Dynatron AG OxySpy User manual

English
Operating Manual
English
R

Important remarks concerning the manual
2
This operating manual makes use of the fol-
lowing icons to indicate especially important
comments:
Information and tips which are important for opti-
mal use of the OxySpy.
Indicates information about details which are
important to prevent a risky situation.
Indicates a potentially hazardous situation
which, if not avoided, could result in death or
serious injury.
The following symbols are used in the operating
manual:
LCD display samples
Grey or red: part of the display actually being
described
Flashing part of the display
Remarks
Attention!
Danger!
English

List of chapters
3
Introduction 4
Warnings and Safety Considerations 5
OxySpy Parts List 6
The OxySpy 6
Set Up Procedure / Maintenance
Before first use 6
Switching the OxySpy on and off 7
Opening the OxySpy 7
Fitting the
O2
sensor 8
Fitting the battery 10
Calibration 11
Measurement 14
Service Indications / Trouble Shooting
CAL symbol is displayed 16
O2
sensor not connected 16
O2
sensor symbol is displayed 17
Battery symbol is displayed 18
O2
% indication not stable in the air 18
O2
sensor 19
Maintenance and Storage 20
Operational limits 20
Technical Information 20
Limited Warranty 21
Notes 22
FCC / Copyright © / Waste disposal 23
Warranty paper 24
English

Introduction
4
The OxySpy is a portable Oxygen Analyser for
the trained and certified mixed gas diver. The
OxySpy can be held directly against the slightly
opened diving tank and displays within seconds
the oxygen percentage of the gas.
By pressing the push button the display can be
frozen. By pressing the push button once more
the maximum operation depth (MOD) is dis-
played. The calibration of the OxySpy is as easy:
The unit detects itself whether the calibration is
made in air or with pure oxygen. A calibration
can only be terminated if the
O2
percentage does
not change during the calibration.
If the OxySpy is not used, it switches off auto-
matically.
Thanks to its ergonomical design and its single
button operation, it can be operated by one
hand.
The OxySpy displays the sensor or battery sym-
bol, if the
O2
sensor or the battery are used up.
Both are easily changed by the user.
English

Warnings and Safety Considerations
5
• If you are unsure about the displayed
O2
per-
centage: Repeat the measurement or check
the value with another system. For measure-
ments after filling the tank and before diving
always use two different measuring units.
• Check the OxySpy in air before each series of
measurements. If necessary, calibrate the
OxySpy. See page 5.
• Using pure oxygen or oxygen enriched gas
needs special safety measures! Please consi-
der the general safety considerations of using
pure oxygen or oxygen enriched gases!
• If the
O2
sensor symbol is shown, the display-
ed values can be wrong.
• If the diver uses the wrong value of the gas
mixture he/she or the dive computer could cal-
culate insufficient decompression time (EAD
etc.) or oxygen toxicity (MOD, CNS% etc.).
Both situations could result in death or serious
injury.
• The maximum partial pressure of oxygen
(pp
O2
limit) necessary for the calculation of the
maximum operation depth (MOD) should not
exceed 1.6 bar (the max. value for sports
divers). Higher values may result in death or
serious injury.
• Avoid risky situations which are marked in the
this manual with and .
• Do not use the OxySpy in pressure chambers.
• Do not submerge the OxySpy in water.
• The battery and the
O2
sensor contain corrosi-
ve liquid. If the battery or
O2
sensor leaks: Do
not let the liquid come into contact with skin,
eyes or mouth. If it does, flush the affected
area with water for at least 15 minutes. Consult
a doctor. Do not inhale vapor. Immediately
clean the leakage from the instrument with a
soft towel and dispose of the towel.
Important
safety con-
siderations
English

English
1 OxySpy with case
1 OxySpy Operating Manual
1 Battery CR 2430
1
O2
Sensor in a sealed package
1 Philips-type Screwdriver
The supplied battery and the
O2
sensor have to
be fitted. Similar to the finite life of a battery, the
O2
sensor has a certain lifetime. From the first
time that a
O2
sensor contacts oxygen it starts to
be used up. To keep the lifetime of the
O2
sensor,
its airtight package should only be opened before
the sensor is used the first time.
OxySpy Parts List / The OxySpy / Set Up
6
OxySpy
Parts List
The
OxySpy
Before first
use
Display
Push button
Handgrip
Gas Input /Exit openings
with rubber rings
Battery symbol
MOD symbol
(maximum operation depth)
Calibration symbol
O2Sensor symbol
Symbols for ppO2,%O
2, mtr. and ft.
O2value, ppO2, MOD, messages

