Engler Piezo-Mate User manual

Piezo-Mate
Piezo-Mate
Veterinary Ultrasonic Dental Scaler / Polisher
Veterinary Ultrasonic Dental Scaler / Polisher
Instruction Manual
Instruction Manual
Engler Engineering Corporation
Engler Engineering Corporation
1099 East 47th Street, Hialeah, Florida 33013
1099 East 47th Street, Hialeah, Florida 33013
800-445-8581 – 305-688-8581 – F X 305-685-7671
800-445-8581 – 305-688-8581 – F X 305-685-7671
Web site:
Web site: www.englerusa.com
www.englerusa.com Support site:
Support site: www.engler411.com
www.engler411.com
REV. B 11 / 22 / 2019
Page 1 of 35

Table of Contents
READ BEFORE YOU START..............................................................................................................
COMPANY PROFILE...........................................................................................................................5
INTRODUCTION..................................................................................................................................7
INSTALLATION INSTRUCTIONS.......................................................................................................9
SCALER OPERATING INSTRUCTIONS..........................................................................................12
SCALER MAINTENANCE..................................................................................................................1
POLISHER INSTALLATION..............................................................................................................15
POLISHER OPERATION...................................................................................................................16
POLISHER MAINTENANCE: PROPHY ANGLE ..............................................................................17
STRAIGHT HANDPIECE MAINTENANCE........................................................................................18
CLEANING and STERILIZATION OF HANDPIECE .........................................................................19
MICROMOTOR MAINTENANCE......................................................................................................20
SCALER TROUBLESHOOTING.......................................................................................................21
POLISHER TROUBLESHOOTING....................................................................................................23
PROPHY ANGLE CLEANING AND MAINTENANCE INSTRUCTIONS...........................................25
WATER FILTER CLEANING INSTRUCTIONS.................................................................................26
SADDLE VALVE ASSEMBLY ...........................................................................................................27
Installation Instructions.......................................................................................................................27
To mount the female quick disconnect, there are two options..........................................................29
WATER TANK INSTRUCTIONS.......................................................................................................30
WATER TANK CARE & MAINTENANCE..........................................................................................31
OPTIONAL Piezo-Mate ACCESSORIES...........................................................................................33
TECHNICAL DATA............................................................................................................................35
DIMENSIONS.....................................................................................................................................36
Page 2 of 35

RE D BEFORE YOU ST RT
The handpiece and tip are water cooled and must always have adequate water flow to function properly. The amount of
water delivered to the handpiece must be regulated according to the power level. If the power level is increased, the
amount of water must also be increased. Not having enough water flow through the scaling handpiece may
permanently damage the handpiece, will cause the handpiece to get hot, and void the warranty. For more information,
please turn to the Scaler Operating Instructions page.
When active, ultrasonic tips vibrate at over a million cycles per minute, if it touches soft tissue, gums or skin, it will
cause burns due the friction of the vibration. The tip does not get hot but the ultrasonic vibration will burn if it makes
contact with soft tissue. This is due to the friction between the skin and the vibrating tip. This is normal for all
ultrasonic scalers. Never let the scaling tip touch soft tissue / skin. Engler Engineering Corporation is not
responsible for any damage caused by improper use of this device or its accessories.
When using a water bottle, it must be kept pumped to at least 30 PSI. The pressure relief valve (if equipped) will slightly
move out showing the yellow interior when pressure builds up. As water is used the pressure will decrease and the
bottle must be pumped to keep adequate pressure. The bottle should be maintained at ½ to ¾ full. Always use distilled
water.
It is recommended to remove the tip and clean / disinfect the handpiece after every use to prevent the spread of germs,
bacteria and disease.
Handpieces, tips, prophy angles, rubber cups, straight handpieces, burrs, and water filters, are normal wear and tear
items. In order to achieve optimal performance they should be replaced regularly.
The power switch is on the left side of the unit.
Page 3 of 35

