EYELA SD-1010 User manual

Tokyo Rikakikai Co., Ltd.
R00
This manual is designed to use this unit safely with the best performance.
Keep this instruction manual beside the unit.
Important
Read carefully the chapter
[For safety operation] before operating
this unit.
SD-1010
Instruction
Manual
Spray Dryer
Spray Dryer

Precautions for your safety
However, if you know about them in advance, you
can prevent most of these accidents.
For that reason, this instruction manual defines
information on matters to be particularly noteworthy
in terms of safety according to their risk, and
indicate them with an alert mark and a signal word.
Please follow these instructions.
1. Warning signal words
This unit is not explosion-proof structured.
When a flammable sample or organic solvent is
used, handle it with care so that the liquid may not
be spilled over.
As hydrogen gas is used, please install and use
the equipment in a fume hood.
This product is used with its part being heated at
high temperature for its function and
characteristics.
A glass component used may get broken, if
mishandled, potentially leading to an unexpected
injury or accident.
We investigate enough possible hazards during the operation, however, it is very difficult
for us to find every hazardous occasion. Therefore this manual cannot describe all
hazardous operations.
However, if you carry out the handling method described in this manual, you can operate
and work more safely. Be sure to pay close attention to the handling of this product so
that accidents and product failures will not occur.
Alert Mark
Signal word Definition
Indicates a potentially hazardous situation which, if you use
incorrectly, could result in death or serious injury.
Indicates a potentially hazardous situation which, if you
use incorrectly, may result in injury or physical damage.
Warning
Caution
Indicates a hazardous situation which, if you use incorrectly,
could result in death or serious injury.
Warning

2.Warning label
A warning label is attached to the unit to refer to the most important clause.
The attached position is shown below.
Be careful to use the unit referring warning messages.
* When the warning label is worn and hardly show the message, change it with a new one.
Please order us a new label.
Explosion
Do not use this unit
to dry any explosive
materials such as
organic solvent.
Warning
Burning
Do not touch
evaporator or cyclone
while drying samples.
Caution
① ②
①
②

Read this manual carefully before operation.
Thank you for choosing products.
Introduct
ion
Table of
contents
This instruction manual describes the procedure of
installation, operation, troubleshooting,
maintenance/check-up, and disposal for
Spray dryer model
SD-1010.
1.For safety operation 1
2.Outline
2-1. Application 1
2-2. Specifications 2
2-3. Description of each part 3
3.Descriptions and functions of the control panel
3-1.Control panel 5
3-2.Safety and alarm functions 7
4.Installation
4-1. Installed place 8
4-2.Environmental conditions 8
4-3.Connection of utility 9
4-4.Connection of utility 10
5.Operation
5-1.Preparation
5-1-1.Attaching evaporating tube 11
5-1-2.Attaching cyclone 12
5-1-3.Attaching outlet temp. sensor 12
5-1-4.Attaching power collection container and
separator 13
5-1-5.Countermeasures against static electricity
13
5-1-6.Attaching silicone tubing 14
5-1-7.Connecting power cord 15
5-1-8.Connecting exhaust hose 15
5-1-9.Connecting air tubing 16
5-1-10.Connecting recorder cable 16
5-1-11.Replacing cap for nozzle 17
5-1-12.Installing agitator 18
5-2.Operation
5-2-1.Preparing sample and distilled water 19
5-2-2.Operation 20
5-2-3.Cleaning spray nozzle 24
5-2-4.Collection of fine powder 25
5-3.End of operation 25
6.Troubleshooting 27
7. Maintenance and Check-up
7-1.Cleaning and care of product 29
7-2.Cleaning and changing filter
7-2-1.Cleaning or changing inlet filter 30
7-2-2.Cleaning or changing outlet filter 31
7-3.Cleaning evaporating tube and cyclone 31
7-4.Cleaning spray nozzle 33
7-5.Replacing the evaporating tube and cyclone
accessories
7-5-1.Replacing parts at the connection 35
7-5-2.Replacing collection bottle fixing parts 36
8.Product disposition 37
9.Post-sale service 38
10.List of Disposable/Replacing Parts/Optional
Products
10-1. Disposable/replacing parts 39
10-2. Optional products 42

