ETH-messtechnik VS-PH10-10000 User manual

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 1 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
Manual
Testing device for
Hydraulic tools
Type VS-PH10-10000
Optional as passive or active transducer

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 2 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
Content: Page:
1. Important notes 3
2. Description 3
3. Features 3
4. Instructions for use 4
5. Conditions at the site 4
5.1 Ambient temperature 4
5.2 Humidity and dust 4
5.3 Chemical influences 4
6. Structure of the measuring chain 5
7. Connecting 5
7.1 Notes for wiring 5
7.2 Connectors 5
7.3 Cable extension 5
7.3.1 Passive version 5
7.3.2 Active version 5
8. Measuring signals 6
9. Quick start guide GMV2 6
10. Maintenance for rotating screwdriving tools 10
11. Installation of spare parts 10
12. Conversion for rotating screw tools 10
13. Conversion for hydraulic screw tools 13
14. Assembly and disassembly instructions counter holder 13
15. Hazard warnings 14
16. Technical data see data sheet
Drawing rotating tools (F.551.00. 005) 15
Drawing hydraulic tools (F.551.00. 006) 16
Drawing screw simulator (F.262.00.005) 1

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 3 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
1. Important notes:
In the event of recognizable damage, the simulator may no longer be used. Use only original spare
parts.
Torque peaks beyond the permissible measuring range can lead to the destruction of the thread.
Contamination of the nut and the M64 screw should be avoided.
Note that the transducers are not designed with the usual safety factors (2...20) in favor of a high
sensitivity.
In particular, consider the specified overload factors.
The torque transducer is not approved for use in explosive areas.
Protect transducers from moisture and dust deposition (protection class IP 40).
The opening of the transducer may only be carried out by specialist personnel.
When opening or dismantling the transducer within the warranty period, the warranty claim
expires.
Only the cables provided for this purpose may be connected to the transducer. Damage caused by
incorrect connection cables is not covered by the warranty.
An extension of the passive transducer cable causes measurement errors, only the original cable
may be used.
2. Description
With the test device VS-PH it is easy to preset rotary screwdrivers and hydraulic screwdriving tools, create torque
tables or check the power range of the tool.
Together with the evaluation device type GMV2, the determined torque values can be displayed and
documented.
Simulator for clockwise screw connections.
Tools with connecting square according to DIN 3121 can be plugged in directly.
For the adaptation of the various tools we supply the desired adapters.
Various adapters are available for tools with hexagonal output. Special adapter on request.
3. Characteristics
•Test equipment monitoring in accordance with DIN EN ISO 9000ff
•Setting and testing of tools before screwing
•Creating Torque Tables
•Standard square according to DIN 3121-H 25 & 40 (1" & 1 1/2 ")
•Replaceable support blocks for the screwdriver's counter-holder
•Portable test facility
•Housing made of high-strength aluminium
•Housing surface is hard anodized for protection.
•EMC-shielded housing

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 4 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
4. Application notes
The torque transducers measure in both right and left loads. For right load, the output signal is positive, with left load
negative. The type plate provides information about the measuring range end value.
The screwing simulator can only be used for clockwise running.
As a precision measuring instrument, the transducer requires careful handling during transport and use, as impacts or
droppings can damage the transducer. Torque peaks beyond the permissible overload can lead to the destruction of
the torsional shaft.
The limits for permissible mechanical, thermal and electrical loads are listed in the Test Device data sheet for
hydraulic tools type PH. It is essential that they are respected.
Please take this into account when planning the measuring arrangement, and finally during operation.
Shielded cables shall be used for electrical connection.
ETH-Torque transducers have been upgraded to their electromagnetic compatibility (EMC) in accordance with EN
55011/03.91 (DIN VDE 0875-11/07.92) or prEN 50082-2/1993 (E DIN EN 50082-2/03.94).
5. Conditions at the site of operation
5.1 Ambient Temperature
In order to achieve optimum results, the nominal temperature range must be maintained.
Constant, at most slowly changing, temperatures are best. The specified temperature errors apply if the
temperature does not change faster than 5°K/h. Unilateral radiant heat or cooling should be avoided and, if
necessary, appropriate measures taken.
5.2 Moisture and dust
The transducers comply with the protection class IP40 according to DIN 40050
If the test device is fogged by large temperature changes, it must not be put into operation, as the moisture is
involved in the measurement.
In the case of active version, the humidity during operation can lead to the destruction of the measuring
amplifier.
Dust, dirt and other foreign objects must not accumulate in such a way that they can penetrate the sensor or the
connector.
Dirt deposits in the plug lead to measurement errors
Note: Do not enter moisture into the transducer's connector!
5.3 Chemical influences
The transducers are not protected against chemical influences. They cannot be used in aggressive environments.
The torque transducer is not approved for use in hazardous areas.