COMP NY PROFILE
Engler Engineering Corporation has been in business since 196 and occupies an 8000 square
foot facility in Hialeah, Florida (USA). Engler manufactures high speed and table top ultrasonic
dental scalers, polishers and combination units. Other devices include electrosurgery equipment
and ultrasonic instruments for the veterinary market as well as a microprocessor controlled
anesthesia delivery system and a respiratory monitor for veterinary use.
Engler Engineering Corp. acquired the exclusive manufacturing and marketing rights of Dynax
products, including stretchers, gurneys, The Cat Grabber, comfort cots, warm water heater /
circulator with pads, and other products. We also acquired the Alpha-Sonic, Ora-Sonic, and Pro-
Sonic line of piezo scalers.
Engler manufactures all of the inserts and tips used in the Engler products as well as many others
on the market today in the 18K and 25K frequency range.
Our repair department has the technical knowledge to repair and maintain a number of dental
devices manufactured by other companies including Shorline.
Engler Engineering Corporation’s foreign sales are handled through a large and growing network
of veterinary distributors. At the present time we are represented throughout the Middle East,
Europe, Central and South America, Canada, Asia, New Zealand, Australia, and most other
countries.
If you have any questions or comments, please contact:
Engler Engineering Corporation
1099 East 47th Street, Hialeah, Florida 33013
800-445-8581 – 305-688-8581 – F X 305-685-7671
Web site: www.englerusa.com Help site: www.engler411.com
Page 4 of 35

Engler Engineering Corporation’s brand name veterinary products proudly include:
•DS 2000, microprocessor controlled anesthesia delivery system / ventilator,
•Excelsior, high speed dental air unit with vacuum / electrosurgery / ultrasonic scaler / high
speed drill / low speed polisher / air - water syringe, and on demand compressor,
•Scale - ire, high speed dental air unit with ultrasonic scaler / high speed drill / low speed
polisher / air - water syringe and on demand compressor,
•Scale - ire Mini, high speed dental air unit with ultrasonic scaler / high speed drill / low
speed polisher / air - water syringe,
•Drill – ire Plus, high speed dental air unit, high speed drill / low speed polisher / air - water
syringe,
•Drill – ire, high speed dental air unit, high speed drill / air - water syringe,
•Son - Mate II, ultrasonic scaler / 35,000 RPM drill / low speed polisher,
•Sonus II, ultrasonic dental scaler,
•Tri - Mate, ultrasonic scaler / 35,000 RPM drill / low speed polisher / electrosurge,
•Electro – Son, touch screen, mono / bi-polar electrosurgical unit,
•Poli - X, 35,000 RPM drill / low speed polisher,
•Sentinel V.R.M., respiratory monitor,
•Engler Veterinary Respiratory Monitor (EVRM)
•More coming soon!
If you have any questions or comments, please contact us!
Engler Engineering Corporation
1099 East 7th Street, Hialeah, Florida 33013
800- 5-8581 – 305-688-8581 – FAX 305-685-7671
Web site: www.englerusa.com Help site: www.engler 11.com
Page 5 of 35

INTRODUCTION
Thank you for purchasing the Engler Piezo-Mate Scaler Polisher combination.
The design of the Piezo-Mate scaler circuitry uses integrated computer technology along with our Time
Remote Feedback Circuitry. This combination produces a powerful and potent tool against periodontal
disease. A reinforced solid aluminum chassis surrounds the internal circuit board, providing a very durable
and reliable unit.
The dental scaler utilizes an ultrasonic principle of operation. The internal circuitry converts ordinary line
voltage to an operating frequency of approximately 30,000 Hz. This frequency is then amplified and
delivered to the scaling tip. As a result, the tip vibrates at this ultrasonic frequency with an amplitude of
0.001 to 0.00 in. (25. um to 102 um).
In designing our unique Engler tips, water flows internally through the tip as it vibrates. As the bubbles in the
lavage are bactericidal, the energy released collapses and destroys the bacterial cell walls. The water
flowing internally through the tip, effectively cools the area and assists in removing any debris from the
operative site.
This device is equipped with a digital readout that provides an indication of the power setting.
PLEASE READ VERY CAREFULLY
Engler Engineering Corporation (EEC) makes every effort to verify that all parts for this
device including any optional accessories ordered with it are included in this shipment. It is
imperative that you inspect the package and if you find any parts damaged or missing, you
must notify us immediately. Claims for damaged or missing parts will only be accepted
within five days of receipt.
EEC makes every effort to verify that our devices are built and tested to approved
standards. ny modification to the device, hoses or power supply initiated by others
nullifies all warranty statements. Engler Engineering Corporation will not be held liable in
any way, for any damage, injury or death due to non-authorized service, improper
installation, or improper use of this device. EEC's liability will not exceed the purchase price
of this machine.
The information contained herein is intended only as a guide. Individuals not properly trained
should not use this equipment. It is intended for professional use only
If you have any questions or comments, please contact:
Engler Engineering Corporation
1099 East 47th Street
Hialeah, Florida 33013
800-445-8581 – 305-688-8581 – F X 305-685-7671
Web site: www.englerusa.com Help site: www.engler411.com
Page 6 of 35

