manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Gossen MetraWatt
  6. •
  7. Measuring Instrument
  8. •
  9. Gossen MetraWatt Z227O User manual

Gossen MetraWatt Z227O User manual

Other Gossen MetraWatt Measuring Instrument manuals

Gossen MetraWatt MAVOLUX 5032C User manual

Gossen MetraWatt

Gossen MetraWatt MAVOLUX 5032C User manual

Gossen MetraWatt SECULIFE IM User manual

Gossen MetraWatt

Gossen MetraWatt SECULIFE IM User manual

Gossen MetraWatt PROFISIX User manual

Gossen MetraWatt

Gossen MetraWatt PROFISIX User manual

Gossen MetraWatt METRACLIP 88 User manual

Gossen MetraWatt

Gossen MetraWatt METRACLIP 88 User manual

Gossen MetraWatt DIGISIX 2 User manual

Gossen MetraWatt

Gossen MetraWatt DIGISIX 2 User manual

Gossen MetraWatt STARLITE User manual

Gossen MetraWatt

Gossen MetraWatt STARLITE User manual

Gossen MetraWatt METRACLIP EARTH User manual

Gossen MetraWatt

Gossen MetraWatt METRACLIP EARTH User manual

Gossen MetraWatt MAVOMASTER User manual

Gossen MetraWatt

Gossen MetraWatt MAVOMASTER User manual

Gossen MetraWatt lunasix Instruction Manual

Gossen MetraWatt

Gossen MetraWatt lunasix Instruction Manual

Gossen MetraWatt mavolog pro User manual

Gossen MetraWatt

Gossen MetraWatt mavolog pro User manual

Gossen MetraWatt METRAFLEX 300MXBL User manual

Gossen MetraWatt

Gossen MetraWatt METRAFLEX 300MXBL User manual

Gossen MetraWatt MAVOWATT 210 ROGOWSKI User manual

Gossen MetraWatt

Gossen MetraWatt MAVOWATT 210 ROGOWSKI User manual

Gossen MetraWatt MAVOWATT 4 User manual

Gossen MetraWatt

Gossen MetraWatt MAVOWATT 4 User manual

Gossen MetraWatt METRA HIT 22S User manual

Gossen MetraWatt

Gossen MetraWatt METRA HIT 22S User manual

Gossen MetraWatt DIGISKY User manual

Gossen MetraWatt

Gossen MetraWatt DIGISKY User manual

Gossen MetraWatt ENERGYSENS User manual

Gossen MetraWatt

Gossen MetraWatt ENERGYSENS User manual

Gossen MetraWatt PROFITEST PV 1500 User manual

Gossen MetraWatt

Gossen MetraWatt PROFITEST PV 1500 User manual

Gossen MetraWatt METRACLIP EARTH User manual

Gossen MetraWatt

Gossen MetraWatt METRACLIP EARTH User manual

Gossen MetraWatt MAVOLUX 5032C USB User manual

Gossen MetraWatt

Gossen MetraWatt MAVOLUX 5032C USB User manual

Gossen MetraWatt MAVO-SPOT User manual

Gossen MetraWatt

Gossen MetraWatt MAVO-SPOT User manual

Gossen MetraWatt mavolog pro User manual

Gossen MetraWatt

Gossen MetraWatt mavolog pro User manual

Gossen MetraWatt ENERGYMID EM2281 User manual

Gossen MetraWatt

Gossen MetraWatt ENERGYMID EM2281 User manual

Gossen MetraWatt METRISO INTRO User manual

Gossen MetraWatt

Gossen MetraWatt METRISO INTRO User manual

Gossen MetraWatt LonWorks ENERGYMID U228 W1 Series User manual

Gossen MetraWatt

Gossen MetraWatt LonWorks ENERGYMID U228 W1 Series User manual

Popular Measuring Instrument manuals by other brands

Suaoki D40 user manual

Suaoki

Suaoki D40 user manual

Elster BK-G1.6 operating instructions

Elster

Elster BK-G1.6 operating instructions

Endress+Hauser Deltabar FMD71 technical information

Endress+Hauser

Endress+Hauser Deltabar FMD71 technical information

Duratool D03128 manual

Duratool

Duratool D03128 manual

Vega VEGAPULS 65 Quick setup guide

Vega

Vega VEGAPULS 65 Quick setup guide

Tonghui TH2618B Operation manual

Tonghui

Tonghui TH2618B Operation manual

Tenmars TM-801 user manual

Tenmars

Tenmars TM-801 user manual

MSR 145 user manual

MSR

MSR 145 user manual

TPS WP-82 manual

TPS

TPS WP-82 manual

PCB Piezotronics 003C30 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 003C30 Installation and operating manual

