DENTSPLY Cavitron G139 Manual

ENGLISH • 1
Cavitron®Built-In G139
Directions For Use
Ultrasonic Scaler with
Cavitron®Steri-Mate®Handpiece
Model

ENGLISH • 1
Introduction
The Cavitron Built-In Ultrasonic Scaler is equipped with a Sustained
Performance System™ (SPS), which offers a constant balance
between scaling efficiency and patient comfort by maintaining the
unit's power level when the insert tip encounters tenacious deposits
allowing the clinician to effectively scale even at a decreased/lower
power setting.
The System operates by converting an SELV source current into
high frequency current. The ultrasonic system consists of two
parts: an insert and the SPS™electronic system. The SPS™system
incorporates two closed loops. One loop provides automatic tuning
(operating frequency is adjusted to be at resonance for each insert),
the second loop automatically controls the tip stroke over different
working conditions. The DENTSPLY Cavitron®Built-In scaler
produces 30,000 microscopically small strokes per second at the
insert’s working tip. This, combined with acoustic effects of the
coolant water, produces a synergistic action that literally “powers
away” the heaviest calculus deposits while providing improved
operator and patient comfort.
ENGLISH
Cavitron®Built-In Ultrasonic Scaler
with Cavitron®Steri-Mate®Handpiece
Model G139
Directions for Use
Please read carefully and completely before operating unit.

ENGLISH • 2
1. Indications For Use.......................................................................................................3
2. Contraindications ..........................................................................................................3
3. Warnings .......................................................................................................................3–4
4. Precautions....................................................................................................................5
5. Adverse Reactions ........................................................................................................5
6. Infection Control...........................................................................................................6
6.1 General Information
6.2 Water Supply Recommendations
7. Water Line Requirements .............................................................................................6
8. Ultrasonic Scaler System Description ..........................................................................7–10
8.1 System Controls
8.2 Steri-Mate®Handpiece
8.3 DENTSPLY Cavitron®30K™ Ultrasonic Inserts
8.4 Foot Control Information & Operation
9. Techniques for Use .......................................................................................................10–12
9.1 Patient Positioning
9.2 Performing Ultrasonic Scaling Procedures
9.3 Patient Comfort Considerations
10. System Care ..................................................................................................................11–12
10.1 Daily Protocol
11. Troubleshooting ............................................................................................................13–16
12. Warranty........................................................................................................................17
13. Specifications................................................................................................................17
14. Classifications...............................................................................................................17
15. Symbol Identification ...................................................................................................18
Table of Contents
Section Section Title/Description of Contents Page Number

ENGLISH • 3
Section 1: Indications for Use
Ultrasonic Procedures
• All general supra and subgingival scaling applications.
• Periodontal debridement for all types of periodontal diseases.
Section 2: Contraindications
• Ultrasonic Systems should not be used for restorative procedures
involving condensation of amalgam.
• Not for use on children under the age of 3.
Section 3: Warnings
• Persons fitted with cardiac pacemakers, defibrillators and other
active implanted medical devices, have been cautioned that
some types of electronic equipment might interfere with the
operation of the device. Although no instance of interference
has ever been reported to DENTSPLY, we recommend that the
handpiece and cables be kept at least 6 to 9 inches (15 to 23 cm)
away from any device and their leads during use. There are a
variety of pacemakers and other medically implanted devices on
the market. Clinicians should contact the device manufacturer
or the patient’s physician for specific recommendations. This
unit complies with IEC 60601 Medical Device Standards.
• It is the responsibility of the Dental Healthcare Professional to
determine the appropriate uses of this product and to understand:
• the health of each patient,
Product Overview
This product is intended for installation into a dental system. The
end users are dental professionals. The devices are used to debride
light to heavy calculus deposits from the tooth and root surfaces.
Technical Support
For technical support and repair assistance in the U.S.,
call a Cavitron CareSM representative at 1-800-989-8826 or
717-767-8502 Monday through Friday, 8:00 AM to 5:00 PM
(Eastern Time). For other areas, contact your local DENTSPLY
representative.
Supplies & Replacement Parts
To order supplies and replacement parts in the U.S., contact your
local DENTSPLY distributor or call 1-800-989-8826 or 717-767-
8502, Monday through Friday, 8:00 AM to 5:00 PM (Eastern Time).
For other areas, contact your local DENTSPLY representative.
Disposal of Unit
Disposal of unit shall be in accordance with local regulations.
Caution: United States Federal Law restricts this device to sale by or
on the order of, a licensed dental professional.

