ELPRO LIBERO CE User manual

PM | WU
1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 1 / 34
LIBERO CE CL CH
Operation Manual

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 2 / 34
Table of Contents
1 Safety Instructions................................................................................................................................... 3
2 Quick Start Guides................................................................................................................................... 4
3 System Overview..................................................................................................................................... 5
3.1 LIBERO CE ........................................................................................................................................ 5
3.2 LIBERO CL......................................................................................................................................... 5
3.3 LIBERO CH........................................................................................................................................ 6
4 Devices - LIBERO CE CL CH....................................................................................................................... 7
4.1 Functionality and Modes................................................................................................................. 7
4.2 Logger service life / battery runtime............................................................................................. 10
4.3 Workflow ....................................................................................................................................... 11
4.3.1 “Alarming ON/OFF” function not configured...................................................................... 11
4.3.2 “Alarming ON/OFF” function configured ............................................................................ 11
4.4 Technical Specifications................................................................................................................. 12
5 Accessories............................................................................................................................................ 13
5.1 External Pt100 Probes for LIBERO CE ............................................................................................ 13
5.1.1 Cryogenic shipments and storage....................................................................................... 13
5.1.2 Dry ice shipments and storage............................................................................................ 15
5.1.3 Freezer / fridge / ambient shipments and storage ............................................................. 16
5.2 Extension of sensor cables ............................................................................................................ 17
5.3 M8 connector incl. mounting service on Pt100 probe.................................................................. 18
5.4 Stainless steel bracket ................................................................................................................... 18
6 Configuration......................................................................................................................................... 19
7 Operation via LIBERO Cx BLE App ......................................................................................................... 24
8 Disposal ................................................................................................................................................. 31
9 Declaration of Conformity..................................................................................................................... 32
9.1 EU Declaration............................................................................................................................... 32
9.2 FCC/ISED Regulatory notices ......................................................................................................... 33

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 3 / 34
1Safety Instructions
Intended Use
LIBERO CE/CH/CL data loggers are exclusively for commercial use (“business to business”) in industrial
environments, representing monitoring solutions for temperature and humidity measurement with
internal and external sensors. LIBERO CE/CH/CL data loggers are not intended for use with children or in
vicinity of children.
If the device is used in a manner not specified by the manufacturer protection provided by the device
may be impaired!
Battery
Material safety data sheets according to provisions of directive 91/155/EEC and shipping information
are available from ELPRO-BUCHS AG. Do not subject the batteries to mechanical stress nor dismantle
them. The leaking battery fluid is highly corrosive and can generate severe heat when it comes into
contact with moister or it can ignite fire.
Environmental Conditions
Temperature
Temperatures exceeding 70°C can damage the battery. For the operation
range see specifications on www.elpro.com.
Water
Device meets requirements of protection class IP54. Only for use in the
specified IP protection class, penetrating water or moister can damage the
device. The degrees of protection apply to any position and orientation of
the device, regardless of the mounting arrangement.
Humidity
Mechanical Force
Operation range 0 … 100 %RH
Violent shocks and impacts can damage the battery (short circuit).
IR and Steam
Infrared radiation (heat) and superheated steam can damage the surface
coating of the casing.
Microwave
There is a risk of battery explosion if the device is exposed to microwave
radiation.
UV Radiation
Pollution
Exposure to UV radiation diminishes the stability of the casing.
Pollution of the device can lead to malfunctions. Maximum permissible
pollution: Grade 2
Cleaning
For cleaning purpose use a slightly wetted cloth. Do not use thinner, fuel, alcohol or agressive cleaning
detergents, they can damage the casing.
Bluetooth
The product operates in 2.40-2.48 GHz band with a maximum radiated output power of +3.7 dBm.
Distance to the body
The device should be installed and operated with a minimum distance of 20 cm between the device and
your body.

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 4 / 34
2Quick Start Guides
LIBERO CE
LIBERO CL
LIBERO CH

