SCREEN ICONS
100% Measured SpO2% Data storage icon
PR Pulse rate display area Weak signal icon
60 BPM Measured Pulse rate
(BPM) 04: 59 Time in informaon area
“hour: minute”
Indicates measured value
is higher than the high
alarm limit
ADU/NEO Paent type:
ADU or NEO
ID:99 Paent ID in
informaon area
Indicates measured value
is lower than the low alarm
limit
SpO2 waveform display Indicates sensor is o or is
not well t on the nger
Pulse amplitude display Sensor unconnected icon
Low baery icon Indicates the memory space
is full
Alarm silence icon Alarm system is o
SpO2SpO2 Display
•To install baeries: Press the baery compartment latch to remove the baery
access door. Place four (4) AA baeries inside using the + and - guides for orientaon.
•Use the System Setup Menu to select paent type: Adult, Pediatric or Neonatal.
•Default SpO2 low/high alarms are set for 90/100 for adults and 90/95 for neonates.
•Default pulse rate low/high alarms are set for 50/120 for adults
and 100/200 for neonates.
•Alarm and pulse tone volumes can be adjusted within a range from one to ve.
•To transfer data to a PC: Turn o the oximeter and push the Power and
Funcon buons at the same me for one second to start data transfer mode.
•To clean: Power o the oximeter and remove the baeries before cleaning with a cloth
that has been dampened with approved 75% ethanol or 70% isopropanol cleaning
soluon. Do not use ammonium disinfectants to prevent damaging the oximeter.
AVAILABLE ACCESSORIES
02.01.210122 Adult Reusable Silicone
So-Tip SpO2 Sensor
02.01.210121 Pediatric Reusable Silicone
So-Tip SpO2 Sensor
02.01.210673 Neonate Reusable Wrap
SpO2 Sensor
01.13.110504 SpO2 Extension Cable (DB9
to DB9)
03.16.113589 Rechargeable Ni-MH Baery
Kit with Charging Stand
TIPS FOR OPERATION