Evco EVC30S40 User manual

Evco S.p.A. • Code 104C30S40E02 • page 1/6
EVC30S40 / EVC70S40 2 output digital modules for electric bread and pizza ovens version 1.02
E ENGLISH
1 IMPORTANT
1.1 Important
Read these instructions carefully before installation and commissioning and follow all recommendations for installation and for the electric
connection; keep these instructions with the instrument for future consultation.
The instrument must be disposed of according to local Standards regarding electric and electronic appliances.
2 DIMENSIONS AND INSTALLATION
2.1 Dimensions and installation
Dimensions in mm (in); back-panel installation using M3 studs.
The diagram illustrates the model with the greatest supply of hardware.
2.2 Recommendations for installation
• make sure that the working conditions (temperature of use, humidity, etc.) lie within the limits indicated in the technical data
• do not install the instrument in proximity of heat sources (heaters, hot air pipes etc.) appliances withe strong magnets (large diffusers etc.), places
subject to direct sunlight, rain, humidity, excessive dust, mechanical vibrations or shocks
• in compliance with Safety Standards, the protection against any contact with the electric parts must be ensured via correct installation of the
instrument. All parts that ensure protection must be fixed in a way such that they cannot be removed without the aid of a tool.
3 ELECTRIC CONNECTION
3.1 Electric connection
The diagram illustrates the model with the greatest supply of hardware.
3.2 Recommendations for installation
• do not operate on the terminal boards using electric or pneumatic
screwdrivers
• if the instrument has been taken from a old place to a hot one, the
humidity could condense inside. Wait about one hour before
applying power
• make sure that the power supply voltage, frequency and operational
electric power correspond to those of the local power supply
•disconnect the power supply before performing any type of
maintenance
• equip the probes with a protection able to insulate them against any
contact with metal parts or use isolated probes
• do not use the instrument as a safety device
• for repairs and information regarding the instrument, contact the
Evco sales network.
4 MODELS AVAILABLE
4.1 Models available
With reference to the electric connection illustrated in paragraph 3.1.
CODE FUNCTION PROBE 1 PROBE 2 LOAD 1 LOAD 2 NUMBER OF KEYS
(ENCODER YES/NO)
EVC30S40J7XXX00 simple regulation chamber probe not used regulation of the not used 0
of the temperature temperature (yes)
EVC30S40J7XXX01 simple regulation chamber probe not used regulation of the auxiliary 4
of the temperature temperature (no)
EVC30S40J7XXX02 top-floor top probe floor probe top floor 4
regulation (no)
of the temperature
EVC30S40J7XXX03 top-floor top probe floor probe top floor 6
regulation (no)
of the temperature
EVC30S40J7XXX04 top-floor top probe floor probe top floor 0
regulation (yes)
of the temperature
EVC30S40J7XXX05 top-floor top probe floor probe top floor 2
regulation (yes)
of the temperature
EVC70S40X7XXX00 cooking timer not used not used acoustics not used 4
(no)
EVC70S40X7XXX01 steam injection not used not used timer 1 timer 2 4
(no)
EVC70S40X7XXX02 steam injection not used not used timer 1 timer 2 6
(no)
EVC70S40X7XXX03 cooking timer not used not used acoustics not used 0
(yes)
4.2 Management of the utilities
Temperature regulation.
The output activity depends mainly on the temperature of the
chamber, the work set point and the parameter P5.
Top.
If functioning with 1 measurement input (parameter P3 = 1):
• the out put is measured in a cyclical mode, preferably when the floor
output is off (the parameter P10 establishes the cycle time. Using the
procedure given in paragraph 6.2.1 it is possible to set the duration
of output switch-on, intended as a percentage of the time established
with the parameter P10)
• the cyclical activation is subordinate to the temperature of the
chamber (chamber probe), at the work set point and parameter P4.
If functioning with 2 measurement inputs (parameter P3 = 0):
• the output activity will mainly depend on the temperature of the top
(top probe), the top set point and the parameter P4.
Floor.
If functioning with 1 measurement input (parameter P3 = 1):
• the out put is measured in a cyclical mode, preferably when the top
output is off (the parameter P10 establishes the cycle time. Using the
procedure given in paragraph 6.2.2 it is possible to set the duration
of output switch-on, intended as a percentage of the time established
with the parameter P10)
• the cyclical activation is subordinate to the temperature of the
chamber (chamber probe), at the work set point and parameter P4.
If functioning with 2 measurement inputs (parameter P3 = 0):
• the output activity will mainly depend on the temperature of the
floor (floor probe), the floor set point and the parameter P4.
Timer 1.
The output is on continuously during the count of the time P0.
Timer 2.
The output is on continuously during the count of the time P6.
Acoustics.
The output is on continuously on the conclusion of the count of
the cooking timer.
Auxiliary.
The activity of the output depends mainly on parameters P6 and P7.
5 USER INTERFACE
5.1 Preliminary considerations
The diagram illustrates the model with the greatest supply of hardware.
For all models.
The following functioning states exist:
• the “on” status (the instrument is powered and “on”: the regulators
can be switched on)
• the “stand-by” status (the instrument is powered but is “off” via software:
the regulators are off
• the “off” state (the instrument “is not powered”).
