Festivo MED User manual

Installaon and operaon manual
Medical and Laboratory Cabinets
• MED, MEDLAB / MED SAVE, MEDLAB SAVE
medicin and laboratory cabinets
• BB bloodbank
Rev 092020b


ENGLISHENGLISH
Contents
Purpose of use 4
Recepon 4
Installaon 4
Introducon 5
Control unit MED XW70L 7
Control unit keys 7
Use of temperature control 8
Funcondisplaylights 9
Alarms 10
Pr1userlevelparametersfortemperatureanddooralarmsandtheirdelay
mesaswellassensortemperaturevalues 11
ControlunitMEDSAVEXW757K 12
Funcondisplaylights 13
Use of temperature control 14
Settheclockme 15
Storageoftemperatureandalarmdata 16
HowtoexportdataandalarmstoUSB 16
Clearingtemperatureandalarminformaonfromthedevicesmemory 16
Alarms 17
Pr1userlevelparametersfortemperatureanddooralarmsandtheirdelay
mesaswellassensortemperaturevalues 18
P4controlsensor 20
Baerybackup(standardonBBmodels,oponalonMEDmodels) 20
Cleaning 21
Noisesgeneratedbythecabinet’srefrigeraonsystem 22
Operaonfailures 23
Disposaloftheequipment 23
Guarantee 24
Changeofthedoor’shandedness,1-doormodels80,120,200 25
Changingthehandednessofdoor,onedoormodels320and400 26
Changeofthedoor’shandedness,2-doormodels160/160and400 28
DIN58345accessorypackageforMEDSAVEcontrolunit 30
AddionalNTCprobeandRJ45cabel(client’sprobe,accessory) 30
3

ENGLISH
Purpose of use
The Med, Medlab and BB refrigerators and freezers and combinaon cabinets are
specically designed for use in the health care and laboratory sectors. They are for the
storage of temperature sensive pharmaceucals or samples. The operator is responsible
that if the cabinet is being used for the storage special items that the equipment is
compliant with the standards required for such storage.
Recepon
Examine the product carefully as soon as you receive it. If you noce that it has suered
from damages during transportaon, you must indicate this in the delivery document and
report to the carrier and the retailer.
Note!
Themanufacturer’sguaranteedoesnotcoverdamagescausedbytransportaon!
Installaon
The ambient temperature must not exceed +32°C or
fall under +16°C. Avoid placing the cabinet close to any
equipment that radiates heat.
Note!
Thecabinetmustnotbeusedoutdoorsnorexposedtorain.
Note!
Accordingtotheelectricsafetyregulaonsnoliquid
materialsmustbekeptabovethecabinet.
Note!
Makesurethereisenoughroomfortheaircirculaon
aroundthecabinet.Nevercovertheairintakeattheboom
ofthecabinet.
The cabinet may be fastened to the back and side
surfaces, but if the side surface is in contact with
another refrigeraon unit or an exterior wall, it is
required to leave a 1-5 cm venlaon gap between the
surfaces. Keep the rear edge gap free for air circulaon.
Note!
Highambienthumiditymaycondenseonthesurface
oftheglassdoor.
Condensaon in the form of ice will be generated on the
evaporator of the refrigerator/freezer, this is defrosted by the cabinet automacally or
manually (oponal feature) by user. Defrost water is lead to the evaporaon vessel in the
compressor secon or to into the bacterial protected water tank inside the cabinet. Water
evaporates from the evaporaon vessel automacally. In the case of a malfuncon of the
unit some water can be spilled onto the oor. We recommend placing a safety container at
the locaon of the compressor.
4