English
Set Up Procedure / Maintenance
After a short push on the button, the OxySpy will
display the measured
O2
percentage. If need be,
the unit indicates that it requires calibration or
indicates that no
O2
sensor is connected.
The OxySpy switches off automatically, at the
latest 30 minutes after switching on. The unit
cannot be switched off manually.
• Open the OxySpy only in a clean and dry area
away from water, dust and other contaminants.
It prevents soiling and damage to the electro-
nics and the mounting ring.
• Open the OxySpy only if the battery or the
O2
sensor must be replaced.
• Do not touch the electronics.
1. Pull the rear rubber ring off the housing.
2. Remove the 7 screws with the enclosed scr-
ewdriver and separate the two parts of the hou-
sing.
Switching
on and off
7
Opening
the Oxy-
Spy
Battery
O2
Sensor
Plug

Set Up Procedure / Maintenance
8
1. Open the rear side of the housing according
to the chapter "Opening the OxySpy" (page 7).
Consider the safety considerations on page 5
and the ones on the
O2
sensor and its package.
2. Take the
O2
sensor (if there is already one) out
of its position and pull off the mounting ring.
If the
O2
sensor is disconnected while the unit is
on, the display shows the word "no" and the
O2
sensor symbol.
In this situation the OxySpy switches off after 5
minutes automatically.
3. Carefully pull the plug out of the socket. The
tab, which stops the plug from being disconnec-
ted, has to be pulled back while pulling out the
plug.
Do not pull on the wires!
• Do not open the airtight container before you
need to use the
O2
sensor.
• After opening the sealed package the
O2
sen-
sor needs 30 minutes start-up time. This
means that calibration cannot be made until 30
minutes after opening the package. See "Cali-
bration" page 11.
• Take care of the environment; properly dispose
of used
O2
sensor! The sensor should be dis-
posed of according to local laws.
Fitting the
O2Sensor
English
O2
sensor
Mounting ring

English
9
Set Up Procedure / Maintenance
4. Take the new
O2
sensor out of its package.
Take the O ring off the
O2
sensor (if there is any
O ring).
Attach the mounting ring by twisting it onto the
O2
sensor, exactly as shown in the picture on the
left. Maneuver until it fits snugly.
Write the date of the fitting onto the
O2
sensor.
5. Insert the plug into the socket of the
O2
sen-
sor. Make sure that the tab, that secures the
plug, is back in place. The lug of the mounting
ring must fit into the notch of the housing.
6. If a battery is mounted: Switch the unit on and
check that the display is on.
The unit will show the CAL symbol, because the
OxySpy requires calibration. The displayed
O2
percentage may be incorrect.
If "no" and the
O2
sensor symbol is displayed
check the chapter "Trouble Shooting", page 16.
7. Carefully refit together the 2 parts of the case.
Make sure that none of the wires get trapped
and damaged by the case. Take care that the
two parts fit together well and carefully tighten
the 7 screws.
8. Locate the rubber ring into the housing so
that its opening does not close the gas flow
to the
O2
sensor.
9. Calibrate the OxySpy or fit a new battery.
See "Calibration" or "Fitting the Battery" on
page 11 or page 10, respectively.
Opening
Mounting ring

