
F1902103.X / 03 / 2008 – updated 01/2014 GB-FR 18 / 38
Step-by-step procedure
A. Devices
3. Contra Angle
2. Implantable devices
1. Instruments Following uses
First use
Operation Operating mode Warning
1. Disassembling - Disassemble the device,
if required
Silicone stops have to be removed. X
2. Pre-Disinfection
- Soak all instruments
immediately after use in a
detergent and disinfecting
solution combined with
proteolytic enzyme if
possible.
- Follow instructions and observe concentrations and
immersion times given by the manufacturer (an excessive
concentration may cause corrosion or others defects on
instruments).
- The disinfecting solution should be aldehyde free (to
avoid blood impurities fixation) and without di- or
triethanolamines as corrosion inhibitor.
- Do not use disinfecting solutions containing Phenol or
any products which are not compatible with the
instruments (See general recommendations).
- For visible impurities observed on instruments a pre-
cleaning is recommended by brushing them manually with
soft material.
X
3. Rinsing - Abundant rinsing (at least
1 min)
- Use quality water in accordance with local regulations.
- If a pre-disinfectant solution contains a corrosion
inhibitor, it is recommended to rinse the instruments just
before the cleaning.
X
4a. Automated
Cleaning
with washer-
disinfector
- Place the devices in a kit,
support or container to
avoid any contact between
instruments or posts.
- Put them in the washer-
disinfector
(Ao value > 3000 or, at
least 5 min at 90 °C).
- Discard any instruments with large obvious defects
(broken, bent).
- Avoid any contact between instruments or posts when
placing in the washer-disinfector use kits, supports or
container.
- Follow instructions and observe concentrations given by
the manufacturer (see also general recommendations).
- Use only approved washer-disinfector according to EN
ISO 15883, maintain and calibrate it regularly.
XXXX
OR
4b Manual Cleaning
or assisted by an
ultrasonic device
- Place the devices in a kit,
support or container to
avoid any contact between
instruments.
- Immerse in the disinfecting
solution with cleaning
properties, assisted by an
ultrasonic device if suitable.
- No visible impurities should be observed on the
instruments.
- Discard any instruments with large obvious defects
(broken, bent, and twisted).
- Follow instructions and observe concentrations and time
given by the manufacturer (see also general
recommendations).
- The disinfecting solution should be aldehyde free and
without di- or triethanolamines as corrosion inhibitor.
XXX
5. Rinsing - Abundant rinsing (at least
1 min)
- Use quality water in accordance with local regulations.
- If a disinfecting solution contains a corrosion inhibitor, it
is recommended to rinse the instruments just before the
autoclaving.
- Dry on a single use non-weaved cloth, or with a drying
machine or filtered compressed air.
XXX
6. Inspection - Inspect devices and sort
out those with defects.
- Assemble the devices
(stops)
- Dirty instruments must be cleaned and disinfected again.
- Discard instruments which show any deformations (bent,
twisted), damages (broken, corroded) or defects (loss of
colour coding or marking) affecting the resistance, the
safety or the performance of the instrument or posts.
- Protect carbon steel bur with corrosion inhibitor before
packaging.
- For Contra Angle : lubricate the device with an adequate
spray before packaging
XXXX