manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hanna Instruments
  6. •
  7. Measuring Instrument
  8. •
  9. Hanna Instruments HI 96770 User manual

Hanna Instruments HI 96770 User manual

1. GLP/ key: press to enter GLPmode. In calibrationmodepress to edit
the date and time.
2. CAL CHECK key: press to perform the validation of the meter, or press
and hold for three seconds to enter calibration mode.
3. ZERO/CFM key: press to zero the meter prior to measurement, to
confirm edited values or to confirm factory calibration restore.
4. READ /TIMER key: In measurement mode, press to make a
measurement, or press and hold for three seconds to start a pre-
programmed countdown. In GLP mode press to view the next screen.
5. ON/OFF key: to turn the meter on and off.
6. Liquid Cristal Display (LCD)
7. Cuvet alignment indicator
8. Cuvet holder
3
DISPLAY ELEMENTS DESCRIPTION:
INSTRUCTION MANUAL
HI 96770HI 96770
HI 96770HI 96770
HI 96770
Silica High Range ISM
This instrument is in
compliance with the
Directives
www.hannainst.com
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not
damaged. If any damage occurred during shipment, please notify your
Dealer.
Each HI 96770 Ion Selective Meter is supplied complete with:
• Two Sample Cuvets and Caps
• 9V Battery
• Instruction Manual
Note: save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.
For more details about spare parts and
accessories see “Accessories”
Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you
with the necessary information for the correct use of the instrument. Please
read it carefully before using the meter. If you need additional technical
information, do not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives.
Technical specifications:
Range 0 to 200 mg/L (as SiO2)
Resolution 1 mg/L for measurement
Precision ±5 mg/L @ 100 mg/L
Typical EMC Dev. ±1 mg/L
Light Source Light Emitting Diode
Light Detector Silicon Photocell with narrow band
interference filter @ 466 nm
Method Adaptation of the USEPA METHOD 370.1
for drinking, surface and saline waters,
domestic and industrial wastes and Standard
Method 4500-SiO2C.
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use in measurementmode;
after 1 hour of non-use in calibrationmode;
with last reading reminder.
Dimensions 192 x 102 x 67 mm (7.6 x 4 x 2.6”)
Weight 290 g (10 oz.).
Functional description:
Light High: There is too much light to perform a
measurement. Please check the preparation of the
zero cuvet.
Light Low: There is not enough light to perform a
measurement. Please check the preparation of the
zero cuvet.
No Light:The instrument cannot adjust the light
level. Please check that the samples does not contain
any debris.
Inverted cuvets: The sample and the zero cuvet
are inverted.
Zero: A zero reading was not taken. Follow the
instructions of the measurement procedure for zero-
ing the meter.
Under range: A blinking “0” indicates that the
silica content in the sample is less than 0 mg/L.
Over Range: A flashing value of the maximum
concentration indicates an over range condition. The
concentration of the sample is beyond the
programmed range: dilute the sample and re-run
the test.
Standard Low: The standard reading is out of
range.
Standard High: The standard reading is out of
range.
Errors and warnings:
Measurement procedure:
1• Turn the meter on by pressing
ON/OFF.
2• When the beeper sounds and the LCD
displays dashes, the meter is ready.
The blinking “ZERO” indicates that
the instrument needs to be zeroed first.
3• Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace the
cap.
4• Place the cuvet into the cuvet holder
and ensure that the notch on the cap
is positioned securely into the groove.
5• Press ZERO/CFM and the instrument
status indicator will appear on the
display, depending on the measurement
phase.
6• After a few seconds the display will
show “-0.0-”. The meter is now
zeroed and ready for measurement.
7• Remove the cuvet.
8• Add the content of one packet of
HI 96770A-0 Silica HR reagent A
and shake until complete dissolution.
9• Remove the cap and add the content
of one packet of HI 96770B-0 Silica
HR reagent B. Replace the cap and
shake vigorously until completevigorously until complete
vigorously until completevigorously until complete
vigorously until complete
dissolutiondissolution
dissolutiondissolution
dissolution.
10• Press and hold READREAD
READREAD
READ /TIMER/TIMER
/TIMER/TIMER
/TIMER for
three seconds. The display will show
the countdown. The beeper indicates
the end of countdown period.
Alternatively, wait for 10 minutes.
Remove the cap.
11• Add the content of one packet of
HI 96770C-0 Silica HR reagent C
and shake vigorously until complete