Part name Qty.
1 Main unit 1
2 Rectification board 1
3 Exhaust hose, cuff (2 pieces) 2.5m
4 Stainless steel diamond clamp (for exhaust hose) 2
5 Air piping connector(1/4') 1
6Air piping tube(OD6mm) 10m
7 Stirring bar 1
8 Silicone tubing for applying sample (3.15 x 5.2) 2m
9Silicone tubing for applying sample (3.7 x 6.1) 1m
10 Aluminum tape 10m
11 Cable for recorder(2m) 3
12 Sensor for outlet temperature 1
13 Evaporator set 1
14 Cyclone set 1
15 Collection bottle set 2
16 Instruction Manual 1
17 Warranty 1
This unit is packed in 2 cartons which are main unit and
accessory parts. Check quantities referring to the below
table.
P a c k i n g
2.Rectification board 3.Exhaust hose,cuffs
4.Stainless steel diamond
clamp 5.Air tube connector
6.Air tube 7.Stirring bar
8,9.Silicone tubing 10.Aluminum tape
11.Cable for recorder 12.Sensor for outlet
temperature
13.Evaporator set 14.Cyclone set
15.Collection bottle set

While operating unit, do not touch evaporator or cyclone
to prevent burning of your hands.
While operating unit, do not touch evaporator or cyclone.
Do not use this unit to fry flammable materials such as organic solvent.
This unit has a built-in heater, so it may ignite.
Do not use flammable materials such an organic solvent.
For safety operation
1
Please pay attention to safety as this product becomes hot
Warning
Caution
-1-
Evaporator
Cyclone
In his unit samples are sprayed through a nozzle and
dry by headed air to make dried fine particle.
Outline
2
Do not use flammable materials
such as organic solvent.
Do not use this unit to dry flammable materials such as
organic solvent.
This unit has a built-in heater, so it may ignite.
Dangerous
2-1 Application
Do not remodel. Do not use out of
applications.
Warning
Remodeling or using out of applications may occur electric
shock hazard or mechanical troubles.

2-2 Specifications
Product name Spray Dryer
Model SD-1010
Performance
Evaporating rate Max 1500ml/h
Drying air flow rate control range and
precision 40~200°C (Outlet temperature) • ±1°C
Drying air flow rate control range 0.2~ 0.75m3/min
Spraying air press. control range 20~250kPa (0.2~2.5kg/cm2)
Flow range of delivery pump 150~1700mL/h
Rotation rate and capacity of sample
agitator 100~1000rpm • 50mL~2L (Water)
Upper agitation available (option)
Function
Inlet temperature display/Outlet
temperature display Digital display
Drying air volume display Digital display
Spraying pressure display Digital display
Safety features
Evaporating tube protection cover, leakage/overcurrent
breaker, overheat protector
Manual return at blackout, self-diagnostic function (
abnormal
temperature, abnormal air volume,
heater disconnected, abnormal pressure for solution being
sent)
Sprayed air line cleaning function Automatic operation with interval timer
Inlet temp. recorder output • Outlet temp.
recorder output 0~10mV (at50 ~ 250°C)
Drying air volume output 0~10mV(at0~1m3/min)
Composition
Heater 3kW
Delivery pump Tube pump
Spray nozzle Twin-fluid nozzle (nozzle hole diameter 0.71 mm in inside
diameter)
Nozzle diameter changeable (option)
Spray air pump Spray air compressor (option)
Evaporating tube, cyclone,
Powder collection container, separator Hard glass
Sample agitator motor, magnet DC brushless motor, cobalt magnet
Suction blower Commutator blower
Standards
Diameter of tube used in delivery pump ID 3.15 xOD 5.2mm
Spray air connection bore, material ID 4x OD 6mm, soft urethane tube Union
Spraying air pressure Pressure:294kPA (3kg/cm2) , Flow rate:25L/min or more
Exhaust port size Exhaust port OD 50mm
Interval timer OFF or 1-20 min (automated intermittent time)
Ambient temperature range 5~35°C
Ambient humidity range 30~80%RH (no condensation)
Overall dimensions(mm)• Net weight 700W x 620D x 1500H mm • Approx.110kg
Power sonsumption and power source 19.1A 4.2kVA・AC220V Single phase50/60Hz
*The performance described demonstrates values "at room temperature 20ºC with rated supply voltage
50Hz using potable water and a silicon tube."
*No power plug is supplied.
-2-

