Epilog Laser Fusion Pro 32 Use and care manual

www.epiloglaser.com
tech@epiloglaser.com
303-215-9171
X-Axis Rail Assembly Replacement
Fusion Pro 32 & 48
Parts Required Tools Required
• Pro 32 - CS1200 - X-Axis Rail Assembly
• Pro 48 - CS0400 -X-Axis Rail Assembly
• 5/32” hex wrench
• 5/16” nut driver or wrench
• Phillips head screwdriver
9/2022

PRO 32 & 48 X-AXIS ASSEMBLY
2
X-Axis Assembly Removal
1. Turn o the engraver.
2. Disconnect engraver from power source.
3. Remove the le and right panels of the
engraver by loosening the 5/32” cam
locks:
4. Manually move the x-axis assembly to
the middle of the engraver:
5. Loosen the Phillips screw on the lens
tube retaining collar. Make sure to hold
the lens tube to ensure that it does not
fall:
6. Remove the lens tube:
7. Remove the le and right bellows
by loosening the six (6) 5/16” nuts and
backing plate:

PRO 32 & 48 X-AXIS ASSEMBLY
3
8. Loosen, but do not remove the four (4)
Phillips screws that hold the x-axis drive
board cover:
9. Remove the x-axis drive board cover.
10. Disconnect the seven (7) electrical
connectors from the drive board:
12. Disconnect the air tubing from the
elbow connector at right side of the
x-axis assembly:
The air tubing is held by a retaining fixture.
To remove the tubing, press the tubing into
the fixture. Then, while holding the orange
retaining ring tight to the fixture pull the air
tubing away from it.
13. Loosen and remove the four (4) 7/64” hex
screws that hold the x-axis drive board
housing:

PRO 32 & 48 X-AXIS ASSEMBLY
4
16. Li the x-axis assembly upwards to clear
the locator pins on the le and right
y-axis bearing blocks:
17. Slide the x-axis assembly out of the
engraver through the right side of
the machine, taking care to clear the
carriage assembly through the opening
without damage:
14. Remove the x-axis drive board housing:
15. Remove and or loosen the four (4)
5/32” fasteners that secure the x-axis
assembly to the y-axis bearings. There
are two (2) on the right side and two (2)
on the le side:

PRO 32 & 48 X-AXIS ASSEMBLY
5
X-Axis Assembly Installation
1. Slide the x-axis assembly into the
engraver through the right side of
the machine, taking care to clear the
carriage assembly through the opening
without damage:
2. Place the x-axis assembly on the y-axis
bearing blocks, ensuring that the
locator pins enter the upper bearing
block assembly:
3. Install and/or tighten the four (4)
5/32” fasteners that secure the x-axis
assembly to the y-axis bearings. There
are two (2) on the right side and two (2)
on the le side:
4. Position the x-axis drive board housing
over the mounting holes:

PRO 32 & 48 X-AXIS ASSEMBLY
6
7. Reconnect the seven (7) electrical
connectors from the drive board:
The electrical connectors on the drive
board are all unique and will only fit into
one receptacle.
8. Install the x-axis drive board cover and
tighten the four (4) Phillips screws:
9. Install the backing plate and bellows
and tighten the six (6) 5/16” nuts that hold
them in place:
5. Install and tighten the four (4) 7/64” hex
screws that hold the x-axis drive board
housing:
6. Reconnect the air tubing to the elbow
fitting at the right side of the x-axis
assembly:
To reconnect the air tubing into the retain-
ing fixture, simply insert the tubing into the
collar. To ensure that it is securely connect-
ed, gently pull on the tubing.