Page 7 of 35
Micromotor
Do not lubricate
Straight handpiece
Lubricate daily
Prophy angle
Clean and lubricate
daily
Micromotor stand
Ultrasonic
handpiece holder
Power control
Selector switch
Scaler handpiece
LED ring light
Micromotor quick disconnect
Scaling tip
Scaler permanent connection
(not a quick disconnect, do not
twist)
Display
Selector indicators
Water line quick
disconnect
Power cord recepticle
Resettable fuse
Foot switch quick disconnect
Multi turn water control,
counterclockwise to open,
clockwise to close.

INST LL TION INSTRUCTIONS
IMPORT NT: This device must be connected to a clean, filtered, water supply, capable of delivering 30 to
60 psi (2.0 to .2 kg/cm2) of water input pressure. This unit comes with an In-Line water filter (P/N: A52030).
When kept clean and free of foreign matter, it will assist in proper water flow to the unit. If the water pressure
in your office is above 60 psi, we recommend you install a water pressure regulator on the supply line to
your scaler.
CONNECTING WATER SUPPLY:
We strongly recommend that a manual shut off valve
be placed prior to the Female Quick Disconnect, so
that the water can be completely shut-off, and line
pressure relieved, when the unit is not in use.
This device comes equipped with an 8 foot (2 cm)
water line, a male quick disconnect fitting, a water
filter and a coupling body.
The water line must be connected to the white fitting
located on the back of the device before a
connection is made to the water source. To connect
the water line, slide the white (male) insert into the
coupling body until the metal latch locks it in place
making a clicking sound. The water line can now be
connected to your water source.
We suggest that you use one of the four methods as
shown on the next two pages.
301 Male quick disconnect
** Please note, male quick disconnect is supplied as
a standard item with all Engler scalers.
PLE SE REMEMBER
It is recommended that you disconnect the device from the water supply when
it is not in operation.
Engler Engineering Corporation will assume no liability for damages due to not
following recommendations in the Engler manuals.
Connecting water supply continued on next page
Page 8 of 35

1. Female Quick Disconnect (P/N: 300) - This is the
female mating connector to the male quick disconnect
supplied with the Piezo-Mate . Use this to create a
custom water installation utilizing ¼ “ I.D. water tubing.
300 Female quick disconnect
2. Saddle Valve Assembly (P/N: A 303) - This kit
contains all parts to quickly and easily connect your unit
to an existing existing 3/8" to 1-3/8" copper tubing cold
water supply line.
A 303 Saddle valve assembly with
female quick disconnect & tubing
3. Faucet Adapter Assembly (P/N: A22303) - This screws
onto an existing faucet and includes a female quick
disconnect.
A22303 Faucet attachment
with female quick disconnect & tubing
. Portable Water Tank (P/N: PT-1) - This is a self-
contained water source, which is ideally suited for
portable operation. We suggest using distilled water and
fill the tank to the water fill line (approximately 2/3).
Tighten the cap, insert the male quick disconnect on the
water line into the female quick disconnect on top of the
tank, pressurize the tank by pumping the handle until
the pressure relief valve’s (if equipped) yellow indicator
begins to show. Lock the handle in place.
IMPORT NT: Engler Engineering recommends the services of a professional plumber. Engler will
not be held liable for any damage including, but not limited to leakage caused by improper
installation of our products.
Page 9 of 35