Successively, the term “switch-on” means that the passage from the
stand-by status to the on status; the term “switch-off” means the passage
from the on status to the stand-by status.
When the instrument is powered it re-proposes the status in which it
found itself at the time when the power supply was disconnected.
If there is a power cut during a time count, the count will start again
from the beginning when the power supply is restored.
5.2 Switching the instrument on
For EVC30S40J7XXX01 / 02 / 03 and for EVC70S40X7XXX00 / 01 / 02.
• make sure no procedure is in progress
• press B1 : the L4 LED will switch-off.
For EVC30S40J7XXX00 / 04 / 05 and for EVC70S40X7XXX03.
• make sure no procedure is in progress
• press the encoder button: the L4 LED will switch-off.
5.3 Switching the instrument off
For EVC30S40J7XXX01 / 02 / 03 and for EVC70S40X7XXX00 / 01 / 02.
• make sure no procedure is in progress
• press B1 for 4 s: the L4 LED will switch-on.
For EVC30S40J7XXX00 / 04 / 05 and for EVC70S40X7XXX03.
• make sure no procedure is in progress
• press the encoder button for 4 s: the L4 LED will switch-on.
5.4 The display
For EVC30S40J7XXX00 / 01.
If the instrument is on, the display will show the size established with
parameter P10:
• if P10 = 0, the display will show the work set point
• if P10 = 1, the display will show the chamber temperature
If the instrument is off:
• the display will be off
• the L4 LED will be on

Evco S.p.A. • Code 104C30S40E02 • page 2/6
For EVC30S40J7XXX02 / 03 / 04 / 05.
If the instrument is on, the display will show the size established
with parameter P16:
• if P16 = 0, if functioning with 2 measurement inputs, the display will
show the temperature of the top (otherwise the temperature of the
chamber)
• if P16 = 1, if functioning with 2 measurement inputs, the display will
show the temperature of the floor (otherwise the temperature of the
chamber)
• if P16 = 2, if functioning with 2 measurement inputs, the display will
show the average temperature (otherwise the temperature of the
chamber)
• if P16 = 3, if functioning with 2 measurement inputs, the display will
show the top set point (otherwise the work set point)
• if P16 = 4, if functioning with 2 measurement inputs, the display will
show the floor set point (otherwise the temperature of the chamber)
If the cooking timer count is in progress, the display shows the value,
the L3 LED will flash and the min LED will be on.
If the instrument is off:
• the display will be off
• the L4 LED will be on
For EVC70S40X7XXX00 / 03.
If the instrument is on, the display will show the value of the cooking
timer and the min LED will be on.
If the cooking timer count is in progress, the display shows the value,
the L3 LED will flash and the min LED will be on
If the instrument is off:
• the display will be off
• the L4 LED will be on
For EVC70S40X7XXX01 / 02.
If the instrument is on, the display will show “ - - - “.
If the steam sequence times count is in progress, the display will show
the value, the L1, L2 or L3 LED will flash and if parameter P9 is set at 1
the min LED will be on.
If the instrument is off:
• the display will be off
• the L4 LED will be on
5.5 Silencing the buzzer
For all models.
• make sure no procedure is in progress
• press a key (or turn the encoder): the first press of the key (or rotation
of the encoder) does not cause the associated effect.
6 SETTINGS
6.1.1 Setting the work set point
For EVC30S40J7XXX00.
• make sure that the instrument is in on status and that no procedure
is in progress
• press the encoder button: the display will show the corresponding
value
• turn the encoder within 15 s; see also parameters P2, P3 and P4
• press the encoder button or do not operate for 15 s: the instrument
will leave the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX01.
• make sure that the instrument is in on status and that no procedure
is in progress
• press B1 : the display will show the corresponding value
• press B2 or B5 within 15 s; see also parameters P2, P3 and
P4
• press B1 or do not operate for 15 s: the instrument will leave the
procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX02 (if the parameter P3 is set at 1).
• make sure that the instrument is in on status and that no procedure
is in progress
• press B2 and B5 for 4 s: the display will show the
corresponding value
• press B2 or B5 within 15 s; see also parameters P5, P6 and
P11
• do not operate for 15 s: the instrument will leave the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX03 (if parameter P3 is set at 1).
• make sure that the instrument is in on status and that no procedure
is in progress
• press B2 : the display will show the corresponding value
• press B2 or B5 within 15 s; see also parameters P5, P6 and
P11
• do not operate for 15 s: the instrument will leave the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX04 / 05 (if parameter P3 is set at 1).
• make sure that the instrument is in on status and that no procedure
is in progress
•for EVC30S40J7XXX05 # press the encoder button: the display will
show the corresponding value; turn the encoder clockwise within
15 s to show in succession the power supplied at the top and the
one supplied at the floor (after 5 s the display will show the work set
point again)
• press the encoder button during the showing of the work set point:
the display will show the corresponding value
• turn the encoder within 15 s; see also parameters P5, P6 and P11
• press the encoder button three times or do not operate for 15 s: the
instrument will leave the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
6.1.2 Setting the top set point
For EVC30S40J7XXX02 (if the parameter P3 is set at 0).
• make sure that the instrument is in on status and that no procedure
is in progress
• press B2 : the display shows the corresponding value and the
L1 LED will flash
• press B2 or B5 within 15 s; see also parameters P5, P6 and P7
• do not operate for 15 s: the L1 LED will switch-off and the instrument
will exit the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX03 (if parameter P3 is set at 0).
