Envirco Hospi-Gard AirCeil User manual

Hospi-Gard®AirCeil™
HEPA Filtration System
INSTALLATION, OPERATION, AND MAINTENANCE MANUAL
ENVIRCO Technical Support: 800-884-0002

2
Installation, Operation, & Maintenance Manual
Innovators in Clean Air Technology | www.envirco.com
Hospi-Gard®AirCeil™
HEPA Filtration System
Table of Contents
Introduction....................................................................................................................... 3
Installation ........................................................................................................................ 4
Operation.......................................................................................................................... 5
Airow Conguration ........................................................................................................ 7
Negative Pressure............................................................................................................ 8
Service.............................................................................................................................. 9
Relocating a Contaminated Unit.......................................................................................11
Unit Decontamination .......................................................................................................11
Troubleshooting............................................................................................................... 12
Wiring Diagram.................................................................................................................13
Specications ...................................................................................................................13
Parts List...........................................................................................................................14
Parts Location...................................................................................................................14
Optional Accessories........................................................................................................15
Appendix...........................................................................................................................16
Limited Warranty ..............................................................................................................17

3
Installation, Operation, & Maintenance Manual
Innovators in Clean Air Technology | www.envirco.com
Hospi-Gard®AirCeil™
HEPA Filtration System
■Introduction
Purpose
The Hospi-Gard®AirCeil™ module is a self-contained, ceiling-mounted HEPA ltration system specically designed to provide
a high velocity airow pattern to lter out unwanted room particles including airborne microbiological contamination. The
AirCeil can either recirculate clean air back into the room or exhaust the air out to create a cost effective, negative pressure,
patient isolation room.
Description
Air enters the module through the intake grille located on the lower section of the unit. The air is evenly drawn by the motor/
blowers rst through a prelter and then through a High Efciency Particulate Air (HEPA) lter before being recirculated back
into the room through the Ceiling Filtered Air Diffuser. Alternatively, the HEPA ltered air can be exhausted to the outside or
returned to the main HVAC system to create negative pressure. Whether used for recirculation or exhaust, the Hospi-Gard
AirCeil moves air at a variable volume ranging from 300 – 600 cubic feet per minute (ft3/min) to carry away airborne particles
generated by the patient or activities within the room.
Features
• The self-contained, ceiling mounted HEPA ltration system is designed to t in a standard 2 ft x 4 ft T-bar ceiling grid or
other ceiling types.
• Designed to effectively create a low-cost negative pressure isolation room to meet CDC and OSHA guidelines.
• Tamper resistant IR hand held remote controller allows for “hands off” controlling of the module.
• The optional Ceiling Filtered Air Diffuser and 12-inch Clear Flexible Ducting allows the unit to be converted from a 100%
exhaust system to a total air recirculation system.
• Total height of the unit does not exceed 19.5 inches.
• Pending FDA and cUL listing.
• HEPA lter and prelter are room-side replaceable.
Applications
Hospital
• Convert Patient Rooms To Negative Pressure Isolation Rooms
• Waiting Rooms
• Bronchoscopy Rooms
• Emergency Rooms
• Intensive Care Units
• Sputum Induction Booths And Rooms
• Aerosol Pentamidine Booths And Rooms
• Renal Dialysis Treatment Rooms
• Hospital Reconstruction Areas
Other
• Clinics
• Physician Ofces
• Correctional Facilities
• Nursing Homes
• Homeless Shelters
• Addiction Recovery Center
CAUTION AND WARNINGS
CAUTION: To reduce the risk of re, electrical
shock, or injury to persons, observe the following.
1. Installation work and electrical wiring
must be done by a qualied person(s)
in accordance with all applicable codes
and standards, including re-rated
construction.
2. When cutting or drilling into a wall or
ceiling, do not damage electrical wiring or
other hidden utilities.
3. Service to and decontamination of this
equipment should be performed by
an authorized technician trained and
experienced in performance evaluation
and maintenance of clean air equipment.
However, certain procedures are outlined
in this manual that can be performed by
the owner.
4. Before servicing the unit, switch power
off at service panel and lock service panel
to prevent power from being switched on
accidentally, and follow proper procedures
as necessary.
5. Use this unit only in the manner intended
by the manufacturer.

