code).
Press YES to accept new setting.
2HOURLY TOTAL **: Press NO then YES to view. Scroll - to
review. Press NO to exit.
3CLEAR HOURLY **: Press NO then YES to view. Press YES to
clear, NO to retain and exit.
4DOWN OCCLUSION: Press NO to change, YES to accept.
5AIR SENSITIVITY: Press NO to change, YES to accept.
6NEXT DOSE: HH:MM until next dose. When correct, press YES to
accept.
7AUDIO: Press 1-9 to hear level, press YES to accept.
8POWER CHECK: Press NO then YES to check level.
9ACCEPT OPT? Press YES to accept any changes made.
HOW TO CHANGE THE LOCK SETTING
1Lock may be initially selected in BIOMED SETUP.
2When infusing, press OPTIONS key.
3Scroll to LOCK and press NO to change.
4Enter access code at prompt.
5Select the appropriate Lock Setting.
6Press YES/ENTER to select.
7Exit Options menu by pressing YES at “ACCEPT OPT?” field.
INSTALLING BATTERIES
1Turn pump OFF or connect to AC adapter.
2Open the battery door on the back of pump by depressing the
release pin with a small coin and sliding it in a downward motion.
3Install 2 new “C” size alkaline batteries with both positive poles at
the top of the battery compartment
4Slide the battery compartment door back into place.
USING THE HELP FUNCTIONS
1When programming, press the HELP key. While running, press
PAUSE, then press HELP key.
2When a HELP screen is displayed and there is more text to read
that is not displayed on the screen, or icons will appear, scroll up
and down the text using the ARROW keys.
3When finished, press the HELP key again to exit the HELP screen.
Reorder No 350-9021
© 2001– 2003 Curlin Medical, LLC. All rights reserved. 08/03 Rev. C
PUMP ALARMS & RESOLUTIONS
* Versions 39 and higher
** Not available on the 2000 Series
LCD DISPLAY RESOLUTION
INFUSION
COMPLETE
Press PAUSE, add another IV bag, select
REPEAT Rx and confirm fields, or turn the pump
off.
ALARM
AIR-IN-LINE
Remove air from administration set and IV bag.
(Disconnect set from patient’s access site prior
to using prime function.)
ALARM
DOWN
OCCLUSION
Check administration set from the pump to
patient’s access site for cause of occlusion.
Correct obstruction per protocol. Consider
changing the pressure setting from Low to High.
When occlusion is resolved, the alarm will stop
and pumping will resume.
ALARM
UP
OCCLUSION
Check administration set from IV bag to pump
for cause of occlusion. When ready to begin
infusion, press PAUSE, select RESUME, and
press RUN.
ALARM
HIGH UP
PRESSURE
Check for excessive pressure on IV bag. When
ready to begin infusion, press PAUSE, select
RESUME, and press RUN.
ALARM
SET NOT
INSTALLED
Install appropriate Curlin Medical administration
set. Use ONLY Curlin Medical administration
sets. See User’s Manual or IV set directions for
proper installation instructions.
ALARM
UNATTENDED
PUMP
To resolve, press RUN and resume the therapy
or continue with operating procedure.
ALARM
REPLACE SET
Replace set, prime set if necessary, select
RESUME and press RUN when ready to begin.
ALARM
DOOR OPEN
Check placement of administration set and close
pump door properly. See User’s Manual for
instructions.
ALARM
EMPTY
BATTERY
Press the pause key and turn pump off. Install 2
new “C” size alkaline batteries.
ERROR CODE If an error code occurs, turn pump off then back
on. If it reoccurs, notify dispensing agency
CURLIN MEDICAL, LLC
15751 Graham Street
Huntington Beach, CA 92649
Tel: (888) 287-5999
www.curlinmedical.co
m
This manual suits for next models
1
Other Curlin Medical Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Otto Bock
Otto Bock Varos Testset Instructions for use

Weinmann
Weinmann VENTIlogic BiLevel ST Description of Unit and Operating Instructions for Patients

Expain
Expain Change Neck & Shoulders user manual

BÜHLMANN
BÜHLMANN Quantum Blue Reader user manual

Boxym
Boxym B-820 mini user manual

Lightmed
Lightmed LightLas 532/670 Operator's manual

Datascope
Datascope Passport 2 Service manual

Enabling Devices
Enabling Devices Push Down Scissors 9074 user guide

NRS
NRS Compact Easy with Armrests & Backrest M48477 User instructions

Winco
Winco HOME CARE CLINER Owner's operating and maintenance manual

Ultra Ankle
Ultra Ankle Ultra Zoom Fitting instructions

NuMED
NuMED MAC 600 Easy user guide