• make sure that the instrument is in on status and that no procedure
is in progress
• press B3 : the display shows the corresponding value and the
L1 LED will flash
• press B2 or B5 within 15 s; see also parameters P5, P6 and P7
• press B3 or do not operate for 15 s: the L1 LED will switch-off and
the instrument will exit the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX04 / 05 (if parameter P3 is set at 0).
• make sure that the instrument is in on status and that no procedure
is in progress
•for EVC30S40J7XXX05 # press the encoder button: the display will
show the corresponding value; turn the encoder clockwise within
15 s to show the floor set point (after 5 s the display will show the top
set point again)
• press the encoder button during the showing of the top set point: the
display will show the corresponding value and the L1 LED will flash
• turn the encoder within 15 s; see also parameters P5, P6 and P7
• press the encoder button twice or do not operate for 15 s: the instru-
ment will leave the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
6.1.3 Setting the floor set point
For EVC30S40J7XXX02 (if the parameter P3 is set at 0).
• make sure that the instrument is in on status and that no procedure
is in progress
• press B5 : the display shows the corresponding value and the L2
LED will flash
• press B2 or B5 within 15 s; see also parameters P5, P6 and P8
• do not operate for 15 s: the L2 LED will switch-off and the instrument
will exit the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX03 (if parameter P3 is set at 0).
• make sure that the instrument is in on status and that no procedure
is in progress
• press B4 : the display shows the corresponding value and the L2
LED will flash
• press B2 or B5 within 15 s; see also parameters P5, P6 and
P8
• press B4 or do not operate for 15 s: the L2 LED will switch-off and
the instrument will exit the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX04 / 05 (if parameter P3 is set at 0).
• press the encoder button during the setting of the top set point: the
display will show the corresponding value, the L1 LED switches-off
and the L2 LED will flash
• turn the encoder within 15 s; see also parameters P5, P6 and P8
• press the encoder button or do not operate for 15 s: the L2 LED will
switch-off and the instrument will exit the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
6.2.1 Setting the power supplied at the top
For EVC30S40J7XXX02 (if the parameter P3 is set at 1).
• make sure that the instrument is in on status and that no procedure
is in progress
• press B2 : the display shows the corresponding value and the L1
LED will flash
• press B2 or B5 within 15 s; see also parameters P9, P10
and P12
• do not operate for 15 s: the L1 LED will switch-off and the instrument
will exit the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX03 (if parameter P3 is set at 1).
• make sure that the instrument is in on status and that no procedure
is in progress
• press B3 : the display shows the corresponding value and the L1
LED will flash
• press B2 or B5 within 15 s; see also parameters P9, P10
and P12
• press B3 or do not operate for 15 s: the L1 LED will switch-off and
the instrument will exit the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX04 / 05 (if parameter P3 is set at 1).
• press the encoder button during the setting of the work set point:
the display shows the corresponding value and the L1 LED will
flash
• turn the encoder within 15 s; see also parameters P9, P10 and P12
• press the encoder button twice or do not operate for 15 s: the
instrument will leave the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
6.2.2 Setting the power supplied at the floor
For EVC30S40J7XXX02 (if the parameter P3 is set at 1).
• make sure that the instrument is in on status and that no procedure
is in progress
•press B5 : the display shows the corresponding value and the L2
LED will flash
• press B2 or B5 within 15 s; see also parameters P9, P10
and P13
• do not operate for 15 s: the L2 LED will switch-off and the instrument
will exit the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX03 (if parameter P3 is set at 1).
• make sure that the instrument is in on status and that no procedure
is in progress
• press B4 : the display shows the corresponding value and the
L2 LED will flash
• press B2 or B5 within 15 s; see also parameters P9, P10 and P13
• press B4 or do not operate for 15 s: the L2 LED will switch-off and
the instrument will exit the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX04 / 05 (if parameter P3 is set at 1).
• press the encoder button during the setting of the power supplied
at the top: the display will show the corresponding value, the L1 LED
switches-off and the L2 LED will flash
• turn the encoder within 15 s; see also parameters P9, P10 and P13
• press the encoder button or do not operate for 15 s: the L2 LED will
switch-off and the instrument will exit the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
6.3 Setting the configuration parameters
For EVC30S40J7XXX00 / 04 / 05 and for EVC70S40X7XXX03.
To access the procedure:
• make sure that the instrument is in off status and that no procedure
is in progress
• press the encoder button for 4 s: the central display will show “PA”
• press the encoder button: the display will show the corresponding
value
• turn the encoder within 15 s to set “-19”
• press the encoder button or do not operate for 15 s
• press the encoder button for 4 s: the display will show “P0”.
To select a parameter:
• turn the encoder.
To modify a parameter:
• press the encoder button: the display will show the corresponding
value
• turn the encoder within 15 s
• press the encoder button or do not operate for 15 s
• turn the encoder.
To exit the procedure:
• press the encoder button for 4 sor do not operate for 60 s (any
modifications will be saved).
For EVC30S40J7XXX01 / 02 / 03 and for EVC70S40X7XXX00 / 01 / 02.