Set Up Procedure / Maintenance
10
1. Open the rear side of the housing according
to the chapter "Opening the OxySpy", (page 7).
Consider the safety considerations on
page 5.
2. Pull the battery (if there is any) out of the bat-
tery holder, by hooking a finger under the bat-
tery.
Take care of the environment: Dispose of the
used battery properly!
3. Wait a minimum of 2 minutes before installing
the battery. Check that it is the correct type of
battery (see "Technical Information" page 20).
Press the battery into the battery holder until it
snaps in place ( + on the top). The battery hol-
der prevents an incorrect fitting.
4.
Switch the unit on and check if the display is on.
If the unit cannot be switched on, repeat step 2
and 3.
Choose the correct depth unit (mtr. or ft.) to dis-
play the maximum operation depth (MOD) by
quickly pressing the push button.
To confirm the depth unit press the push button
for at least two seconds.
Check the chosen metric system before
refitting the parts of the case!
The unit will show the CAL symbol, because the
OxySpy requires calibration. The displayed
O2
percentage may be incorrect.
5. Carefully refit together the 2 parts of the case.
Make sure that none of the wires get trapped
and damaged by the case. Take care that the
two parts fit together well and carefully tighten
the 7 screws.
6. Calibrate the OxySpy or mount a new sensor.
See "Calibration" or "Fitting the
O2
sensor" on
page 11 or page 8, respectively.
Fitting the
battery
English

Calibration
11
The calibration can be made either in air
(20.9%
O2
) or pure oxygen (100%
O2
). The Oxy-
Spy will automatically detect which gas is being
used for calibration.
The
O2
sensor measures the partial pressure of
the oxygen and the OxySpy calculates the
O2
percentage (
O2
%) from it. Changes of the
ambient pressure or temperature will influence
the
O2
measurement values. This is the reason
why the OxySpy has to be calibrated frequently.
Once the
O2
sensor has been calibrated, the
maximum allowed partial pressure of oxygen
(pp
O2
limit)
can be set. This value is needed by
the OxySpy to calculate the maximum operation
depth (MOD).
A new calibration is required when:
• The measured
O2
percentage in the air is
below 20.7%
O2
.
• The measured
O2
percentage in the air
exceeds 21.1%
O2
.
• The CAL symbol is displayed.
• If you are unsure about the displayed
O2
per-
centage.
The
O2
percentage of the gas mixture (used for
the calibration) must not change during the cali-
bration.
The duration of an OxySpy calibration is a maxi-
mum of 30 seconds.
The calibration can be interrupted at any time by
pressing the button.
In hot or humid climates, the OxySpy should be
calibrated either in pure oxygen (100%) or in
compressed air from a tank.
What you
have to
know
about
calibration
English

Calibration
12
1. Before the calibration of the OxySpy in air
(20.9%
O2
) it has to be moved slowly back and
forth, through the air, for about 1 minute to flush
the
O2
sensor. Without this flushing, the calibra-
tion may not be successfully completed. Do not
blow through the OxySpy as the C
O2
in your
breath would cause an incorrect calibration.
2. Press the button until "CAL" and "run" appear.
"CAL" and "run" will flash during the calibration.
Move the OxySpy back and forth, slowly, during
the entire 30 second calibration.
3. After the
O2
sensor has been calibrated, the
OxySpy will display the pp
O2
limit (necessary for
calculating the MOD). By quickly pressing the
push button the setting may be changed in the
range of 1.2 to 1.8 bar, displayed in 0.05 bar
increments.
To confirm the indicated pp
O2
setting press the
push button for at least 2 seconds.
If neither changes nor confirmations are made
within 30 seconds, the OxySpy automatically
switches back to the default setting of 1.4 bar.
If the selected pp
O2
limit is higher than 1.6 bar,
the pp
O2
symbol will flash to warn the diver.
• The maximum partial pressure of oxygen
(pp
O2
limit) for calculating the maximum opera-
tion depth (MOD) should not exceed 1.6 bar
(the max. value for sports divers). Higher
values may result in death or serious injury.
• For strenuous dives or diving in cold water the
chosen pp
O2
limit should not exceed 1.4 bar.
The OxySpy switches automatically into the nor-
mal measurement mode.
Calibrating
in
air
English