2-3 Description of each part
[Front view]
-3-
[Rear view]
Evaporator
Separator
Powder collection container
Cyclone
Connecting tube
Duct
Sensor for outlet temperature
Protection cover
Blower exhaust port
Maintenance port
(Inside: Inlet filter)
Caster Adjuster
Maintenance port
(Inside: Outlet filter)
Cooling fan for control part
Cooling fan for main unit
For cooling blower
Air intake
Outside air inlet
((Silicon tube)
Connection collar

Spray nozzle
Pressure sensor
Delivery pump Sample agitator
(Magnetic stirrer)
Pole fixing screw
Recorder output terminals
Operation lamp of over temperature protector Reset switch of over temperature protector
Power switch
Ground terminal
Air connecting port
Electrical leakage breaker
Power cord
[Top face of operating portion]
[Right side of operating portion]
[Lower right side of housing]
-4-
Operating panel
Screw cover

Descriptions and functions of the control panel
3
3-1 Control panel
No. Description Function
1 Pump line alarm lamp It illuminates when the pressure of flow line attains to a certain pressure.
2 Pump operation lamp It illuminates while the pump is running.
3 Pump ON/OFF key Delivery pump Run/Stop
4 Jet cleaner operation lamp It illuminates when the jet cleaner is in operation
5 Jet cleaner manual operation key To operate manually a cycle of jet cleaning
6 Interval time setting dial To set interval time of jet cleaning
7 Spraying air supply lamp It illuminates when spraying air is supplied
8 Spraying air supply switch To open or close the air supply valve
9 Spraying pressure control knob To control spraying pressure
10 Drying air flow rate control knob To control drying air volume
11 Drying air flow alarm lamp It illuminates when the drying air volume is lower than a certain value.
12 Temperature controller/blower RUN/STOP
key To run/stop temperature controller/blower
13 Temp. control and blower operation lamp To control temperature. It illuminates when the blower is in operation
14 SET key To determine setting values for temperature control and various parameters
15 UP key To change setting values for temperature control and various parameters
16 DOWN key To change setting values for temperature control and various parameters
17 Heater alarm lamp It illuminates when the heater is disconnected
18 Temp. alarm lamp It illuminates when the inlet temp. exceeds the band of set point
19 Heating lamp It illuminates when the heater is active
20 Set value indication lamp It illuminates when the set value is displayed on the inlet temp. indicator.
21 Measured value indication lamp It illuminates when the measured value is displayed on the inlet temp.
indicator
22 Inlet temp. indicator It displays the inlet temperature of evaporating tube
23 Outlet temp.indicator It displays the outlet temperature of evaporating tube
24 Drying air flow rate indicator It displays drying air volume into the evaporating tube
25 Spraying pressure indicator It displays the spraying pressure of sample.
(19)
(18)
(17)
(22)
(21)
(20)
(23)
(1)
(2)
(3)
(4)
(5)
(6)
(24) (25)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
(14)
(15)
(16)
-5-

No. Description Function
26 Stirrer operating lamp It illuminates when the stirrer operates
27 Stirring speed control dial To control stirring speed
28 Flow rate control knob To control the flow rate of sample.
29 Flow rate lock lever To fix the flow rate of sample
(27)
(26)
(29)
(28)
-6-

