DRC DR 30000 PK User manual

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 1
Table of Contents
1 Abstract ............................................................................ 4
1.1 Introduction .................................................................. 4
1.2 Fields of application ...................................................... 5
1. Technical Specifications ................................................ 5
1.5 Care and Maintenance .................................................. 6
1.6 Liability .......................................................................... 9
2 Introduction to Product .................................................. 10
2.1 Product Appearance ................................................... 10
2.2 Button Board Instructions .......................................... 11
2. Accessories Instructions ............................................. 12
3 Product Instructions ....................................................... 12
.1 Boot Instructions ......................................................... 12
.2 Operating Instructions ................................................ 12
.2.1 Start Measuring ....................................................... 1
.2.2 Rebar Position & Thickness Measuring ............... 18
.2. Rebar Diameter Measurement(Limited to Different
Models)...................................................................................... 21

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 2
.2.4 JGJ Measurement (Limited to Different Models) ... 21
.2.5 Data Management (Limited to Different Models) .. 2
.2.6 System Setting ...................................................... 25
.2.7 Power Off .............................................................. 26
Appendix Ⅰ: On-site Notice ............................................. 27
Appendix Ⅱ: Menu Quick Index ...................................... 28
Appendix Ⅲ Measurement and Calibration .......... Errore. Il
segnalibro non è definito.
Appendix Ⅳ: Relevant Standards ..................................... 29
Manufacturer CertificateErrore. Il segnalibro non è definito.
Packing List of ZD 2X Rebar DetectorErrore. Il segnalibro non
è definito.
Product Warranty CardErrore. Il segnalibro non è definito.

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page
The Operating Manual Instructions:
1. The symbol【】in this manual represents a button on the product panel,
such as【 】.
2. The symbol in this manual represents supplementary
specifications.
. In addition to the contents of this manual, there are some other
prompts displayed automatically when operating, please follow the
prompts.
4. The software interface and figure in this manual are just for illustrative
purpose and may be subjected to change without prior notice as
software upgrade and product improvement.

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 4
1 Abstract
1.1 Introduction
The analyzes that can be performed on structures in order to obtain
information on the armatures, the presence of irons or carpentry, is
called the magnetometric or pacometric survey. The aim of this
survey is mainly to locate the presence of bars and / or reinforcement
brackets on structural elements. Once the localization has been
carried out and the same are highlighted on the element, it will be
possible to carry out non-destructive surveys on the concrete thus
avoiding to have results falsified by the presence of irons or even to
cut them in the case of performing cores. All the Non-Destructive
investigations that have the objective of the mechanical
characterization of the building materials can and must be performed
after a pacometric survey.
Through the pacometric survey, in addition to the identification and
location of the reinforcements, it is possible to measure the diameter
of the iron and the thickness of the cover. This measurement must
always be verified by making an opening (removal of the concrete
cover) near the localized iron for checking or calibrating the
instrument.

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 5
1.2 Fields of application
The fields of application are:
Rebars Identification;
Cover Measure;
Dimensions measurements of rebars in structural elements;
Localization of hidden metal objects.
1.3 Technical Specifications
Items Technical Spec.
Diameter Measuring
Range(mm) Ф6~Ф50
Measuring Range(mm)
Location Range: 1~120
First Range: 1~60
Second Range: 0~120
Thickness Measuring
Accuracy(mm)
1~40 ±1
41~60 ±2
61~80 ±
81~100 ±5
101~120 ±8
Rebar Location Accuracy 1~60 ±

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 6
61-120 ±6
Diameter Measuring
Modes
Estimation of the diameter of the
investigated component rebar
JGJ Measurement Possibility to re-measure the
single point of each bar
Data Storage Over 10.000 values
Off Time Automatic
Power Supply Rechargeable Lithium Battery
Working Temperature -10
℃
~+42
℃
Size(mm) 170×78× 8
Weight(kg) 0.28
1. Care and Maintenance
1
.
Operation Notice
Do not operate the button with too much force and stain it with oil and
mud so as not to affect the button board life.
2. Power Supply
This product built-in rechargeable lithium battery is left in standby model
for at least 20h if fully charged. Please pay attention to battery indicator
when using, if the battery is low, you should turn off the power as soon as
possible and charge timely, otherwise the sudden power failure may

