Gossen MetraWatt Z227O User manual

Operating Instructions
Z227O
Probes for 4-Wire Measurement 3-447-097-03
2/3.21
1
Table of Contents
1 Safety Instructions............................. 1
2 Applications ....................................... 2
2.1 Intended Use /
Use for Intended Purpose .............. 2
2.2 Use for Other than Intended
Purpose ......................................... 2
2.3 Liability and Guarantee................... 2
2.4 Opening the Instrument / Repairs... 2
3 The Instrument................................... 3
3.1 Scope of Delivery ........................... 3
3.2 Symbols on the Instrument and the
Included Accessories ..................... 3
3.3 Technical Data ............................... 3
4 Operation ........................................... 3
5 Maintenance ...................................... 3
5.1 Cleaning......................................... 3
5.2 Replacing the Spring-Loaded
Contact Pins .................................. 4
6 Repair and Replacement Parts
Service, Calibration Center and
Rental Instrument Service................. 4
7 CE Declaration ................................... 4
1 Safety Instructions
Observe this documentation, in particular all
included safety information, in order to pro-
tect yourself and others from injury, and to
prevent damage to the instrument.
• Carefully and completely read and
adhere to these operating instructions.
The respective document can be found
at http://www.gossenmetrawatt.com.
Retain the document for future refer-
ence.
• Read and adhere to the operating
instructions included with your measur-
ing instrument.
• Observe and comply with all safety regu-
lations which are applicable for your
work environment.
• Wear suitable and appropriate personal
protective equipment (PPE) whenever
working with the instrument.
• Use the instrument for milliohm mea-
surements only (four-wire measurement)
in measuring category CAT II.
• Only use the instrument and its accesso-
ries in compliance with the specified
technical data and under the specified
conditions (ambient conditions, IP pro-
tection code, measuring category etc.).
• The instrument may only be used as long
as it’s in good working order.
Inspect the instrument before use. Pay
particular attention to damage, broken
insulation or kinked cables.
Damaged components must be
replaced immediately.
2
Z227O Gossen Metrawatt GmbH
• If the instrument or an accessory is dam-
aged during use, e.g. through falling,
remove it from operation and secure it
against inadvertent use.
• Do not use the instrument and its acces-
sories after long periods of storage under
unfavorable conditions (e.g. humidity,
dust or extreme temperature).
• Do not use the instrument and the
accessories after extraordinary stressing
due to transport.
• Plugging in the measurement cables
must not necessitate any undue force.
• Never touch conductive ends (e.g. of
test probes).
• Fully unreel all measurement cables
before starting a test/measurement.
Never perform a test/ measurement with
the measurement cable rolled up.
• Ensure that the alligator clips, test
probes or Kelvin probes make good con-
tact.
• Do not move or remove test probes, alli-
gator clips or Kelvin probes until the test/
measurement has been completed.
Unwanted sparking may otherwise occur
due to test current.
2 Applications
Please read this important information!
2.1 Intended Use /
Use for Intended Purpose
The probes are connected to compatible
measuring instruments with four-wire mea-
suring function and are used for four-wire
measurements. They connect low-imped-
ance resistors to the measuring instrument
by means of four-pole technology and com-
pensate for erroneous influences during
measurement resulting from cable and con-
tact resistances.
Safety of the operator, as well as that of the
instrument, is only assured when it’s used
for its intended purpose.
2.2 Use for Other than Intended
Purpose
Using the instrument for any purposes other
than those described in these operating
instructions is contrary to use for intended
purpose.
2.3 Liability and Guarantee
Gossen Metrawatt GmbH assumes no liabil-
ity for property damage, personal injury or
consequential damage resulting from
improper or incorrect use of the product, in
particular due to failure to observe the prod-