ENGLISH
Adjust the cabinet to ensure it is horizontal in all
direcons, by using the adjustable feet, check with
spirit level if necessary, always ensure that all feet
are in contact with the ground.
Fasten a double-door refrigeraon cabinet, a
combinaon cabinet, glass door cabinet and
MEDLAB 320 to the wall so as to prevent it from
falling forward. There is a fastening brackets at the back edge of the unit (photos 1 and 2).
Screw a screw through the bracket into the wall. Note! There is no bracket installed in LAB
400 models.
When cabinet is delivered and if temperature is cold, condensaon may form on the
exterior surface. Do not connect power unl unit is dry and at room temperature!
Connect the connecon cord of the cabinet to an earthed 230 V / 50 Hz wall socket with
its own 10 A safety fuse. Check the voltage and frequency of the cabinet from the rang
plate. The power socket must be posioned such that it is possible to pull the unit’s plug
from the socket without having to move the unit.
Note!
Thedamagedpowersupplycordmustonlybechangedbyaprofessionalelectrician,
authorizedbythemanufacturer.
Note!
Donotuseelectricalequipmentinsidetheunit,unlessitisrecommendedbytheunit’s
manufacturer.
Introducon
Before using it, clean the cabinet carefully with mild cleaning agent, e.g. liquid dishwasher
detergent. Rinse and wipe dry. Leave the door open for airing for a moment.
Put the shelves and drawers in place. The high edge of the shelf is placed against the back
wall of the cabinet.
Before loading with goods, ensure the cabinet has reached the required holding
temperature (See chapter ”Use of temperature control”).
Cabinet is equipped with automacal defrost water evaporaon system.
1
2
5

ENGLISH
Note!
Cabinetsareveryhermec.Freezer(F)cabinetdoordoesnotopenjustaerclosing.Wait
foramomentunlthepressureisequalizedanddooropensslightly.
Note!
Toavoidevaporatorcorrosion,itisimportanttokeepgoodswithhighacidcontent/
vinegaretc.inairghtcontainers.
Note!
Donotpreventvenlaonontheoor.Productsmustneverbeplaceddirectlyonthe
oorofthecabinet,butonlyontheboomshelf.Whenplacinggoodsontotheshelves
ensuretheydonotoverhangthefrontoftheshelves,asthiswillalsointerruptair
circulaon.Nevercoverthefanonthebackwall,ortheventsonthebothsidesofthe
backwall.
6

ENGLISH
Control unit MED XW70L
SIngle- and double door refrigerators, single door freezer cabinets
Double door combinaon cabinets
Control unit keys
Not in use
Temperature control key
Temperature control key
manual defrost
To display target set point:
in programming mode it selects a parameter or conrm an operaon
LED light
This switch can be used in models with glass doors if having the light on when
the door is closed is preferred. Normally, the light switches on and o
automacally with the door switch. A signal light displays on the screen when
the light has been switched on from the controller switch .
ON/OFF controller
Lower part chiller
or freezer
Upper part chiller
MED XW70L
7

ENGLISH
Use of temperature control
Switch on the cabinet
• Switch the power on by pressing the key .
Adjust temperature
• Press once key , then the display will show the set temperature.
• Adjust temperature with keys and .
• Press again key or wait for 15 seconds and the display will revert to the normal
state.
Additional defrost (manual defrost)
• Press key , for more than 3 seconds to start addional defrost.
Addional defrost cannot be terminated by pressing the key. The defrost funcon
stops automacally.
Note!
Addionaldefrostdoesnotturnonifthereisnoneedfordefrosng.
To lock and unlock the keyboard
• Keep pressed together for more than 3 s the and keys. The “POF” message will
be displayed and the keyboard will be locked. If the keyboard is locked it is impossible
to adjust the temperature or switch of the power of the unit.
• Keep pressed together for more than 3 s the and keys ll the “Pon” message
will be displayed.
Control of the minimum and maximum temperatures recorded
• Press buon once. ”Lo” will appear on the display for a moment followed by the
minimum temperature logged.
• Press buon . once. ”Hi” will appear on the display for a moment followed by the
maximum temperature logged.
• The equipment will revert to the normal operaon in 5 seconds.
Reset the the minimum and maximum temperatures recorded
• Press buon ”Lo” or ”Hi” and the recorded temperature will appear on the
display. Immediately press and hold buon more than 3 s and the ”rst” will blink
on the display. Recorded temperature will be reset during blinking. The equipment will
revert to the normal operaon automacally.
Reset both minimum and maximum recorded temperature separately.
MED XW70L
8

ENGLISH
Funcondisplaylights
Light on Compressor is running
Light ashing Time delay to protect compressor aer start-up
Light on The evaporator fan is on
Light ashing Fans delay aer defrost in progress.
Light on Defrost is on
Light ashing Time delay aer defrost
Light on Internal LED light is on
Light on Alarm; see chapter ”Alarms”
MED XW70L
9