dissolution.
12• Replace the cuvet into the cuvet holder
and ensure that the notch on the cap
is positioned securely into the groove.
Measurement
1) Instrument status indicator appears during different phases of zero
or reading measurement
2) Icons displayed in this area are warning or action icons, signaling
next operation
3) The battery icon shows the charging level of the battery
4) The hourglass appears when the instrument is performing an internal
check up
5) CAL: appears when the instrument operates in the calibration mode
6) The month, day and date icons appear when a date is displayed
7) The chronometer appears when the reaction timer is running
8) Four digit main display
9) Measuring units
10) Four digit secondary display
ON ZERO READING:
ON SAMPLE READING:
DURING CALIBRATION PROCEDURE:
OTHER ERRORS AND WARNINGS:
Dead battery: This indicates that the battery is
dead and must be replaced. Once this indication
is displayed, the meter will lock up. Change the
battery and restart the meter.
Cap error: Appears when external light enters in
the analysis cell. Assure that the cuvet cap is present.
2
4, 12
5-6
10
8, 9
Cooling lamp: The instrument waits for the lamp
to cool down.
Battery low: The battery voltage is getting low
and the battery needs to be replaced.
11
To save the battery, the instrument shuts down after 10 minutes of non-use
in measurement mode and after 1 hour of non-use in calibration
mode.
If a valid measurement was displayed before auto-shut off, the value is
displayed when the instrument is switched on. The blinking “ZERO”
means that a new zero has to be performed.
One fresh battery lasts for around 750 measurements, depending on the
light level.
The remaining battery capacity is evaluated at the instrument startup and
after each measurement.
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements
can’t be taken any more, the instrument shows
“deaddead
deaddead
dead” and turns off.
To restart the instrument, the battery must be
replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
• Turn the instrument off by pressing ON/OFF.
• Turn the instrument upside down and
remove the battery cover by turning it
counterclockwise.
• Extract the battery from its location and
replace it with a fresh one.
• Insert back the battery cover and turn it
clockwise to close.
FACTORY CALIBRATION RESET
It is possible to restore the factory calibration.
1• Press GLP/ to enter GLP mode.
REAGENT SETS
HI 96770-01 Reagents for 100 SILICA HIGH RANGE tests
HI 96770-03 Reagents for 300 SILICA HIGH RANGE tests
OTHER ACCESSORIES
HI 96770-11 CAL CHECK™ Standard Cuvets (1 set)
HI 721310 9V battery (10 pcs)
HI 731318 Tissue for wiping cuvets (4 pcs)
HI 731331 Glass cuvets (4 pcs)
HI 731335 Caps for cuvets (4 pcs)
HI 93703-50 Cuvets cleaning solution (230 mL).
2• Press READ /TIMER to enter in the
factory calibration restore screen. The
instrument asks for confirmation before
deleting the last calibration.
3• Press ZERO/CFM to restore the factory
calibration or press GLP/ again to
abort factory calibration restore.
4• The instrument briefly notifies “done” when
restores factory calibration and returns to
measurement mode.
LAST CALIBRATION DATE
1• Press GLP/ to enter GLP mode. The
calibration month and day will appear on
the main display and the year on the
secondary display.
2• If no calibration was performed, the factory
calibration message, “F.CAL” will appear
on the main display and the instrument
returns to measurement mode after three
seconds.
12• Press GLP/ to edit the desired year
(2000-2099). If the key is kept pressed,
the year number is automatically increased.
13• When the correct year has been set, press
ZERO/CFM or READ /TIMER to
confirm. Now the display will show the
month blinking.
14• Press GLP/ to edit the desired month
(01-12). If the key is kept pressed, the
month number is automatically increased.
15• When the correct month has been set, press
ZERO/CFM or READ /TIMER to
confirm. Now the display will show the
day blinking.
16• Press GLP/ to edit the desired day
(01-31). If the key is kept pressed, the
day number is automatically increased.
Note: It is possible to change the editing
from day to year and to month by
pressing READ /TIMER.
17• Press ZERO/CFM to save the calibration
date.
18• The instrument displays “Stor” for one
second and the calibration is saved.
19• The instrument will return automatically to
measurement mode by displaying dashes
on the LCD.
3
Warning: do not validate or calibrate the instrument with standard
solutions other than the HANNA CAL CHECK™ Standards, otherwise
erroneous results will be obtained.
For accurate validation and calibration results, please perform tests at