Calibration
13
Calibration
was not
successful
English
Calibrating
with pure
oxygen
The OxySpy can also be calibrated with pure
medical oxygen (100%
O2
).
• Please consider the general safety con-
siderations of using pure oxygen!
• The calibration gas has to be 100%
medical oxygen.
• At the gas input make sure that there is
a good seal and that air cannot be
sucked in.
The calibration procedure with pure oxygen is
similar to the one described in the chapters
"Calibrating in air" (page 12) and "Measure-
ment" (page 14).
After the pure oxygen calibration: "Flush"(move
it back and forth slowly) the OxySpy for 1 minu-
te in air and check the indicated value. The
value must be between 20.7 and 21.1%
O2
. If
not, redo the calibration.
If the calibration has not been successfully com-
pleted or the calibration has been interrupted by
pressing the button switch, the display will show
a "- - -", for a short time. The OxySpy will then
change automatically into the measurement
mode. No new calibration values have been
entered and the indicated
O2
% values are still
based on the old calibration data. The CAL sym-
bol is displayed to inform you about the unsuc-
cessful calibration.
Repeat the calibration if the CAL symbol is still
displayed after the calibration. Several calibrati-
ons without success could be an indication that
the
O2
sensor needs to be replaced or a fluctua-
tion of the mixture of the calibration gas during
the calibration. The fluctuation of the gas mixtu-
re might have been caused by air leaking in at
the gas input due to a bad seal.

Measurement
14
Consider the following points if you measure
with the OxySpy:
•
The gas must be able to flow through both elas-
tic rings without resistance.
• Check the OxySpy in air before each measu-
rement by "flushing" it (moving it back and forth
for 1 minute in the air). If the displayed value is
not between 20.7 and 21.1%
O2
, the OxySpy
has to be calibrated.
• The OxySpy must have a steady stream of the
gas until the displayed
O2
% value has stabili-
zed.
• At the gas input (the rubber ring) make sure
that there is a good seal and that air can not be
sucked in.
1. Press one of the two rubber rings against the
tank valve.
2. Open the valve until you hear a gentle, uni-
form hiss and wait until the displayed
O2
% value
has stabilized.
3. "Freeze" the display by briefly pressing the
push button. For confirmation the "%
O2
" starts
flashing.
4. Close the tank valve.
What you
have to
under-
stand
English
Procedure
to measu-
re on
diving
tanks

Measurement
15
5. If the displayed value does not match with
your expectation, repeat the measurement or
calibrate the OxySpy. See "Calibration" page 11.
6. Record the results of your measurement.
7. After quickly pressing the push button, the
OxySpy will calculate the MOD based on the
measured
O2
values and the chosen pp
O2
limit.
The indicated maximum operation depth will be
rounded to 0.1 mtr. or 1 ft.
If the selected pp
O2
limit is higher than 1.6 bar,
the pp
O2
symbol will flash to warn the diver.
8. By pressing the push button once more the
OxySpy returns to measuring mode.
Repeat the measurement until you are sure
that
you measured the actual oxygen percentage
of
the gas mixture.
English

Service Indications / Trouble Shooting
16
If the CAL symbol appears, the OxySpy has to
be calibrated. The displayed
O2
% value could be
wrong.
If the OxySpy does not "see" an
O2
sensor, the
display shows "no" and the
O2
sensor symbol.
The CAL
symbol is
displayed
English
Possible reasons:
New battery has
been installed.
New
O2
sensor has
been installed.
Calibration was not
successful.
How to solve:
Calibrate the OxySpy.
See page 11
Calibrate the OxySpy.
See page 11
Calibrate the OxySpy.
See page 11
Possible reasons:
No
O2
sensor
connected
Water in the
housing
Plug or wire defect
Electronic defect.
How to solve:
Install an O2sensor and
plug in. See
"
Fitting the O2
sensor
"
, page 8.
Remove the battery and
the
O2
sensor. Dry the
case with a towel and
warm air (max. 50 °C /
122 °F).
Take the OxySpy to the
store that it was purchased
from.
Take the OxySpy to the
store that it was purcha-
sed from.
The O2
sensor not
connected

Service Indications / Trouble Shooting
17
If the
O2
symbol appears, the indicated
O2
% value could be incorrect.
The
O2
sensors are available at OxySpy sales
points. There is more information, regarding the
O2
sensor, on page 20.
The
O2sensor
symbol is
displayed
English
Possible reasons:
The ambient air
pressure is excep-
tionally low. (for
example in high
altitude)
The
O2
sensor is
used up
Water in the hou-
sing
Electronic defect.
How to solve:
At high altitudes the
O2
sensor symbol could be
displayed by mistake.
If the
O2
sensor symbol
had not been displayed
before ascending (to the
higher altitude), than it is
almost certain, that the low
ambient pressure of the
altitude is the reason for
displaying the O2sensor.
Replace the O2 sensor. See
"
Fitting the O2sensor
"
,
page 8.
Remove the battery and
the
O2
sensor. Dry the
case with a towel and
warm air (max. 50 °C /
122 °F).
Take the OxySpy to the
store that it was purcha-
sed from.
Check the
OxySpy before
ascending to
altitude.