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 5 / 34
3System Overview
3.1 LIBERO CE
LIBERO CE is a multi-level and multi-use PDF Logger with USB and Bluetooth®
interface for external Pt100 probe and the ideal PDF Logger to simplify storage
and shipment process for products with known stability data.
It covers a wide measurement range from -200 °C to +400 °C and monitors
temperature of cell and drug deliveries with cryogenic containers, dry ice or
other applications using an external Pt100 probe.
Basically it can be used in all cases where the logger is to be mounted outside,
also for equipment, e.g. refrigerators. For easy mounting, a stainless steel
bracket can be ordered as an option.
Supporting up to eight temperature alarm zones, MKT and duration as alarm
criteria –LIBERO CE allows you to create temperature profiles for individual
products. In addition, you have the option to switch the alarming function off
and on again if alarming criteria is required. This can be useful in cases where a
transportation box has been cleaned, or liquid nitrogen has been refilled in a
cryogenic container.
LIBERO CE is reconfigurable and reusable for up to 3 years (depending on
usage of Bluetooth® interface). The external Pt100 probe and M8 connector
enables reliable, superior temperature data transfer to the data logger. Up to
75’500 temperature values can be stored on the data logger.
Release products with confidence based on the OK or ALARM status on the
display, the PDF report can be easily downloaded via the PC's USB interface or
read out to a smart device via the LIBERO Cx BLE app (available for iOS and
Android), thanks to the logger's Bluetooth® interface. The app can be used to
start or stop the logger without having to remove it from your equipment or
consignment.
3.2 LIBERO CL
LIBERO CL is a multi-level and multi-use PDF Logger with USB and Bluetooth®
interface with internal temperature sensor for shipment monitoring as well as
monitoring of site and storage conditions.
The internal temperature sensor is highly accurate and comes with a 100%
sensor calibration. Supporting up to eight temperature alarm zones, MKT and
duration as alarm criteria –LIBERO CL allows you to create temperature
profiles for individual products. In addition, you have the option to switch the
alarming function off and on again if alarming criteria is required. This can be
useful in cases where for example a transportation box or refrigerator has to
be cleaned.
LIBERO CL can store up to 75’500 temperature values and has a battery
lifetime of up to 3 years (depending on usage of Bluetooth® interface).
It can be ordered with an optional stainless steel bracket for easy mounting on
your equipment.
LIBERO CE
with external
probe
LIBERO CL

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 6 / 34
Release products with confidence based on the OK or ALARM status on the
display, the PDF report can be easily downloaded via the PC's USB interface or
read out to a smart device via the LIBERO Cx BLE app (available for iOS and
Android), thanks to the logger's Bluetooth® interface. The app can be used to
start or stop the logger without having to remove it from your equipment or
consignment.
3.3 LIBERO CH
LIBERO CH is a multi-level and multi-use PDF logger with USB and Bluetooth®
interface with internal temperature and relative humidity sensor for shipment
monitoring as well as monitoring of site and storage conditions.
The internal combined temperature and relative humidity sensor comes with a
100% sensor calibration. Supporting up to eight temperature alarm zones, two
humidity thresholds, MKT and duration as alarm criteria –LIBERO CH allows
you to create temperature/humidity profiles for individual products. In
addition, you have the option to switch the alarming function off and on again
if alarming criteria is required. This can be useful in cases where for example a
transportation box or storage container has to be cleaned.
LIBERO CH can store up to 75’500 measurement values, 37’750 values each for
temperature and humidity, and has a battery lifetime of up to 3 years
(depending on usage of Bluetooth® interface). It can be ordered with an
optional stainless steel bracket for easy mounting on your equipment.
Release products with confidence based on the OK or ALARM status on the
display, the PDF report can be easily downloaded via the PC's USB interface or
read out to a smart device via the LIBERO Cx BLE app (available for iOS and
Android), thanks to the logger's Bluetooth® interface. The app can be used to
start or stop the logger without having to remove it from your equipment or
consignment.
LIBERO CH

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 7 / 34
4Devices - LIBERO CE CL CH
4.1 Functionality and Modes
Unless otherwise noted the following information applies equally to all three LIBERO Cx models.
After configuration of the logger measured values for temperature and relative humidity (LIBERO CH
only) are recorded, stored and evaluated with regard to the defined alarm criteria. The display shows
the current mode.
In configuration mode the device can be configured using the free liberoCONFIG software, this is shown
on the display as follows:
or
After configuration, the logger is ready to start recording, this is indicated in the display accordingly.
In this start mode the configuration of the device can be changed again if required. When plugged into
the USB port of a PC, the device is automatically recognized by the liberoCONFIG software and set to
"ConF"-mode without the need to press any keys.
The logger can be started by pressing and holding (3 seconds) the "Start/ON" key, the display shows the
current measured value and the flashing "Run" symbol.
If a temperature- or time-dependent delayed alarm activation has been configured a corresponding
note appears in the display after the device is started:
- Temperature-based delay: - Time-based delay:
The delay until the configured temperature The display shows the remaining time in
threshold is reached is indicated by "dELAY". The minutes before the device automatically
device then automatically switches to alarm activated switches to alarm activated measuring mode
measuring mode (see below). (see below).
(Countdown to zero takes place in measuring interval steps,
e.g. measuring interval = 10 min. >> recounting 60 - 50 - 40 - …)

