19
MODEL RDO PROGRAMMING THE ANALYZER
as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the
arrow keys to change the setting.
3. Select Temperature.
a. Press the Select key to toggle between enable
and disable. If the box is checked, temperature
has been selected.
b. Scroll to Units and select it. Choose between °C
and °F.
c. Scroll to Resolution. Press Select repeatedly to
change the resolution. The setting will be stored
as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the
arrow keys to change the setting.
4. Select Saturation.
a. Press the Select key to toggle between enable
and disable. If the box is checked, percent satu-
ration has been selected.
b. Scroll to Units. There is no selection to make.
The units are % saturation.
c. Scroll to Resolution. Press Select repeatedly to
change the resolution. The setting will be stored
as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the
arrow keys to change the setting.
5. Select Pressure.
a. Press the Select key to toggle between enable
and disable. If the box is checked, partial pres-
sure has been selected.
b. Scroll to Units. There is no selection to make.
The units are torr.
c. Scroll to Resolution. Press Select repeatedly to
change the resolution. The setting will be stored
as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the
arrow keys to change the setting.
5.3.3 Replacing a Probe
If you replace the RDO probe with another RDO probe,
use Replace Probe to transfer the existing setting to
the new probe.
1. Disconnect the existing RDO probe and replace it
with the new probe. See Section 3.6.
2. Select Menu, then Settings, then Probes. Select
RDO Probe A or RDO Probe B, if it is present, then
Select Replace Probe. The analyzer will automati-
cally configure the new probe.
3. Press and hold Back to return to the main display.
5.3.4 Deleting a Probe
If you plan to permanently remove a probe, follow the
steps below.
1. Select Menu, then Settings, then Probes.
2. Highlight the probe you wish to delete and press
Select.
3. Scroll to Delete Probe.
4. Press Select, then Enter to confirm.
5. The probe has been deleted from the configuration.
5.3.5 Adding a Probe
1. Select Menu, then Settings, then Probes.
2. Select Probe A or B, then Add Probe.
3. The Add Probe screen appears with the name of the
sensor that is wired in this position. If there is no
Probe B, the display will say no probe connected.
Select Enter to add the probe.
5.4 CONFIGURING THE ANALYZER
The analyzer (called the Con TROLL PRO®) measures
two parameters, barometric pressure and temperature.
There are four settings to make for each parameter.
• Enable or disable the measurement. If the meas-
urement is enabled, the analyzer will make the
measurement and will display the result. If the
measurement is disabled, the analyzer will not
make the measurement.
• Select units in which measurement results will be
displayed
• Set the display resolution
• Set a sentinel. The sentinel is the value that will be
displayed when the analyzer returns an error condi-
tion for the parameter.
1. Select Menu, then Settings, then Probes. Select
Con TROLL PRO.
2. Select Barometer.
a. Press the Select key to toggle between enable
and disable. If the box is checked, barometric
pressure has been selected.
b. Scroll to Units and select it. Choose the desired
units for barometric pressure.
c. Scroll to Resolution. Press Select repeatedly to
change the resolution. The setting will be stored
as soon as you press any key to leave.
d. Scroll to Sentinel and press Select. Use the
arrow keys to change the setting.
Con TROLL PRO is a registered trademark of In-Situ Inc. All rights reserved.