Tenma 72-10465 manual

Tenma

Tenma 72-10465 manual

Endress+Hauser Proline Promag 50 Brief operating instructions

Endress+Hauser

Endress+Hauser Proline Promag 50 Brief operating instructions

Siemens SITRANS F Coriolis FCT030 Function manual

Siemens

Siemens SITRANS F Coriolis FCT030 Function manual

KLINGER CMF V Series instruction manual

KLINGER

KLINGER CMF V Series instruction manual

Keysight M8290A Getting started guide

Keysight

Keysight M8290A Getting started guide

Acterna MS1400 Operation manual

Acterna

Acterna MS1400 Operation manual

R&S ZPH manual

R&S

R&S ZPH manual

Datacolor ColorReader DC10-2 user guide

Datacolor

Datacolor ColorReader DC10-2 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Operating Instructions
Z227O
Probes for 4-Wire Measurement 3-447-097-03
2/3.21
1
Table of Contents
1 Safety Instructions............................. 1
2 Applications ....................................... 2
2.1 Intended Use /
Use for Intended Purpose .............. 2
2.2 Use for Other than Intended
Purpose ......................................... 2
2.3 Liability and Guarantee................... 2
2.4 Opening the Instrument / Repairs... 2
3 The Instrument................................... 3
3.1 Scope of Delivery ........................... 3
3.2 Symbols on the Instrument and the
Included Accessories ..................... 3
3.3 Technical Data ............................... 3
4 Operation ........................................... 3
5 Maintenance ...................................... 3
5.1 Cleaning......................................... 3
5.2 Replacing the Spring-Loaded
Contact Pins .................................. 4
6 Repair and Replacement Parts
Service, Calibration Center and
Rental Instrument Service................. 4
7 CE Declaration ................................... 4
1 Safety Instructions
Observe this documentation, in particular all
included safety information, in order to pro-
tect yourself and others from injury, and to
prevent damage to the instrument.
• Carefully and completely read and
adhere to these operating instructions.
The respective document can be found
at http://www.gossenmetrawatt.com.
Retain the document for future refer-
ence.
• Read and adhere to the operating
instructions included with your measur-
ing instrument.
• Observe and comply with all safety regu-
lations which are applicable for your
work environment.
• Wear suitable and appropriate personal
protective equipment (PPE) whenever
working with the instrument.
• Use the instrument for milliohm mea-
surements only (four-wire measurement)
in measuring category CAT II.
• Only use the instrument and its accesso-
ries in compliance with the specified
technical data and under the specified
conditions (ambient conditions, IP pro-
tection code, measuring category etc.).
• The instrument may only be used as long
as it’s in good working order.
Inspect the instrument before use. Pay
particular attention to damage, broken
insulation or kinked cables.
Damaged components must be
replaced immediately.
2
Z227O Gossen Metrawatt GmbH
• If the instrument or an accessory is dam-
aged during use, e.g. through falling,
remove it from operation and secure it
against inadvertent use.
• Do not use the instrument and its acces-
sories after long periods of storage under
unfavorable conditions (e.g. humidity,
dust or extreme temperature).
• Do not use the instrument and the
accessories after extraordinary stressing
due to transport.
• Plugging in the measurement cables
must not necessitate any undue force.
• Never touch conductive ends (e.g. of
test probes).
• Fully unreel all measurement cables
before starting a test/measurement.
Never perform a test/ measurement with
the measurement cable rolled up.
• Ensure that the alligator clips, test
probes or Kelvin probes make good con-
tact.
• Do not move or remove test probes, alli-
gator clips or Kelvin probes until the test/
measurement has been completed.
Unwanted sparking may otherwise occur
due to test current.
2 Applications
Please read this important information!
2.1 Intended Use /
Use for Intended Purpose
The probes are connected to compatible
measuring instruments with four-wire mea-
suring function and are used for four-wire
measurements. They connect low-imped-
ance resistors to the measuring instrument
by means of four-pole technology and com-
pensate for erroneous influences during
measurement resulting from cable and con-
tact resistances.
Safety of the operator, as well as that of the
instrument, is only assured when it’s used