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 8 / 34
Depending on the device’s configuration we can distinguish between the following two types of
measurement mode:
- Measurement mode without "Alarming ON/OFF" function:
Display of current temperature measured value (in the lower section of the display), the alarm status (in
the upper right section of the display) and the flashing measurement mode- / "Run"- indicator (top left
on the display).
LIBERO CH alternatively displays the humidity value after briefly pressing the "Stop/OFF" key. Press the
key again briefly to display the temperature value again.
- Measurement mode with "Alarming ON/OFF" function:
Display of current temperature measured value (in the lower section of the display), the alarm status (in
the upper right section of the display), the flashing measurement mode- / "Run"- indicator (top left on
the display) and the status "Alarming ON" (top center section of the display), which indicates that the
measured values are evaluated according to set alarm criteria.
LIBERO CH alternatively displays the humidity value after briefly pressing the "Stop/OFF" key. Press the
key again briefly to display the temperature value again.
Users who wish to take advantage of pausing the alarm (e.g. when refilling or cleaning) can do so by
pressing and holding (3 seconds) the "Stop/Alarming OFF" button in measurement mode.
In this case the note "AL on" appears in the lower half of the display every 2 seconds alternating with
the currently measured value, to inform the user that the alarm is suspended and can or should be
reactivated. This is done very simply by pressing and holding (3 seconds) the "Start/ON" button.
From measuring mode without configured function "Alarming ON/OFF" or from measuring mode with
status "Alarming OFF", the device can be stopped by pressing and holding (3 seconds) the "Stop/OFF"
button.

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 9 / 34
Two different display states are possible in stop mode:
Immediately after stopping the logger "Make PdF" appears on the display. This message ensures that
the user does not forget to read out the device and it disappears after the PDF report has been created
by plugging the device into the USB port of a computer.
After the device has been read out, "Stop" appears on the display. In this situation, the device can either
be restarted (based on the existing configuration) or reconfigured via liberoCONFIG software by
connecting it to a USB port on the computer without pressing any more buttons.

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 10 / 34
4.2 Logger service life / battery runtime
PDF Loggers LIBERO Cx BLE are designed for an extended usage period of up to 3 years. The actual
service life is reduced by operation at low temperatures (especially below 0 °C) and frequent use of
Bluetooth® connection to a smart device.
Shortening the period of use
Should the period of use for the devices be deliberately limited to 12 or 24 months, the user can order
this option from ELPRO as a factory-made configuration.
Note: This factory-configured expiry date cannot be reset by the user by reconfiguring with
liberoCONFIG.
Reaching the end date is indicated in the display by the message "EoL" (End-of-Life), the logger can no
longer be used.
Reduction of battery runtime
Depending on the Bluetooth® mode selected, there is a reduction in the service life due to increased
energy consumption, according to the following table:
Bluetooth® mode
Runtime
BLE permanently OFF
36 months
BLE temporarily ON (by pressing "Start" button)
24 months
BLE permanently ON
14 months
Conditions: Measuring interval: 10 minutes, operating temperature range,
Customer usage behavior: average assumptions.
Battery warning
If the capacity of the battery has reached a minimum level, this is indicated by a warning symbol in the
logger display.
From this moment on, the logger should no longer be used and can therefore no longer be started. If
the logger is in "Run" mode, a remaining period of use of approx. 8-10 days can be assumed.

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 11 / 34
4.3 Workflow
4.3.1 "Alarming ON/OFF" function not configured
The following figure shows the sequence of modes if "Alarming ON/OFF" function has not been
configured (see chapter 6, section Alarm Conditions).
4.3.2 "Alarming ON/OFF" function configured
The following figure shows the sequence of modes if the "Alarming ON/OFF" function has been
configured (see chapter 6, section Alarm Conditions).

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 12 / 34
4.4 Technical Specifications
LIBERO CE https://shop.elpro.com/daten/img/Documents/TechSpecs/LIBERO/TS_LIBERO_CE_BLE_EN_web.pdf
LIBERO CL https://shop.elpro.com/daten/img/Documents/TechSpecs/LIBERO/TS_LIBERO_CL_BLE_EN_web.pdf
LIBERO CH https://shop.elpro.com/daten/img/Documents/TechSpecs/LIBERO/TS_LIBERO_CH_BLE_EN_web.pdf

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 13 / 34
5Accessories
5.1 External Pt100 probes for LIBERO CE
LIBERO CE can be used for different applications, depending on the sensor element. ELPRO offers
standard probes for three main applications:
Cryogenic shipments and storage
Dry ice shipments and storage
Freezer (-25 °C..-15°C, typ.) / fridge (+2 °C..+8 °C) / ambient (+15 °C..+25 °C) shipments and
storage
5.1.1 Cryogenic shipments and storage
For cryogenic applications the LIBERO CE logger is usually mounted directly to the container or the
container lid, using the optionally available stainless steel bracket, with the sensor leading into the tank.
ELPRO offers an easy, turnkey service for mounting the assembly and calibration.

