DentalVibe GEN 4 User manual

User Manual
Injection Technique: Page 6

Charging
DentalVibe® is equipped with a Nickel Metal
Hydride (NiMH) rechargeable battery.
Only use provided power supply.
1. Insert power supply cord into receptacle on charger
base, plug power supply into an electrical outlet. Use
international plug adaptors as needed.
2. Charge the DentalVibe at least 15 hours before initial use.
3. After each patient use, place the DentalVibe back in its
charging base.
Note: Battery does not experience the “memory eect”.
DENTALVIBE LED INDICATOR LIGHT
BLINKS AMBER (o charger) while button is pressed: needs
a new tip.
BLINKS AMBER (on charger) battery is charging.
BLINKS AMBER (o charger) battery needs to be charged.
BLINKS GREEN (on charger) when ready for use.
Note: DentalVibe needs to be placed on charger until light
blinks green.
2
Model # GTM 41076-0615
Input 100-240V, 50-60Hz, 0.5A
Output 15V, 0.4A
3
1. Place a new comfort tip onto a fully charged
DentalVibe® and turn it on.
2. Rest the vibrating tip above your upper lip. Apply
diering pressure to the skin and you will notice
that a light touch delivers the greatest amount of
stimulation.
Learn the correct pressure that stimulates
the sensory receptors by experiencing the
DentalVibe on yourself rst.
Get to know your
DentalVibe®GEN 4

Align Arrows and
Push Straight Down.
(Do not twist)
Twist the clear plastic portion
180° counter-clockwise and
Lift Up to Remove.
TIP ON TIP OFF
REVIEW BEFORE USE.
Placing Comfort Tip
On and Off
Watch instructional videos at www.DentalVibe.com/training
WARNING: Do Not re-use the Tip.
Comfort Tips must be disposed of after each patient use to
prevent cross-contamination and ensure FDA compliance
of infection-control protocols. These tips are porous,
contain microlaments and will harbor blood, saliva
and bacteria. Dispose of used White Tip in red bag
(hazardous waste).
Infection-control protocol for comfort tips.
BIOHAZARD AFTER USE
DO NOT RE-USE
5
Note: Comfort Tips are Latex-Free.
Installing the Comfort Tip for use
1. Remove a fresh Comfort Tip from its sealed bag in
plain view of your patient. Align the arrow on the
Comfort Tip with the arrow on the DentalVibe® unit.
Push straight down until it snaps on securely (See
photos).
2. Each Comfort tip is programmed to be turned on
and o a total of 6 times and will run for a total of 6
minutes.
Removing the Comfort Tip after use
1. While holding the DentalVibe unit in one hand, grasp
the clear plastic portion of the Comfort Tip with
your other hand, rotate 180° counter-clockwise and
lift up to remove. The tip will tear, preventing cross
contamination.
2. After each patient use, return the DentalVibe unit
back to its charging base.
Note: Safe with pacemaker/debrillator when used normally
in accordance with intended use.
Comfort Tip
4
INSTALLING / REMOVING COMFORT TIP
INSTALLING / REMOVING COMFORT TIP

For instructional videos, visit:
http://www.dentalvibe.com/instructional-videos/
6
Injection Technique
with DentalVibe®
1. Activate the DentalVibe by fully depressing the On/
O button; and then introduce the vibration to your
patient by gently touching the back of their hand or on
the outside of their cheek before placing in the mouth.
2. Hold the DentalVibe as a retractor in a similar fashion
as you would hold a dental mirror. (It is designed with
the same angles for ease of use.)
3. Position DentalVibe’s Comfort Tip on the oral
mucosa at the injection site.
4. Vibrate 10 seconds before inserting the needle.
5. Maintain contact and vibrate the tissue during the
entire injection, plus 2 seconds after the injection.
6. Insert the needle as close as possible to one of the
vibrating prongs.
7
Sweet Spot
Block
Palatal
Inltrations