PRO 32 & 48 X-AXIS ASSEMBLY
7
10. Position the lens tube in the lens
assembly, ensuring that the notch in
the tube faces the rear of the engraver.
Make sure to hold the lens tube aer
inserting to ensure that it does not fall:
11. Tighten the Phillips screw on the lens
tube retaining collar:
12. Replace both the le and right panels
of the engraver.
13. Reconnect the engraver to power and
boot up the machine, ensuring proper
boot sequence.
Updates and Calibrations
The x-axis assembly contains many
components, including circuit boards,
cameras, and a motor. Further procedures
are required.
Updating the Carriage Camera
1. On the display, press the gear icon to
enter the Settings menu:
2. Press and hold the word “Settings” that
appears at the touch of the touchpad
to gain access to the Advanced
Settings menu:

PRO 32 & 48 X-AXIS ASSEMBLY
8
3. Enter the Diagnostics menu and scroll to
the Camera tab. Press Test & Update:
4. Press Update:
5. Allow the engraver to complete the
update. Once done, a success message
will appear:
Updating the Peripherals
1. At the display panel, press the gear icon
in the upper right-hand corner of the
display to open the Settings menu:

PRO 32 & 48 X-AXIS ASSEMBLY
9
2. Press Program to enter the Peripheral
Update menu:
3. Once in the Peripheral Update menu,
select Motor Drives and FPGA. Press
Update:
When running a peripheral update, do not
power o the engraver. During a peripher-
al update it is common for the display to
power o. Once the update has been ap-
plied, the engraver will prompt to restart.
4. Once the engraver prompts you to
restart, power cycle the engraver.
Calibrating the Focus
Focus Calibrations for CO2- Only
Engravers
Focus calibration must be completed in
the order outlined in this procedure.
Table Calibration
The table calibration is a crucial step in
the laser focus calibration. During this
procedure, the table moves from the top
to the bottom of the z-axis, locating the
magnetic sensors along the z-axis PCB. This
step must be completed for a successful
and accurate focus calibration.
Before completing the table calibration
install the raster table.
Since the table will rise to depress the
autofocus plunger, use the Jog feature
to position the lens carriage in a position
which will prevent the autofocus plunger
and lens cone from coming into contact
with the le and or top ruler guides.
1. On the display, press the gear icon to
enter the Settings menu:

PRO 32 & 48 X-AXIS ASSEMBLY
10
2. Once in the Settings menu, press
the Focus button to enter the Focus
Commands/Calibrations menu:
3. Press the Table button:
4. Press the Start button to start the table
calibration:
4. Allow the engraver to complete the
calibration. Once complete, a success
message should appear.
Move to the CO2 Focus Procedure.
CO2 Focus Calibration
The CO2 focus calibration establishes
the focal height of the laser and manual
focus gauge. You will run a small job while
adjusting the table height to find the
correct focal height.
A small piece of anodized aluminum (at
least 3” x 3” or 75mm x 75mm) is required to
complete this procedure.
1. Open your preferred illustrating
program and create a black, raster box
measuring ~2” x 2” or 50mm x 50mm.
2. Send the job to the Soware Suite.
3. Ensure that the process shows as an
engraving job and set the both power
and speed at 5-10%.
If completing the procedure on the Fusion
Pro 32, place the artwork and material
in the upper le corner of the engraving
table, ensuring that the artwork appears
over the anodized aluminum.
If completing the procedure on the Fusion
Pro 48, place the artwork and material in
the center of the engraving table, ensuring
that the artwork appears over the anod-
ized aluminum.
4. Send the job to the engraver.

PRO 32 & 48 X-AXIS ASSEMBLY
11
5. Once the job is running, press the Focus
button and slightly raise and lower the
table with the joystick while observing
the spark coming from the laser hitting
the anodized aluminum:
6. While raising and lowering the table,
find the height where the spark appears
the brightest. If it is diicult to determine,
turn o any overhead lights in the room.
7. Once the appropriate table height is
found, use the touchpad to lower the
power to 3-5%:
8. Continue to raise and lower the table
height with the lower power setting to
find the height where the spark appears
the brightest.
Using the lower power setting allows you to
more finely dial in the correct focal height
as the spark fades very quickly when the
material becomes out of focus.
9. Once the correct focal height has been
found, let go of the joystick and press
the Go/Stop button to stop the job. Then
press the Reset button to return the
laser head back to the home position.
10. Remove the anodized aluminum from
the engraver.
11. On the display, press the gear icon to
enter the Settings menu:
12. Once in the Settings menu, press
the Focus button to enter the Focus
Commands/Calibrations menu:

PRO 32 & 48 X-AXIS ASSEMBLY
12
13. Press the CO2 Focus button:
14. Ensure that the material thickness used
during the previous step is entered in
the Thickness box. If using anodized
aluminum provided by Epilog Laser, the
default value of 0.025” corresponds to
the material.
15. Press the Start button. This calibration is
very fast as the engraver simply notes
and saves the current position of the
table.
16. Press Done.
Move to the Autofocus Calibration
procedure.
Autofocus Calibration
The autofocus calibration takes note of,
saves, and uses the CO2 focus calibration
value to ensure that the table moves to
the correct focal height when running a
job with the autofocus enabled. During this
calibration, the table will rise up, depress
the autofocus plunger, and then move
back to the appropriate focal height.
Before completing the table calibration
install the raster table.
Since the table will rise to depress the
autofocus plunger, use the Jog feature
to position the lens carriage in a position
which will prevent the autofocus plunger
and lens cone from coming into contact
with the le and or top ruler guides.
1. On the display, press the gear icon to
enter the Settings menu:

PRO 32 & 48 X-AXIS ASSEMBLY
13
2. Once in the Settings menu, press
the Focus button to enter the Focus
Commands/Calibrations menu:
3. Press the AF Plunger button:
4. Press the Start button.
5. Allow the engraver to complete the
calibration. Once complete, a success
message should appear.
Focus Calibrations for
Fiber- Only Engravers
Focus calibration must be completed in
the order outlined in this procedure.
Table Calibration
The table calibration is a crucial step in
the laser focus calibration. During this
procedure, the table moves from the top
to the bottom of the z-axis, locating the
magnetic sensors along the z-axis PCB. This
step must be completed for a successful
and accurate focus calibration.
Before completing the table calibration
install the raster table.
Since the table will rise to depress the
autofocus plunger, use the Jog feature
to position the lens carriage in a position
which will prevent the autofocus plunger
and lens cone from coming into contact
with the le and or top ruler guides.
1. On the display, press the gear icon to
enter the Settings menu:

PRO 32 & 48 X-AXIS ASSEMBLY
14
2. Once in the Settings menu, press
the Focus button to enter the Focus
Commands/Calibrations menu:
3. Press the Table button:
4. Press the Start button to start the table
calibration:
5. Allow the engraver to complete the
calibration. Once complete, a success
message should appear.
Move to the Fiber Focus Procedure.
Fiber Focus Calibration
The CO2 focus calibration establishes
the focal height of the laser and manual
focus gauge. You will run a small job while
adjusting the table height to find the
correct focal height.
A small piece of anodized aluminum (at
least 3” x 3” or 75mm x 75mm) is required to
complete this procedure.
1. Open your preferred illustrating
program and create a black, raster box
measuring ~2” x 2” or 50mm x 50mm.
2. Send the job to the Soware Suite.
3. Ensure that the process shows as an
engraving job and set the both power
and speed at 5-10%.
If completing the procedure on the Fusion
Pro 32, place the artwork and material
in the upper le corner of the engraving
table, ensuring that the artwork appears
over the anodized aluminum.
If completing the procedure on the Fusion
Pro 48, place the artwork and material in
the center of the engraving table, ensuring
that the artwork appears over the anod-
ized aluminum.
4. Send the job to the engraver.