18
At a low battery voltage, the OxySpy shows the
battery symbol.
If the battery symbol appears, the indica-
ted
O2
% value could be wrong.
The battery is sold in camera and electronic sto-
res. For more information about the battery, refer
to page 20.
The displayed oxygen percentage of the air
goes up and down, even if the actual percenta-
ge does not change.
The
Battery
Symbol is
displayed
English
Possible reasons:
Extremely low
temperature
The battery is
weak.
How to solve:
Warm up the OxySpy
and check again.
Replace the battery. See
page 10.
Possible reasons:
After a measure-
ment of an
O2
enri-
ched gas, the high
concentration of
oxygen in the Oxy-
Spy has first to be
reduced.
O2sensor is defec-
tive or used up.
How to solve:
"Flush" (move it back
and forth) the OxySpy in
air for approximately 30
seconds. Try to measure
again.
Replace the O2sensor.
O2% indi-
cation not
stable in
the air
Service Indications / Trouble Shooting
Possible reasons:
The OxySpy has
been calibrated in
extremely humid air.
How to solve:
Calibrate the OxySpy in
pure oxygen (100%) or
in compressed air from
a tank!
The indi-
cated
O2-value is
inaccurate
The measured O2-value deviates from the actual
O2-content by more than 1 or 2 percent.

O2Sensors
19
The
O2
sensors have, like batteries, a certain
lifetime. From the first time that a sensor contacts
oxygen it starts to be used up. They can be
replaced by the user (see "Fitting the
O2
sensor",
page 8).
Disconnecting the
O2
Sensor from the socket
does not extend it's life. On page 20 you will find
more information about storing the OxySpy.
After opening the airtight package the
O2
sensor
requires 30 minutes to stabilize in air. Calibration
should be made only after this "warm-up" time.
The battery and the O2sensor contain corrosive
liquid. If the battery or O2sensor leaks: Do not
let the liquid come into contact with skin, eyes
or mouth. If it does, flush the affected area with
water for at least 15 minutes. Consult a doctor.
Do not inhale vapor. Immediately clean the
leakage from the instrument with a soft towel
and dispose of the towel.
English

Maintenance and Storage
20
• Protect the unit from impact and intensive sun
and heat.
• Store the unit in a cool, not too dry place. This
helps to preserve the O2sensor life.
• Ideal storage temperature: 5 °C up to +15 °C
(41 °F up to 59 °F)
• Do not leave the OxySpy in an oxygen enriched
atmosphere except when it is necessary.
• Disconnecting the
O2
Sensor from the socket
does not extent its life.
• Replace the battery and
O2
sensor early.
• Do not clean the unit with any solvent (except
water) or detergent.
• Clean the OxySpy with a damp, soft towel.
• Dry the instrument with a soft towel.
• Do not scratch the display window.
• Maximum altitude: 4000 m (13,000 feet)
above sea level
• Temperature range: 0 °C up to +50 °C
(32 °F to 122 °F)
• Relative air humidity: 10-90%,
non-condensing
•O
2Display: • Resolution: 0.1% O2
• Display range: 0.0 up to 199.9%O2
•O
2Sensor: • Type Dynatron P/N DiveceL3.D1
or other O2sensors if licensed by
the manufacturer.
• Lifetime: minimum 1 year*
• Battery: • Type: CR2430
• Lifetime: min. 3 years at
500 on/ off cycles per year
• Linearity: <3% (at constant pressure and
temperature)
Maintenance
and Storage
English
Operational
limits
Technical
Information
*
Lifetime depending on environment conditions and the used product.
Table of contents