To access the procedure:
• make sure that the instrument is in off status and that no procedure
is in progress
• press B2 and B5 for 4 s: the central display will show “PA”
• press B1 : the display will show the corresponding value
• press B2 or B5 within 15 s to set “-19”
• press B1 or do not operate for 15 s
• press B2 and B5 for 4 s: the display will show “P0”.
To select a parameter:
• press B2 or B5
To modify a parameter:
• press B1: the display will show the corresponding value
• press B2 or do not operate for 15 s
• press B1 or do not operate for 15 s
• press B2 or B5
To exit the procedure:
• press B2 and B5 for 4 s or do not operate for 60 s (any
modifications will be saved).
7 COOKING TIMER (for EVC30S40J7XXX02 / 03 /
04 / 05 and for EVC70S40X7XXX00 / 03)
7.1 Preliminary considerations
For EVC30S40J7XXX02 / 03 / 04 / 05.
The cooking timer allows to start the countdown.
The display shows the value during counting (for EVC30S40J7XXX05
# the display will show the size established with parameter P24), the
L3 LED will flash and the min LED is on.

Evco S.p.A. • Code 104C30S40E02 • page 3/6
Before the conclusion of the count (of the time established with
parameter P21) the buzzer is activated, for the time established with
parameter P20.
If there is a power cut during the count, the same will start again from
the beginning when the power supply is restored.
For EVC70S40X7XXX00 / 03.
The cooking timer allows to start the countdown.
The display shows the value during counting, the L3 LED will flash
and the min LED is on.
Before the conclusion of the count (of the time established with
parameter P4) the buzzer is activated, for the time established with
parameter P3.
The acoustic output is activated at the conclusion of the count until the
B6 key is pressed (or the encoder button is pressed).
If there is a power cut during the count, the same will start again from
the beginning when the power supply is restored.
7.2 Setting and start-up of the cooking timer
For EVC30S40J7XXX02 / 03.
• make sure that the instrument is in on status and that no procedure
is in progress
• press B6 : the display shows the corresponding value and the
min LED will flash
• press B2 or B5 within 15 s; see also parameters P17,
P18 and P19
• press B6 : the instrument will exit the procedure, the L3 LED will
flash and the min LED will switch-on and the timer will be
started.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX04.
• make sure that the instrument is in on status and that no procedure
is in progress
• turn the encoder: the display shows the corresponding value and
the min LED will flash
• turn the encoder within 15 s; see also parameters P17, P18 and P19
• press the encoder button: the instrument will exit the procedure, the
L3 LED will flash and the min LED will switch-on and the timer will
be started.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC30S40J7XXX05.
• make sure that the instrument is in on status and that no procedure
is in progress
• press B1: the display shows the corresponding value and the min
LED will flash
• turn the encoder within 15 s; see also parameters P17, P18 and
P19
• press B1 : the instrument will exit the procedure, the L3 LED will
flash and the min LED will switch-on and the timer will be
started.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC70S40X7XXX00.
• make sure that the instrument is in on status and that no procedure
is in progress
• press B1 : the display shows the corresponding value and the min
LED will flash
• press B2 or B5 within 15 s; see also parameters P0,
P1 and P2
• press B6 : the instrument will exit the procedure, the L3 LED will flash
and the min LED will switch-on and the timer will be started.
Alternatively:
• make sure that the instrument is in on status and that no procedure
is in progress
• press B1 : the display shows the corresponding value and the min
LED will flash
• press B2 or B5 within 15 s; see also parameters P0,
P1 and P2
• press B1 or do not operate for 15 s: the instrument will leave the
procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
Successively:
• press B6 : the L3 LED will flash and the min LED will switch-on and
the timer will be started.
For EVC30S40X7XXX03.
• make sure that the instrument is in on status and that no procedure
is in progress
• turn the encoder: the display shows the corresponding value and
the min LED will flash
• turn the encoder within 15 s; see also parameters P0, P1 and P2
• press the encoder button: the instrument will leave the procedure.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
Successively:
• press the encoder button: the L3 LED will flash and the min LED will
switch-on and the timer will be started.
7.3 Interrupting the cooking timer
For EVC30S40J7XXX02 / 03 and for EVC70S40X7XXX00.
• press B6 : the L3 LED and the min LED will switch off.
For EVC30S40J7XXX04 and for EVC70S40X7XXX03.
• press the encoder button: the L3 LED and the min LED will switch
off.
For EVC30S40J7XXX05.
• press B1 : the L3 LED and the min LED will switch off.
8 STEAM SEQUENCE (for EVC70S40X7XXX01 / 02)
8.1 Preliminary considerations
The steam sequence allows to start the countdown of three times.
On the conclusion of the count of a time, the instrument automatically
passes to the next count.
During the countdown of the first three times (successively also called
“time t1”) the display shows the value, the L1 LED flashes. If the param-
eter P9 is set at 1 the min LED is on and the timer 1 output is on.
During the countdown of the second three times (successively also
called “time t2”) the display shows the value, the L2 LED flashes. If the
parameter P9 is set at 1 the min LED is on and the outputs are switched
off.
During the countdown of the third three times (successively also called
“time t3”) the display shows the value, the L3 LED flashes. If the param-
eter P9 is set at 1 the min LED is on and the timer 2 output is on.
