BATTE Y INSTALLATION
The unit is shipped with a separate 9 volt alkaline battery which must be installed
before operation. Remove the two screws holding the bottom endcap in place and
remove the endcap. Connect the battery to the enclosed battery clip, making sure
to observe correct polarity. Be careful not to trap wires between the battery, case,
or foam pads which retain the battery. This could make it difficult to install the battery
or remove it later for replacement. Be sure that the rubber gasket is properly seated
in the gasket channel of the endcap and replace endcap. Note that the endcap will
only fit one way because the holes are slightly off-center. Place the Z shaped wrist
strap clip in one of the screw recesses and replace the screws. Do not over tighten
the screws. Attach the wrist strap to the clip. Note: On – KIT option 490A, the
rubber boot will have to be pulled back to have access to the battery compartment.
When battery replacement becomes necessary, use only a 9 volt alkaline type such
as a Duracell®MN 1604, Eveready®522, or equivalent. Zinc-carbon types, often
labeled heavy-duty are not recommended because of the increased potential for
leakage. Alkaline batteries are also a better value because they last up to three
times longer in this device.
Low Battery Indicator
A weak battery can cause improper operation or inaccurate measurements. A low
battery indicator is provided on the display to show when the battery needs
replacement.
ON-OFF OPE ATION
Press the ON/OFF button once to turn the unit on. Once on, press the ON/OFF
button for more than 5 seconds to turn the unit off. If the manometer is left on with
no activity for approximately 20 minutes, the unit will turn itself off to conserve the
battery.
ZE OING P ESSU E EADING
Potential inaccuracy due to temperature effects can be minimized by re-zeroing
immediately before use. To zero the display, vent both ports to the atmosphere so
no pressure is applied to either port. Press the ZERO/STORE button and ---- will
be momentarily displayed as zeroing occurs. Zeroing is not possible when the
memory mode is in use. It must be done before selecting that function.
If the unit is accidentally zeroed with pressure applied to one of the ports, the
pressure reading might display incorrectly. To correct, vent the pressure ports to
atmosphere and press the ZERO/STORE button to zero the unit.
BACKLIGHT
The Model 490A includes a display backlight to allow use in the dark or in poor
lighting conditions. Pressing the ON/OFF button toggles the backlight on or off. With
the 490A on, press the ON/OFF button to toggle the backlight on or off. The
backlight will remain lit for approximately 2 minutes after which it will turn itself off
to conserve battery life.
P ESSU E CONNECTION
Measuring Single Pressure
To measure single positive pressure, connect the tubing to the port marked “ igh
Pressure” and vent the opposite port to the atmosphere.
Measuring Differential Pressure
To measure differential pressure, connect the higher positive pressure to the port
marked “ igh Pressure” and the lower positive pressure to the port marked “Low
Pressure”. The manometer will indicate the difference between the two.
Overpressure Alarm
A visual indicator and audible alarm are provided to alert the user that pressure has
exceeded the operating range of the unit. Exceeding the range will not damage it
or affect calibration as long as the maximum rated pressure is not exceeded. The
maximum pressure for all units is shown on the rear label and on page 1 of these
instructions.
Do not exceed the maximum rated pressure of the manometer.
Doing so will cause permanent damage to the sensor, may
rupture the housing, and/or cause personal injury.
CAUTION
Although the unit might appear to function and indicate properly,
the accuracy of readings cannot be guaranteed when the LOW
BAT indicator is illuminated. Replace the battery with a new one. Do not leave an
exhausted battery in the unit due to potential leakage.
NOTICE
AUTOMATIC ESOLUTION ADJUSTMENT
For all ranges, the series 490A will automatically move the decimal place one point
to the left when reading in the bottom 10% of its calibrated range. This allows the
user to gain a higher resolution reading for lower static or differential pressure
readings.
SELECTING P ESSU E UNITS
Up to seven pressure units are available. The display will indicate the current
selection. To change to different units, use the UNITS/LOC button. Each press of
the button will cause an advance to the next choice. The selected units will remain
in the memory even when the power is shut off. This way, the user preferences will
always be displayed after the initial selection.
DISPLAY HOLD
In situations where a reading needs to be temporarily retained, the Model 490A
includes a display hold feature which freezes the current reading and holds it on
the display until it is cleared. To activate this operation, press the OLD/MEMORY
button when the desired pressure is displayed. A OLD indicator will appear on the
display to indicate that the reading shown is frozen. To return to normal operation,
press the OLD/MEMORY button again. The OLD indicator will disappear and
the current pressure will be shown.
MEMO Y FUNCTION
A memory function is included in the Model 490A that allows users to store up to 40
pressure readings for later review or recording. The readings are stored in non-
volatile memory so they will be retained even if the unit is shut off or the battery is
removed.
Storing Pressure eadings
T
o store a reading, press and hold the OLD/MEMORY button until ST01 displayed.
Next, press the ZERO/STORE button to safe the current reading to the ST01
memory location. A beep will sound indicating that the reading has been saved. As
each reading is saved, the memory location display will advance to the next number.
To resume pressure measurement, press the OLD/MEMORY button again. Note
that in the memory mode, the display zero function is not available. To zero the
display, exit the memory mode and then press the ZERO/STORE button.
Viewing Stored eadings – Selecting a Location
To view the contents of the memory, press and hold the OLD/MEMORY button
until RD01 is displayed. Next, press UNITS/LOC to view other memory locations.
To resume pressure measurement, press the OLD/MEMORY button again.
Clearing Memory
To clear the contents of the memory, press and hold the OLD/MEMORY button
until CLR is displayed. Next, press the ZERO/STORE button to clear all previously
stored readings. During this operation --- will be displayed. Once the memory is
cleared, the current pressure will be displayed.
Exiting Memory Mode
To exit the memory mode, press the OLD/MEMORY button again and the unit will
return to normal operation.