This unit has safety and alarm functions as below
When an abnormal operation occurs, solve it referring to [Troubleshooting] on P.18.
3-2 Safety and alarm functions
Safety features
Safety device Function Cause Measures
Mains switch /
Electrical leakage
breaker
It detects an excess
current of electric leakage
and turns off to shut down
the power.
Excess current.
Or electric leakage occurred due
to insulation deterioration etc. of
electric parts.
Immediately stop operation and contact the
company from which you purchased the unit or the
nearby service center.
Over temperature
protector
It detects over heating of
heater, and shuts down
the power.
Heater temperature is too high
because of low volume of drying
air
• Increase dry air volume.
• Clean the inlet filter or replace it.
If the above method does not cure the problem,
immediately stop the operation and contact the
company from which you purchased it or the
nearby service center.
For blower
protection
Thermal relay
It detects an excess
current of blower and
shuts down the power for
blower.
Excess current flows into the
blower due to blockage, suction
of rubbish, etc. of outdoor air
inlet at the back side of the unit,
• Remove what blocks the outdoor air inlet.
• Clean the exit filter or replace it.
-7-
Alarm function
Alarm name Alarm operation Alarm cause
Temp. alarm Temp. Alarm
Temp.alarm lamp illuminates Inlet temperature is 10°C or higher for 3 minutes or longer time.
Sensor alarm [_ _ _][⁻⁻⁻]blinks on the temperature
indicator of which sensor alarm occurs. Sensor input value is abnormal.
Outlet temp. upper
limit
alarm
[Outlet temperature] blinks The outlet temperature exceeded100°Cfor more than 5 min.
Outlet temperature becomes 120ºC or higher.
Drying air volume
alarm Flow Alarm
Drying air volume lamp illuminates Drying air volume is 0.1m3/min
Heater alarm Heater Alarm
Heater alarm lamp illuminates. Current of heater lowers unusually.
SSRalarm [SS r]is indicated on the inlet temp.
indicator Heater current exists while the heater output is OFF.
Pump line pressure
alarm Pump Line Alarm
Pump line alarm lamp blinks Pump line pressure exceeds 100kPa for1min. or more.
Spraying pressure
alarm [Spraying pressure] blinks. Spraying pressure at 0kPa lasts more than 30 sec.
Spraying pressure exceeds 300kPa.
*The alarm will be canceled by turning Off the power switch and turning it On again.
*In the event of alarm, all key operations become invalid.
*In the event of alarm, buzzer sounds.
To stop the buzzer sound, press "Temp. control/blower RUN/STOP key."
(This just stops the buzzer sound but the alarm state continues until the unit is rebooted)

4-1 Installed place
Installation
4
Do not install in a dangerous circumstance.
As this unit equips a heater, there is the fear of
fire in the dangerous circumstance.
Caution
4-2 Environmental conditions
Main unit approx. 110kg
Caution
Place to be installed;
●The part where no hazardous materials such as
flammable materials, pyrotechnic materials nearby
●Ambient temperature between5and35°C.
●The part where no condensation is.
●Places with little humidity, where water drops do not
reach.
●Places with little dust.
●Places not exposed to direct sunlight.
●Well ventilated place.
●A level, stable and robust place.
Keep good ventilation space.
The following space (See the left figure)
from the ceiling surface, ceiling, etc. is
required as a minimum to maintain excellent
performance of the unit.
Caution
Be careful to transport unit.
-8-
30cm
50cm
70cm
70cm

4-3 Connection of utility
1) Push up the stopper levers (2 places in front
only) to unlock.
2) Move the unit to the installation place.
*If you move the unit on an uneven or ribbed
place, excessive shock may apply to the casters
and damage the unit: in this case, lift it up to move
it.
3) Once the installation place is determined, push
down and lock the stopper levers of the casters.
4) Adjust the level of adjusters (4) using a wrench
to place the unit horizontally.
*Float the casters from the ground by lifting up
the unit to fixate the unit.
Caster (with stopper)
Adjuster
The unit goes downThe unit goes up
-9-
Lock
Stopper lever

