5
MEMORY FUNCTIONS
After measurement, the last set of measured values will be saved automatically, when the
unit is turned o by pressing“POWER” button or by automatic power-o after 3 minutes.
View past readings
1. Press the“memo s”button to enter memory function.
2. If the User Page is not what you want to view, press the “paget”button to select
average function.
3. Press the“memo s”button to display the next set of values. Past reading are displayed
from the most recent to the oldest.
4. This can store up to 60 past readings for each user (2 x 60)
TROUBLESHOOTING
Note: If your problem cannot be solved by the above, consult your store of purchase.
Do not disassemble the unit.
Problem Cause Solution
No display when you press Are the batteries done? Replace with new batteries.
the Power/Start button Are the battery polarities wrong? Re-align batterieswith correct polarities.
“Err P”displayed Fails to inate
“Err 1”displayed Deates too rapidly Return for servicing.
“Err 2”displayed Movement during measurement Do not move during measurement
Signal interference
“Err 3”displayed Incorrect results
Check if the cu is properly connected.
Replace the air tube if it is broken.
Remove interfering source e.g. mobile
phones, magnets. Measure again.
STORAGE & MAINTENANCE
• Keep away the unit from direct sunlight, extreme temperatures, humidity or moisture.
• Use a dry, soft cloth to clean the unit, or if desired, use a cloth lightly dampened with water.
• Do not use alcohol, benzene, thinner or other volatile liquids to clean the unit.
• Do not wash or expose the wrist cu to liquid.
• Remove batteries from the unit when not in use for more than 3 months.
SPECIFICATIONS
Measurement method Oscillometric
Measurement range Pressure: 20 – 280 mmHg
Pulse Rate: 40 – 195 beats/min
Accuracy Pressure: +/- 3 mmHg
Pulse Rate: +/- 5%
Ination Pump driven
Pressure detection Semi-conductor
Power supply 4 x AA batteries
Auto power o 3 minutes after last operation
Storage environment -20°C to 60°C, 10% to 95% RH
Operation environment 10°C to 40°C, 30% to 85% RH
External dimensions 195 (L) x 165(W) x 87 (H) mm
Net weight 602g (batteries excluded)
Cu dimensions Arm circumference of 22~34cm