PRO 32 & 48 X-AXIS ASSEMBLY
15
5. Once the job is running, press the Focus
button and slightly raise and lower the
table with the joystick while observing
the spark coming from the laser hitting
the anodized aluminum:
6. While raising and lowering the table,
find the height where the spark appears
the brightest. If it is diicult to determine,
turn o any overhead lights in the room.
7. Once the appropriate table height is
found, use the touchpad to lower the
power to 3-5%:
8. Continue to raise and lower the table
height with the lower power setting to
find the height where the spark appears
the brightest.
Using the lower power setting allows you to
more finely dial in the correct focal height
as the spark fades very quickly when the
material becomes out of focus.
9. Once the correct focal height has been
found, let go of the joystick and press
the Go/Stop button to stop the job. Then
press the Reset button to return the
laser head back to the home position.
10. Remove the anodized aluminum from
the engraver.
11. On the display, press the gear icon to
enter the Settings menu:
12. Ensure that the material thickness used
during the previous step is entered in
the Thickness box. If using anodized
aluminum provided by Epilog Laser, the
default value of 0.025” corresponds to
the material.
13. Press the Start button. This calibration is
very fast as the engraver simply notes
and saves the current position of the
table.
14. Press Done.
Move to the Autofocus Calibration
procedure.

PRO 32 & 48 X-AXIS ASSEMBLY
16
Autofocus Calibration
Before completing the table calibration
install the raster table.
Since the table will rise to depress the
autofocus plunger, use the Jog feature
to position the lens carriage in a position
which will prevent the autofocus plunger
and lens cone from coming into contact
with the le and or top ruler guides.
1. On the display, press the gear icon to
enter the Settings menu:
2. Once in the Settings menu, press
the Focus button to enter the Focus
Commands/Calibrations menu:
3. Press the AF Plunger button:
4. Press the Start button.
5. Allow the engraver to complete the
calibration. Once complete, a success
message should appear.
Focus Calibration for Dual
Source Engravers
Focus calibration must be completed in
the order outlined in this procedure.
Table Calibration
The table calibration is a crucial step in
the laser focus calibration. During this
procedure, the table moves from the top
to the bottom of the z-axis, locating the
magnetic sensors along the z-axis PCB. This
step must be completed for a successful
and accurate focus calibration.

PRO 32 & 48 X-AXIS ASSEMBLY
17
Before completing the table calibration
install the raster table.
Since the table will rise to depress the
autofocus plunger, use the Jog feature
to position the lens carriage in a position
which will prevent the autofocus plunger
and lens cone from coming into contact
with the le and or top ruler guides.
1. On the display, press the gear icon to
enter the Settings menu:
2. Once in the Settings menu, press
the Focus button to enter the Focus
Commands/Calibrations menu:
3. Press the Table button:
4. Press the Start button to start the table
calibration:
5. Allow the engraver to complete the
calibration. Once complete, a success
message should appear.
Move to the Fiber Focus Procedure.
Fiber Focus Calibration
The CO2 focus calibration establishes
the focal height of the laser and manual
focus gauge. You will run a small job while
adjusting the table height to find the
correct focal height.
A small piece of anodized aluminum (at
least 3” x 3” or 75mm x 75mm) is required to
complete this procedure.

PRO 32 & 48 X-AXIS ASSEMBLY
18
1. Open your preferred illustrating
program and create a black, raster box
measuring ~2” x 2” or 50mm x 50mm.
2. Send the job to the Soware Suite.
3. Ensure that the process shows as an
engraving job and set the both power
and speed at 5-10%.
If completing the procedure on the Fusion
Pro 32, place the artwork and material
in the upper le corner of the engraving
table, ensuring that the artwork appears
over the anodized aluminum.
If completing the procedure on the Fusion
Pro 48, place the artwork and material in
the center of the engraving table, ensuring
that the artwork appears over the anod-
ized aluminum.
4. Send the job to the engraver.
5. Once the job is running, press the Focus
button and slightly raise and lower the
table with the joystick while observing
the spark coming from the laser hitting
the anodized aluminum:
6. While raising and lowering the table,
find the height where the spark appears
the brightest. If it is diicult to determine,
turn o any overhead lights in the room.
7. Once the appropriate table height is
found, use the touchpad to lower the
power to 3-5%:
8. Continue to raise and lower the table
height with the lower power setting to
find the height where the spark appears
the brightest.
Using the lower power setting allows you to
more finely dial in the correct focal height
as the spark fades very quickly when the
material becomes out of focus.
9. Once the correct focal height has been
found, let go of the joystick and press
the Go/Stop button to stop the job. Then
press the Reset button to return the
laser head back to the home position.
10. Remove the anodized aluminum from
the engraver.

