Druck DPI 705 IS User manual

K240 Issue No. 2
Druck Limited, Fir Tree Lane, Groby, Leicester, LE6 0FH, England
Tel: +44 (0) 116 231 7100 Fax: +44 (0) 116 231 7103
Druck DPI 705 IS Intrinsically Safe
Digital Pressure Indicator User Guide a
d
Calibration
Refer to the Calibration and Configuration Instructions, Druck publication K215 or
contact the nearest Druck Subsidiary or Agent.
Specification
Accuracy:
Combined non-linearity, hysteresis and repeatability ..................................... ±0.1%FS
Temperature effects: Span ................................................................... ±0.02%rdg/°C
Zero................. <= 1 bar ±0.05%FS/°C (absolute ranges only)
....................... > 1 bar ±0.02%FS/°C (absolute ranges only)
Maximum safe working pressure ......................................................................... 2 x FS
Pressure connector ..................... 6 mm o/d and 4 mm i/d hose or G1/8 female thread
Environmental ......................................................................................................... IP54
Electrical safety ...................................................................BS EN 61010 as applicable
Electromagnetic compatibility ...................................................EN50081-1 (emissions)
.................................................... EN50082-2 (immunity)
Batteries
Pressure
Cleaning
C E
This pressure indicator meets the essential protection requirements of the
relevant EEC directives.
This intrinsically safe pressure indicator is certified by BASEEFA and CSA
and has sucessfully met the examination and test requirements:
EN50014 : 1992 and EN50020 : 1994
EEx ia IIC T4 (-10°C Tamb +50°C)
This pressure indicator may be used in zones 0, 1 or 2 for industries with any gas group.
Maximum component temperature class T4 (135°C).
Power Supply ................. only use 3 x LR6 (AA) made by Ever Ready, Varta, Duracell or Procell
Do not apply a pressure greater than 1.1 bar absolute to the -ve pressure port of differential
pressure indicators. CAN/CSA-E79-0-95 and CAN/CSA-E79-11-95
Ex ia IIC T4, Class 1, Div.1 Groups A, B, C & D
Ambient Temp. -10°C to +50°C
Safety
This symbol, on the pressure indicator, tells the user to refer to the user
guide or manual.
WARNING: ONLY REMOVE THE BATTERIES IN A SAFE AREA.
Remove batteries immediately when discharged and before storage.
Disposeinaccordancewithlocalregulationsandbatterymanufacturers'instructions.
When storing and transporting batteries make sure they cannot be short circuited.
Do not apply pressure greater than the maximum safe working pressure.
Clean the pressure indicator with a damp cloth.

Operation
K240 Issue No. 2
TIME OUT
If a key is not pressed within 10 minutes then the instrument times out and switches off.
Todisablethisautomatictimeout,holdtheLEAKkeywhenswitchingonthepressureindicator.
LEAK TEST
Toperformaleaktest,presstheLEAKkey3times. The symbolflashesonthedisplay
with the number 60. To start the leak test press the LEAK key again. The instrument
countsdown60secondsdisplayingtheleakageattheendofthe60secondperiod. Press
the LEAK key at any time during the leak test to quit and return to normal measurement.
ZERO
A zero should be performed on gauge and differential instruments before measuring
pressure. To perform a zero: Open all pressure ports to atmospheric pressure. Press
the and TARE keys together, the display briefly shows ZErO and the instrument
calculates a new zero.
Note: A zero can only be performed on absolute pressure indicators if a vacuum is
first applied to the pressure port.
Alarm
A single alarm can be set to operate when the displayed pressure value rises above the
alarmsetting. Thealarmcausesthedisplaytoflashandabeepertosoundforoneminute.
PressingTAREandUNITSkeystogetherdisplaysthe alarm value, pressing the FILTER
keyincreasesthealarmvalue,pressingtheUNITSkeydecreasesthealarmvalue. When
the display shows the required alarm value, press the TARE key to set the alarm.
Druck DPI 705 IS User Guide
Display Symbols
Switch Tare on or off (changes the
display reading to zero)
Switch filter on or off
Change units
(1 of 16
pressure units,
°C or °F)
Switch on or off
Viewmaximum,
minimum or perform leak
test (max/min is reset at
switch on)
Batteries are low, replace, in a safe
area, observing polarity shown on case
Displayed reading - maximum value
Displayed reading - minimum value
TARE
Filter applied
(10 reading rolling average)
Tare applied
Displayed reading - leakage (per minute) Leak test in progress
(count down)
A
(flashing) Alarm - pressure value
more than alarm setting
s
t
Software Version This guide contains operating instructions for pressure indicators
with software version 1.XX. Further changes to the pressure indicator's software may require a change to the
operating instructions and an issue number change of the guide.