for its intended purpose.
2.2 Use for Other than Intended
Purpose
Using the instrument for any purposes other
than those described in these operating
instructions is contrary to use for intended
purpose.
2.3 Liability and Guarantee
Gossen Metrawatt GmbH assumes no liabil-
ity for property damage, personal injury or
consequential damage resulting from
improper or incorrect use of the product, in
particular due to failure to observe the prod-
uct documentation. Furthermore, all guaran-
tee claims are rendered null and void in such
cases.
Nor does Gossen Metrawatt GmbH accept
any liability for data loss.
2.4 Opening the Instrument / Repairs
The instrument may only be opened by
authorized, trained personnel in order to
ensure flawless, safe operation and to
assure that the guarantee isn’t rendered null
and void. Even original replacement parts
may only be installed by authorized, trained
personnel.
Unauthorized modification of the instrument
is prohibited.
If it can be ascertained that the instrument
has been opened by unauthorized person-
nel, no guarantee claims can be honored by
the manufacturer with regard to personal
safety, measuring accuracy, compliance
with applicable safety measures or any con-
sequential damages.
3
Gossen Metrawatt GmbH Z227O
3 The Instrument
3.1 Scope of Delivery
Please check for completeness.
3.2 Symbols on the Instrument and
the Included Accessories
3.3 Technical Data
4 Operation
Make sure that the probes are compatible
with the utilized measuring instrument. Com-
pare the technical data and all other specifi-
cations included in the documentation for
the measuring instrument to this end.
Connect the probes as specified in the doc-
umentation for the measuring instrument.
Perform measurement as described in the
documentation for the utilized measuring
instrument. Contact must be established by
applying an appropriate amount of pressure.
Attention!
!
The probes may only be used for
four-wire measurements (milliohm
measurements) in measuring cate-
gory CAT II.
If necessary, use the included S hooks for
cable management. They can be used to
hold 2 measurement cables together.
5 Maintenance
5.1 Cleaning
No special maintenance is required. Keep
outside surfaces clean. Use a cloth for
cleaning, which has been slightly dampened
with water.
Attention!
!
Avoid the use of cleansers, abrasives
or solvents.
1 Set of test probes, 1 red and 1 black
3Shooks
1 Operating Instructions (this document)
Warning concerning a point of
danger (attention, observe
documentation!)
Double insulation
(protection category II)
European conformity marking
The device may not be disposed of
with household trash. Current and
further information is available on
our website at http://www.gossen-
metrawatt.com under the search
terms “WEEE” and “environmental
protection”.
Ambient Conditions
Operating
temperature:
-20 ... +50 °C
Storage
temperature:
-25 ... +75 °C
Relative
humidity:
Max. 75%
No condensation
allowed
Elevation: Max. 2000 m
Electrical Safety
Measuring
category:
CAT II 300 V 10 A
Pollution degree: 2
!
Mechanical Design
Plug: 4 mm,
90° angled
safety plug
Length (overall): 120 cm
Weight: Approx. 180 g
Strain relief: > 100 N
© Gossen Metrawatt GmbH
Edited in Germany • Subject to change, errors excepted • PDF version available on the Internet
All trademarks, registered trademarks, logos, product names and company names are the property of
their respective owners.
Gossen Metrawatt GmbH Z227O
Gossen Metrawatt GmbH
Südwestpark 15
90449 Nürnberg • Germany
Phone: +49-911-8602-111
Fax: +49-911-8602-777
e-mail: info@gossenmetrawatt.com
www.gossenmetrawatt.com
Attention!
!
Never touch or clean the conductive
ends of the test probes!
5.2 Replacing the Spring-Loaded
Contact Pins
The spring-loaded contact pins are subject
to wear and damage. The spring-loaded
contact pins can be replaced.
Replacement part: Z227F spring-loaded
contact pins for Kelvin probes.
Orders can be placed by contacting our
sales department, See below for contact
details. Information concerning replacement
can be found in the supplementary sheet
included with the replacement part.
6 Repair and
Replacement Parts Service,
Calibration Center and
Rental Instrument Service
If required please contact:
This address is only valid in Germany.
Please contact our representatives or sub-
sidiaries for service in other countries.
7 CE Declaration
The instrument fulfills all requirements of
applicable EU directives and national regula-
tions. We confirm this with the CE mark. The
CE declaration is available upon request.