ENGLISH • 4
• the dental procedures being undertaken,
• and applicable industry and governmental agency
recommendations for infection control in dental
healthcare settings,
• requirements, and regulations for safe practice of
dentistry; and
• these Directions for Use in their entirety, including
Section 4 Precautions, Section 6 Infection Control, and
Section 10 System Care.
• This product is designed to assist in eliminating susceptibility
to retraction of oral fluids. To ensure adequate protection
from cross-contamination from other devices connected to the
Dental Unit, it is highly recommended that the Dental Unit be
installed with anti-retraction features. Additionally, the Dental
Unit, including the anti-retraction features, must be properly
maintained and periodically tested. For more information,
please contact your Dental Unit manufacturer.
• Where asepsis is required or deemed appropriate in the best
professional judgment of the Dental Healthcare Professional,
this product should not be used, unless used in combination with
a Sterile Lavage Kit, (P/N 81340).
• During boil-water advisories, this product should not be operated
as an open water system (e.g. connected to a public water
system). Dental Healthcare Professionals should discontinue use
on patients and contact the local water authority to determine
when it is safe to continue use of this product. When the
advisory is cancelled, the local water authority should provide
guidance for flushing of waterlines. All incoming waterlines
from the public water system inside the dental office (e.g.,
faucets, waterlines, and dental equipment) should be flushed in
accordance with manufacturer’s instructions for a minimum of
5 minutes.
• Prior to beginning treatment, patients should rinse with a known
antimicrobial such as Chlorhexidine Gluconate 0.12%. Rinsing
with an antimicrobial reduces the chance of infection and
reduces the number of microorganisms released in the form of
aerosols during treatment.
• The use of High Volume Saliva Evacuation to reduce the
quantity of aerosols released during treatment is highly
recommended.
• Failure to follow recommendations for environmental operating
conditions, including input water temperature, could result in
injury to patients or users.
• Handle Cavitron insert with care. Improper handling of insert,
specifically the insert tip, may result in injury and/or cross
contamination.
• Failure to follow properly validated sterilization processes and
approved aseptic techniques for Cavitron inserts may result in
cross contamination.
• DO NOT mount the Cavitron Built-In Scaler outside of the
Dental Unit; doing so will void the warranty.

ENGLISH • 5
• If excessive wear is noted, or the insert has been bent, reshaped
or otherwise damaged, discard the insert immediately.
• Ultrasonic insert tips that have been bent, damaged, or reshaped
are susceptible to in-use breakage and should be discarded and
replaced immediately.
• Retract the lips, cheeks, and tongue to prevent contact with the
insert tip whenever it is placed in the patient’s mouth.
• Water should be used for all scaling procedures. Water flow is
adjustable from less than 10 to greater than 60 ml per minute.
• Only adjust the systems’s power adjustment knob with the insert
outside the patients mouth.
• As with all dental procedures, use universal precautions (i.e.,
wear face mask, eyewear, or face shield, gloves, and protective
gown).
Section 5: Adverse Reactions
None known.
Section 4: Precautions
• Equipment flushing and dental water supply system
maintenance are strongly recommended. See Section 10:
System Care.
• Verify handpiece fits properly in the dental delivery system
holder. If not properly seated, there is a potential for the
handpiece to slip out of the holder and be damaged or
contaminated. Contact a service technician for proper fit.
• Close manual shut-off valve on the dental office water supply
every night before leaving the office.
• The use of an in-line water filter is recommended.
• Never operate system without fluid flowing through handpiece.
• Always ensure that the electrical connections on the handpiece
cable and the Steri-Mate® Handpiece are clean and dry before
assembling them for use.
• Cavitron ultrasonic scaling units and Cavitron ultrasonic inserts
are designed and tested for safety and efficacy as a system.
Use of any other brand ultrasonic equipment with Cavitron
equipment has not been tested and may have a negative effect
on the safety and effectiveness of your Cavitron equipment and
their warranties. Please review the warranty statement enclosed
before use.
• Like bristles of a toothbrush, ultrasonic insert tips “wear” with
use. Inserts with just 2 mm of wear lose about 50% of their
scaling efficiency. In general, it is recommended that ultrasonic
inserts be discarded and replaced after one year of use to
maintain optimal efficiency and avoid breakage. A DENTSPLY
Professional Insert Wear Indicator is enclosed for your use.