On conclusion of the countdown of the time t3, the buzzer is activated
for the tie established with parameter P12.
The functioning of the instrument on conclusion of the count of time
t3 depends on parameter P10.
If there is a power cut during the count, the same will start again from
the beginning when the power supply is restored.
8.2 Setting times t1, t2 and t3
For EVC70S40X7XXX01.
• make sure that the instrument is in on status and that no procedure
is in progress
• press B1 : the display shows the time t1 and the L1 LED will flash
• press B2 or B5 within 15 s; see also parameters P0,
P1 and P2
• press B1 : the display will show the time t2, the L1 LED switches-
off and the L2 LED will flash
• press B2 or B5 within 15 s; see also parameters P6, P7 and P8
• press B1 : the display will show the time t3, the L2 LED switches-off
and the L3 LED will flash
• press B2 or B5 within 15 s; see also parameters P3,
P4 and P5
• press B1 : the instrument will exit the procedure and L3 LED will
switch-off.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
For EVC70S40X7XXX02.
To set time t1:
• make sure that the instrument is in on status and that no procedure
is in progress
• press B3 : the display shows the time t1 and the L1 LED will flash
• press B2 or B5 within 15 s; see also parameters P0,
P1 and P2
• press B3 : the instrument will exit the procedure and L1 LED will
switch-off.
To set time t2:
• make sure that the instrument is in on status and that no procedure
is in progress
• press B1 : the display shows the time t2 and the L2 LED will flash
• press B2 or B5 within 15 s; see also parameters P6, P7 and
P8
• press B1 : the instrument will exit the procedure and L2 LED will
switch-off.
To set time t3:
• make sure that the instrument is in on status and that no procedure
is in progress
• press B4 : the display shows the time t3 and the L3 LED will flash
• press B2 or B5 within 15 s; see also parameters P3,
P4 and P5
• press B4 : the instrument will exit the procedure and L3 LED will
switch-off.
To exit the procedure in advance:
• do not operate for 15 s (any modifications will be saved).
8.3 Starting the steam sequence
• make sure that the instrument is in on status and that no procedure
is in progress
• press B6 : the L1 LED will flash and if the parameter P9 is set at 1 the
min LED will switch on.
8.4 Interrupting the steam sequence
press B6 : the L1, L2 or L3 LED will switch off and if parameter P9
is set at 1 also the minLED.
9 QUICK HEATING (for EVC30S40J7XXX02 /
03 / 05)
9.1 Preliminary considerations
The quick heating allows to reach the work set point as quickly as
possible, supplying 100% power at the top and the floor (i.e. excluding
switch-on of the top and floor outputs in a cyclical mode with the
benefit of continuous switch-on).
When the chamber temperature reaches the “work set point -
temperature established with parameter P23”, the function is
interrupted.
If there is a power cut during the quick heating, the same will start-up
again when the power supply is restored.
9.2 Activation of quick heating
• make sure that the instrument is on, that there is no procedure in
progress and that the parameter P22 is set at 1
• press B6 for 4 s.
9.3 Interruption of quick heating in manual mode
• make sure no procedure is in progress
• press B6 for 4 s.
10 AUXILIARY OUTPUT (for EVC30S40J7XXX01)
10.1 Preliminary considerations
The functioning mode of the auxiliary output depends on
parameters P6 and P7.
If parameter P6 is set at 0, pressing the B6 key when the instrument is
on causes the output to switch on and when pressed again, causes it
to switch off.
If parameter P6 is set at values different to 0 and parameter P7 is set at 0,
pressing the B6 key when the instrument is on causes the output to
switch on for the time established with parameter P6.
If parameters P6 and P7 are set at values different to 0, the output will
be switched on cyclically (parameter P6 establishes the duration of
switch-on and parameter P7 that of switch-off). Pressing the B6 key
when the instrument is on causes the cyclical activity to start and pressing
the B5 key causes its interruption.
The L2 LED supplies indications regarding the state of the auxiliary
output, see chapter 11.
11 SIGNALS
11.1 Si gnals
LED MEANING
L1 For EVC30S40J7XXX00 / 01.
temperature regulation LED
if on, the output for temperature regulation will be off.
if flashing, the output for temperature regulation will be on.
For EVC30S40J7XXX02 / 03 / 04 / 05.
top LED
if it is on, the top output will be on
if flashing, the top set point setting is in progress or that of
the power supplied at the top
For EVC70S40X7XXX01 / 02.
time t1 LED
if on, the setting of time t1 is in progress
if flashing, the time t1 count is in progress
L2 For EVC30S40J7XXX01.
Auxiliary output LED
if it is on, the auxiliary output will be on
For EVC30S40J7XXX02 / 03 / 04 / 05.
floor LED
if it is on, the floor output will be on
if flashing, the floor set point setting is in progress or that
of the power supplied at the floor
For EVC70S40X7XXX01 / 02.
time t2 LED
if on, the setting of time t2 is in progress
if flashing, the time t2 count is in progress
L3 For EVC30S40J7XXX02 / 03 / 04 / 05 and for
EVC70S40X7XXX00 / 03.
cooking timer LED
if it is on, the cooking timer count will be concluded
if flashing, the cooking timer count is in progress
For EVC70S40X7XXX01 / 02.
time t3 LED
if on, the setting of time t3 is in progress
if flashing, the time t3 count is in progress
L4 For all models.
on/stand-by LED
if it is on, the instrument is in the stand-by status
For all models.