uct documentation. Furthermore, all guaran-
tee claims are rendered null and void in such
cases.
Nor does Gossen Metrawatt GmbH accept
any liability for data loss.
2.4 Opening the Instrument / Repairs
The instrument may only be opened by
authorized, trained personnel in order to
ensure flawless, safe operation and to
assure that the guarantee isn’t rendered null
and void. Even original replacement parts
may only be installed by authorized, trained
personnel.
Unauthorized modification of the instrument
is prohibited.
If it can be ascertained that the instrument
has been opened by unauthorized person-
nel, no guarantee claims can be honored by
the manufacturer with regard to personal
safety, measuring accuracy, compliance
with applicable safety measures or any con-
sequential damages.
3
Gossen Metrawatt GmbH Z227O
3 The Instrument
3.1 Scope of Delivery
Please check for completeness.
3.2 Symbols on the Instrument and
the Included Accessories
3.3 Technical Data
4 Operation
Make sure that the probes are compatible
with the utilized measuring instrument. Com-
pare the technical data and all other specifi-
cations included in the documentation for
the measuring instrument to this end.
Connect the probes as specified in the doc-
umentation for the measuring instrument.
Perform measurement as described in the
documentation for the utilized measuring
instrument. Contact must be established by
applying an appropriate amount of pressure.
Attention!
!
The probes may only be used for
four-wire measurements (milliohm
measurements) in measuring cate-
gory CAT II.
If necessary, use the included S hooks for
cable management. They can be used to
hold 2 measurement cables together.
5 Maintenance
5.1 Cleaning
No special maintenance is required. Keep
outside surfaces clean. Use a cloth for
cleaning, which has been slightly dampened
with water.
Attention!
!
Avoid the use of cleansers, abrasives
or solvents.
1 Set of test probes, 1 red and 1 black
3Shooks
1 Operating Instructions (this document)
Warning concerning a point of
danger (attention, observe
documentation!)
Double insulation
(protection category II)
European conformity marking
The device may not be disposed of
with household trash. Current and
further information is available on
our website at http://www.gossen-
metrawatt.com under the search
terms “WEEE” and “environmental
protection”.
Ambient Conditions
Operating
temperature:
-20 ... +50 °C
Storage
temperature:
-25 ... +75 °C
Relative
humidity:
Max. 75%
No condensation
allowed
Elevation: Max. 2000 m
Electrical Safety
Measuring
category:
CAT II 300 V 10 A
Pollution degree: 2
!
Mechanical Design
Plug: 4 mm,
90° angled
safety plug
Length (overall): 120 cm
Weight: Approx. 180 g
Strain relief: > 100 N
© Gossen Metrawatt GmbH
Edited in Germany • Subject to change, errors excepted • PDF version available on the Internet
All trademarks, registered trademarks, logos, product names and company names are the property of
their respective owners.
Gossen Metrawatt GmbH Z227O
Gossen Metrawatt GmbH
Südwestpark 15
90449 Nürnberg • Germany
Phone: +49-911-8602-111
Fax: +49-911-8602-777
e-mail: info@gossenmetrawatt.com
www.gossenmetrawatt.com
Attention!
!
Never touch or clean the conductive
ends of the test probes!
5.2 Replacing the Spring-Loaded
Contact Pins
The spring-loaded contact pins are subject
to wear and damage. The spring-loaded
contact pins can be replaced.
Replacement part: Z227F spring-loaded
contact pins for Kelvin probes.
Orders can be placed by contacting our
sales department, See below for contact
details. Information concerning replacement
can be found in the supplementary sheet
included with the replacement part.
6 Repair and
Replacement Parts Service,
Calibration Center and
Rental Instrument Service
If required please contact:
This address is only valid in Germany.
Please contact our representatives or sub-
sidiaries for service in other countries.
7 CE Declaration
The instrument fulfills all requirements of
applicable EU directives and national regula-
tions. We confirm this with the CE mark. The
CE declaration is available upon request.
GMC-I Service GmbH
Service Center
Beuthener Str. 41
90471 Nürnberg, Germany
Phone +49-911-817718-0
Fax +49 911-817718-253
e-mail service@gossenmetrawatt.com
Web www.gmci-service.com