4
Installation, Operation, & Maintenance Manual
Innovators in Clean Air Technology | www.envirco.com
Hospi-Gard®AirCeil™
HEPA Filtration System
■Installation
Unpacking Instructions
1. The Hospi-Gard AirCeil unit is shipped fully assembled in an open crate.
2. Inspect the AirCeil crating for any external shipping damage immediately upon arrival. If any damage is observed, a freight
claim report should be completed and promptly led with the responsible carrier.
3. Carefully remove the open crating from around the AirCeil unit.
4. Remove the plastic wrapping from the AirCeil unit and inspect for any undisclosed damage that may have occurred during
transportation (Figure 1).
5. If any damage is discovered, le a claim with the responsible carrier immediately.
Assembly
Note: The Hospi-Gard AirCeil unit is completely assembled at the factory with the exception of the ¼-inch-20 eyebolts that can
be used when hanging the unit from an overhead structure.
• Insert four (4) ¼-inch-20 eyebolts (not included) into the nutserts located on the top-side of the AirCeil unit before lifting
into an overhead position (Figure 2, page 5).
• Raise the unit and secure it into place using “S” hooks and chain to suspend it from an overhead structure (supplied by
others).
• The AirCeil unit may be used to exhaust all of its air directly to the outside or to the main HVAC system by attaching 12
inch exhaust ducting to the duct collar, located on the side of the AirCeil unit.
Note: Verify that the ducting and the external exhaust motor/blower can handle the air volume exhausted by the unit.
• The optional Ceiling Filtered Air Diffuser (Part No. 63841) and 12-in. clear exible ducting (Part No. 63842) will convert
the AirCeil unit for use as a 100% Air Recirculation Unit.
• If purchased with the optional power cord, plug the cord into a properly grounded three-pronged receptacle, otherwise
hard wire according to the wiring diagram (Figure 8).
• Conrm the circuit to be used is sized to provide a minimum of 5 Amps and conforms to NEC (National Electrician Code).

5
Installation, Operation, & Maintenance Manual
Innovators in Clean Air Technology | www.envirco.com
Hospi-Gard®AirCeil™
HEPA Filtration System
■Operation
Airow Indicator Light
The Airow Indicator light on the AirCeil control panel (Figure 3 below) glows red when the unit is running and green when in
the standby mode or when adjusting the speed with the hand held IR Remote Controller.
Note: When the AirCeil is “OFF” both the red and green indicator lights are not lit.
AirCeil Flow-Set
The Flow-Set is a hand held Infrared Remote Controller congured to adjust the airow of the AirCeil unit. The combined use
of the Flow-Set Transmitter and the Infrared Sensor mounted on the control panel of the AirCeil, allows the motor to be turned
“ON” or “OFF”, adjustments to the airow index (from 1-100) and conrms the current airow setting.
Note: When any of the Flow-Set buttons are depressed, the Flow-Set Indicator Light will ash, indicating that a change has occurred.
During an adjustment session, the green Setting Indicator Light on the AirCeil Control Panel ashes each time a valid entry is made.
• The button to the left of the up arrow quickly increases to the airow index by increments of ten (see Figure 4).
• The button to the left of the down arrow quickly lowers to the airow index by increments of ten.

6Innovators in Clean Air Technology | www.envirco.com
Installation, Operation, & Maintenance Manual
Hospi-Gard®AirCeil™
HEPA Filtration System
• Similarly, using the set of buttons to the right of the arrows 1 allows ne, single digit adjustments to the airow index.
• If the airow index is already set to its maximum value of 100, any further attempts to raise the airow index using either
the 10 button or the 1 button will not be allowed. The green Setting Indicator Light will not blink, indicating an invalid entry.
• With an airow index of 91, depressing the 10 button will be considered an invalid entry. The green Setting Indicator Light
will not blink and the increase will not occur because this entry would take the airow index above 100.
• The 1 and 10 buttons respond in a similar manner during any attempts to set the airow index below 1 because 0 is not
a valid airow index.
Turning the AirCeil “ON” (Run Mode)
1. Point the Flow-Set Transmitter toward the Infrared Sensor on the AirCeil control panel.
2. Press the ON/OFF button once and the both the red operation light and the green setting light will glow conrming the unit is “ON.”
3. Press the Enter button to conrm that the unit is in the Run Mode and to display the current airow setting.
4. In the Run Mode only the red operation light will remain lit.
Temporarily Turning the AirCeil ”OFF” (Standby Mode)
1. With the AirCeil “ON”, press the ON/OFF button once to temporarily stop the unit.
2. The red operation will turn “OFF”, but the green Setting Indicator Light will remain lit indicating the unit is in the Standby Mode.
3. After 15 minutes of inactivity, the unit will resume operation at its current airow setting.
4. The Airow Indicator Light on the AirCeil control panel will now glow red,conrming that the unit is now back in the Run Mode.
Turning the AirCeil “OFF”
1. With the AirCeil “ON”, press the ON/OFF button once and the unit will stop temporarily and the Setting Indicator Light will
glow green, conrming that the unit is in the Standby Mode.
2. Press the Enter button once to prevent the unit from resuming after 15 minutes of inactivity.
3. The green Setting Indicator Light will ash, showing the last programmed airow setting.
4. The motor will ramp up temporarily and then turn “OFF.”
5. When the AirCeil is not in operation, both the red and green indicator lights will be turned “OFF.”

