dispomed Moduflex COMPACT User manual


805-0975-030
2DISPOMED Ltd. / 1 800 363-1746 / [email protected] / www.dispomed.com
TM/MC
TABLE
OF CONTENTS
User Responsibility .......................... 3
Standard and Optional Features ............... 4
Receiving , Assembly and Instructions .........5-9
Assembled View ............................10
Optional accessories Assembly Instructions ...11-14
Assembled View With Optional Accessories .....15
Daily Leak Check Procedure ............... 16-17
Moduflex Compact ..........................18
Annual Maintenance.........................19
Consumable Materials List...................20
Warranty...................................21
Order Part Numbers ..................... 22-23
ModuflexTM is a trademark of Dispomed Ltd.

805-0975-030 3
TM/MC
Read this manual before operating the unit. Pay special attention to all
warnings and cautions.
• This product will perform in conformity with the description
contained in this operating manual when assembled,
operated, maintained and repaired in accordance with the
instructions provided.
• This product must be verified periodically.
• A defective product should not be used.
• Parts that are broken, missing, plainly worn, distorted or
contaminated should be replaced immediately.
• Should a repair or replacement become necessary, Dispomed
recommends placing a request by phone or in writing to the
nearest Dispomed regional service center.
• This product, or any of its parts, should not be repaired other
than in accordance with written instructions provided by
Dispomed and by Dispomed trained personnel.
• The product may not be altered without prior written approval
from Dispomed.
• The user of this product shall have the sole responsibility
for any malfunction which results from improper use, faulty
maintenance, improper repairs, damage or alteration by
anyone other than a Dispomed trained technician.
USER
RESPONSIBILITY

805-0975-030
4DISPOMED Ltd. / 1 800 363-1746 / [email protected] / www.dispomed.com
TM/MC
STANDARD FEATURES
• 0.1 to 4 LPM O2 flowmeters (up to 2 flowmeters can be installed).
• Rotary float and large scale for easy readability.
• Recessed flush valves to prevent accidental activation.
• Quick connect fresh gas outlet for easy and secure switching
between breathing circuits.
• SELECTATEC™ compatibility manifold for easy and quick installation of vaporizer.
• Five caster spider bases.
• 1500 ml CO2 absorber designed for easy and fast replenishing of soda lime.
• Condensation trap to prevent water build-up in soda lime.
Easy and efficient gas sealing.
• Momentary closed features incorporated into our pop-off valve.
• 2.5 meter oxygen supply hose (D.I.S.S.)
• 5-year warranty.
OPTIONAL FEATURES
• Small cylinder holder
• Bain circuit adaptor
• 0-1 LPM O2 flowmeter
• Up to 2 flowmeters can be installed
• Circuit pressure alarm
• Shelf assembly
• Wall mount unit
THE COMPACT
ANESTHESIA MACHINE

805-0975-030 5
TM/MC
Your Moduflex Compact anesthesia machine re-
quires some assembly before use.
RECEIVING
In order to reduce the possibility of damage to
a minimum during shipping, certain components
have been detached from the unit. These com-
ponents are packed separately within the ship-
ping crate.
INSPECTION
When received your unit, inspects the shipping
container and the enclosed components for
signs of damage. Check your Materials List to
ensure that all the necessary components are
included.
RECEIVING AND ASSEMBLY
INSTRUCTIONS
x1
x1
x1
x1 x1
x1
x1
x1
x1x1
x1

805-0975-030
6DISPOMED Ltd. / 1 800 363-1746 / [email protected] / www.dispomed.com
TM/MC
Remove the head assembly from the inner box
and insert into the top of the post (there is a
large hole under the head assembly manifold)
and tighten the head assembly by securing the
screw onto the post using the key supplied.
RECEIVING, ASSEMBLY
INSTRUCTIONS CONTINUED
Remove the spider base from the box and slide
the tapered end of the post onto it.
1 2

805-0975-030 7
TM/MC
Install a breathing bag on the rotary bag mount
of the absorber.
RECEIVING, ASSEMBLY
INSTRUCTIONS CONTINUED
Remove the accessories from the bag.
34