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 5 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
6. Structure of the measuring chain
In order to be able to measure with the transducer, the construction of a complete measuring chain is required.
This consists of:
Torque transducer
Connecting cable
Supply and evaluation device
a registration device, e.g. printer (if required)
7. Connecting
7.1 Notes for wiring
Electric and magnetic fields often cause the coupling of interference voltages into the measuring chain.
These faults are primarily caused by high-voltage power lines parallel to the measuring line, but also from nearby
contactors or electric motors.
In addition, interference voltages can be coupled by galvanic means, in particular by grounding the measuring
chain at several points, so that potential differences occur.
Observe the following notes
Use only shielded and low-capacity measuring cables
(ETH measuring cables meet these conditions).
do not install the measuring cable in parallel with high current or control lines.
Scattering fields of transformers, motors and fuses must be avoided.
transducer, evaluation and display device do not erase multiple times.
Connect all devices in the measuring chain to the same protective conductor.
7.2 Connectors
The transducers are equipped with a 14-pin (passive version) or 12 pin (active version) built-in plug type Binder
Series 680.
7.3 Cable Extension
Extension cables must be shielded and low in capacity.
We recommend the use of ETH-cables that meet these requirements.
For cable extensions, care must be taken to ensure proper connection and good insulation. Care must be taken to
ensure that the cable cross-section is sufficiently large to ensure sufficient supply voltage on the transducer.
7.3.1 Passive version
A cable extension is involved in the measurement, since the internal resistance of the measuring cable and
that of the transducer form a voltage divider. This can be avoided by using a supply and evaluation device
with sensor lines.
7.3.2 Active version
A cable extension is not involved in the measurement if the supply and evaluation device have a galvanic
separation of the supply voltage of the transducer and the measuring signal amplification.
(The supply and evaluation device type GMV2 has a galvanic isolation).

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 6 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
8. Measurement signals
The transducer provides a torque-proportional voltage signal. The respective measurement signal size can be found in
the type plate or data sheet.
For right load, the output voltage is positive, with left load negative.
9. Description GMV2
General information
The GMV2 is a microprocessor-controlled measuring, control and evaluation device for the most common applications
in bolting technology. Thanks to its compact design and optional battery operation, it is very portable.
Measurements can be saved and printed with date and time. The automatic shutdown of the screwdriver is possible
via a separate power part.
Characteristics
•Menu-guided operation and setting
•Detection of torque, angle of rotation, speed and power
•For torque transducers from 1 Ncm to 100 kNm final value
•Shutdown according to torque and angle of rotation
•1000 measured value memory
•50 parameter set memory
•Optional 8 hours battery operation, charger built-in or mains operation
•RS-232 Interface for Printers
•Measured value transfer to PC
•EMC-dense housing
Commissioning
Important notes
By opening or dismantling the device within the warranty period, the warranty claim expires.
The device may only be opened by qualified personnel.
Only the cables provided for this purpose may be connected to the device. Damage caused by incorrect connecting
cables is not covered by the warranty.
The device is not grounded when running on a battery. If current-carrying devices are connected, they must be
properly grounded.
The final value of the transducer shown in the display of the device must always match the transducer used.
Where people and property can be harmed, appropriate safety measures (e.g. covers, overload protections) must be
taken by the user (observe relevant accident prevention regulations!).
The device is not approved for use in explosive areas.