4-4 Connection of utility
Check the voltage, phase and
capacity of power and correctly
connect the unit.
Wrong power connection may cause fire or
an incident of electrical shock.
Do not use a branch socket or
table tap.
Excess current may burn the cable or cause
fire.
Never connect a ground wire to the gas piping
or water piping to prevent an incident of
electrical shock.
Connect a ground wire correctly.
Warning
WarningWarning
1)Check the voltage, phase and capacity of power to
which the product model is connected.
Power to which the product is connected is as follows.
2)No power plug is supplied.
Directly connect to the distribution board.
*Do not connect power yet here.
Connect the ground wire of the power cord to the
ground.
Product model Power necessary for
connection
Power Capacity
SD-1010 AC220Vsingle
phase 30A
-10-
If a ground wire is not connected, an incident of
electrical shock may occur.
Be sure to connect a ground wire.
Warning
Ground (green)
Power (white, black)
To AC220V single phase power

5-1 Preparation
Operation
5
-11-
Handle glass wares carefully.
Handle glass wares carefully not to break down.
Caution
5-1-1 Attaching evaporating tube
1) Place the rectification plate to the evaporating tube.
*There is no restraint etc. on direction of the current
plate.
2) Install the evaporating tube on the main unit.
Match the position of engraved mark ○on the front
of the evaporating tube with the position of
engraved mark ○on the evaporating tube
mounting portion of the main unit and insert the
evaporating tube into the main unit.
*During installation, you may put the evaporating
tube temporarily on the evaporating tube
receiving cradle of the main unit.
**When insertion becomes heavy (tight), replace O
ring inside the main unit.
3) After insertion into the main unit, rotate the
evaporating tube leftward at 45º.
A click on the evaporating tube mounting portion is
clutched and the evaporating tube is temporarily
fastened on the main unit.
4)Tighten the lock knob with the hand.
The evaporating tube is fastened on the main unit.
Rectification board
Evaporator
Evaporating tube
mounting portion
Engraved mark ○
Evaporator
Evaporating tube
receiving cradle
①Rotate
Lock knob
Insert
Installation
②Fasten
Code No. Part name Remarks
180210 O ring containing Si oil G175

5-1-2 Attaching cyclone
1) Fit the packing (SC) into the groove of the
connector of the evaporating tube.
2) Connect connectors of the cyclone and evaporating
tube, screw the connectors and fasten the
cyclone on the evaporating tube.
3) Insert the connection collar in the cyclone.
*When insertion becomes loose, replace two O rings
inside the connection collar.
CycloneConnector
Evaporator
Cyclone
5-1-3 Attaching outlet temp. sensor
1) Connect the connecting portion of outlet
temperature sensor to the connector on the main
unit.
2) Slightly loosen the cap on the sensor mounting port
of the evaporating tube and insert the outlet
temperature sensor.
3) Make adjustments so that the tip of the outlet
temperature sensor is positioned in the branched
tube center of the evaporating tube and then
fasten the cap to fix the sensor.
*Be careful not to touch the cable of the outlet temp.
sensor to the evaporating tube.
②Fasten
Connector
Sensor for outlet
temperature
Branch tube of evaporating tube
Cap
③Fasten
-12-
Insert
Connection collar
①Connect
Packing (SC)
②Insert ①Connect
Code No. Part name Remarks
275510 O ring for connection
collar