Power supply cut off LED
if flashing, if a power cut should occur when the instrument
is on or (during a time count), press a key (or turn the
encoder) to switch the LED off.
min For EVC30S40J7XXX02 / 03 / 04 / 05 and for
EVC70S40X7XXX00 / 03.
minutes LED
if on, the cooking timer count is in progress
if flashing, cooking timer setting is in progress
For EVC70S40X7XXX01 / 02.
minutes LED
if it is on, the unit of measurement of the time will be the
minute (parameter P9)
°CFor EVC30S40J7XXX00 / 01 / 02 / 03 / 04 / 05.
Degrees Celsius LED
if it is on, the unit of measurement of the temperatures will
be the degree Celsius (parameter P0)
°FFor EVC30S40J7XXX00 / 01 / 02 / 03 / 04 / 05.
Degrees Fahrenheit LED
if it is on, the unit of measurement of the temperatures will
be the degree Fahrenheit (parameter P0)
12 ALARMS
12.1 Alarms
CODE MEANING
H-t For EVC30S40J7XXX00 / 01.
use temperature alarm
Solutions:
• press B2 (or turn the encoder) to display the use tem-
perature
• see parameters P8 and P9

Evco S.p.A. • Code 104C30S40E02 • page 4/6
PT • 39/10
Consequences:
• all outputs will be deactivated
H-t For EVC30S40J7XXX02 / 03 / 04 / 05.
use temperature alarm
Solutions:
• press B2 to display the use temperature
• see parameters P14 and P15
Consequences:
• all outputs will be deactivated
When the cause of the alarm has disappeared, the instrument restores
normal functioning, except for the use temperature alarm
(code “H-t”), which requires the B1 key to be pressed or the encoder
button). This switches the instrument off.
13 INTERNAL DIAGNOSTICS
13.1 Internal diagnostics
CODE MEANING
AL1 For EVC30S40J7XXX00 / 01.
chamber probe error
Solutions:
• check that the probe is a J thermocouple
• check the integrity of the probe
• check the instrument-probe connection
• verify the chamber temperature
Main consequences:
• the temperature regulation output will be deactivated
AL1 For EVC30S40J7XXX02 / 03 / 04 / 05 (if parameter P3 is set
at 0).
top probe error
Solutions:
• the same as the previous case but relative to the top probe
Main consequences:
• the top output will be deactivated
For EVC30S40J7XXX02 / 03 / 04 / 05 (if the parameter P3
is set at 1).
chamber probe error
Solutions:
• the same as the previous case but relative to the chamber
probe
Main consequences:
• the top output and the floor output will be deactivated
AL2 For EVC30S40J7XXX02 / 03 / 04 / 05 (if parameter P3 is set
at 0).
floor probe error
Solutions:
• the same as the previous case but relative to the floor
probe
Main consequences:
• the floor output will be deactivated
When the causes of the alarm have disappeared, the instrument will
go back to normal functioning.
14 TECHNICAL DATA
14.1 Technical data
With reference to the model with most hardware.
Container: open frame board
Front panel protection rating: IP 00.
Connections: 6.3 mm faston (0.248 in, power supply and outputs),
screw terminal board (measurement inputs), 4-pole connector (to the
encoder).
Temperature of use:from 0 to 55 °C (from 32 to 131 °F,
10 ... 90% relative humidity without condensate).
Power supply: 230 Vac, 50/60 Hz, 1.5 VA.
Alarm buzzer: incorporated.
Measurement inputs: 2 for J thermocouple; the instrument also
incorporates a sensor for the detection of the use temperature.
Range of measurement:from 20 to 500 °C (from 99 to 999 °F).
Resolution: 1 °C/1 °F.
Digital outputs: 2 5 A res. relays @ 250 Vac (NO).