Hospi-Gard®AirCeil™
HEPA Filtration System
7Innovators in Clean Air Technology
Installation, Operation, & Maintenance Manual
Motor Speed or Airow Adjustment Session
1. Point the Flow-Set Transmitter toward the Infrared Sensor.
2. Press the ON/OFF button or any of the four arrow buttons and the Setting Indicator Light will glow green, indicating an
airow adjustment session is in process.
3. Continue to press the appropriate arrow buttons to reset the AirCeil to the new airow setting.
4. Press the Enter button to accept the new airow setting and to exit the adjustment session.
5. Pressing the Clear button prior to hitting the Enter button to delete the new setting and revert back to the original airow setting.
Note: If the Controller’s buttons are inadvertently depressed or you enter an adjustment session but do not modify the
airow setting, the adjustment session automatically terminates after 15 minutes of inactivity.
Viewing the Current Motor Speed or Airow Setting
1. With the unit in the Run Mode, depress the Clear button once.
2. The green Airow Setting Indicator Light begins to ash indicating the signal was received.
3. The green ash sequence indicates the current airow index (1-100).
4. The green ash sequence occurs in two sets, with rst set of long ashes representing the tens digit setting and the
second set of short ashes representing the single digit setting.
5. An extra long ash replacing either the long ash(es) or the short ash(es) indicates the value of the corresponding tens
digit or single unit digit setting is zero.
• Example 1: Two long ashes followed by four short ashes represents an airow index of 24. The motor will be operating
at approximately 24% of its maximum capacity.
• Example 2: Eight long ashes followed by nine short ashes represents an airow index of 89. This setting indicates that
the motor is operating at approximately 89% of its maximum capacity.
• Example 3: An extra long ash followed by four short ashed represents an airow index of 04 or that the motor is running
at 4% of capacity.
• Example 4: Nine long ashes followed by an extra long ash represents a ow index of 90 or that the motor is running at
90% of capacity.
■Airflow Configuration
Note: Turn the AirCeil “ON” and let it run for a 15-30 minutes to ensure a clean environment is established when running
the unit in the recirculation mode. Alternatively, a permanently mounted or handheld room pressure monitor can be used to
conrm room conditions meet the latest CDC guidelines for a negative pressure isolation room.
Complete Air Recirculation
1. Install the optional Ceiling Filtered Air Diffuser (Part No. 63841) in a standard 2’ x 4’ T-bar ceiling grid with a 2’ x 2’ blank
off panel.
2. Use clear, exible 12-inch ducting (Part No. 63842) to connect the optional Ceiling Filtered Air Diffuser to the AirCeil
exhaust collar.
3. The AirCeil has the capacity to recirculate up to 600 ft3/min of air.
Note: The actual capacity of the AirCeil will depend on several factors unique to each facility and may include, but shall
not be limited to the length and number of bends in the exible ducting, the type of prelter used and the voltage supplied.
4. The room air change rate the AirCeil unit will generate depends on both the AirCeil airow index setting and the size of
the room. See Table 1 in the Appendix for representative room air changes per hour (ACH).
5. The exibility of placement of the AirCeil unit relative to the Ceiling Filtered Air Diffuser can optimize the needs of the
patient, healthcare workers and the specic activities within the room.
6. Positioning the AirCeil away from the foot of the bed and the Ceiling Filtered Air Diffuser directly above the bed or area of
activity (see Figure 11) allows clean air to be constantly swept across the patient. This may be benecial for an immune
compromised patient.
7. To minimize the spread of airborne contaminants generated by the patient, position the AirCeil toward the head of the bed
and the Ceiling Filtered Air Diffuser downstream from the foot of the bed (see Figure 12). This allows clean air to sweep
past healthcare workers and visitors while removing airborne contaminants generated by the patient.

