Eagle Eye SG-1000 User manual

SG-1000
Specific Gravity Tester
User Manual
The BEST Battery Testing Equipment. The BEST Customer Support. PERIOD.
www.eepowersolutions.com
Copyright © 2012 Eagle Eye Power Solutions, LLC
V1.0


Eagle Eye SG-1000C User Manual
2
Trouble?
If abnormal measurements are experienced,
It may be due to low battery power. Replace
with new alkaline battery.
CAUTION
Care should be taken to avoid damage to the prism. Use soft tissue paper when
wiping the prism surface to avoid scratching. Never make contact with any metal
tools or abrasive materials.
The instrument is designed to measure the electrolyte solution of lead acid
batteries. Do not use the measurement of other electrolyte solutions such as in
alkaline batteries.
The electrolyte contains sulfuric acid and is toxic. Burns may occur if contact is
made on the skin or clothes. Blindness may occur if contact is made to the eyes.
Always wear protective glasses and rubber gloves. If the electrolyte is splashed
onto the skin or clothing, rinse with sufficient amounts of cold water. If it gets
into an eye, flush with cold clean water for about 15 minutes and seek immediate
medical attention.
Remove the battery from the Unit for long periods of non-use and storage.
Temperature
Attention to thefollowing points is needed during the measurement when the
temperature oftheelectrolyte varies greatly with theambient temperature.
The instrument compensates for the temperature ofthesample automatically
according to the temperature detected on the prism in the Sample Chamber. The
instrument always compensates to a temperature of 68°F (20°C). If the temperature of
the sample isdifferent fromthe prism temperature, wait approximately 3 seconds for
every 2 degree F (3 seconds for every 1 degree C) of difference between temperatures
before pressing the Start Switch. (If any difference between temperatures issuspected,
wait some time before pressing the Start Switch.)

Eagle Eye SG-1000C User Manual
3
Table of Contents
CAUTION........................................................................................................ 2
1. Names and Functions of Parts ................................................................ 4
2. Inserting the Nozzle ................................................................................. 6
3. Inserting the Battery ................................................................................ 7
4. Attaching the Wrist Strap........................................................................ 8
5. Zero Setting............................................................................................... 9
6. Measurement of Sample........................................................................ 11
7. Explanation of Error Messages.............................................................. 13
8. Temperature Display .............................................................................. 13
9. Maintenance............................................................................................ 14
10. Trouble? ................................................................................................. 17
11. Specifications ........................................................................................ 20

Eagle Eye SG-1000C User Manual
4
1. Names and Functions of Parts

Eagle Eye SG-1000C User Manual
5
1. Display
The LCD provides a digital readout of the measured values.
2. SELECT key
Used to select a measurement item.
The item to be displayed switches between TEMPERATURE and MEASURE each time this
key is pressed.
(Specific gravity of electrolyte)
3. ZERO key
Used for zero-setting
4. START/OFF key
Press to take measurement. To turn off the instrument, press and hold down the key for
2 seconds.
5. Battery Cover
Remove to replace battery.
6. Nozzle Cap
Used to protect tip of nozzle.
7. Tube Pump
The Tube Pump is made of rubber. Squeeze and release the Tube Pump to remove the
sample.
8. Sample Camber Hook
Used to hold the Sample Chamber in place.
9. Sample Chamber
The removed sample is stored in the Sample Chamber for measurement.
10. Nozzle
The Sample is extracted through the Nozzle.

Eagle Eye SG-1000C User Manual
6
2. Inserting the Nozzle
The Nozzle is packed separately from the Main unit. Insert the Nozzle to Main Unit
following the procedure described below.
(1) Push the Nozzle into the hole located (2) Then, fit the nozzle into the slit by
at the bottom of the Sample Chamber pushing with the thumb.
until it stops.

Eagle Eye SG-1000C User Manual
7
3. Inserting the Battery
(1) Using a screwdriver, turn the locking
pin to the “OPEN” position and open the
Battery Cover.
(2) Insert the battery as indicated, take
note of the polarity.
Note: The battery (006P) provided with
the Unit as an accessory is to be used to
check that the Unit operates properly.
The battery itself is not guaranteed, and
be sure to use a new battery if its voltage
is low.
(3) Close the Battery Cover and using a
Screwdriver, turn the locking pin to the
“CLOSE” position.
Note: The SG-1000 does not have a
power switch. Once the battery is
inserted the electric circuit of the unit is
on. Remove the battery from the unit for
long periods of non-use and storage.
Replace with a new 9V alkaline
battery when power is low.
Use of an alkaline battery can
provide twice as many
measurements compared to an
ordinary 9V battery.
(approximately 2000)

Eagle Eye SG-1000C User Manual
8
4. Attaching the Wrist Strap
Attach the wrist strap to the hook located on the lower side of the Main Unit. When
measuring samples, loop the strap around your wrist to avoid any accidents.