Evco S.p.A. • Code 104C30S40E02 • page 5/6
15 CONFIGURATION PARAMETERS
15.1 Configuration parameters
PAR. MIN MAX U. M. EVC30..00 EVC30..01 EVC30..02 EVC30..03 EVC30..04 EVC30..05 EVC70..00 EVC70..01 EVC70..02 EVC70..03
P001---- 000000----------------
P0 P1 P2 min - - - - - - - - - - - - - - - - - - - - - - - - 20 - - - - - - - - 2 0
P0 P1 P2 s/min (2) - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1 0 10 - - - -
P1 -9 9 °C 0 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P1 -9 9 °C --------0000----------------
P1 0 P2 min. - - - - - - - - - - - - - - - - - - - - - - - - 0 - - - - - - - - 0
P1 0 P2 s/min (2) - - - - - - - - - - - - - - - - - - - - - - - - - - - - 0 0 - - - -
P2 P3 P4 °C 10 10 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P2 -9 9 °C --------0000----------------
P2 P1 999 min - - - - - - - - - - - - - - - - - - - - - - - - 199 - - - - - - - - 199
P2 P1 999 s/min (2) - - - - - - - - - - - - - - - - - - - - - - - - - - - - 999 99 9 - - - -
P3 20 P4 °C 0 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P301---- --------0000----------------
P3 0 60 s - - - - - - - - - - - - - - - - - - - - - - - - 20 - - - - - - - - 2 0
P3 P4 P5 s/min (2) - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1 0 10 - - - -
P4 P3 450 °C 45 0 45 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P4 2 9 °C --------2222----------------
P4 0 240 s - - - - - - - - - - - - - - - - - - - - - - - - 1 0 - - - - - - - - 1 0
P4 0 P5 s/min (2) - - - - - - - - - - - - - - - - - - - - - - - - - - - - 0 0 - - - -
P5 2 9 °C 2 2 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P5 0 P6 °C --------0000----------------
P5 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - • - - - - - - - - •
P5 P4 999 s/min (2) - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9 9 9 999 - - - -
P6 0 1 - - - - 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P6 0 600 - - - - - - - - 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P6 P5 450 °C - - - - - - - - 400 40 0 400 40 0 - - - - - - - - - - - - - - - -
P6 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - • - - - - - - - - •
P6 P7 P8 s/min (2) - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1 0 10 - - - -
P7 55 65 °C 6 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P7 0 600 s - - - - 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P7 P5 P6 °C - - - - - - - - 200 20 0 200 20 0 - - - - - - - - - - - - - - - -
P7 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - • - - - - - - - - •
P7 0 P8 s/min (2) - - - - - - - - - - - - - - - - - - - - - - - - - - - - 0 0 - - - -
P8 0 1 - - - - 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P8 0 1 - - - - - - - - 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P8 P5 P6 °C - - - - - - - - 200 20 0 200 20 0 - - - - - - - - - - - - - - - -
P8 P7 999 s/min (2) - - - - - - - - - - - - - - - - - - - - - - - - - - - - 9 9 9 999 - - - -
P9 - - - - - - - - - - - - • - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P9 55 65 °C - - - - 6 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P902---- --------0000----------------
P9 0 1 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 0 0 - - - -
P10 - - - - - - - - - - - - • - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P10 0 1 - - - - - - - - 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P10 1 999 s - - - - - - - - 6 0 6 0 6 0 60 - - - - - - - - - - - - - - - -
P10 0 2 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 0 0 - - - -
P11 - - - - - - - - - - - - • - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P11 - - - - - - - - - - - - - - - - • - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P11 P5 P6 °C - - - - - - - - 2 0 0 200 20 0 20 0 - - - - - - - - - - - - - - - -
P11 1 1 0 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1 1 - - - -
P12 - - - - - - - - - - - - - - - - • - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
P12 0 100 % - - - - - - - - 70 7 0 7 0 70 - - - - - - - - - - - - - - - -
P12 0 6 0 s - - - - - - - - - - - - - - - - - - - - - - - - - - - - 0 0 - - - -
P13 - - - - - - - - - - - - - - - - • - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
DESCRIPTION
temperature unit of measurement (1)
0=°C
1=°F
predefined cooking timer value
time t1 predefined value
chamber probe offset
if P3 = 0, top probe offset
if P3 = 1, chamber probe offset
cooking timer minimum value
time t1 minimum value
work set point predefined value
floor probe offset (only if P3 = 0)
cooking timer maximum value
time t1 maximum value
minimum work set-point
type of functioning
0 = with 2 measurement inputs (allows to set the top work temperature
independently from that of the floor )
1 = with 1 measurement input (allows to set the power supplied at the
top independently from that supplied at the bottom)
during activation of the buzzer on conclusion of the cooking timer count;
also see P24
time t3 predefined value
maximum work set-point
if P3 = 0, top set point differential and of the floor set point
if P3 = 1, work set point differential
time that passes between the activation of the buzzer and the conclusion
of the cooking timer count, see also P3
time t3 minimum value
work set-point differential
if P3 = 0, minimum top set point and floor set point
if P3 = 1, minimum work set point
firmware code (3)
time t3 maximum value
enabling of the use temperature alarm (code “H-t”)
1 = YES
duration of auxiliary output switch-on; see also P7
if P3 = 0, maximum top set point and floor set point
if P3 = 1, maximum work set point
firmware version code (3)
time t2 predefined value
temperature over which the use temperature alarm is activated
(code “H-t”); see also P6
duration of auxiliary output switch-off; see also P6
predefined value of the top set point (only if P3 = 0)
firmware revision code (3)
time t2 minimum value
size displayed when the instrument is on
0 = work set point
1 = chamber temperature
enabling of the use temperature alarm (code “H-t”)
1 = YES
predefined value of the floor set point (only if P3 = 0)
time t2 maximum value
firmware code (3)
temperature over which the use temperature alarm is activated
(code “H-t”); see also P8
effect between the power supplied at the top and that at the floor
0 = no effect
1 = the modification of the power supplied at an output automatically
causes the supply of the maximum power at the other
2 = the modification of the power supplied at an output causes an
automatic adaptation of the power supplied at the other in order to
guarantee that he sum of the two percentages is always 100
unit of measurement of times P0, P1, P2, P3, P4, P5, P6, P7 and P8
0 = second
1 = minute
firmware version code (3)
size displayed when the instrument is on
0 = work set point
1 = chamber temperature
cycle time for switch-on of the top output and the floor output; see also
P12 and P13 (only if P3 = 1)
functioning of the instrument at the conclusion of the time t3 count.