Hospi-Gard®AirCeil™
HEPA Filtration System
8Innovators in Clean Air Technology | www.envirco.com
Installation, Operation, & Maintenance Manual
Total Air Exhaust Mode (Negative Pressure)
1. Attach any 12-inch ducting to the AirCeil exhaust collar.
2. The HEPA ltered exhaust air can be vented either to the outside or returned to the main HVAC system2.
Note: Veried that the ducting and the external exhaust motor/blower can handle the air volume exhausted by the unit.
3. The AirCeil has the capacity to exhaust up to 600 ft3/min of air.
Note: The actual capacity of the AirCeil will depend on several factors unique to each facility and may include, but shall
not be limited to the length and number of bends in the exible exhaust ducting, the type of prelter used and the voltage
requirements.
Initial Start-Up Testing and Certication
Note: The CDC requires that quantitative leakage and lter performance tests be performed at initial installation and every
time the lter is changed or the AirCeil unit moved. These tests should be repeated every 6-12 months. The factory tests
should not be substituted for the initial test.
1. Initial on-site testing and certication of the Hospi-Gard AirCeil is included with the purchase of each unit within the
Continental United States.
Note: Operation of the unit prior to authorized certication may void the warranty.
2. The following services shall be performed at the time of initial on-site testing and certication.
3. Test the HEPA lter quantitatively for leakage and to verify lter performance. Any leaks detected will be repaired or the
lter will be replaced as required.
4. Conrm the maximum capacity of the AirCeil unit is 600 ft3/min ± 10% when the airow index is set on its maximum value
of 100 and the unit free of optional carbon or anti-microbial prelters and any ducting attached to the AirCeil collar.
5. Test ground leakage and ground continuity for electrical safety.
6. Contact an Authorized Service Contractor to arrange a certication date.
Authorized Service Contractor: ENV Services, Inc. | 800-345-6094
■Negative Pressure
CDC guidelines2recommend that rooms used for negative pressure isolation should be single patient rooms with negative
pressure relative to the corridor or other areas connected to the room. The minimum pressure difference necessary to achieve
and maintain negative pressure that will result in airow into the room is 0.001 inch W.G. (a higher pressure is preferred). To
achieve this, the exhaust ow should be 10% or 50 CFM greater than the room supply air, whichever is greater.
Achieving Negative Pressure
Note: It is the responsibility of the purchaser and not ENVIRCO Corporation to ensure the room in which the AirCeil unit
is used to create the negative pressure is in compliance with CDC guidelines2for a negative pressure isolation room and
monitored to ensure its continuing performance.
1. Each AirCeil has the capacity to exhaust up to 600 ft3/min of air.
2. The attachment of ductwork to the exhaust port decreases the air handling capacity of the AirCeil unit. The magnitude of
the decrease will vary depending on the length and type of duct as well as its conguration. For example, a 10-foot run of
standard exible duct containing two 90-degree bends will reduce the exhaust capacity by approximately 100 cfm.
3. A room pressure monitor can be used to conrm the room conditions meet the latest CDC guidelines for a negative
pressure isolation room.
2 CDC. Guidelines for Preventing the Transmission of Tuberculosis in Health-Care Facilities, 1994. MMWR, October 28, 1994; 43 (no. RR-13.)

Hospi-Gard®AirCeil™
HEPA Filtration System
9Innovators in Clean Air Technology
Installation, Operation, & Maintenance Manual
Difculty Achieving Negative Pressure
Note: The AirCeil has the capacity to exhaust up to 600 ft3/min of air, however the actual capacity will depend on several
factors unique to each facility and may include, but shall not be limited to the length and number of bends in the exible
exhaust ducting, the type of prelter used and the voltage requirements.
1. The equipment may not be able to achieve or maintain negative pressure in conformance with the CDC guidelines due to
issues with the existing building ventilation system or air leaks within the negative pressure room.
2. The room should be inspected for any possible source of air leakage through doors, windows, plumbing and equipment
wall penetrations.
3. Take appropriate corrective action should be taken to seal the leaks.
4. If negative pressure cannot be attained, an additional AirCeil may be required.
■Service
Note: Proper protective equipment and measures must be used at all times during unit cleaning. Check with your Safety Ofce
to assure the cleaning solutions, protective equipment (disposable hospital gown, NIOSH approved respirator, protective
gloves) and protocols followed comply with your facility’s guidelines.
Cleaning the Unit
1. Periodic Cleaning of the unit may be done with a mild disinfecting solution or a mild bleach solution.
2. Cleaning must be performed with the unit running. Wipe the exterior of the unit with a damp sponge or cloth. Be careful
not to saturate the unit or the lter. Carefully clean the unit, paying particular attention to the prelter intake grille.
Prelter Replacement (Refer to CDC Guidelines)
NOTE: The unit must be running during prelter replacement.
1. The prelter should be replaced every 60-90 days, depending on the environment. Regular replacement of the prelter
will maximize the life of the HEPA lter.
2. Wipe the unit as described in the Section 6.1.1.
3. Spray a mist of phenolic disinfectant such as, “Amphyl” or “Lysol” into the air intake grille, covering the entire prelter area.
4. Open the room-side Filter Access Panel using the latch key to unlock the three tamper resistant panel latches by turning
in the direction indicated in Figure 5 below.
5. Remove the three (3) Prelter Retainer Slats by shifting them through the cutouts in the prelter screen (Figure 6).
6. Insert hand into red plastic medical waste disposal bag. Grasp the top corner of the prelter with bagged hand and gently
pull outward.
7. Turn bag inside out over prelter and close bag for containment. Use closure strip tie to seal bag.
8. Position new prelter in the prelter screen. Ensure the edges of the prelter are tucked neatly under screen anges.
9. Replace the Prelter Retainer Slats.