805-0975-030
8DISPOMED Ltd. / 1 800 363-1746 / [email protected] / www.dispomed.com
TM/MC
1. Loosen the absorber sealing screw and re-
move the soda lime canister. 2. Affix soda lime
expiration label to the soda lime container. 3.
Fill the canister with absorbent using loose-fill
granules. (Caution: do not overfill the canister;
leave approximately ¼” (5 mm) from the top
empty.) 4. Reinstall the canister and tighten the
sealing screw.
RECEIVING, ASSEMBLY
INSTRUCTIONS CONTINUED
Install the breathing circuit on the front connec-
tors of the absorber.
65
1
2

805-0975-030 9
TM/MC
Connect the other end of the oxygen tube to 50
p.s.i. (3 bars) oxygen supply.
Install the oxygen supply hose to the rear of the
machine.
RECEIVING, ASSEMBLY
INSTRUCTIONS CONTINUED
87

805-0975-030
10 DISPOMED Ltd. / 1 800 363-1746 / [email protected] / www.dispomed.com
TM/MC
ASSEMBLED
VIEWS 1
Absorber container
Post
Flowmeter
Flush valve
Base

805-0975-030 11
TM/MC
1. Insert the vaporizer into the bracket. 2. Secure
the evaporator in place with the locking mecha-
nism.
OPTIONAL ACCESSORIES
ASSEMBLY INSTRUCTIONS
1. Install the universal bracket (not included) at
the desired height on the anesthesia machine
post. 2. Screw the interface to the bracket using
the screws provided in the package.
During surgery unused gas should be evacuated
outside of the building to avoid it being inhaled
by your employees and patients. Connected to
a gas evacuation ventilator, the scavenger inter-
face controls the flow of suction and maintains
the balance with the quantity of unused gas.
IMPORTANT : Always refer to the user manual of the option you are installing for more detailed instructions.
SCAVENGER INTERFACEVAPORIZER
2
1
1
2

805-0975-030
12 DISPOMED Ltd. / 1 800 363-1746 / [email protected] / www.dispomed.com
TM/MC
1. Screw the Bain adapter holder onto the anes-
thesia machine. 2. Insert the Bain adapter into
the bracket previously attached to the unit.
IMPORTANT: Always refer to the user manual of the
option you are installing for more detailed instruc-
tions.
*Bain type open circuit on an anesthesia ma-
chine gives you better control during surgery.
The breathing bag is always inflated, no need
to wait for it to have reached a certain size be-
fore ventilating the patient. Moduflex Bain is
equipped with a pressure gauge that gives you
the possibility to check the animal’s lung pres-
sure. The Bain adaptor is equipped with a mo-
mentary valve to avoid forgetting the valve in a
closed position, which could cause pulmonary
trauma to the animal.
OPTIONAL ACCESSORIES
ASSEMBLY INSTRUCTIONS
1. Screw the circuit alarm holder on the anes-
thesia machine. 2. Screw the circuit alarm to the
support previously screwed to the device.
During assisted ventilation, the alarm will alert
you if the predetermined pressure level is
reached. For example, if you forget the evacua-
tion valve in a closed position, the circuit alarm
will alert you with a warning sound at the prede-
termined pressure.
BAIN CIRCUIT ADAPTOR PATIENT CIRCUIT ALARM
11
2
2

805-0975-030 13
TM/MC
Installation of dual cylinder holders.
The cylinder holder can be helpful in many situ-
ations. It can be very useful when there is a lack
of oxygen in the central system. In the occasion-
al case where you need to move an anesthetized
animal to a room where there are no oxygen out-
lets, the cylinder holder enables you to keep the
patient in an ideal condition.
IMPORTANT: Always refer to the user manual of the option you are installing for more detailed instructions.
1. Screw the wall bracket to the reinforced wall*.
2. Insert the wall arm to the wall bracket, then
insert the anesthesia machine.
*The support wall must be reinforced 5 feet high
by 3 feet wide to accommodate the anesthesia
machine.
OPTIONAL ACCESSORIES
ASSEMBLY INSTRUCTIONS
CYLINDER HOLDER WALL MOUNT
12