CONNECTING WATER SUPPLY (cont.)
PLEASE NOTE: Minerals and foreign particles in the water may cause a buildup or blockage of
internal hoses and parts.
The water filter supplied with this device should be opened and inspected on a regular basis.
Engler suggests that the filter disc and O-ring be changed out at least twice per year. See WATER
FILTER CLEANING INSTRUCTIONS.
CONNECTING THE FOOT SWITCH:
To connect the footswitch to the unit, insert the male quick disconnect into the female quick
disconnect at the rear of the device and then tighten the securing nut by turning it clockwise.
CONNECTING POWER SUPPLY:
First plug the power cord into the unit and then plug the male end of the power cord into a
grounded power outlet. DO NOT remove or bypass the ground pin from the power cord of this
device. Doing so will void the warranty.
IMPORTANT: Your Piezo-Mate has been equipped with a universal switching power supply and will
automatically adjust the input voltage. It will not require any adjustments in this regard. See
technical data for specifications.
Do not alter the scaling tip. The tip is shaped to deliver the optimal vibrating power and frequency.
Deforming the tip in any way will cause the handpiece to get hot, degrade vibration power and void
the warranty.
Page 10 of 35

SC LER OPER TING INSTRUCTIONS
Initial procedures at the start of every day:
1. The power switch is located on the left side of the unit. Switch it ON.
2. Make sure the water is open (counter-clockwise) and flowing to the device. Rotate the Selector
Switch to the “SCALER” position, the red LED indicator should light up, showing that you have power
to the unit.
3. Adjust the power control knob to the minimum power setting fully counter-clockwise. The digital
readout will read 1.
. Install handpiece and tip. Use the tip wrench to gently tighten the tip.
5. With handpiece and tip installed, set the water control to its maximum setting by rotating it
counterclockwise, (knob will rotate up to 3 and a half turns for maximum water) hold the handpiece
over a sink and press the footswitch until water comes out in a stream. This should take no more
than 30 seconds. This step is done to insure proper operation of the delay cavitation feature by
removing air that may be trapped in the water lines.
6. Always keep the power control at the lowest setting and the water control at a level where you have a
fine mist at the tip. Higher power settings will result in hotter water.
NOTE: Tips sent from our facility are not sterilized.
IMPORT NT: Keep in mind that higher power levels tend to heat the out-flowing water.
Temperature control can be achieved by balancing the power with water flow volume. Thus, high
power settings require high water flow rates and conversely low power requires low water flow rates.
7. The scaler is now ready for use.
8. Do not alter the scaling tip. The tip is shaped to deliver the optimal vibrating power and frequency.
Deforming the tip in any way will cause the handpiece to get hot, degrade vibration power and void
the warranty.
9. An anti fatigue mat is recommended to prevent the handpiece from contacting the floor should the
handpiece get dropped. Piezo handpieces work with a crystal which vibrates when energized,
transmitting energy to the tip. Dropping the handpiece could break the crystal. A dropped handpiece
is not covered by the warranty.
Page 11 of 35