Hospi-Gard®AirCeil™
HEPA Filtration System
10 Innovators in Clean Air Technology | www.envirco.com
Installation, Operation, & Maintenance Manual
HEPA Filter Replacement
Note: Decontaminate the unit before replacing the HEPA lter. Only technicians trained and experienced in decontamination
and performance evaluation of clean air equipment should perform this procedure.
Note: A procedure for formaldehyde gas decontamination is provided in the Unit Decontamination section; however, contact
your Safety Ofce to conrm that this procedure complies with your facility’s guidelines.
Note: Always use a NIOSH approved respirator and other protective equipment and clothing before replacing the HEPA lter
using your Safety Ofce approved procedures.
Note: Turn the unit off and disconnect from power source before HEPA lter replacement or before attempting service.
1. Open the Filter Access Panel as described in the previous section.
Note: If the unit includes an optional UV Germicidal Light with Reector, it must be removed before the HEPA lter can
be accessed.
2. While supporting the HEPA lter by its aluminum frame, remove the six (6) lter clips and bolts and lower the HEPA lter
(Figure 7) from the AirCeil unit. Slip a red plastic medical waste disposal bag over the HEPA lter from the top down. The
top of the HEPA lter downward and wrap the bag over the bottom end and seal the bag.
3. Remove the new HEPA lter from its container and carefully install in reverse order of removal.
Caution: Handle the HEPA lter with extreme care, touching only the sides of the HEPA lter’s aluminum frame. Never
touch the white HEPA lter pleats. Even a light touch on the lter surface may damage the lter.
4. A quantitative leakage and lter performance test should be performed each time the lter is changed.
Motor/Blower Lubrication
The AirCeil motor/blower comes equipped with sealed for life bearings and requires no lubrication.

Hospi-Gard®AirCeil™
HEPA Filtration System
11Innovators in Clean Air Technology
Installation, Operation, & Maintenance Manual
■Relocating a Contaminated Unit
Note: Decontaminate before relocating a contaminated AirCeil unit. Always use a NIOSH approved respirator and other
protective equipment and clothing when relocating the AirCeil unit following your Safety Ofce approved procedures. See
CDC recommendations for unit relocation.
1. With the AirCeil unit “ON”, remove and replace the prelter as described in the Service section.
2. Clean the exterior surfaces of the unit as described in the Service section.
3. Turn the unit “OFF” and disconnect it from the power source.
4. Decontaminate the AirCeil as described in the section below.
5. Perform a leakage and lter performance test prior to using the relocated unit.
Note: CDC guidelines require that a quantitative leakage and lter performance test should be performed each time the unit
is relocated. A list of licensed testing organizations is available on request from ENVIRCO Corporation.
■Unit Decontamination
Formaldehyde Decontamination
Should decontamination of the unit be required, the following is an adaptation by Annex G. of the Recommended Microbiological
Decontamination Procedure from NSF 49, May 1992. Conrm with appropriate Biosafety and Industrial Safety Professionals
that the procedures meet your facility’s guidelines.
1. Transport the unit as described in the Service section to a controlled access, non-public area with a non-porous oor, good
ventilation and a dedicated exhaust directly outside the building. Attach and seal a exible hose to a dedicated exhaust.
Place the other end of the exhaust hose near the unit.
2. The total volume of the AirCeil is 13.85 ft3 (24 in. x 52 in. x 20 in.)
3. Multiply the total volume of the unit by 0.3 g/ft3 to determine the gram weight of paraformaldehyde required. Decontamination
of the AirCeil requires 4.2 g (13.85 ft3 x 0.3 g/ft3) paraformaldehyde.
4. Wipe down the lter access panel with an appropriate surface decontaminant.
5. Open the hinged lter access panel and place an unplugged heating device, such as a commercially available electric
frying pan with the thermostat set at 232.2° to 246.1° C (450° to 475° F) inside the unit through the opened lter access
panel. Spread the paraformaldehyde evenly over the heating surface of the electric frying pan.
Caution: The auto-ignition temperature of paraformaldehyde is 300° C (572° F).
6. Place a hot plate, beaker of water, temperature and humidity indicators in the unit next to the pan that will contain the
paraformaldehyde.
7. Enclose the unit completely with heavy gauge plastic lm and tape. The optional power cord should be coiled and taped
to the unit and sealed under the plastic lm. Seal the lm to the oor on which the unit stands.
8. Determine the temperature and humidity inside the unit.
9. The temperature should be 21.1° C (70° F) or higher with a humidity level between 60% to 85%. Use the hot plate to heat
the beaker of water unit the desired temperature and humidity are achieved. Disconnect the hot plate.
10. Carefully seal around the power cord extending from the electric frying pan so that formaldehyde gas will not leak out.
11. Plug the cord of the electric frying pan into an outlet.
12. After the paraformaldehyde has de-polymerized, disconnect the frying pan from the electrical outlet.
13. Allow the unit to stand for a minimum of two hours or overnight.
14. Attach the exible hose to the unit and allow it to draw from inside the unit.
15. After 15 minutes of visible exhaust activity, small openings in the lm sealing the unit may be made over to improve ventilation.
16. Allow the unit to ventilate overnight.
17. Remove the unit from exhaust ventilation when the formaldehyde gas has been exhausted.
18. During unit decontamination, respiratory protection for service personnel is recommended. Only National Institute for
Occupational Safety and Health (NIOSH) approved respirators should be used.
19. Remove the lter as described in the Section 6.3. The lter can now be disposed of as general waste.