ENGLISH • 6
Section 6: Infection Control
6.1 General Information
•For operator and patient safety, carefully practice the
infection control procedures detailed in the Infection Control
Information Booklet accompanying your System. Additional
booklets can be obtained by calling Customer Service at
1-800-989-8826, Monday through Friday, 8:00 A.M. to 5:00
P.M. (Eastern Time). For areas outside the U.S., contact
your local DENTSPLY Professional representative.
• As with high speed handpieces and other dental devices,
the combination of water and ultrasonic vibration from the
Cavitron Built-In Scaler will create aerosols. Following
the procedural guidelines in Section 9 of this manual can
effectively control and minimize aerosol dispersion.
6.2 Water Supply Recommendations
• It is highly recommended that all dental water supply systems
conform to applicable CDC (Centers for Disease Control
and Prevention) and ADA (American Dental Association)
standards, and that all recommendations be followed in terms
of flushing, and general infection control procedures (See
Sections 7 and 10). Knowledge of and compliance with
agency guidelines, standards and recommendations is the sole
responsibility of the Dental Healthcare Professional.
• As a medical device, Cavitron®products need to be installed
in accordance with local or national regulations, including
guidelines for water quality (e.g. drinking water). As an open
water system, such regulation may require your Cavitron
product to be connected to a centralized water control device
that prevents water containing contaminates from back-flow
into the water supply.
Section 7: Water Line Requirements
•Incoming water supply line pressure to the ultrasonic
scaler must be 25 psi (172kPa) minimum to 60 psi (414
kPa) maximum. If your dental water system's supply line
pressure is above 60 psi, install a water pressure regulator on
the water supply line to your Ultrasonic Scaler.
• Incoming water temperature to the Cavitron System should
not exceed 25˚C (77˚F). If needed device should be installed
to maintain a temperature within this specification, or a
Cavitron DualSelect Dispensing System attached to allow
this system to be operated as a closed water system.
• A manual shut-off valve on the dental water system supply
line should be used so that the water can be completely shut-
off when the office is unoccupied.
• A filter in the dental water system supply line is
recommended so that the particles in the water supply will
be trapped before reaching the ultrasonic scaler.
• After the above installations are complete on the dental
water supply system, the dental office water line should
be thoroughly flushed prior to connection to the ultrasonic
system.
• After flushing system, verify there are no leaks.

ENGLISH • 7
Section 8: Ultrasonic Scaler System
Description
8.1 System Controls
The DENTSPLY Cavitron® Built-In module is enabled
whenever the clinician removes the handpiece from the
handpiece holder.
ON/OFF Function
The Ultrasonics is switched ON and OFF using the treatment unit’s
foot control.
Power Adjustment
There are 2 options for adjusting the power to your Cavitron Built-In
Ultrasonic Scaler.
Option 1: Power is adjusted by turning the Power
Control Knob located on the treatment unit head
that has been installed by your service technician.
Stickers have been provided for application to the
treatment unit to denote the Cavitron Ultrasonic
Scaler power range (see image). These stickers
are not required and may or may not have been applied by your
technician. If they have not been applied or left behind by the
technician and you would like to apply them to your treatment unit,
contact DENTSPLY Customer Service to order them.
Option 2: If you are using a treatment unit equipped with a touch
pad, your service technician will wire the power adjustment to the
appropriate buttons on the treatment unit. Power can be adjusted
using these buttons.
Lavage Flow Adjustment
Water flow through the handpiece is adjusted by rotating a control
element on the cable connector (blue).