IMPORT NT: Operating this device with hot water may cause damage to gums, lips and tongue. If the
handpiece begins to get warm, stop and check water temperature. If it is hot, make sure that the power is at
the lowest setting and the water is set at a high enough setting to provide a lukewarm mist.
Engler Engineering Corporation will not be liable for damage due to improper use of this device. Do not use
this device if the water temperature is too hot.
Call Engler Engineering Corporation technical support if further help is needed.
ULTR SONIC SC LING PROCEDURES
1. Before placing tip into patient’s mouth, activate the scaler over a sink by depressing the
footswitch. A fine mist, with the temperature between cool to lukewarm to the touch is
recommended. This should be accomplished with minimal power and proper water flow.
2. It is recommended that when a tip is placed into a patient’s mouth, the lips, cheek and tongue
be retracted to prevent unwanted contact.
3. Place the tip into the patient’s mouth and press the footswitch to activate the scaler.
. Bring the tip lightly up to the tooth and gently move it over the surface of the tooth with an erasing
motion.
5. A saliva ejector or HVE is recommended.
IMPORT NT: Do not leave the vibrating tip in one place as it can cause serious damage to the tooth
or surrounding tissues. Engler Engineering Corporation will not be liable for damage due to
improper use of this device.
Note: We have designed the this device with a feature called Delayed Cavitation. This function purges the
tip of water after releasing the footswitch to prevent bacteria from entering the tip.
IMPORT NT: Pressure on the tip is not necessary to remove calculus or tartar. Enamel on the teeth
may be damaged or removed if using excessive pressure. It can also be damaged if the scaling tip is left to
rest in one spot for even a few seconds. Using the tip as a pry to remove calculus or tartar will likely damage
the tooth and may change the shape of the tip, which in-turn, changes the frequency. The normal power
setting for most procedures should be LOW range. Since every operator has a different technique and
some patient’s are more sensitive than others, the power may be adjusted to satisfy specific requirements.
The use of a face mask is recommended when operating the scaler, to avoid inhalation of bacterially
contaminated aerosol (water mist) by the operator.
Only individuals properly trained in modern dental cleaning techniques
should be allowed to perform dental procedures.
Improper use of dental equipment can permanently and severely harm a
patient's teeth.
This manual does not attempt to teach dentistry.
Do not use this equipment without proper training.
Page 12 of 35

SC LER M INTEN NCE
FINAL PROCEDURES AT THE END OF EACH DAY:
1. Make sure the unit is turned off.
2. Detach the tip and ultrasonic handpiece and sterilize.
3. Disconnect the unit from its water source or turn off the water supply.
. Clean and disinfect all surfaces.
Instrument Tips:
IMPORT NT: The scaling tips should be thoroughly cleaned and free of blood, tissue, or any
other debris before sterilization.
The scaling tips,and ultrasonic handpiece may be sterilized by autoclave or chemiclave, always
following the manufacturer’s instructions and recommendations. Do not autoclave over 270
degrees F or more than twenty (20) minutes..
It is recommended to remove tips daily for sterilization. The tip may “weld” itself in the handpiece if
left in for weeks on end.
Chassis:
The chassis should be cleaned at the end of every operating day with a chemical sterilization
solution. This procedure could be done by spraying a fine mist of sterilization solution on the unit,
allowing it to remain on the chassis for the length of time recommended by the manufacturer. The
surface should be wiped with a clean damp cloth or as suggested by the chemiclave manufacturer.
Dry completely.
SCALER HANDPIECE, FOOTSWITCH AND POWER CABLES:
After each procedure, or at least once a day, it is suggested that the handpiece and its cable be
thoroughly cleaned and sterilized. The recommended procedure is as follows:
1. Remove tip, and ultrasonic handpiece- sterilize these items as listed above.
2. Clean the outer surface of the handpiece and its cable with an antiseptic soap, rinse off with
water and sterilize with a chemical sterilization solution.
Note: If any chemicals are allowed to get into the handpiece it must be flushed out with
clean water.
3. Place cleaned tip and into handpiece for next patient.
. The footswitch and power cables should be cleaned periodically by spraying a mist of
sterilization or cleaning solution on the cables. It should remain on the cables for the length
of time recommended by the manufacturer. Wipe the surface with a damp cloth and dry the
cables completely.
Page 13 of 35

POLISHER INST LL TION
1. Connect the micromotor into the front of the control box. This is done by inserting the male connector at
the end of the micromotor cable into the female receptacle on the front panel of the unit and rotating the
lock collar clockwise.
2. Slide the straight handpiece down over the top of the micromotor.
3. Line up the notch of the prophy angle with the aligning pin on the straight handpiece, and then push the
shaft of the prophy angle into the chuck of the straight handpiece.
. Rotate the lock ring clockwise, until it clicks and locks the prophy angle in place.
5. Place a disposable rubber-polishing cup on the end of the prophy angle by pushing / snapping it on. The
prophy angle is now secured and ready for operation.
6. Rotate the selector knob on the front of unit to either forward or reverse operation.
NOTE 1: When using the polisher the green led indicator should be on.
NOTE 2: High speed will damage the gears in the prophy angle, splatter the polishing compound and
overheat the teeth, possibly burning them. Always start at the lowest setting, then increase speed as
necessary.
SAFETY INFORMATION:
Never turn the lock ring while the handpiece is in operation.
1. Never reverse the direction of the micromotor while it is in operation. Damage to the unit may occur.
2. Take care not to drop the micromotor, handpiece, or prophy angle on the floor or other hard surface.
3. Do not lock or run the micromotor / straight handpiece assembly without a prophy angle, cutting disc,
contra angle, or test shaft installed, doing so could damage the straight handpiece and / or micromotor.
. Never oil the micromotor.
Polisher continued on next page
Page 14 of 35