Hospi-Gard®AirCeil™
HEPA Filtration System
12 Innovators in Clean Air Technology | www.envirco.com
Installation, Operation, & Maintenance Manual
Phenolic Disinfectant
In facilities where formaldehyde gas decontamination is prohibited or not feasible, the unit can be disinfected with the following
phenolic procedure.
Note: The unit must be running during phenolic disinfection.
1. With the AirCeil ”ON”, remove the prelter as described in Service section.
2. Close the lter access panel and with the unit running, spray a mist of phenolic disinfectant such as Amphyl or Lysol into
the intake screen.
3. Continue spraying the entire lter access area until the HEPA lter is saturated.
4. Remove the HEPA lter as described in the Section 6.3. Since the lter has not been decontaminated by the formaldehyde
procedure, it must be placed in a red medical waste bag and disposed of properly.
■Troubleshooting
Note: Disconnect the unit from the electrical power source before attempting service.
Note: Always use a NIOSH approved respirator and other protective equipment and clothing before attempting to service or
repair the unit using your Safety Ofce approved procedures.
Inoperative Air Flow
1. Conrm that the optional power cord is plugged into a properly functioning building receptacle or that the unit is properly
hard-wired to the main electrical system and that power is present.
2. Conrm that the batteries to the Flow-Set or hand held IR Remote Controller are charged.
3. Conrm that the Flow-Set Transmitter is pointed directly toward and is no more than 10 feet from the Infrared Sensor
mounted on the control panel of the AirCeil.
4. Conrm that the unit is turned “ON” and that the red Operational Light is lit.
5. If the green Setting Light is lit, the unit is in the Standby Mode. Press the Enter button to resume operation, or if nothing
is done, the unit will resume operation at its last accepted setting within fteen minutes.
6. Ensure that the airow index is properly set from 1-100 and press the Enter button to accept the setting if it is at the
desired rate or reset to the correct setting.
Low Air Capacity
1. Check prelter for obstruction and remove obstruction as necessary.
2. Replace the prelter media.
3. Check power supply for proper voltage, amperage and distribution frequency.
4. With the unit in the Run Mode, depress the Clear button once to ensure that the airow index is properly set from 1-100.
Press the Enter button to accept the setting if correct or reset if incorrect.
High Air Capacity
1. Check power supply for proper voltage, amperage and distribution frequency.
2. With the unit in the Run Mode, depress the Clear button once to ensure that the airow index is properly set from 1-100.
Press the Enter button to accept the setting if correct or reset the index is set too high.
For further help with troubleshooting, please contact ENVIRCO Technical Support:
ENVIRCO
101 McNeill Road
Sanford, NC 27330
Tel: 800.884.0002
Fax: 919.774.8771

Hospi-Gard®AirCeil™
HEPA Filtration System
13Innovators in Clean Air Technology
Installation, Operation, & Maintenance Manual
■Wiring Diagram
■Specifications*
Dimensions AirCeil Unit: 23.62 in. D x 47.62 in. W x 19.5 in. H
Exhaust Collar: 2.0 in H x 12.0 in. Diameter.
Weight 78 lbs.
Power Requirements
115 V, Single Phase, 60 Hz
1.4 Amps at 600 ft3/Minute
Full Load Amps: 2.25
Power Connection Standard 7/8-inch Conduit Knock-Out
HEPA Filter Media 99.99% at 0.3 Micron Minimum Efciency
Prelter Woven Polyester Media in a Removable Frame
Motor 2 ea. 128 Watts, Direct Drive, Continuous Duty with Sealed-for-Life Bearings
Sound Level 50-55 dBA (Ambient Sound Level < 40 dBA), Measured 30 Inches from the Filter Face
*Designs and specications are subject to change without notice and without incurring liability or modications to equipment previously sold.