Other Gossen MetraWatt Measuring Instrument manuals

Gossen MetraWatt MINITEST PRO User manual

Gossen MetraWatt

Gossen MetraWatt MINITEST PRO User manual

Gossen MetraWatt METRAOHM 413 User manual

Gossen MetraWatt

Gossen MetraWatt METRAOHM 413 User manual

Gossen MetraWatt MAVOWATT 230 Operating instructions

Gossen MetraWatt

Gossen MetraWatt MAVOWATT 230 Operating instructions

Gossen MetraWatt SECUTEST PRO User manual

Gossen MetraWatt

Gossen MetraWatt SECUTEST PRO User manual

Gossen MetraWatt Metraline ZCheck User manual

Gossen MetraWatt

Gossen MetraWatt Metraline ZCheck User manual

Gossen MetraWatt DIGISKY User manual

Gossen MetraWatt

Gossen MetraWatt DIGISKY User manual

Gossen MetraWatt METRACLIP EARTH PE User manual

Gossen MetraWatt

Gossen MetraWatt METRACLIP EARTH PE User manual

Gossen MetraWatt METRACLIP 87 User manual

Gossen MetraWatt

Gossen MetraWatt METRACLIP 87 User manual

Gossen MetraWatt ENERGYMID EM2281 User manual

Gossen MetraWatt

Gossen MetraWatt ENERGYMID EM2281 User manual

Gossen MetraWatt DIGISIX 2 User manual

Gossen MetraWatt

Gossen MetraWatt DIGISIX 2 User manual

Gossen MetraWatt MAVOPAD User manual

Gossen MetraWatt

Gossen MetraWatt MAVOPAD User manual

Gossen MetraWatt LINAX 4000L User manual

Gossen MetraWatt

Gossen MetraWatt LINAX 4000L User manual

Gossen MetraWatt METRAFLEX 3000 User manual

Gossen MetraWatt

Gossen MetraWatt METRAFLEX 3000 User manual

Gossen MetraWatt MAVOWATT 4 User manual

Gossen MetraWatt

Gossen MetraWatt MAVOWATT 4 User manual

Gossen MetraWatt U1281 User manual

Gossen MetraWatt

Gossen MetraWatt U1281 User manual

Gossen MetraWatt MAVOLUX COMPACT User manual

Gossen MetraWatt

Gossen MetraWatt MAVOLUX COMPACT User manual

Gossen MetraWatt MAVO-MONITOR User manual

Gossen MetraWatt

Gossen MetraWatt MAVO-MONITOR User manual

Gossen MetraWatt METRA HIT 22S User manual

Gossen MetraWatt

Gossen MetraWatt METRA HIT 22S User manual

Gossen MetraWatt lunasix Instruction Manual

Gossen MetraWatt

Gossen MetraWatt lunasix Instruction Manual

Gossen MetraWatt PhaseCop2 User manual

Gossen MetraWatt

Gossen MetraWatt PhaseCop2 User manual

Gossen MetraWatt Starlite 2 User manual

Gossen MetraWatt

Gossen MetraWatt Starlite 2 User manual

Gossen MetraWatt METRAmax 2 User manual

Gossen MetraWatt

Gossen MetraWatt METRAmax 2 User manual

Gossen MetraWatt SECULIFE IM User manual

Gossen MetraWatt

Gossen MetraWatt SECULIFE IM User manual

Gossen MetraWatt METRISO G1000 User manual

Gossen MetraWatt

Gossen MetraWatt METRISO G1000 User manual

Popular Measuring Instrument manuals by other brands

Innovative THE PROTECTOR instructions

Innovative

Innovative THE PROTECTOR instructions

TM 9-1829A maintenance

TM

TM 9-1829A maintenance

elsner elektronik KNX SO250 Installation and adjustment

elsner elektronik

elsner elektronik KNX SO250 Installation and adjustment

Azbil mF MCF user manual

Azbil

Azbil mF MCF user manual

Lumel N32H Service manual

Lumel

Lumel N32H Service manual

DeFelsko PosiTector6000 Guide

DeFelsko

DeFelsko PosiTector6000 Guide

Sonel CMP-402 user manual

Sonel

Sonel CMP-402 user manual

Rycom 8890 user manual

Rycom

Rycom 8890 user manual

MSA PrimaX I operating manual

MSA

MSA PrimaX I operating manual

Anritsu MS2690A Operation manual

Anritsu

Anritsu MS2690A Operation manual

EML ARG100 user manual

EML

EML ARG100 user manual

Orbit Merret OM 402UNI manual

Orbit Merret

Orbit Merret OM 402UNI manual

SENTRON TITAN Operation manual

SENTRON

SENTRON TITAN Operation manual

Siemens 3WN1 3ZX1812-0WX31-9BN1 operating instructions

Siemens

Siemens 3WN1 3ZX1812-0WX31-9BN1 operating instructions

Ista ultego III eco manual

Ista

Ista ultego III eco manual

Chauvin Arnoux CA 8345 user manual

Chauvin Arnoux

Chauvin Arnoux CA 8345 user manual

Intek Rheotherm 210 quick start guide

Intek

Intek Rheotherm 210 quick start guide

Binder COMBIMASS compact Assembly and operating manual

Binder

Binder COMBIMASS compact Assembly and operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.