elysator PUROTAP profi Parts list manual

kompakt
leistungsfähig
sicher
PUROTAP profi
www.elysator.com
Minerals and salts in technical water circulationslead
to corrosion and mineral deposits.
PUROTAP filters aggressive materials out of the water
and enables trouble-free operation.
Installation
Function
Operation
Service
Demineralised heating water
perfect for every system
PUROTAP profi
fast
professional
safely monitored

Content
4
4
5
6
7
7
8-9
10
10
10
11
11
12
PUROTAP profi function
according to the latest standards
Connection
The instruments
Determining water hardness
Calculating capacity
Replacing ion exchanger resin
TDS meter malfunction
Changing batteries
Flow rate measurement device malfunction
Reduced flow rate
Drain sieve for residual water
Journal

4
Heating and cooling systems use water as a medium for heat transfer. The water circulates from where the
warm water is generated to the user and back. Even if the system always re-uses the same water, lime and
aggressive materials enter the closed water system when first filling the system, which can damage modern
components.
The filling apparatus filters lime and aggressive dissolved substances such as sulfate, nitrate and chloride out
of the fill-up water. The device operates on the basis of a mixed bed ion exchanger and provides demineral-
ised water that has been completely desalinated. Thus, damage caused by lime and corrosion can be effec-
tivelypreventedintheheatingsystem.
The filling apparatus is equipped with precise meters for monitoring pure water production in terms of qual-
ityandquantity.
Thismethoddoesnotreleaseanychemicaladditivesintothewater.
Thedeviceoperateswithoutanexternalpowersupply.
If theion exchangecapacityis exhausted, thenthe ionexchange resincan be easilyreplaced anddisposed of
withthehouseholdwaste.
Protective
measures
Onlytrainedpersonnelshouldusethefillingapparatus.
Operatingspecificationsmustbeadheredtoaccordingtothismanual.
Localguidelinesaretobefollowedfortheconnectionbetweensanitaryandheatingsys-
tems. The device already has a return and pressure-reducing valve. In the DIN EN 1717
(Germany)fieldof application,a pipe disconnectormustalso be installedbeforethe fill-
upstation.
The system is not suitable for unmonitored, permanent connection under pressure. The
valves in the input and output are to be kept closed and only to be opened for the dura-
tionofthesystemfilling.
Even demineralised water contains dissolved gasses, including oxygen and carbon diox-
idethatcan causethebeginningofacorrosionprocess.Thegasesarepurgedbyheating
the water, therefore we recommend carrying out a heating test run for the system as
soonaspossibleafterfill-up.
Leading boiler manufacturers and component suppliers approve and recommend the procedure for purging
make-upwaterviatheionexchanger.
The demineralised make-up water as a result of desalination also fulfils the requirements for heater fill-up
waterqualityaccordingtoVDIguideline2035(AssociationofGermanEngineers),theSWKIguideline
BT102-01(SwissHeatingandAirConditioningEngineers)andtheÖ-NormH5195.
PUROTAP profi function
According to the latest standards

5
Connection
Theinputfortheuntreatedwaterisontheleft,theoutputforthedemineralisedwaterfortheheatingsys-
temisontheright.
The equipment contains a non-return valve. When the connections are turned, the untreated water can-
notflowthroughtheapparatus.
In the DIN EN 1717 (Germany) field of application, a pipe disconnector must also be installed before
thefill-upstation.Followtheguidelinesfromthewatersupplycompany.
PUROTAP profi may only be under pressure for the duration of the filling process. A permanent con-
nectionunderpressureisnotpermitted.
Temp.max.60°C
Pressuremax.6bar
Cleanthefinesieveperiodically
Open the drain valve on the cartridge only when replacing the resin and only if a hose is connected
tothecollectionsack.Undernocircumstances,donot connectthedrainvalvetoawatersupplysys-
tem.
Empty the water content of the cartridge once before each use, since bacteria can build up after a
longidleperiod.Alsocheckwhetherthewateraccidentallycontainsresinpellets.