Hospi-Gard®AirCeil™
HEPA Filtration System
14 Innovators in Clean Air Technology | www.envirco.com
Installation, Operation, & Maintenance Manual
■Parts List
Description Qty. Required Part No.
Blower/Motor 2 63829
Capacitor 1 63223
HEPA Filter (19.88” x 43.88” x 3”) 1 69391-033
Prelter Slat 3 38168-001
Prelter Media 1 24375-001
Flow-Set, IR Remote Controller 1 63011
Panel Latch 3 63843
Panel Latch Key 1 63844
Optional Accessories Part No.
Eyebolts 4 62529
Hospital Grade Power Cord (12 ft/115v) 1 23888
Charcoal Prelter 1 24365-001
Ceiling Filtered Air Diffuser with 12-in Collar 1 63841
Clear, 12-in, Flexible Ducting (per/ft) 1 63842
UV-C Germicidal Light with Reector, 30 Watt 1 23466-001
Room Pressure Monitor 1 63337
Annunciator 1 63338
■Parts Location

Hospi-Gard®AirCeil™
HEPA Filtration System
15Innovators in Clean Air Technology
Installation, Operation, & Maintenance Manual
■Optional Accessories
Hospital Grade Power Cord
If purchased with an optional power cord, insert the plug into a properly grounded three pronged receptacle.
Charcoal Prelter
Carbon prelters integrate odor control capabilities to the AirCeil Unit. Carbon prelters help control the odors from cigarette
smoke, perfume, urine and much more. Carbon impregnated lters generally last 30-60 days.
Note: AirCeil Unit can accommodate either the optional charcoal prelter or the optional antimicrobial prelter, but is not
designed to accommodate both.
Antimicrobial Prelter
Antimicrobial treated polyester prelters inhibit the growth of microorganisms (85-90% resistance) on the primary HEPA lters.
These lters should be changed every 30-60 days.
Note: AirCeil Unit can accommodate either the optional antimicrobial prelter or the optional charcoal prelter, but is not
designed to accommodate both.
Ceiling Filtered Air Diffuser
• Required for use in the Total Clean Air Recirculation Mode.
• Fits in a standard 2’ x 2’ ceiling grid section.
UV-C Germicidal Light with Reector
Warning: The Charcoal Prelter must be used with the UV Germicidal Light.
Warning: Never override the safety switch located behind the front panel.
Note: As a safety precaution, the UV Light Safety Switch located just inside the Filter Access Panel door frame will automatically
inactivate the UV Light whenever the Filter Access Panel is opened.
Operating the UV Germicidal Light
• For units with the optional UV Germicidal Light with Reector, locate the “ON/OFF” toggle switch located on the Control
Panel (Figure 3.)
• Press “ON/OFF” toggle switch to the “1” position to turn the UV Light “ON”.
• Press the “ON/OFF” toggle switch to the “0” position to turn the UV Light “OFF.”
Removing the UV Germicidal Light
Caution: The unit must be decontaminated prior to opening the door panel to remove or service the UV Germicidal Light.
Caution: Conrm that the AirCeil Unit is disconnect from its power source before attempting to remove the UV Germicidal Light.
Caution: When removing the UV Light, avoid touching the white HEPA lter pleats. Even a light touch on the lter surface
may damage the lter.
1. Open the Filter Access Panel as described in the Section 6.2.4.
2. Remove the white, green and black wires from the routing clamps on the light reector.
3. Remove the all three UV light wires (white, green and black) leading from the power source to the UV ballast at the
insulated wire connectors (located near the ballast.)
4. Remove the nuts and washer from the end of the light reector channel located on the end opposite the Control Panel.
5. Remove the hardware (nut, screw and washers lock and at) from the Lamp Mount Channel.
6. Carefully, shift the UV Germicidal Light with Reector towards the lamp mount angle until it can be slid to the right (away
from the door hinge.)
Caution: Ensure that the HEPA lter media is not damaged by the UV Germicidal Light as it is removed from the AirCeil Unit.