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 14 / 34
ELPRO offers two Pt100 standard probes for cryogenic applications with M8 connector in different
lengths:
PRO_PT100_ST300D3_M8_CRYO (part number 802287)
PRO_PT100_ST350D3_M8_CRYO (part number 802288)

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 15 / 34
5.1.2 Dry ice shipments and storage
Also in dry ice applications, the LIBERO CE is usually attached to the outside of the container using the
optionally available stainless steel bracket and the sensor leads into the tank. ELPRO offers an easy,
turnkey service for mounting the assembly and calibration.
For this application, ELPRO offers two standard probes with a probe length of 10 cm and Teflon cable in
different lengths:
PRO_PT100_ST100D4_PTFE1_M8 (part number 802284)
PRO_PT100_ST100D4_PTFE2.65_M8 (part number 802285)

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 16 / 34
5.1.3 Freezer / fridge / ambient shipments and storage
For temperature monitoring of freezers, refrigerators or rooms, ELPRO offers two waterproof silicon
Pt100 probes with different cable lengths as standard articles:
PRO_PT100_P20D5_PLA1_M8 (part number 802290)
PRO_PT100_P20D5_PLA2.65_M8 (part number 802291)

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 17 / 34
5.2 Extension of sensor cables
An extension cable with two M8 connectors at a length of 1m is also available to attach the PDF data
logger and the probe.
ATTENTION:
Total cable length (including sensor and cable tail on the data logger) must not exceed 3 m!
ECA_PLA_1M_M8 (part number 802282)

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 18 / 34
5.3 M8 connector incl. mounting service on Pt100 probe
ELPRO offers a mounting service, adding an M8 connector to a Pt100 temperature sensor in order to use
any 4-wire Pt100 probe in combination with LIBERO CE.
CTR_M8_SER (part number 802289)
5.4 Stainless steel bracket
ELPRO offers an optional stainless steel bracket for mounting of LIBERO CE/CL/CH loggers if required, i.e.
to containers for cryogenic applications.
BRA_LIBERO CE CL CH (part number 802286)

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 19 / 34
6Configuration
liberoCONFIG is the free software to configure LIBERO PDF data loggers. It allows users to define all
necessary configuration parameters and to save them as profiles. A profile contains all settings for the
monitoring task and is summarized in the PDF report generated by the logger or in a special
configuration report that can be generated in liberoCONFIG.
The configuration of a single LIBERO Cx is done with liberoCONFIG.
With SmartStart Pack & Go a profile can be assigned to a larger number of LIBEROs quickly and safely.
SmartStart Pack & Go .exe files can be used on any PC without installation and without special drivers.
System requirements
- Operating system: Windows 7, 8 or 10
- CPU clock frequency: 1.5 GHz
- Memory/RAM: 512 MB
- Hard disk: 100 MB
- Monitor resolution: 800 x 600 pixel
Details regarding configuration of LIBERO Cx can be found in the corresponding manual
(https://shop.elpro.com/daten/img/Documents/Operation%20Manuals/LIBERO/OM_LIBEROC_EN_web.
pdf).
In the following section only differences or additional configuration options for LIBERO CE / CL / CH are
described.

1000475V01E DOC_Operation Manual LIBERO Cx BLE.docx | Page 20 / 34
Logging
"Loop" has been added as an additional recording mode, since LIBERO CE/CL/CH loggers are for
multiple use.
NOTE: When memory capacity of the logger is reached, newly measured values continuously
overwrite the oldest stored values!
For temperature- or time-delayed alarm activation the option has been added that such activation is
also possible in case the alarm is subsequently switched on again after a phase of "alarming OFF" mode.
Other manuals for LIBERO CE
1
This manual suits for next models
8
Table of contents
Other ELPRO Data Logger manuals

ELPRO
ELPRO HAMSTER-E EHT1 User manual

ELPRO
ELPRO HAMSTER-E ET1-D User manual

ELPRO
ELPRO Libero Cx User manual

ELPRO
ELPRO Libero Ti1 User manual

ELPRO
ELPRO ECOLOG-PRO Series User manual

ELPRO
ELPRO LIBERO CE User manual

ELPRO
ELPRO ECOLOG TN Series User manual

ELPRO
ELPRO LIBERO G Series User manual

ELPRO
ELPRO ECOLOG-NET LP4 User manual

ELPRO
ELPRO ECOLOG-PRO G User manual