6
The pressure may not be adjusted higher than the
pressure that the safety valve of the heating or air
conditioning system responds to. In case of doubt,
donotsethigherthan2.5bar.
The instruments
The pressure reduction valve regulates the pressure of the
untreatedwaterwithregardtothesystem.
Operation: Use the screw driver to turn the adjusting screw
inthecentreof thevalve:To theleft: pressureis reduced; to
theright,pressureisincreased.The higherthe pressure,the
faster the water flows through the fill-up station. The ad-
justedpressureiscontrolledviathepressuregauge.
The battery-powered meter shows the content of the dis-
solvedsubstancesinthewater(totaldissolvedsolids).
Operation: Switch on using the red button on the top left,
then select the measuring point, in the input or output. As
soon as the value on the output is > 20 ppm, then the resin
in the cartridge is exhausted. The untreated water hardness
can be determined via the measuring value in the input and
thecapacityoftheresincanbecalculated.
In order to conserve batteries, the meter turns off
automaticallyafter20seconds.
Resetifnecessary.
The current fill-up station capacity and the total quantity of
the hitherto treated quantity can be read on the flow and
quantitymeter.
Operation: The places after the comma show the litres, the
places in front of the comma show the cubic metre (1,000
litres).Example:0001.248is1,248litres.
The top flow meter shows the quantity of completely
demineralised water, the bottom meter shows the total
amountofwater,includingmixedwater.
It is recommended to log the last meter reading
whenchangingtheresin;themetercannotbereset.
Ifdemineralisationonlysufficesup toa certainhard-
ness, then untreated water can be added. The mix-
ing valve is used as well when cleaning the heating
system: Rinse with untreated water and then fill the
heatingcirculationwithdemineralisedwaterbymov-
ingthevalvelever.
Pressure reduction valve
TDS measuring device
Flow and quantity meter
Mixing valve
100 %
demineralised
Total litres H2O
Litre H O
2
demineralised
Capacity 25/l min

0 l
750 l
1’500 l
2’250 l
3’000 l
3’750 l
4’500 l
5’250 l
6’000 l
6’750 l
7’500 l
8’250 l
9’000 l
9’750 l
10’500 l
11’250 l
12’000 l
12’750 l
13’500 l
14’250 l
15’000 l
°fH
°dH
5°
10°
20°
30°
35°
40°
45°
50°
25°
15°
3°
6°
11°
17°
20°
22°
25°
28°
14°
8°
0 l
500 l
1’000 l
1’500 l
2’000 l
2’500 l
3’000 l
3’500 l
4’000 l
4’500 l
5’000 l
5’500 l
6’000 l
6’500 l
7’000 l
7’500 l
8’000 l
8’500 l
9’000 l
9’500 l
10’000 l
°fH °dH
5°
10°
20°
30°
35°
40°
45°
50°
25°
15°
3°
6°
11°
17°
20°
22°
25°
28°
14°
8°
7
3 3
140 m à 1°fH 80 m à 1 °dH
3 3
70 m à 1°fH 40 m à 1 °dH
Determining water hardness
Water is electrically conductive because it contains dissolved solids (salts). It is mainly lime in drinking
water. Therefore the water hardness can be read from the salt content.
The more the
water conducts So hard
the water
Level of hardness
French / German Gram per 1’000 l
So much lime
is in the water Classification
1.7
4.2
6.4
8.4
10.5
12.6
16.0
16.9
19.0
21.1
23.3
25.3
°dH
°dH
°dH
°dH
°dH
°dH
°dH
°dH
°dH
°dH
°dH
°dH
38
75
113
150
188
225
285
300
338
375
415
450
very soft
very soft
soft
soft
moderately hard
moderately hard
slightly hard
slightly hard
hard
hard
very hard
very hard
75
150
225
300
375
450
575
600
675
750
825
900
50
100
150
200
250
300
350
400
450
500
550
600
3.8
7.5
11.3
15.0
18.8
22.5
28.5
30.0
33.8
37.5
41.5
45.0
g
g
g
g
g
g
g
g
g
g
g
g
This value
is measured
dissolved solids
(salt content)
TDS ppm
°fH
°fH
°fH
°fH
°fH
°fH
°fH
°fH
°fH
°fH
°fH
°fH
The conversion of TDS into water
hardness only provides an esti-
mated value and is only possible
in untreated drinking water.
Other water contains other dis-
solved solids in addition to lime.
Since PUROTAP removes all salts,
capacity in such cases is much
lower.
Whythecapacitycalculation?
1. In order to know the amount of resin
needed for the demineralisation of the
systemwater.
2. In order to know the capacity of a
resin fill-up that does not have to be
monitored.
3
A capacity of 140 m at 1°fH means that
forawaterhardnessof20°fH,forexam-
ple, thevalue of140must bedivided by
20, in order to determine a capacity of
3
7m . At 40°fH, the capacity would still
3
be3.5m .
Calculating
capacity
PUROTAP profi 50 capacity
Litres of demineralised (completely desalinated) water
per cartridge
Total hardness
Total hardness
PUROTAP profi 25 capacity
Litres of demineralised (completely desalinated) water
per cartridge
Total hardness
Total hardness
electric conductivity
µS