PRO 32 & 48 X-AXIS ASSEMBLY
19
11. On the display, press the gear icon to
enter the Settings menu:
12. Ensure that the material thickness used
during the previous step is entered in
the Thickness box. If using anodized
aluminum provided by Epilog Laser, the
default value of 0.025” corresponds to
the material.
13. Press the Start button. This calibration is
very fast as the engraver simply notes
and saves the current position of the
table.
14. Press Done.
The CO2 focus must be set to the same
height as the fiber focus before moving
onto the CO2 Focus Calibration. Failure to
complete this step will result in an inaccu-
rate focus calibration.
15. Press the CO2 Focus button:
16. Ensure that the material thickness used
during the previous step is entered in
the Thickness box. If using anodized
aluminum provided by Epilog Laser, the
default value of 0.025” corresponds to
the material.
17. Press Done.
CO2 Focus Calibration
The CO2 focus calibration establishes
the focal height of the laser and manual
focus gauge. You will run a small job while
adjusting the table height to find the
correct focal height.
A small piece of anodized aluminum (at
least 3” x 3” or 75mm x 75mm) is required to
complete this procedure.
1. Open your preferred illustrating
program and create a black, raster box
measuring ~2” x 2” or 50mm x 50mm.
2. Send the job to the Soware Suite.
3. Ensure that the process shows as an
engraving job and set the both power
and speed at 5-10%.
If completing the procedure on the Fusion
Pro 32, place the artwork and material
in the upper le corner of the engraving
table, ensuring that the artwork appears
over the anodized aluminum.

PRO 32 & 48 X-AXIS ASSEMBLY
20
If completing the procedure on the Fusion
Pro 48, place the artwork and material in
the center of the engraving table, ensuring
that the artwork appears over the anod-
ized aluminum.
4. Send the job to the engraver.
5. Once the job is running, press the Focus
button and slightly raise and lower the
table with the joystick while observing
the spark coming from the laser hitting
the anodized aluminum:
6. While raising and lowering the table,
find the height where the spark appears
the brightest. If it is diicult to determine,
turn o any overhead lights in the room.
7. Once the appropriate table height is
found, use the touchpad to lower the
power to 3-5%:
8. Continue to raise and lower the table
height with the lower power setting to
find the height where the spark appears
the brightest.
Using the lower power setting allows you to
more finely dial in the correct focal height
as the spark fades very quickly when the
material becomes out of focus.
9. Once the correct focal height has been
found, let go of the joystick and press
the Go/Stop button to stop the job. Then
press the Reset button to return the
laser head back to the home position.
10. Remove the anodized aluminum from
the engraver.
11. On the display, press the gear icon to
enter the Settings menu:
12. Once in the Settings menu, press
the Focus button to enter the Focus
Commands/Calibrations menu:
Other manuals for Fusion Pro 32
5
This manual suits for next models
1
Other Epilog Laser Industrial Equipment manuals
Popular Industrial Equipment manuals by other brands

Lissmac
Lissmac FBM 300 P operating manual

HPSi
HPSi 500 Operator, maintenance and parts manual

Dustcontrol
Dustcontrol RAF 2501 Original instructions

ABB
ABB HT609904 Operation manual

Syntec
Syntec R41-LXC100-917Y product manual

Maguire Products
Maguire Products GRAVIMETRIC AUGER FEEDER MGF-ST Original instruction manual

Challenge
Challenge CMT 330 Installation and operator's manual

morse
morse CP-305-1-220-50 quick start guide

Smake
Smake EPS Jump Assembly manual

Minebea Intec
Minebea Intec PR 6246/12 installation manual

KTR-Group
KTR-Group KTR-STOP S F B Series Operating & assembly instructions

stellar labs
stellar labs TMAX1 owner's manual