room temperature (18 to 25°C; 64.5 to 77.0°F).
Use the HANNA CAL CHECK™ cuvets (see
“Accessories”) to validate or calibrate
instruments.
Validation and Calibration procedures
VALIDATION
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
3• Place the CAL CHECK™ Standard
Cuvet A into the cuvet holder and ensure
that the notch on the cap is positioned securely
into the groove.
4• Press ZERO/CFM and the instrument status
indicator will appear on the display,
depending on the measurement phase.
5• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for validation.
6• Remove the cuvet.
7• Place the CAL CHECK™ Standard
HI 96770-11 Cuvet B into the cuvet
holder and ensure that the notch on the
cap is positioned securely into the groove.
8• Press CAL CHECK key and the instrument
status indicator together with “CAL
CHECK” will appear, depending on the
measurement phase.
9• At the end of the measurement the display
will show the validation standard value.
The reading should be within specifications
as reported on the CAL CHECK™ Standard
Certificate.
Validation
Calibration
If the value is found out of specifications,
please check that the cuvets are free of
fingerprints, oil or dirt and repeat
validation. If results are still found out of
specifications then recalibrate the
instrument.
GLP
Battery management
Warranty
HI 96770 is warranted for two years against defects in workmanship and
materials when used for its intended purpose and maintained according
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Department
and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for your specific
application and for the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic
equipments, this requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instrument’s EMC performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and
the instrument safety do not use or store the instrument in hazardous environments.
HANNA instruments® reserves the right to modify the
design, construction and appearance of its products
without advance notice.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find the Hanna Office in your area,
visit our web site
www.hannainst.com
CALIBRATION
Note: It is possible to interrupt the
calibration procedure at any time
by pressing CAL CHECK or
ON/OFF.
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
3• Press and hold CAL CHECK for three seconds
to enter calibration mode. The display will
show “CAL” during calibration procedure.
The blinking “ZERO” asks for instrument
zeroing.
4• Place the CAL CHECK™ Standard Cuvet A
into the cuvet holder and ensure that the
notch on the cap is positioned securely into
the groove.
6• Press ZERO/CFM and the instrument status
indicator will appear on the display,
depending on the measurement phase.
7• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for calibration. The blinking “READ”
asks for reading calibration standard.
8• Remove the cuvet.
9• Place the CAL CHECK™ Standard
HI 96770-11 Cuvet B into the cuvet
holder and ensure that the notch on the
cap is positioned securely into the groove.
10• Press READ /TIMER and the instrument
status indicator will appear on the display,
depending on the measurement phase.
11• The instrument will show for three seconds
the CAL CHECK™ standard value.
Note: If the display shows “STD HIGH”,
the standard value was too high.
If the display shows “STD LOW”,
the standard value was too low.
Verify that both CAL CHECK™
Standard HI 96770-11 Cuvets, A
and B are free from fingerprints
or dirt and that they are inserted
correctly.
Then the date of last calibration (e.g.:
“01.08.2005”) appears on the display,
or “01.01.2005” if the factory calibration
was selected before. In both cases the year
number is blinking, ready for date input.
IST96770R1 02/07
4-5
7
8
3
4
6-7
9
10-11
12-13
or
14-15
or
16
17
18
Last
Calibration
Date
In GLP mode, the last calibration date can be consulted and the factory
calibration can be restored.
1
2
2
3-4
Accessories:
13•Wait for 2 minutes and then press
READREAD
READREAD
READ /TIMER/TIMER
/TIMER/TIMER
/TIMER.
The instrument status indicator will appear
depending on the measurement phase.
The instrument directly displays
concentration in mg/L of silica (SiO2) on
the LCD.
To convert the reading to mg/L of silicon
(Si) multiply the reading by 0.467.
14• 1
9
Factory
Calibration
Reset
13
INTERFERENCES
• Iron (Fe2+ and Fe3+) interferes at high levels.
• Sulfide (S2-) interferes at high levels.
• Phosphates does not interfere up to 50 mg/L PO4
3-.
Read the relevant SDS sheets before performing the test.
Safety Data Sheets are available at: www.hannainst.com
Safety Data Sheets