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 7 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
The front page at a glance
E1 1 CW peak SR: 995
↓8,10
Nm
P: 1 ST: 5 DR: 1 10 Nm
Zeroadj
Test
Menu
NIO ↓12 Grad
Error:
Charg
ON/
OFF
Back
Menu
buttons
Button
without
function
Transducer end
value
Torque
Rotation angle or speed
Parameter set
number
Field for
error messages
Data
record
Memory number
Type of
measuremen
t
Direct function
Zeroadj / Clear
Direct function
Transducer test
Menu tree
Remaining Memory
(number of free memory spaces)
Assessment of the
Assessment of torque
Overall
t
Peak value status
Password level
n= 5
Number of measured
values

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 8 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
The back at a glance
Connection of the torque transducer
For different types of torque transducers there are the appropriate connection cables!
This GMV2 version is intended for the use of active transducers with ± 5V. A passive input is optionally available.
Please use only the designated cables that you can obtain from ETH.
Connection of the power section
On the back of the device is the 25-pin socket "IN / OUT" which is intended, among other things, for connecting the
power section. Before switching on the device, the wrench should be switched off.
Turning on
After checking the correct connection of the cables, you can turn on the device with the button. After a short self-
test, all settings that were valid before switching off are restored. Now a zeroing of the transducer is carried out. After
that, the device is ready for measurement.
If the error message "charge!" appears, please connect the power cord. You can now work in mains or turn off the
device during the charging period of about five hours.
Notes:
−When the device is switched on, the outputs switch on briefly. Therefore, the hand switch of the wrench should be
switched off.
−Before the first measurement, you should check whether the basic settings of the device meet your requirements.
−If the battery is not charged 15 minutes after the first charge notification, the device itself will turn off.
Turning off
The device can only be switched off in level 0 (when the torque is displayed) with the button !
If necessary, the back button must be pressed beforehand until the torque is displayed.
If a defect causes the device to stop being operated or switched off, the fuse on the back can be removed. In this case,
however, data loss is not excluded.

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 9 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
Battery operation
The device is optionally equipped with a battery. Since the torque sensor is also supplied, it is possible to carry out
measurements independently of the power grid.
The built-in lead-gel battery is maintenance-free, overcharge protected and is automatically recharged in (indefinite)
mains operation. A full charge takes a maximum of 8 hours during the measurement operation, and a maximum of 5
hours when switched off. After that, a continuous measurement operation of at least 9 hours is possible.
The GMV2 has a battery monitor that displays the error message "Charge!" every minute and a beep also sounds. At
the bottom right of the display, the message remains constantly displayed. You should then charge the battery as
soon as possible (further operation is possible with power cord).
Notes:
−If this does not happen and the device is in measuring mode, it switches off after 15 minutes!
−Caution: If the device is in the menu level, it can no longer turn itself off, which can cause the battery to be deeply
discharged and its performance compromised.
−In case of prolonged storage or non-use (1/2 year), the battery must be recharged.

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 10 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
10. Maintenance for rotating screw tools
(Drawing-No.: F. 551.00.005; F. 262.00.005)
In the event of stiffness or a noticeable lack of grease, the thread of the
Screw M64 [pos. 16], the bearing [pos. 25], nut [pos. 18] and the guide pins [pos. 20] are to be greased with a MoS2-
Grease, suitable for high pressure & shock loads.
•To remove the hood [pos. 02] first loosen the hook clamp [pos. 03], then remove the hood [pos. 02]
and the machine support ring [item 01].
•Unscrew the M64 screw [pos. 16] counterclockwise and grease.
•When installing the screw M64 [pos. 16], care must be taken that the thread is not bent, therefore
always turn by hand.
•Loosen the fastening screws [pos. 27] for removing the bearing, remove the disc [pos. 29] and the
spacer sleeve [pos. 21], grease the housing locating washer [pos. 26] and the bearing [pos. 25].
•Remove the housing [pos. 17] and grease the guide pins [pos. 20], make sure that
adapter-4kt [pos. 19] does not fall out.
•Grease the nut [pos. 18], paying attention to its installation position.
11. Installation of spare parts
•When installing the disc springs, care must be taken, to ensure that the disc springs are layered as described in
drawing F.262.00.005. Incorrect layering can damage the disc springs.
•When replacing the M64 screw [pos. 16], the thread must be greased with a mounting paste.
12. Conversion for rotating screw tools
(Drawing-No.: F.551.00.005; F. 551.00.006 & F.262.00.005)
•Disassemble the counter holder [pos.
13].
(See assembly and disassembly
instructions for counter holders on
page 14.)
→
Image 1
•Mount the base plate including gear
[pos. 07] to VS-PH10.
•When setting up, pay attention to the
position of the milling (image 2).
•Screws M14x30 (4x) [pos. 06]
•Tightening torque 135 Nm.
Image 2