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 7
cause loss of test data or even damage to the product.
Do not use other batteries or power supply for charging, otherwise it may
cause the damage to the product, battery leakage, fire etc. If you have
any problems, please contact us.
Tips: The battery capacity will be insufficient after it is used for
some time, which the symbol will be displayed in the top right
corner of the screen. The more the gray part, the more battery, the
symbol shows that the product need charging as the little gray part
displayed.
. Battery Charging
This product built-in lithium battery uses Micro USB standard port. It is
recommended that the battery should be charged under the shutdown
state. Please connect the power outlet to AC220±10%V power socket, the
other end Micro USB port access to the bottom of the product, or
connected directly on a computer with the USB cable.
If the indicator light turns red when charging, it shows the battery is
being charged. When the red light turns green or off, it shows that
battery has completely charged. The charger or USB cable should be
pulled out in time so as not to overcharge the battery and affect the
battery life. During the charging, there will be the normal phenomenon
that battery and charger will produce a certain quantity of heat. It is
recommended to place the product in a ventilated place.
Tips: In order to ensure the battery completely being charged one

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 8
time, please keep charging for hours or so, and do not charge the
battery in an environment exceeding 50℃. It is recommended that you
should use the manufacturer's standard charger and USB cable, or it may
cause damage to the product.
4. Lithium Battery
General life of rechargeable battery is for charging and discharging 500
times or so. If the battery does not work, no power and low power when
charging, or working time is short every time, it may have been damaged.
Please contact our after-sales service department to replace new battery.
It is forbidden to short-circuit the battery and place it near the heat
source. The battery capacity is 2000mAH (7.4Wh).
5. Preserve & Clean
When the instrument is not used, please place it in a ventilated, cool and
dry (relative humidity less than 90%) place of room temperature. If you
do not use for a long time, please charge it before using, and regularly
turn on it, and charge it once a month at least as well because
rechargeable battery will discharge automatically resulting in power loss.
After using the instrument each time, it should be properly cleaned to
prevent water, oil, mud and dust from the connector for fear of affecting
the test performance.
Tips: Do not immerse the instrument and accessories in water or
scrub them with a damp cloth! Do not use any organic solvents or acid
and alkaline liquid! Please use a clean, dry cloth to scrub the instrument

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 9
and clean the jack with a soft brush.
6. Fault and Handling Method
Instrument cannot boot: please check the battery power whether it is
sufficient or directly insert it into the charger, turn on the soft switch. If
the above method is invalid, connect the charger to charge the battery
for half an hour before switching on.
Instrument automatic shutdown: the product has the function of the
detecting power capacity. The product will automatically shut down
when the power is too low, charge the battery for some time, or directly
insert it into the charger before switching on.
Tips: we provide a year warranty and life-long maintenance services.
Please contact us if you need maintenance service. Repair it by yourself is
not suggested.
1.6 Liability
The product is a precise instrument. In the following cases, the company
does not undertake any liability.
1. Abnormal operation
2. Disassembling, repairing and refitting by the organizations or
personnel that are not authorized by our company.
. Do not follow the requirements for working and storage environment
mentioned above.
4. Serious damage caused by man-made or accidental impact.

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 10
2 Introduction to Product
2.1 Product Appearance and buttons
(a) Top View
(b) Bottom View
Fig. 2.1 General View
Tips: The actual instrument may differ from the schematic, please in
kind prevail.

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 11
2.2 Button Board Instructions
Each button is on the host panel and the function specifications are
shown in the table 2.2.
Table 2.2 List of Function Buttons
Button Symbol Function Spec.
【 】
Power On/Off, Confirm, Move Right
Option
【▲】
Move Up Option, Switch Numbers and
Letters, set “second range” in the
measure inteface
【▼】
Move Down Option, Switch Numbers
and Letters
【 】 Return, Move Left Option
【Side Upper
Button】
Shortcut key, Return
【Side Lower
Button】
Delete Data, Diameter Measuring,
Return
Tips: Some buttons have different functions in different interfaces
as shown in the relevant introduction.

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 12
2.3
Accessories Instructions
1. USB Cable
Micro USB cable is used to connect the instrument and the charger for
charging as well as connect the instrument and the computer for data
upload as shown in Figure 2. .
Fig. 2. USB Cable General View
3 Product Instructions
3.1 Boot Instructions
Press the button 【 】to power on, seconds later, the main menu is
displayed.
3.2 Operating Instructions
The main menu includes Start Measuring, Data Management (limited to
different models), System Setting, and Power Off.
In the main menu, Start Measuring is by default, and the current
function menu displays in gray background. The different function menus
can be chosen through pressing the 【▲】&【▼】 and confirming to

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 1
enter the chosen function menu.
3.2.1 Start Measuring
The start measuring menu mainly realized these functions such as
measurement setting, rebar diameter setting, component name setting,
stirrup spacing setting, JGJ scanning setting, instrument calibration
setting, as shown in figure .2 (a, b, c, d, e, f).
a. Start Measuring Interface b. Rebar Diameter Setting

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 14
c. Component Name Setting d. Stirrup Spacing Setting
e. JGJ Measurement Setting f. Instrument Calibration
Fig. .2 Start Measuring Menu Interface

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 15
In the start measuring interface, please operate the following buttons to
select the function.
【▲】: Select the function upwards to set and adjust the value of
parameters (recyclable used)
【▼】: Select the function downwards to set and adjust the value of
parameters (recyclable used)
【 】: Enter selected setting and set cursor parameter
【 】: Return to previous menu
Use the rebar diameter setting menu as an example in the following
operation.
When you press【 】to select Rebar Diameter Setting option, it will be
displayed in gray background. Then adjust the value to change the
diameter size by pressing the【▼】,【▲】until appropriate value appeared.
Finally, press 【 】to return to rebar diameter setting option. The same
operation applies to Component Name Setting, Stirrup Spacing Setting,
JGJ Measuring Setting.