Other manuals for HI 96770

1

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI99165 User manual

Hanna Instruments

Hanna Instruments HI99165 User manual

Hanna Instruments HI 96712 User manual

Hanna Instruments

Hanna Instruments HI 96712 User manual

Hanna Instruments HI 931101 User manual

Hanna Instruments

Hanna Instruments HI 931101 User manual

Hanna Instruments HI96786C User manual

Hanna Instruments

Hanna Instruments HI96786C User manual

Hanna Instruments HI 93700 User manual

Hanna Instruments

Hanna Instruments HI 93700 User manual

Hanna Instruments HI97700 User manual

Hanna Instruments

Hanna Instruments HI97700 User manual

Hanna Instruments HI 93707 User manual

Hanna Instruments

Hanna Instruments HI 93707 User manual

Hanna Instruments HI 96804 User manual

Hanna Instruments

Hanna Instruments HI 96804 User manual

Hanna Instruments HI 9143 User manual

Hanna Instruments

Hanna Instruments HI 9143 User manual

Hanna Instruments HI 9812 User manual

Hanna Instruments

Hanna Instruments HI 9812 User manual

Hanna Instruments HI 96721C User manual

Hanna Instruments

Hanna Instruments HI 96721C User manual

Hanna Instruments HI 96104C User manual

Hanna Instruments

Hanna Instruments HI 96104C User manual

Hanna Instruments HI 9146 User manual

Hanna Instruments

Hanna Instruments HI 9146 User manual

Hanna Instruments HI98193 User manual

Hanna Instruments

Hanna Instruments HI98193 User manual

Hanna Instruments HI96802 User manual

Hanna Instruments

Hanna Instruments HI96802 User manual

Hanna Instruments HI 93738 User manual

Hanna Instruments

Hanna Instruments HI 93738 User manual

Hanna Instruments HI 93727 User manual

Hanna Instruments

Hanna Instruments HI 93727 User manual

Hanna Instruments HI96707 User manual

Hanna Instruments

Hanna Instruments HI96707 User manual

Hanna Instruments HI 747 User manual

Hanna Instruments

Hanna Instruments HI 747 User manual

Hanna Instruments HI 83205 User manual

Hanna Instruments

Hanna Instruments HI 83205 User manual

Hanna Instruments HI96733C User manual

Hanna Instruments

Hanna Instruments HI96733C User manual

Hanna Instruments HI96000 Series User manual

Hanna Instruments

Hanna Instruments HI96000 Series User manual

Hanna Instruments HI 96771C User manual

Hanna Instruments

Hanna Instruments HI 96771C User manual

Hanna Instruments HI 96707C User manual

Hanna Instruments

Hanna Instruments HI 96707C User manual

Popular Measuring Instrument manuals by other brands

Biogal CombCam user guide

Biogal

Biogal CombCam user guide

Megger ATEC 210415 Series manual

Megger

Megger ATEC 210415 Series manual

ACS A1270 EMAT Operation manual

ACS

ACS A1270 EMAT Operation manual

Vanguard Instruments ATRT-01 S3 user manual

Vanguard Instruments

Vanguard Instruments ATRT-01 S3 user manual

Altronics Redback A 2067 instructions

Altronics

Altronics Redback A 2067 instructions

GQ Electronics EMF-360V2 quick start guide

GQ Electronics

GQ Electronics EMF-360V2 quick start guide

Schonstedt Instrument TraceMaster instruction manual

Schonstedt Instrument

Schonstedt Instrument TraceMaster instruction manual

Primes Cube M operating manual

Primes

Primes Cube M operating manual

Otodata TM6030 Installation procedures

Otodata

Otodata TM6030 Installation procedures

VOLTCRAFT IR-SCAN-350RH/2 operating instructions

VOLTCRAFT

VOLTCRAFT IR-SCAN-350RH/2 operating instructions

HT Italia HT52 user manual

HT Italia

HT Italia HT52 user manual

Thermo Scientific LevelPRO user manual

Thermo Scientific

Thermo Scientific LevelPRO user manual

Tecnosystemi Galaxy Proair GAZ100001 manual

Tecnosystemi

Tecnosystemi Galaxy Proair GAZ100001 manual

Irwin Geiger Counter instructions

Irwin

Irwin Geiger Counter instructions

Applent Instruments AT682 user manual

Applent Instruments

Applent Instruments AT682 user manual

Marantz professional PMD580 user guide

Marantz professional

Marantz professional PMD580 user guide

Keysight Technologies U4205A user guide

Keysight Technologies

Keysight Technologies U4205A user guide

HP 4154A Operation and service manual

HP

HP 4154A Operation and service manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.