ENGLISH • 8
8.2 Steri-Mate®Handpiece
The Steri-Mate handpiece accepts all Cavitron 30K Ultrasonic Inserts. The Cavitron Steri-Mate 360 Handpiece accepts all
Cavitron 30K plastic grip type Ultrasonic Inserts. It is not compatible with the Cavitron 30K Slimline metal grip type inserts
(30K SLI-10S, SLI-10L and SLI-10R).
Cable Assembly
360 Rotating
Handpiece Nose
Handpiece Connector
(Handpiece & Mating
Assembly are keyed)
Steri-Mate®360 Handpiece
Date Codes
Lavage Flow Adjustment
Holding open end of handpiece upright, lavage can be increased by
holding the handpiece and gently twisting the blue lavage adjustment
knob clockwise. The rate of flow through the Handpiece determines the
temperature of the lavage. Low flow rates produce warmer temperatures,
high flow rates produce cooler temperatures.
360 Rotating Handpiece Nose
To rotate the insert, place fingers on the nose of the handpiece and
rotate to desired position. This allows adjustable hand positioning,
free flowing movement and access within the anterior and posterior
of the oral cavity.
BLUE Lavage Knob
Insert Port

ENGLISH • 9
Hold the handpiece in an upright position. Activate to fill the handpiece with water. Lubricate the rubber O-Ring on the insert
with water before placing it into the handpiece. Fully seat insert with a gentle push-twist motion. DO NOT FORCE.
Magnetostrictive Stack
Converts energy provided by the
handpiece into mechanical oscillations
used to activate the insert tip.
Insert Tip
Shape and size of tip
determine access and
adaptation. Preheated
lavage is directed to
the tip.
O-Ring
Provides seal for handpiece coolant
Connecting Body
Transfers and amplifies mechanical motion
of the stack to insert tip.
Finger Grip
Insert Marking
Manufacturer, Date (YDDD = single year and
three digit day of year), Frequency, and Type
(e.g., DENTSPLY 7346 30K FSI-SLI-10S)
8.3 DENTSPLY Cavitron® 30K™ Ultrasonic Inserts
The many styles of DENTSPLY Cavitron®Ultrasonic Inserts are easily interchangeable for various procedures and applications.

ENGLISH • 10
Section 9: Techniques for Use
9.1 Patient Positioning
• The backrest of the chair should be adjusted for optimal
access to both the upper and lower arches. This assures
patient comfort and clinician visibility.
• Have the patient turn head to the right or left.
• Position chin up or down depending on the quadrant and
surface being treated.
• Evacuate irrigant using either a saliva ejector or High
Vacuum Evacuator (HVE).
9.2 Performing Ultrasonic
Scaling Procedures
• The edges of DENTSPLY Cavitron®Ultrasonic Inserts
are intentionally rounded so there is little danger of tissue
laceration with proper ultrasonic scaling techniques. The
lips, cheek and tongue should be retracted to prevent
accidental prolonged contact with the activated tip
whenever the insert is placed in the mouth.
• Always adjust the lavage so adequate fluid will be
available to cool the tip-tooth interface.
• In general, it is suggested that a “feather light touch” be
used both supra- and subgingivally. The motion of the
activated tip and acoustic effects of the irrigating fluid, in
most cases, is adequate to remove even the most
tenacious calculus.
8.4 Foot Control Information & Operation
• The dental unit manufacturer provides the Foot
Control. Refer to their directions for use for operating
characteristics.

ENGLISH • 11
assembly from handpiece holder and set power
adjustment to 30% of maximum.
3. Connect Steri-Mate®handpiece to the cable connector
assembly (refer to the Infection Control information
booklet for sterilizing instructions).
4. Adjust the lavage control to maximum.
5. Hold the handpiece (without an insert installed) over a
sink or drain. Activate the foot control and flush water
through handpiece for at least two minutes.
6. Place a sterilized insert into the handpiece using a gentle
push-twist motion. Refer to the Cavitron Ultrasonic
Inserts Directions for Use for cleaning and sterilizing
instructions.
7. Activate the ultrasonics and adjust the power and lavage
control to your preferred operating positions.
Between patients
1. Remove any ultrasonic inserts and the Steri-Mate®
handpiece. Clean and sterilize the handpiece and all
inserts used during the procedure (refer to the Infection
Control information booklet and the Cavitron Ultrasonic
Inserts Directions for Use for cleaning and sterilizing
instructions).
2. Clean and disinfect the handpiece cable assembly by
applying a medically approved non-immersion type
disinfectant solution* carefully following the instructions
• Set the System’s power adjustment knob to the lowest
power setting for the application and the selected insert.
• Only adjust the System’s power adjustment knob with
the insert outside the patients mouth.
9.3 Patient Comfort Considerations
Reasons for sensitivity
• Incorrect tip placement. Point should be directed away
from root surfaces.
• Not keeping the tip in motion on the tooth. Do not allow
the insert to remain in a static position on any one area
of the tooth. Change the insert’s path of motion.
• Applying pressure. Use extremely light grasp and
pressure, especially on exposed cementum.
• If sensitivity persists, decrease power setting and/or
move from the sensitive tooth to another and then return.
Section 10: System Care
10.1 Daily Protocol
Start-up procedures at the beginning of the day
1. Switch ON the dental treatment unit (refer to instruction
manual of treatment unit).
2. Remove the DENTSPLY Cavitron®cable connector

