Dear Practitioner,
As the recent purchaser of an , I would like to thank
you for your condence in our product.
Habits are changing, and is a practical example
of this fact. Your patients can now enjoy the benets of a
new treatment for meibomian blepharitis (dry-eye syn-
drome) using state-of-the-art pulsed light technology,
with complete safety and in a compact format.
You have my personal assurance that the greatest care
has been taken in the manufacture of each ; as you
will see for yourself, is an exceptional product.
I hope that your device will give you complete
satisfaction, and that you will choose to share your
experience on the website www.esw-vision.com.
With sincerest regards,
Yves Vincent Brottier
Creator of
18
26
32
36
38
40
42
4
8
12
14
44
45
49
50
51

4 5
mask - Protective glasses.
gel
(ashing intended gel – 85 ml).
user manual. extractor
Extraction tool for the mobile head.
NOTE: E-Swin SAS holds exclusive rights for the marketing of
and its accessories:
optic, clean, patches, stop, gel, mask, eye mask etc.
IMPORTANT: YOU MUST KEEP ALL THE ITEMS OF ORIGINAL PACKAGING (BOXES, PACKING BLOCKS, ETC.).
The manufacturer’s guarantee for parts and labour will only be valid if your device is returned in its complete
original packaging. All items of original packaging in which your
was delivered should therefore be
carefully stored.
base unit and applicator.
ACCESSORIES
optic - Consumable
cartridge (included with your
applicator).
activation card .
consents to be completed
by patients.
pairs of eye mask protective
shells (in special packaging).
clean
Optical spray.
patches
Self-adhesive patches
(per sheet).
power cord.
stop
Guide plate.


8 9
PRECAUTIONS FOR USE
❙ FOR YOUR SAFETY
• E-Swin SAS cannot accept any liability for damage resulting from the incorrect use of the device, or from the use of the device
other than as described in the present user manual.
• Attention – Using controls or setting or performance of procedures others than those specied in this user manual may result in
exposure to dangerous optical radiation.
• Each of your patients must be supplied with a pair of eye mask protective shells. The protective shells eye mask must be
worn throught the operation of the device by the patient.
• One pair of glasses mask is supplied with the device: these glasses should be worn throughout the operation of the device by
the professional user of .
• Once is in service, it must not be brought into contact with the hair, the area covered by the eye mask protective
shells, clothing or any other object, in order to avoid any risk of injury, obstruction or deterioration.
• The tip of the applicator must not be in contact with the eye mask shells at the time of administration of the ash.
• The device must never be used in the proximity of an aerosol, in the proximity of a heat source, or in the proximity of other
electronic devices.
• The device must not be stacked on other electronic devices.
• Your is supplied with a mains lead (2 metres in length) and an applicator lead (2 metres in length). No extension cable,
or any cable other than those supplied by E-Swin, should be used – the device may otherwise be rendered non-compliant for
electromagnetic purposes.
• Wireless communication devices must be kept at a distance (approximately 3 metres), as these may impair the correct operation
of the device.
• The device must be installed in an environment which is substantially free of electromagnetic interference (from computers,
electrical treatment appliances, etc.).
• Earth conductor is only used as functional earth.
The professional user of must ensure that only persons who full the criteria for use described in the present
manual will receive treatment using .
Before using your , the following information should be read carefully:
• Service voltage: 100 - 240 V a.c.
• Rated frequency: 50-60 Hz.
• Maximum power consumption: 540 VA.
• Check that this voltage is consistent with the voltage rating of your installation. Any connection error may cause irreparable damage
which is not covered by the guarantee.
• Check that the power socket to which your is connected is not damaged, is in perfect working order, and will not impair the
correct use of your (the power socket must be easily accessible).
• This device is reserved exclusively for the use of ophthalmological specialists.
• Never disconnect by pulling on the power cord in normal use.
• In case of emergency, may be separated from mains by disconnecting the power cord on either side.
• Never use with wet hands.
• must not be used in any location in which the device might be exposed to water spray (on either the applicator or the base
unit).
• Never immerse the base unit, its applicator or the Optic cartridge in water.
• must not be used if it is damaged, in case of the display of a functional anomaly on the touch screen, or if the device has
been dropped (concealed damage may jeopardize your safety and the safety of your patient).
• This device can only be repaired using special tools. In case of any problems after delivery, contact us by filling out the contact
form on
www.esw-vision.com/contact
.
• optic cartridges, and all accessories in general, must be ordered exclusively by visiting
www.esw-vision.com/contact
and
choosing the applicable contact for your country. We cannot accept any liability associated with the use of consumable cartridges
and accessories supplied by another manufacturer.
• During operation, the device must be positioned on a at supporting surface, in a location where there is no risk of the device being
dropped.
can be used to treat the condition of meibomian blepharitis (dry-eye syndrome) in phototypes I - V. VI phototypes must not
be treated using .
Table of contents
Popular Medical Equipment manuals by other brands

Getinge
Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics
Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care
Pressalit Care R1100 Mounting instruction

Denas MS
Denas MS DENAS-T operating manual

bort medical
bort medical ActiveColor quick guide

AccuVein
AccuVein AV400 user manual