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 11 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
Image 3
•Set up the adapter 4kt [pos. 05].
Image 4
•Put the screwing simulator [pos. 04]
on completely.
•Tools: Eye bolt [pos. 28] with spacer
bolt [item 22]. Dismantling both after
set up!
Image 5

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 12 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
•Put on cover [pos. 02]
•Align the hook clamps [pos. 03] to the
long side of the case so that they reach
into the hook pocket on the bottom of
the base plate.
•Close the hook clamps.
Image 6
•Set up the machine support ring
[pos. 01].
Image 7

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 13 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
•Put on the screwdriver, let it run
clockwise until the teeth and square
drive match and the machine fully
engages the teeth.
Image 8

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 14 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
13. Conversion for hydraulic screw tools
(Drawing-No.: F.551.00.005 & F.551.00.006)
•Remove the machine support ring [pos. 01] (page 12, image 7).
•Open the hook clamps [pos. 03] and remove the cover [pos. 02] (page 11 image 6).
•Remove the screwing simulator [pos. 04], taking care that the
Adapter 4kt [pos. 05] does not fall out (page 11 images 4 & 5).
•Remove the adapter-4kt [pos. 05] (page 11 fig. 4).
•Loosen the screws M14x30 [pos. 06] and remove the base plate including gear [pos. 07] (page 10,
image 2& 3). Note! The gear is screwed and pinned to the base plate. Please do not dismantle them
from each other.
•Mount the counter holder [pos. 13] (page 10, image 1).
(See assembly and disassembly instructions for counter holders, page 14)
14. Assembly and disassembly instructions counter holders
(Drawing-No. F.551.00.006)
Assembly:
1. Grease cylinder pins 20m6x100 DIN 6325 [pos. 14] and drive them into counter holder [pos. 13] until stopper.
Beware! Do not mount in the PH first.
2. Grease the headed press fit bushes.
3. Put on the counter holder [pos. 13] and drive it evenly with a plastic hammer.
4. Put the washers M16 DIN433 [pos. 12] on cap screws M16x80 DIN 912 8.8 [pos. 11], grease the threads, turn
evenly and tighten with 200Nm.
5. Close all holes with the appropriate cover caps [pos. 09 & 10].
Disassembly:
1. Remove the 3 large cover caps [pos. 09].
2. Loosen and remove cap screws M16x80 DIN 912 8.8 [pos. 11].
3. Screw a cap screw M16x80 DIN 912 8.8 [pos. 11] into the middle thread, whereby the counter holder is
pressed. In case of heavy-duty or edging, it may be easy to knock off with a plastic hammer.

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 16 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01
15. Warnings
! The screw M64 [pos. 16] has no stop in the dissolving mode (left-hand rotation),
i.e. the machine support ring [pos. 01] is pressed upwards.
Remedy: Stop the machine before or turn it by hand until the gear is flat before the
new measurement process.
! When removing the transducer [pos. 04] using the eye bolt [pos. 28] it is essential to ensure
that the adapter square [pos. 05] only rests loosely and has to be removed separately.
!If the test specimen or test device is touched during the operation of the test facility, there is
an increased risk of injury!
!Before each commissioning, the fixed seat of the test specimen shall be checked on the test
device. Furthermore, care must be taken to ensure that the counter-holder finds a correct grip
on the start-up block. If this is not the case, a suitable start-up block must be installed.
!Where people and objects can be harmed in the event of a break, appropriate safety measures
(e.g. covers, overload protections) must be taken by the user (observe relevant accident
prevention regulations!).
!Note that the transducers are not designed with the usual safety factors (2...20) in favor of
high measurement sensitivity. In particular, consider the overload factors specified in the
datasheet.
!If the test device is fitted by large temperature changes, it must not be put into operation, as
the moisture is involved in the measurement. In the case of active version, the humidity during
operation can lead to the destruction of the measuring amplifier.
10. Technical data:
Please refer to the technical data in the data sheet
Test device for hydraulic tools type PH.

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 17 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 19 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01

ETH-messtechnik gmbh D-74417 Gschwend Hagstr.10
N°: 094 Date: 12.05.2022 Page 20 Of 17
Subject to technical changes BA094_VS-PH10_EN_Rev01