8
…
Replacing ion exchanger resin
a) Empty the used resin
If the content of dissolved solids in the treated water exceeds 30 TDS (approx. 2 °fH) according to the
meter, then the resin must be changed. We first recommend that you briefly interrupt the water supply,
shake the cartridge and then slowly turn the water supply back on and fill-up. Thus the residual capacity
canbeasexhaustedaswellaspossible.
Close ball valve 1) and 2) in the cartridge inlet and outlet. Leave PUROTAP profi connected to the
untreated water tap 3).
Connect hose to the drain valve 4) and insert it into the supplied water-permeable collection
sack. Tightly connect the sack to the top of the hose with a cable binder or tape.
Open the drain valve 4).
Open ball valve 1) in the inlet.
Let the resin drain out and then close the ball valve in inlet 1) again.
Let the cartridge have a dry run and close the drain valve 4).
If the collection sack is dripping with resin, then dispose of it in the household waste.
1.
2.
3.
4.
5.
6.
7.
Open the drain valve on the cartridge
only when replacing the resin and only
if a hose is connected to the collection
sack. Under no circumstances, do not
connect the drain valve to a water sup-
plysystem.
Empty

9
b) Fill with new resin
Loosen reinforced hose screws 1) from the apparatus (not at the top on the head).
Hold head 2) by hand and unscrew carefully.
Remove head together with the spray hose 3).
Fill original PUROTAP resin 5) with funnel 4). PUROTAP profi 25 = 1 sack with 25 l, PUROTAP
profi 50 = 2 sacks per 25 l.
Replace head with spray hose. Make sure that the spray hose is pressed into the head until it is
tight and sealed; otherwise, resin will leak into the system.
1.
2.
3.
4.
5.
Make sure that no resin ends up on the floor. This represents a significant slipping hazard. Carefully
andthoroughlyremoveleakedresinoffthefloor(e.g.withthevacuumcleaner).
Checkthatthesprayhose ispressedtightlyintotheheadand issealed;otherwise,resinwillleakinto
thesystem.
Theresinpelletistobestoredinamoistplace,whichiswhyitissuppliedinsealedplasticbags.Once
thebags areopened, usetheresinimmediately. Theresincanbestoredinacooldarkplaceforupto
12months.Thereistheriskofcapacitylossandbacteriaformationifincorrectlystored.
new resin