805-0975-030
14 DISPOMED Ltd. / 1 800 363-1746 / [email protected] / www.dispomed.com
TM/MC
Install the shelf as shown in the illustrations
above.
1. Insert the two shelf brackets around the posts
under the soda lime container. 2. Screw the two
shelf supports onto each of the previously in-
stalled brackets. 3. Place the shelf on the two
installed brackets and screw together.
IMPORTANT: Always refer to the user manual of the option you are installing for more detailed instructions.
OPTIONAL ACCESSORIES
ASSEMBLY INSTRUCTIONS
2
3
1

805-0975-030 15
TM/MC
ASSEMBLED VIEW
WITH OPTIONAL ACCESSORIES
Bain Circuit Adaptor
965-0500-000 p.12
Reusable Rebreathing Bag
940-0200-020 p.7
Patient Circuit Alarm
987-0010-000 p.12
Vaporizer
990-VI5P-SVEEZ p.11
Shelf
975-0700-700 p.14
Y Patient Circuit
940-3704-800 p.8
Cylinder Holder
975-0300-400 p.13
Semi-active Scavenger Interface
985-0035-000 p.11

805-0975-030
16 DISPOMED Ltd. / 1 800 363-1746 / [email protected] / www.dispomed.com
TM/MC
Connect the circuit to be verified to the fresh
gas outlet of the anesthesia machine.
Obstruct the patient’s circuit outlet with a plug
or your thumb.
Open flowmeter to approximately 200 ml. Close the gas evacuation valve.
The unit’s integrity should be verified every day, and at each re-
placement of patient circuit components such as breathing bags,
tubes, etc. A proper maintained anesthesia machine reduces gas
consumption and operating room pollution.
Prerequisites:
Oxygen supply of 40 to 60
psi (3 bars) Breathing circuit
and bag.
DAILY LEAK
CHECKS PROCEDURES
21
3 4

805-0975-030 17
TM/MC
If you encounter an excessive leak, it is rec-
ommended that you proceed by elimination
to find the source of the problem by replac-
ing a single component at a time. The most
frequent leaks can be found at the fresh gas
outlet o’ring, the patient circuit and the bag.
Once the leak is identified, dispose of the de-
fective item to avoid it being reused by error.
At the end of each day, it is recommended
that you close the oxygen tank and fill the
anesthetic agent vaporizer.
DAILY LEAK
CHECKS PROCEDURES
8
Press on the oxygen flush button and hold
until the pressure on the manometer indicates
approximately 30 cm H2O.
Check the manometer. The pressure must
not drop (but it can increase). If the pressure
drops, replace the patient circuit and the bag
and try again.
Caution: Open the evacuation valve before
using the anesthesia machine again.
5 6
7

805-0975-030
18 DISPOMED Ltd. / 1 800 363-1746 / [email protected] / www.dispomed.com
TM/MC
MODUFLEX COMPACT
ANESTHESIA MACHINE

805-0975-030 19
TM/MC
ANNUAL
MAINTENANCE
Prerequisite:
Preventative maintenance kit
# 975-0900-900
The maintenance kit contains
all the necessary parts to per-
form the annual maintenance.
Refer to this technical draw-
ing the parts are identified.
It is important to perform the
daily leak check procedure
(page 17) once the mainte-
nance has been completed.
980-0300-300 x1
845-S019 x2
845-S019 x2
845-S011 x2
845-S106 x2
845-S114 x2
845-S113 x2
845-S011 x1
980-0300-180 x2
845-S227 x2
980-0300-100 x1
980-0300-105 x1

805-0975-030
20 DISPOMED Ltd. / 1 800 363-1746 / [email protected] / www.dispomed.com
TM/MC
CONSUMABLE MATERIALS
LIST
7g : # 850-0052-000
Y” Patient Circuit 40″ (101.6cm)
long: # 940-3704-800
152.4cm / 5’ :
# 985-0065-000
1.6 liters : # 905-8855-000
18.9 liters : # 905-8862-000
KRYTOX
Y PATIENT CIRCUIT
GAS EVACUATION
TUBING
CARBON DIOXIDE
ABSORBENT
0.5 liter: # 940-0200-005
1 liter: # 940-0200-010
2 liters: # 940-0200-020
3 liters: # 940-0200-030
REBREATHING BAGS