ENGLISH • 12
provided by the solution manufacturer. To clean the cable,
generously spray disinfectant solution on a clean towel
and wipe the cable and connector. Discard used towel. To
disinfect the system, generously spray disinfectant on a
clean towel and wipe the cable and connector. Allow the
disinfectant solution to air dry.
3. Clean and disinfect all surfaces of the dental treatment
control head according to the manufacturer’s instructions.
4. Connect a sterilized Steri-Mate®handpiece onto its
mating cable connector.
5. Hold the handpiece (without an insert installed) over a
sink or drain. Activate the foot control and flush water
through handpiece for at least thirty seconds.
6. Place a sterilized insert into the handpiece.
Shut down at the end of the day
1. Flush the DENTSPLY Cavitron Built-In ultrasonic system
according the dental treatment unit manufacturer’s
instructions.
2. Remove any ultrasonic inserts and the Steri-Mate®
handpiece. Clean and sterilize the handpiece and
all inserts used during the procedure (refer to the
Infection Control information booklet and the Cavitron
Ultrasonic Inserts Directions for Use for cleaning and
sterilizing instructions).
3. Clean and disinfect the handpiece cable assembly
by applying a medically approved non-immersion
type disinfectant solution* carefully following the
instructions provided by the solution manufacturer.
To clean the cable, generously spray disinfectant
solution on a clean towel and wipe the cable and
connector. Discard used towel. To disinfect the system,
generously spray disinfectant on a clean towel and
wipe the cable and connector. Allow the disinfectant
solution to air dry.
4. Clean and disinfect all surfaces of the dental treatment
control head according to the manufacturer’s
instructions.
5. Place the cable connector into the handpiece holder for
storage.
6. Turn OFF the power and water to the dental treatment
unit.
*NOTE: Water-based disinfection solutions are preferred. Some alcohol-
based disinfectant solutions may discolor the handpiece cable
and connector.

ENGLISH • 13
wear indicator
Section 11: Troubleshooting
Some basic troubleshooting procedures that will help avoid
unnecessary service calls are listed below.

ENGLISH • 14

ENGLISH • 15

ENGLISH • 16
Verify input
water temperature
is not higher than
77OF (25OC)

ENGLISH • 17
Section 12: Warranty Period
The Cavitron®Built-In G139 Ultrasonic Scaler is warranted
for TWO YEARS from date of purchase. The Steri-Mate®
Handpiece enclosed with your system is warranted for SIX
MONTHS from date of purchase. Refer to the Warranty
Statement Sheet furnished with your system for full Warranty
Statement and Terms.
Section 13: Specifications
Operating Conditions:
Ambient temperature: 15 – 40°C
Humidity: 30 – 75 %RH
Input for Optional Power Supply if used:
APX Model AP7948DR
Voltage 120 VAC
Power 60 Watts
Frequency 60 Hz
Output Voltage 24 VAC
Output Current 2.5 Amps
Input for Model G139 Scaler Unit
Voltage: 24 VAC
Current 2.5 Amps
Rated Power 60 VA
Frequency 50/60 Hz
Water Temperature 41 – 77°F (5 – 25°C)
Water Pressure 25 – 60 psi (172 – 414 kPa)
Output Frequency 30 kHz
Power 3 – 30 Watts
Water Flow Rate 10 > flow > 60 ml/min
Storage and Shipping Conditions:
Ambient temperature range: -40 – 75°C
Relative Humidity range: 10 – 95 %RH
(non-condensing)
Atmospheric pressure range: 7 – 15 psi (50 – 106 kPa)
Section 14: Classifications
Input Voltage supplied by SELV
Degree of protection against electric shock: Type B
Degree of protection against harmful ingress of water:
Ordinary
Mode of operation: Continuous
Equipment is not suitable for use in the
presence of flammable mixtures
Medical Device Directive Classification: IIa
• Input Voltage to the Cavitron Built-In Ultrasonic Scaler G139, when
not supplied by the dental unit, may only be supplied by the specified
power supply from APX, Model AP7948DR.