10
ININ OUTOUT
000PPM
ININ OUTOUT
000PPM
ININ OUTOUT
ININ OUTOUT
Dual TDS meter 2Dual TDS meter 2
Dual TDS meter 2Dual TDS meter 2
DI G I T AL
DI G I T AL
OUT IN
+
-
+
-
OUT IN
TDS meter malfunction
The display goes off after
approx. 20 seconds.
The inlet value (IN) shows -
000-
The outlet value (OUT) shows
-000-
The display remains empty,
even after pressing the
"POWER" button.
The display flutters, is diffi-
cult to read or contains illegi-
ble characters.
The display shows -ERR-
The resin seems to be used
faster (increased TDS) than
the water hardness calcu-
lated.
Pull the measuring
device upward out
of the holder.
First, loosen the
metal housing Loosen and remove
the four metal
screws on the
back
Remove
and replace 2 AA bat-
teries.
back side
Thisisnotamalfunction;itautomaticallyswitchesofftoconservethe
batteries.Pressthe"POWER"buttonagain.
Thesensorintheinlethasnomeasuringvaluesorisdefective.Usethe
flush valve to check whether the refill station is full of water. If this is
thecaseandthemalfunctionpersists,thenthemeasuringdeviceand
sensormustbereplaced.
Checkthewatermetertoseewhethertherefillstationisfullofwater.
If this is the case, then this is not a malfunction, instead it is acorrect
measurement of the desalinated water. If it can be assumed that the
cartridge is exhausted (due to the amount of treated water) and the
display remains on -000-, then the measuring device should be
replacedalongwiththesensor.
Changebatteries.
Changebatteries.
Measuring devicemalfunction. Reset by switching on and off quickly
2-3 times until the display is normal again. Otherwise, remove and
reinsertthebatteries.
Thereprobablyisnotamalfunction,insteadtheuntreatedwatercon-
tains dissolved solids (sulfate, nitrate, chloride) in addition to lime,
whichareremovedandthusreducethecapacity.
Otherwisethereisaflowmeasurementmalfunction(seebelow)
Malfunction
Batteriewechsel
Solution
Flow rate measurement device malfunction
The water meters appears to
be too slow or do not run at
all.
A fine sieve is installed in front of the outlet valve in order to prevent
resinfromleakingintothesystemfromtheapparatusduetoanopera-
tionfailure.Ifresindoesleakoutofthecartridge,itiscaughtinthefine
sieveandinterruptsthefunctionofthearmatures.Solution:Rinseand
cleanarmaturesout.

11
Reduced flow rate
The flow drops below
15 l/min
The flow is weak despite the
unrestricted supply line
Thepressuredifferencebetweenthewatersupplylineandthecoun-
ter-pressurefromthesystemistoolow.Openthe pressurereduction
valveagain.
Thefinesieveintheoutlet(rightside)ofthecartridgeisblocked.
Loosenscrew,rinseoutfinesieveandreattach.
Malfunction Solution
Plastic - ring pressed in
Plastic - ring pressed in
Stainless steel sieve
Das Restwasser im
Harzbehälter und in den
Leitungskomponeten kann
gefrieren, dadurch können
die Komponenten bersten!
Damage
Drain sieve for residual water
Frost damage can be prevented whereby the residual water is blown
outoftheresincontainerwiththesupplieddrainsieve.
Inordertochangetheresin,thedrainsievemustberemoved.
Solution
Drain sieve

12
Journal
Date
Resin changed 3
top meter, m TDS in TDS out
This manual suits for next models
2
Table of contents
Other elysator Water Filtration System manuals

elysator
elysator PUROTAP profi 1 Series User manual

elysator
elysator PUROTAP 1000 Parts list manual

elysator
elysator PUROTAP compenso 2 User manual

elysator
elysator 50c User manual

elysator
elysator PUROTAP easy Installation and user guide

elysator
elysator PUROTAP easy II User manual

elysator
elysator PUROTAP expert Owner's manual

elysator
elysator PUROTAP profi 1 Series Parts list manual

elysator
elysator PUROTAP expert Parts list manual

elysator
elysator PUROTAP micro User manual