Read all instructions before use.
INTENDED USE:
DentalVibe®is a vibrating retractor and is intended for use
by dental professionals in a dental oce setting. It has been
carefully designed to oer your patients a more comfortable
dental injection experience that is both safe and eective.
It does not modify the tissue or impart energy and is not
intended for therapeutic, diagnostic or any other purpose.
• The use of the DentalVibe varies between 10-120 seconds per application.
• Contraindication for use: Do not use in an area where an abscess is present.
DO NOT OPEN THE DENTALVIBE OR
CHARGING BASE.
The internal circuit boards are manufactured
in a static-free environment. For service or
battery replacement, please
call 1-877-503-VIBE (8423) for an authorized Bing Innovations,
LLC Service Center closest to you.
Caring for your
DentalVibe®
SUPPLIED PARTS: DentalVibe hand piece,
Charging base with Power Supply and Comfort Tips.
CLEANING INSTRUCTIONS: Wipe handle clean with
plastic safe surface disinfectant. Thoroughly wet the
handle and ensure that the surface remains visibly wet
for 1 minute at room temperature (68°F/20°C).
•Do Not place DentalVibe or Comfort Tip in an autoclave.
•Do Not immerse DentalVibe or Charger in water.
•Remove the power cord plug from the electrical outlet
before cleaning the charger.
WARNING:
IMPORTANT:
8 9
Do not place or store the charger where it can fall or be pulled into a sink.
Do not place the hand held unit or charger in water or other liquid.
Do not reach for a charger that has fallen into water. Unplug immediately.
Use this type B applied product only for its intended use as described in the manual.
Do not use attachments or chargers not recommended by the manufacturer.
Position the charger so the power supply is easy to unplug.
Never insert any objects into any opening of the unit or charger base.
Never operate the product if it has a damaged cord or plug, not working properly,
has been dropped, damaged or submerged into water. If any of these occur, contact
a service center and return the product for examination and repair.
Periodically check the cord for damage. If cord is damaged, ship the charging unit
to a Bing Innovations Service Center. A damaged or non-functioning unit should
no longer be used.
Do not touch the power cord with wet hands as this can cause electric shock.
When unplugging, always hold the power plug instead of the cord.
Do not put metal objects such as coins or clips on the charger. This may generate
heat, damaging the unit and heating the metal item presenting potential burn
hazards.
No User Serviceable Parts. Do not modify this equipment without authorization
of the manufacturer. If equipment is modied, appropriate inspection and testing
must be conducted to ensure continued safe use.
Batteries are not user serviceable.
Equipment protected from AC power by double insulation or reinforced insulation
by use of the supplied power supply. Will not cause electromagnetic interference
and complies with 601-1-2.
.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
Operating:
Storage:
Permissible environmental conditions during normal use.
Temperature Humidity Barometric Pressure
+18°C to +30°C 30-85% RH 100kpa-50kpa
-20°C to +45°C 10-85% RH 100kpa-50kpa

OBTAINING SERVICE:
Should this product require service, please call
1-877-503-VIBE (8423) to be referred to the authorized
Bing Innovations, LLC Service Center closest to you.
Bing Innovations, LLC warrants this appliance to be free of defects in
material and workmanship for a period of one year from the date of
original purchase. If appliance exhibits such a defect, Bing Innovations,
LLC will, at its option, repair or replace it without cost for parts and labor.
The product must be shipped, prepaid and insured (recommended),
to an authorized Bing Innovations, LLC Service Center. Proof of date
of original purchase is required. The cost to ship the product to an
authorized Service Center is the responsibility of the dental practice;
the cost to return the product to the dental practice is the responsibility
of Bing Innovations. Warranty does not cover nishes or normal wear,
nor does it cover damage resulting from accident, misuse, dirt, water,
tampering, unreasonable use, use of any tip other than those supplied
by Bing Innovations, LLC or covering the tip with a plastic sleeve, service
performed or attempted by unauthorized service agencies, failure to
provide reasonable and necessary maintenance, including recharging,
or units that have been modied or used for commercial purposes.
ALL IMPLIED WARRANTIES, INCLUDING ANY IMPLIED WARRANTY
OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE,
ARE LIMITED IN DURATION TO ONE YEAR FROM DATE OF ORIGINAL
PURCHASE. IN NO EVENT WILL BING INNOVATIONS, LLC BE RESPONSIBLE
FOR CONSEQUENTIAL DAMAGES RESULTING FROM THE USE OF THIS
PRODUCT. Some states do not allow the exclusion or limitation of
incidental or consequential damages, so the above limitation may not
apply to you. This warranty gives you specic legal rights, and you may
have other legal rights which vary from state to state.
Limited 1 year
warranty
EXPLANATION OF SYMBOLS
Do not reuse
Biohazard after use
Warning
Type B applied part
Consult User Manual Electrical hazard
Double insulated
Manufacturer
10 11
Non sterile

Manufactured by
Accumedix, Inc.,
888 E. Belvidere Road, Ste. 212 Grayslake, IL 60030
Marketing by
BING INNOVATIONS, LLC.
990 South Rogers Circle, Suite 4 Boca Raton, FL 33487
877.503.VIBE (8423)
www.DentalVibe.com [email protected]
© 2015Bing Innovations, LLC. All rights reserved. Under the copyright laws, this manual
may not be copied, in whole or in part, without the express written consent of Bing
Innovations, LLC. Every eort has been made to ensure that the information in this manual
is accurate. Bing Innovations, LLC is not responsible for printing or clerical errors. This
product may be covered by one or more U.S. patents or pending patent applications.
See www.DentalVibe.com/patents for details. FDA Registered. Reverse engineering or
disassembly is prohibited.
Model Name: DentalVibe GEN 4
Model # VIBE 4
Manufactured in the USA.
BI-LS-0001 Rev 0 (2016-04)