0 = the instrument switches off
1 = the steam sequence is repeated automatically for the number of times
established with parameter P11
2 = the steam sequence is repeated automatically until it is interrupted by
pressing key B6
firmware revision code (3)
firmware code (3)
predefined value of the work set point (only if P3 = 1)
number of steam sequence (only if P10 = 1)
firmware version code (3)
predefined value of the power supplied at the top (percentage of P10;
only if P3 = 1); see also P9
duration of the activation of the buzzer on conclusion of the time t3 count
firmware revision code (3)

EVCO S.p.A. This document belongs to Evco; unless you are authorized by Evco, you can not publish this document.
Via Mezzaterra 6, 32036 Sedico Belluno ITALY Evco does not take any responsibility about features, technical data and possible mistakes related in this document.
Phone +39-0437-852468 • Fax +39-0437-83648 Evco does not take any responsibility about damages coming by the non-observance of additional information.
[email protected] • www.evco.it Evco reserves the right to make any change without prior notice and at any time without prejudice the basic safety and operating features.
Evco S.p.A. • Code 104C30S40E02 • page 6/6
PAR. MIN MAX U. M. EVC30..00 EVC30..01 EVC30..02 EVC30..03 EVC30..04 EVC30..05 EVC70..00 EVC70..01 EVC70..02 EVC70..03
P13 0 100 % - - - - - - - - 30 3 0 3 0 30 - - - - - - - - - - - - - - - -
P13 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - • • - - - -
P1401---- --------0000----------------
P14 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - • • - - - -
P15 55 65 °C - - - - - - - - 6 0 6 0 6 0 6 0 - - - - - - - - - - - - - - - -
P15 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - • • - - - -
P1604---- --------2222----------------
P17 P18 P19 min - - - - - - - - 20 2 0 2 0 20 - - - - - - - - - - - - - - - -
P180P19min --------0000----------------
P19 P18 9 9 9 min - - - - - - - - 19 9 199 199 1 9 9 - - - - - - - - - - - - - - - -
P20 0 6 0 s - - - - - - - - 2 0 2 0 20 2 0 - - - - - - - - - - - - - - - -
P21 0 240 s - - - - - - - - 1 0 1 0 1 0 10 - - - - - - - - - - - - - - - -
P22 0 1 - - - - - - - - - - - - 0 0 - - - - 0 - - - - - - - - - - - - - - - -
P22 - - - - - - - - - - - - - - - - - - - - - - - - - - - - • - - - - - - - - - - - - - - - - - - - -
P23 0 99 °C - - - - - - - - 0 0 - - - - 0 - - - - - - - - - - - - - - - -
P23 - - - - - - - - - - - - - - - - - - - - - - - - - - - - • - - - - - - - - - - - - - - - - - - - -
P24 - - - - - - - - - - - - - - - - - - - - • • - - - - - - - - - - - - - - - - - - - - - - - -
P24 - - - - - - - - - - - - - - - - - - - - - - - - - - - - • - - - - - - - - - - - - - - - - - - - -
P24 0 1 - - - - - - - - - - - - - - - - - - - - - - - - 0 - - - - - - - - - - - - - - - -
P25 - - - - - - - - - - - - - - - - - - - - • • - - - - - - - - - - - - - - - - - - - - - - - -
P25 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - • - - - - - - - - - - - - - - - -
P26 - - - - - - - - - - - - - - - - - - - - • • - - - - - - - - - - - - - - - - - - - - - - - -
P26 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - • - - - - - - - - - - - - - - - -
P26 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - • - - - - - - - - - - - - - - - -
(1) the modification of the parameter affects all parameters whose unit of measurement is the degree Celsius or the degree Fahrenheit
(2) the unit of measurement depends on parameter P9
(3) the parameter cannot be modified.
DESCRIPTION
predefined value of the power supplied at the floor (percentage of P10;
only if P3 = 1); see also P9
firmware code (3)
enabling of the use temperature alarm (code “H-t”)
1 = YES
firmware version code (3)
temperature over which the use temperature alarm is activated
(code “H-t”); see also P14
firmware revision code (3)
size displayed when the instrument is on
0 = if P3 = 0,, top temperature
if P3 = 1, chamber temperature
1 = if P3 = 0, floor temperature
if P3 = 1, chamber temperature
2 = if P3 = 0, average temperature
if P3 = 1, chamber temperature
3 = if P3 = 0, top set point
if P3 = 1, work set point
4 = if P3 = 0, floor set point
if P3 = 1, chamber temperature
predefined cooking timer value
cooking timer minimum value
cooking timer maximum value
during activation of the buzzer on conclusion of the cooking timer count;
also see P21
time that passes between the activation of the buzzer and the conclusion
of the cooking timer count, see also P20
enabling of the quick heating function (only P3= 1)
1 = YES
firmware code (3)
temperature of the chamber above which the quick heating function is
interrupted (relative to the work set point, i.e. “work set point - P23”; only
if P3 = 1)
firmware version code (3)
firmware code (3)
firmware revision code (3)
size displayed when the cooking timer count is in progress
0 = size established with parameter P16
1 = cooking timer value
firmware version code (3)
firmware code (3)
firmware revision code (3)
firmware version code (3)
firmware revision code (3)
This manual suits for next models
1
Table of contents
Other Evco Control Unit manuals