5. When installing the prophy angle or other accessory, make sure that the lock ring is rotated fully in the
unlock position, otherwise the accessory can not be installed and the straight handpiece will not operate.
6. Do not rotate the selector switch on the front of the control box between forward and reverse rapidly.
always allow a moment between the two selections.
7. As this is a precision instrument, always return it to Engler Engineering Corporation for maintenance and
repair.
POLISHER OPER TION
1. Dampen the rubber cup and place a small amount of polishing paste onto it.
2. Rotate the POWER control to the minimum setting in the Prophy Range.
3. Press the footswitch and the rubber cup will begin to rotate. The speed of rotation may be adjusted to
your desired level by re-adjusting the POWER control.
. To keep the paste from flying off the cup, maintain a low speed and bring the cup up to the tooth before
depressing the footswitch. Then gently rotate the spinning cup against the tooth.
Caution: It is very easy to overheat and cause permanent damage the teeth. Use the lowest
speeds possible with minimal pressure when polishing.
IMPORT NT: The prophy angle is only rated for up to 5000 RPM - therefore, in order to prevent
premature failure of the bearings and gears, keep the unit set in the prophy range whenever the
prophy angle is attached to the straight handpiece.
5. High-speed settings can overheat the tooth being polished and throw the polishing paste off of
the rubber cup. Always start with a lower speed and then adjust to a higher speed as required.
6. Place the end of the angle into the patient's mouth and gently apply the rubber cup to the
surface of the tooth with a circular motion. Do not allow the rubber cup to remain stationary on
one area for an extended period of time.
Page 15 of 35

POLISHER M INTEN NCE: PROPHY NGLE
The Prophy Angle is a precision engineered dental device. All of the gears and shaft assemblies
are made of high-grade stainless steel, which when kept clean and properly lubricated, will provide
trouble free service.
Daily Cleaning and Lubrication:
1. Remove prophy angle from straight handpiece.
2. Discard used rubber cup.
3. Follow the cleaning instructions that were supplied with the prophy angle.
. Place a new rubber cup onto the angle.
5. Slide the prophy angle down over the straight handpiece and lock it in place.
IMPORT NT: For a long dependable life, the prophy angle should be cleaned and lubricated daily,
if possible after each use.
Keep hair away from prophy cup and head cap.
SUGGESTION: To keep hair from being tangled in the angle, we recommend using a gentle
adhesive tape such as masking tape around the lips, keeping hair in place and away from
treatment area.
Page 16 of 35
P-MM-E 35,000 RPM
Micromotor W/Cord
P-SH-A1:1 Straight
handpiece
P-A1-B Prophy
Angle (Sealed
Bearings)
Snap on rubber cups
Polishing
paste

STR IGHT H NDPIECE M INTEN NCE
LUBRICATION:
1. The spray nozzle oiling method is optional but highly recommended because it cleans as well as
lubricates. It forces lubricant into the smallest spaces and forces dirt out. The alternate method
is to place 1 drop of approved oil in the chuck hole. Do not lubricate the handpiece while it is on
the micromotor.
2. Lubrication of the straight handpiece is required at least once a day when used. Preferably by the end of
the day.
3. Make sure that the straight handpiece is in the unlocked position prior to lubricating.
. Install the E-Type nozzle by pushing it into the top of the spray can. Insert the E-Type nozzle
into the bottom of the handpiece (where the micromotor goes). Holding the two together tightly,
with can in the upright position, push spray button for 2 to 3 seconds.
NOTE: If spray time is too short oil may not be propelled into all areas of the handpiece.
Page 17 of 35