Hospi-Gard®AirCeil™
HEPA Filtration System
16 Innovators in Clean Air Technology | www.envirco.com
Installation, Operation, & Maintenance Manual
■Appendix
AirCeil Room Changes per Hour with Varying Room Sizes
Room Size (Room Air Volume)
Fan Speed 10’ x 10’ x 8’
(800 ft3)
12’ x 12’ x 8’
(1150 ft3)
12’ x 15’ x 8’
(1440 ft3)
12’ x 15’ x 9’
(1620 ft3)
15’ x 15’ x 8’
(1800 ft3)
15’ x 20’ x 8’
(2400 ft3)
Index 30
(300 ft3/min) 22.5 ACH 15.6 ACH 12.5 ACH - - -
Index 65
(450 ft3/min) 33.8 ACH 23.5 ACH 18.8 ACH 16.7 ACH 15.0 ACH -
Index 100
(600 ft3/min) 45 ACH 31.3 ACH 25 ACH 22.2 ACH 20 ACH 15 ACH
Table 1: Representative Room Air Changes per Hour (ACH)
Calibrating the AirCeil
1. The airow index does not represent the actual airow capacity of the unit. The index represents only the percentage of
capacity at which the motor is operating.
2. The actual air volume generated by the AirCeil should be veried using a calibrated airow measurement device.
3. Determine the room air volume by multiplying the width, length and height of the room in feet.
4. In the air recirculation mode, the CDC recommends that 12 or more air changes per hour (ACH) be established.
5. Multiply the desired ACH (12 or more) by the volume of the room and divide by 60. This is the CFM setting that will provide
the desired ACH for the room.
6. Using a calibrated airow measurement device, set the AirCeil at the airow capacity.
Note: If the desired air capacity cannot be attained, an additional AirCeil may be required.
Air Volume Measurement
• Test Equipment: ShortRidge with 24” X 48” ow hood.
• Test Procedure: Place the ow hood over the intake screen. Take three sets of readings with the airow index set to 100
(full speed.)
Full Speed (airow index at 100)
Uncorrected Corrected
____________ __________
____________ __________
____________ __________
Average of Corrected __________
Acceptable Range: 600 cfm ± 10% on full speed
HEPA Filter Leak Test
Test Equipment
• “DOP” generator with Laskin type nozzles or equivalent
• Aerosol photometer with a linear or expanded logarithmic scale (calibrated)
Test Procedure
1. Set the speed setting index to 55% (5 long ashes – 5 short ashes.)
2. Place the photometer probe in the exhaust collar.
3. Move the aerosol generator hose to all possible leak locations (check the perimeter of the lter seal, the entire lter face
and the back panel of the unit).
4. Acceptance: DOP penetration not to exceed 0.03%.

Hospi-Gard®AirCeil™
HEPA Filtration System
17Innovators in Clean Air Technology
Installation, Operation, & Maintenance Manual
Positioning the AirCeil within a Patient Room
■Limited Warranty
ENVIRCO warrants that all equipment manufactured by it and bearing its name will be free from defects in materials and
workmanship under normal use. The obligation of ENVIRCO under this warranty is limited to repair or replacement of any parts
that are defective for a period of one year after invoice date, provided that ENVIRCO receives written notice of such defect. For
a period of 90 days after invoice date ENVIRCO may effect such repairs or replacement, via ENVIRCO Service personnel, at
the equipment installation site provided that the equipment is located in the continental United States and that Envirco receives
written notice of such defect. Excluded from this warranty are certain expendable items such as light tubes, lters, etc., as well
as damage due to abuse or accident. Except for this warranty ENVIRCO makes no warranty, expressed or implied, including but
not limited to, those of description, quality merchantability, sample, tness for a particular purpose or productiveness.

Clean air solutions built for you.
MORE INFORMATION AVAILABLE AT WWW.ENVIRCO.COM
Copyright © 2017 ENVIRCO. All Rights Reserved.
Form P/N#: 63963 Rev. 02/17
ENVIRCO MAC 10 is a registered trademark of ENVIRCO, USA. Specications subject to modications or changes without notice. US patents 4,560,395 and 5,470,363. Other patents issued and pending in foreign countries.
Industrial
Cleanroom Products
»MAC 10®Original
»MAC 10®IQ™
»MAC 10®IQ™ LAF
»MAC 10®LEAC™
»MAC 10®LEAC2™
»MAC 10®LEDC™
»Ducted Ceiling Module:
DCM & RSR
»AC or DC Control Systems
»MAC-T Ceiling Grid System
Hospital & Healthcare
Hospital & Healthcare
»IsoClean®and IsoClean
with Ultraviolet Light
»AirCeil®
»Hospi-Gard®Room
Pressure Monitor
Enviramedic Products
»HOR Horizontal Flow
Enclosure
»HCF Horizontal Flow
Surgery Room
»VOR Vertical Flow Surgery
Isolator
Pharmaceutical
Pharmaceutical &
Medical Device
»Unimodule M2 Vertical
Laminar Flow Workstation
»METD 100% Exhausted
Vertical Flow Workstation
»Laminar Downow Module
Laboratory & Research
Laboratory & Research
»100-Plus Horizontal or
Vertical Flow Clean Bench
»Unidirectional Flow
Horizontal Flow Bench (LF)
»TT Table Top Horizontal
Flow Clean Bench
»EnviraLab Sterility Module:
ESM
ENVIRCO®
101 McNeill Road | Sanford, NC 27330, USA
tel: 919.775.2201 | toll free: 800.884.0002 | fax: 800.458.2379
Table of contents
Other Envirco Air Cleaner manuals