ENGLISH • 18
Type B Equipment AC power
(AC 24~)
MEDICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK, FIRE AND
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH
UL 60601-1CAN/CSA-C22.2 No. 601.1,
ANSI/AAMI ES60601-1 (2005, 3rd ed.)
CAN/CSA-C22.2 No. 60601-1 (2008),
13VA
Consult Instructions
For Use
Section 15: Symbol Identification
This symbol is a mandatory marking for devices enter-
ing the European market to indicate conformity with the
essential health and safety requirements set out in European
Directives. The symbol may be accompanied by a four-digit
identification number of the notified body.

PrintedinU.S.A.Form#88016Rev.1(07/15)
Worldwide Service Centers
Manufactured by:
DENTSPLY Professional
1301 Smile Way
York, PA 17404-1785
Imported and
Distributed by:
DENTSPLY Canada
Woodbridge, Ontario
L4L 4A3
DENTSPLY DeTrey GmbH
De-Trey-Str. 1
78467 Konstanz
Germany
DENTSPLYAsia
25/F,PacificLinkTower
Southmark,11YipHingStreet
WongChukHang,
HongKong
DENTSPLYDental(Tianjin)Co.,Ltd
No.151,6thAvenue,
TEDA300457
Tianjin,China
DENTSPLYInt’lInc.(Taiwan)
7F-11,351Sec.2
Chung-ShanRoad
Chung-Ho,TaipeiCounty
Taiwan
DENTSPLYThailandLimited
23/F,PanjathaniTower
127/128RatchadapisekRoad
Chongnonsee,Yannawa
Bangkok10120
DENTSPLYMexico,S.A.deC.V.
CalzadaVallejoNo.846
ColIndustrialVallejo
02300Mexico,D.F.
Manufacturedby:
DENTSPLYProfessional
DENTSPLYLLC
1301SmileWay
York,PA17404
DENTSPLYJapanK.K.
TsunashimaNo.2Building
20-12Yushima3-Chrome
Bunkyo-Ku,Tokyo113Japan
DENTSPLY(Australia)Pty,Ltd.
11-21GilbyRoad
MountWaverley,Victoria3149
Australia
DENTSPLYDeTrey(Switzerland)
Oberdorfstrasse11
CH-6342Baar,Switzerland
DENTSPLYCanada
161VinylCourt
Woodbridge,Ontario
CanadaL4L4A3
DENTSPLYIHLimited
BrunelWay
StroudwaterBusinessPark
Stonehouse,Gloucestershire
GL103SX
UnitedKingdom
DENTSPLYFranceS.A.S.
17,rueMichaelFaraday
78180MontignyleBretonneux
France
DENTSPLYDeTreyGmbH
DeTrey-Strasse1
D-78467Konstanz,Germany
DENTSPLYDeTreyGmbH
ServiceCenter
GuttingerStrasse35
D-78315Radolfzell,Germany
DENTSPLYDeTreyItaliaS.r.l.
ViaCurtatone,3
00185Rome,Italy
Other manuals for Cavitron G139
1
Table of contents
Other DENTSPLY Media Converter manuals
Popular Media Converter manuals by other brands

LEAB
LEAB PP 48-12/29 Assembly instructions

M-system
M-system M8CT instruction manual

Elsner
Elsner GPS-DCF Converter UTC± Technical specifications and installation instructions

Steren
Steren 207-340 user manual

SeeEyes
SeeEyes SC-IPH3002DC user manual

Speedaire
Speedaire Chisel Scaler Operating instructions & parts manual