Eagle Eye SG-1000C User Manual
9
5. Zero Setting
(1) Remove the Nozzle Cap
(3) Confirm that the Sample Chamber
has been filled with water. If bubbles are
present in the Sample Chamber, operate
the Tube Pump again to refill water into
the Sample Chamber.
(2) Dip the tip of the Nozzle in water,
and squeeze and release the Tube
pump. The water will be drawn into the
Sample Chamber.

Eagle Eye SG-1000C User Manual
10
(4) Press the ZERO key. “000” will blink
three times and “000” will be displayed.
(5) Zero-Setting is now complete.
Remove the Nozzle from the water and
squeeze the Tube Pump to drain the
water from the Sample Chamber.
Allow the unit to equilibrate to the ambient temperature before taking
measurements.
Zero-setting can be performed anytime. Using water, the unit will be calibrated to
zero.
Use distilled water for zero-setting. Tap water may be used if the impurity level of
the water is less than 0.01%.
Before zero-setting, the water used should also be at the ambient temperature.
Zero-setting should be performed whenever the battery is replaced.

Eagle Eye SG-1000C User Manual
11
6. Measurement of Sample
(1) Press the SELECT key to move the
triangle mark to the desired
measurement item on the display.
(2) Dip the tip of the Nozzle into the
electrolyte sample and squeeze and
release the Tube Pump a few times to
draw the sample into the sample
chamber.
Once the sample is drawn into the unit,
Drain it completely and then refill the
sample into the unit again. This process
will help prevent cross contamination
from the previous sample measured.
(3) Press the START key to take a
measurement. The triangle mark will
blink three times and the measurement
result will be displayed.

Eagle Eye SG-1000C User Manual
12
(4) The displayed value will remain on
the LCD for 5 minutes.
(5) To reset the display, press and hold
down the START switch for 2 seconds.
(6) Before storing the SG1000F, squeeze
and release the Tube Pump several
times to drain the sample from the
Chamber. Then, repeat filling and
draining the Sample Chamber several
times with water. Drain the water well
before storing.
If electrolyte is left in the Sample Chamber for a few days, the prism surface will be
soiled and consequently cause errors during measurement. To clean the prism
surface, use a soft tissue soaked with ethyl or isopropyl alcohol. If the electrolyte is
left in the Sample Chamber for a long period of time, corrosion of the prism occurs
and will cause permanent damage to the unit.
The Sample Chamber should be shaded when taking measurements. Erroneous
measurements may result if the Sample Chamber is exposed to too much external
light.
When sample liquid cannot be removed from the Nozzle by operating the Tube
Pump, apply tissue paper to the tip of Nozzle and manually squeeze it out.

Eagle Eye SG-1000C User Manual
13
7. Explanation of Error Messages
The SG-1000 will indicate an error message when there is a problem with the operation
on the instrument. Error messages that may be displayed are explained below.
“AAA” Zero-setting error
Zero-setting was performed
without water in the Sample
Chamber.
Zero-setting was performed with
a sample other than water in the
Sample Chamber.
Note: This error message may not occur
when the concentration of the sample is
low (close to that of water). Be sure
distilled or tap water is used when
performing zero-set.
“LLL” Sampling error
Electrolyte in the Sample
Chamber is insufficient to cover
the prism surface or air bubbles
may be present.
“HHH” Range-over error
The sample measured is out of
range of the SG-1000.
“HHH” Battery error
This error message is displayed
after zero-setting and
measurement when the battery is
low. Replace the battery with a
new one.
The SG-1000 may operate in an
abnormal manner without
displaying the “HHH” when the
battery is low. Replace the battery
when abnormal operation is
noticed.
“. . .” Temperature Error
The temperature of the
sample in the Sample
Chamber is below 14 degrees
F or above 122 degrees F.
This error will also appear
after zero-setting if the
temperature is out of range.
8. Temperature Display

Eagle Eye SG-1000C User Manual
14
(1) Press the SELECT key to move the
triangle mark to Temperature, and the
temperature will be displayed.
(2) The temperature indicated is the
temperature of the Sample Chamber
and is not a direct measurement of the
temperature of the sample.
9. Maintenance
(1) To replace the Tube Pump, slide it to
the left to release and replace it with a
new one.
(2) After measurement, fill and drain the
Sample Chamber with distilled or tap
water several times to clean. Completely
drain before storing the Unit.
(3) Do not store in a location of high
humidity or in direct sunlight.
(4) The body casing of the unit is made
of plastic. Do not use, or make contact
with any organic solvents on the unit.
If the SG-1000 is dropped, check for any of the below symptoms.