CLE NING and STERILIZ TION OF H NDPIECE
CLEANING:
1. Wipe the handpiece clean with an alcohol-soaked soft tissue.
2. Never clean the handpiece with boiling water, chemical solutions, ultrasonic cleaner, or with
wire brushes.
STERILIZATION:
1. Autoclaving is recommended for the Engler straight handpiece.
2. Clean the handpiece as described above.
3. Lubricate the handpiece as described above .
. Place the in an autoclaving pouch and seal it in accordance with the instructions on the pouch.
5. Autoclave the handpiece for no longer than 20 minutes at 121 C (250 F), or 15 min. at 132 C (270 F).
Keep the straight handpiece away from water vapor or mist that may settle and cause premature damage to
the bearings.
IMPORT NT: DO NOT UNDER ANY CIRCUMSTANCE, attempt to repair, disassemble or unscrew the
straight handpiece. Doing so may shift the internal springs, causing permanent damage to the unit and will
void the warranty. If you experience problems during operation, call our repair department.
Page 18 of 35

MICROMOTOR M INTEN NCE
The E-Type Micromotor is capable of speeds up to 35,000 RPM, for use in cutting, sectioning and drilling. It
contains sealed bearings and does not require any lubrication. The Micromotor has cooling vents at the
back of the unit, Do not allow water, oil, or any other substance to enter these vents. Failure to keep debris
out of the Micromotor will shorten the life of the unit and cause permanent damage.
IMPORTANT:
1. NEVER change the direction of the micromotor while it is in operation.
2. ALWAYS wait until it has come to a full stop.
3. NEVER oil, or allow oil or water to get into the micromotor.
Page 19 of 35

SC LER TROUBLESHOOTING
“ON” LED INDICATOR DOES NOT LIGHT UP:
1. Verify that unit is switched ON, the ON / OFF switch is located on the left hand side.
2. The unit is not plugged in to a power outlet: verify that the unit is plugged in.
3. Power outlet not active: try another outlet.
. The power cord is not connected to the device.
5. Contact Engler Engineering Corporation.
“ON” LED INDICATOR LIGHTS UP, NO WATER FLOW:
1. Verify that the selector switch is in scaling mode.
2. Verify water line is connected and water is flowing to unit.
3. Verify that the waterline is correctly connected to the coupling insert at the back of the unit.
. Check if water line is kinked or twisted.
5. Check Water Filter and Disk: clean disk with plain water and a toothbrush. If clogged, replace O-Ring
and Disc.
6. If using Portable Water Tank: Verify you have the correct water level and sufficient pressure.
7. Water blockage in tip: replace the tip. (Clean with 0.012” piano wire)
8. Contact Engler Engineering Corporation.
“ON” LED INDICATOR LIGHTS UP, LITTLE OR NO VIBRATION / AT THE TIP:
1. Tip loose: tighten the tip.
2. Tip damaged: replace the tip.
3. Handpiece dropped or the crystal is otherwise damaged.
. Contact Engler Engineering Corporation.
Scaler troubleshooting continued on the next page
Page 20 of 35
Table of contents
Other Engler Dental Equipment manuals
Popular Dental Equipment manuals by other brands

Belmont
Belmont MCX-II technical information

Dentsply Sirona
Dentsply Sirona CEREC SpeedFire operating instructions

NSK
NSK Viva Ace Quick operation guide

B.A. International
B.A. International Ultimate 1400 Instructions for use

3M ESPE
3M ESPE Pentamix Lite operating instructions

Eighteeth
Eighteeth Fast-Fill user manual

KaVo
KaVo SMARTmatic PROPHY S31 Instructions for use

MIHM-VOGT
MIHM-VOGT HT-2/M/Zirkon-120 Original operating manual

Reborn Endo
Reborn Endo Expert instruction manual

Anthogyr
Anthogyr Implanteo instruction manual

Schick Dental
Schick Dental S2 Master Original operating instructions

Belmont
Belmont tbCompass operating instructions