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 16
Fig. . Rebar Diameter Size Setting
Calibration: When the measuring environment is changed or error
measurement result is appeared, you should perform zeroing, as shown
in figure .4.
In calibrating, keep the instrument away from ferromagnetic metal
objects, just in the air. Select the Calibration by pressing 【▼】,【▲】, and
press the 【 】 to calibrate. When the interface display Calibration
Completed, press any key to return. Once the exception has been
displayed, please recalibrate.

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 17
Fig. . 4 The Instrument Calibration
The parameters setting of each option as follows:
(1) Rear Diameter
A range of rebar specifications: 6, 8, 10, 12, 14, 16, 18, 20, 22, 25, 28,
2, 6, 40, 50.
(2) Component Name
Component name consists of two letters and three digits, which are
A, B, C, D, E, F, G, H, I, J, K, L, M, N, O, P, Q, R, S, T, U, V, W, X, Y, Z, 1, 2,
, 4, 5, 6, 7, 8, 9, 0 respectively, its default setting is HD001.
The default setting is HD001, which automatically plus one after the
confirming, and the users manually adjust as needed.
(3) Stirrup Spacing
The stirrup spacing parameter setting: if the on-site stirrup spacing is
less than 125mm, the parameter is set to <125, compensated
automatically. If the on-site spacing is greater than 125mm, the

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 18
parameter is set to >125.
( ) JGJ Measurement (limited to different models)
No: unused, Yes: used.
3.2.2 Rebar Position & Thickness Measuring
Select the Start Measuring (press 【Side Upper Button】if you want to
select “second range” H) and move the instrument rightwards slowly and
evenly along the concrete surface until the alignment mark displayed as
square-shaped frame is changed from (hollow square) to
(cross-line in solid square) , indicator light in the front of screen
turns red accompanied by the beep prompt tone, which shows the
instrument is precisely aboove the rebar and the rebar has been detected
yet. The current thickness value is displayed in the upper right corner of
the screen, as shown in figure .4.

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 19
a. Measuring Interface . Measuring Location Interface
c. Measuring Location Interface(Second Range)
Fig. .4 Rebar Position & Thickness Measurement (a, b,c)
As the instrument moves away from the rebar, the square-shaped frame
moves away from the centerline again and gradually moves towards the
edge of the screen. When the measuring signal completely disappears,
the square-shaped frame and the centerline overlap without displaying
cross-line in the square-shaped frame.
In the start measuring interface, the scope by default is 600mm each

Rebar Locator DR 30000 PK
DRC s.r.l. – info@drcitalia.it Page 20
page, and automatically increase or decrease in multiples of 600 each
time. In the process of scanning, the system will automatically page if the
display is beyond the scope, press the 【▼】& 【▲】to view the previous
screen data, the before and after data in every screen will change
accordingly, except the middle data. Besides, during the measurement,
data thickness interval is automatically increased by 100mm each time.
NOTE: the instrument has not any device to measure distance; the
thickness interval is only indicative.
In the measuring interface, please operate the following buttons to select
the functions:
【 】: Save the measuring data and restart the next measuring
【 】: Clear the measuring data and restart measuring.
【▲】, 【▼】: Page up and down to view the before and after data page.
【Side Upper Button】: Switch the large & small range.
【Side Lower Button 】: The rebar diameter & the corresponding
thickness measuring.
Tips: The measuring modes are divided into the large range and
small range (by default). If the thickness is less than 80mm, please use
the small range. If the thickness is more than 80mm, please use the large
range.
Table of contents
Popular Measuring Instrument manuals by other brands

Endress+Hauser
Endress+Hauser Liquiline Mobile CML18 operating instructions

Mitutoyo
Mitutoyo D/E user manual

Balluff
Balluff BMP 01-EL1PP 1A Series user guide

RHS
RHS SF6 Analyzer 973 Operation manual

Rothenberger
Rothenberger ROLeak Aqua 3Plus Instructions for use

Geophysical Survey Systems, Inc
Geophysical Survey Systems, Inc StructureScan Mini LT quick start guide