Eagle Eye SG-1000C User Manual
15
Symptom
Remedy
The unit will not turn on. The contact plate
in the battery compartment may have
been bent creating a gap between the
plate and the battery.
Restore contact between the contact plate
and battery by using a screwdriver to
return the plate to the normal position.
The Sample Chamber, Sample Hook
and/or Tube Pump may have been freed
from the normal position.
Make sure the Sample Chamber Hook is
properly attached to the unit. Adjust the
Sample Chamber and Tube Pump into
position and lock in place with the Hook.
Unit is cracked. Any cracks in the casing cannot be
repaired. Return the unit to the
manufacturer.
The SG-1000 is a precision optical instrument and should be handled with care. Do not
drop or subject to any heavy impact or shock.
Refer to the images on the next page.

Eagle Eye SG-1000C User Manual
16

Eagle Eye SG-1000C User Manual
17
10. Trouble?
If any problem in operation is suspected, check the following reference list for repair
before contacting the manufacturer.
Symptom
What to Check
1.”HHH” is displayed at
zero-setting.
Does the battery need replacement?
2. “AAA” is displayed at
zero-setting.
Is there water in the Sample Chamber?
Are there bubbles in the Sample Chamber?
Is the zero-setting being performed?
Does the batteryneed replacement?
Is the O-ring damaged in the Sample Chamber?
Is the Tube Pump damaged?
Isthe Nozzle damaged?
Is the cover of the Sample Chamber damaged?
Is the prism surface strained or blurred?
3. Measurement is
unstable.
Is the ambient temperature and the sample temperature close to
equilibrium?
Does the battery need replacement?
Is the O-ring damaged in the Sample Chamber?
Is the Tube Pump damaged?
Is the Nozzle damaged?
Is the cover of the prism surface stained or blurred?
Is the Nozzle loose at the insertion point of the Sample Chamber?
4. Measured values are
lower than those measured
by other hydrometers.
Check the values of the other hydrometers again.
Does the battery need replacement?
Does the battery need replacement?
Is the O-ring damaged in the Sample Chamber?
Is the Tube Pump damaged?
Is the Nozzle damaged?
Is the cover of the prism surface stained or blurred?
5. Nothing is displayed
when a key is pressed.
Check that the battery is properly inserted.
Does the battery need replacement?
6. Other symptoms not
mentioned above.
Contact your supplier or manufacturer.

Eagle Eye SG-1000C User Manual
18
*Is the power level of the
battery low?
The unit does not measure if the battery
is low. The battery level can be checked
by using a multimeter. If the mattery is
low, replace it with a new battery. (Refer
to page 8 on how to replace the battery
and other remarks.)
The unit uses a 9V DC battery (006P)
*Is the O-ring damaged in the
Sample Chamber?
Remove the Sample Chamber cover
from the unit (see page 4), and check
the O-ring for deformation or damage.
If the O-ring has been damaged replace
it with a new O-ring which can be
purchased from your dealer.
After replacing the O-ring, properly
replace the Sample Chamber cover.

Eagle Eye SG-1000C User Manual
19
*Is the Tube Pump damaged?
*Is the Nozzle damaged?
*Is the cover of the Sample
Chamber damaged?
*Is the Nozzle loose at the
insertion point of the Sample
Chamber?
Squeeze the Tube Pump to draw water
into the Sample Chamber. At the same
time check the Tube Pump, Nozzle, and
cover of the Sample Chamber for
damage. Make sure the Nozzle is
securely attached. (Leakage of liquid or
air can be observed if the Tube Pump,
Nozzle or cover has been damaged, or
the Nozzle is loose.)
If damage is found, replace the
damaged part. (Two spare Nozzles have
been provided as accessories.) Parts can
be purchased from your dealer or
manufacturer. See page 4 and 5.
*Is the prism surface stained or
blurred?
Remove the cover of the Sample
Chamber from the unit and check the
prism surface for stains or blurs.
If stained or blurred, using a soft tissue,
wipe with water or ethyl or isopropyl
alcohol.
Table of contents
Other Eagle Eye Test Equipment manuals
Popular Test Equipment manuals by other brands

Tanita
Tanita Breath Checker instruction manual

SIGLENT TECHNOLOGIES
SIGLENT TECHNOLOGIES SDS1202X-E user manual

Moldex
Moldex BITREX Test instructions

Kewtech
Kewtech Fuse Finder 2 user manual

Elliott
Elliott RECON 2500 Operating and maintenance instructions

Keysight Technologies
Keysight Technologies X Series Installation note