5-1-4 Attaching power collection container and separator
1) Place the upper connecting packing in the groove
of the upper cap of collection bottle (powder
collection container, separator).
*The collection container and upper connection
packing used in powder collection container are
common parts.
2) After connecting the collection bottle below each of
the evaporating tube and cyclone, fasten the cap
screw to fixate the collection bottle.
*The capacity of the supplied collection container is
600 mL. If the capacity is not enough, use an
optional powder collection container (1.2 L).
Cyclone
Evaporator
Cap screw
Upper connection packing
Powder collection container
-13-
Separator
Code No. Part name Remarks
146750 Powder collection
container 600mL Standard
accessory
147940 Powder collection
container 1.2L
5-1-5 Countermeasures against static electricity
When spraying and drying a sample vulnerable to
electrification of static electricity including salts in
general such as sodium chloride and calcium
chloride, paste aluminum tape (supplied with the
product) onto the cyclone and collection bottle
and connect aluminum tape to the ground
terminal.
Cyclone
Collection bottle
Aluminum tape
Ground terminal
Insert
Fasten

5-1-6 Attaching silicone tubing
1) After raising the lever, pull up the tube retainer
immediately above and remove it from the
sample delivery pump main.
Use a proper tubing.
Use the tubing diameter and material
designated by our company for the tubing
used to deliver a solution.
The use of non-designated tubing may cause
degradation of performance or trouble.
Caution
Do not press the pressure sensor too
much or do not pour water on it.
Applying unreasonable force to the pressure
sensor main or cable or pouring water on it
may result in failure of the pressure sensor.
Caution
Tube retainer
Lever
Delivery pump main
Length reaching sample
container
Length reaching pressure sensor
(approx. 100 mm)
Pull
Pull
Pressure sensor
Spray nozzle
Silicon tube
(ID3.7 x OD6.1 mm)
-14-
Code No. Part name Qty.
125510 Silicon tube 3.15 x 5.2 5m
125530 Silicon tube 3.7×6.1 5m
2) Set silicon cut in an appropriate length on the
delivery pump main and fix the tube retainer.
3) Push down the lever of tube retainer and fixate the
tube.
4) Slightly pull the base of tube to bring it out about 2
mm.
5) Connect one side of silicon tube to the pressure
sensor.
6) Joint the pressure sensor and spray nozzle with
silicon tube cut in an appropriate length.
*Set the silicon tube (supplied with the product)
"3.15×5.2 [mm] (ID×OD)" with the delivery
pump.
[Tube used]
*Use a silicon tube (supplied with the product) in
"3.7×6.1 [mm] (ID×OD)" between the pressure
sensor and spray nozzle.

5-1-7 Connecting power cord
1) Connect the power cord of the unit to the breaker of
switch board.
* The unit must be connected to AC200V, single
phase (or three phase), and needs 30A or more
as the capacity of breaker.
If you use three phase source, connect R phase
(red) to S phase (white) or T phase (black) to
phase (white).
In either case, securely connect the ground wire
of power cord to the ground terminal via type D
grounding work.
Check the voltage, phase and capacity of
power and correctly connect the unit.
Wrong power connection may cause fire or
an incident of electrical shock.
Never connect a ground wire to the gas piping
or water piping to prevent an incident of
electrical shock.
Connect a ground wire correctly.
Warning
Warning
-15-
5-1-8 Connecting exhaust hose
1) Attach cuffs at both ends of exhaust hose.
2) Attach exhaust hose to the blower exhaust port and
fix it with stainless steel diamond clamp.
3) Put another side of exhaust hose into a draft
chamber or outside of laboratory.
*The length of supplied exhaust hose is 2.5 m.
If the length is not enough, use the optional exhaust
hose (10 m).
Code No. Part name Remarks
179980 Exhaust hose 2.5mStandard
accessory
179990 Exhaust hose 10m
Blower exhaust port
Exhaust hose
Stainless steel diamond clamp
*Install the exhaust
hose outlet in a fume
hood or outdoor.
Cuff
Table of contents
Other EYELA Dryer manuals
Popular Dryer manuals by other brands

LG
LG DL X748 E Series owner's manual

Electrolux Professional
Electrolux Professional TD6-24S installation manual

Beko
Beko DRY934CI user manual

SMC Networks
SMC Networks IDFB22E-11 Operation manual

Beko
Beko DRYPOINT M eco control DEC 1-30S Original Installation And Operation Manual

Casselin
Casselin CB2INOX manual