ENGLISH
Alarms
HA High temperature alarm
The temperature inside the cabinet has exceeded the set alarm temperature. The alarm
can be insgated by sudden icing of the evaporator. Iniate addional manual defrost.
See instrucons on the previous page. Ensure the cabinet has not been loaded with warm
products. Check that a door has not been le open. Check that any product is not blocking
air circulaon within the cabinet. The cabinet is meant to keep ready chilled products cold.
LA Low temperature alarm
The temperature inside the cabinet has fallen below the set value. Ensure that the products
put into the cabinet are not too cold.
The temperature alarm will reset automacally aer the interior of the cabinet returns to
the set temperature range. If this does not happen, move the products to another cabinet
and contact your service provider.
dA Door open alarm
Close the door carefully and the alarm will reset automacally.
HA2 Condenser alarm (MEDLAB 400 models only)
MEDLAB 400 cabinet alarms of dirty condenser / condenser lter. Check that ambient
temperature is not more than +32°C and clean condenser / lter. If alarm recurs call to
service.
P1 Thermostat probe failure*
P2 Evaporator probe failure*
P4 Condenser probe failure (MEDLAB 400 models only)*
*Note! The cabinet operates on its standby system and will hold normal temperature, but
reset the alarm signal and immediately contact your service provider.
EE Data failure
To reset an ‘EE’ Alarm on the display and resume normal funcon of the equipment press
any key for 3 seconds. Aer this ‘rSt’ message will be displayed again for 3 seconds and the
equipment should return to normal funcon.
MED XW70L
10

ENGLISH
Pr1 user level parameters for temperature and door alarms and
theirdelaymesaswellassensortemperaturevalues
Press simultaneously + A text “ALC” will appear on the screen.
To change the parameter ALC, ALU, ALL, ALd or did search with the arrow buons or
and press at the desired parameter. Adjust the value with the arrow buons or
and conrm with selecon buon .
The easiest way to turn the alarm on when using factory sengs is by leaving the door ajar
(door switch acvates) for one minute, which results in open door alarm (voltage free alarm
relay control)
Alarms
ALC alarm mode (rE or Ab)
rE (default factory seng)
• temperature alarm limit changes when adjusng the temperature
> for example with the sengs Set=4, ALU=4 and ALL=2 the alarm thresholds will be
at temperatures of +8°C and +2°C aer ALd delay.
ALU maximum temperature alarm
• seng 4 in both C and F models
ALL minimum temperature alarm
• seng 2 in C models and 3in F models
Ab(opon):
• xed alarm threshold with ALU and ALL remaining unchanged while adjusng
temperature
• ALU and ALL are set to preferred alarm threshold
Delays
• Ald; temperature alarm delay (seng: 20 min)
• did; door alarm delay (seng 1 = 1 minute, 0 = immediate)
db1-db4 sensor temperature values
Move with the arrow buons and to the desired sensor, and press , The
screen will display the temperature of the selected sensor. If the sensor is not in use, a text
”noP” appears on the screen. Return to the menu using the buon . The device will
return automacally to normal mode within 15 seconds.
MED XW70L
11

ENGLISH
ControlunitMEDSAVEXW757K
SIngle- and double door refrigerators, single door freezer cabinets
Double door combinaon cabinets
USB connector for copying data stored in the controller
LED light switch (standard in GD models)
Copying the recorded data to a memory sck
Temperature and alarm recording buon
Temperature control key
In programming mode it browses the parameter codes increases
the displayed value.
Temperature control key
in programming mode it browses the parameter codes or decreases the
displayed value.
To display target set point:
in programming mode it selects a parameter or conrm an operaon.
Push for 3s to access SET point programming
Push for 5s to start a manual defrost
ON/OFF controller
Lower part chiller
or freezer
Upper part chiller
MED SAVE XW757K
12

ENGLISH
Funcondisplaylights
Light on Compressor is running
Light ashing Time delay to protect compressor aer start-up
Light on The evaporator fan is on
Light ashing Fans delay aer defrost in progress
Light on Defrost is on
Light ashing Drip me in progress
Light on An alarm is occurring, see ”Larms”
Light on Data recording (REC) is on
Light on Baery status OK (baery backup)
Light ashing Baery is being charged (baery backup)
Light ashing Baery is faulty or empty (baery backup)
MED SAVE XW757K
13

ENGLISH
Use of temperature control
1. Switch on the cabinet
• Switch the power on by pressing the key 3 seconds.
2. Adjust temperature
• Checking temperature sengs: Press the buon shortly, and the temperature seng
displays on the screen.
• Press key for 3 seconds and ”SEt” message will be displayed. Release the buon
immediately then the display will show the set temperature.
• Adjust temperature with keys and .
• Press again key or wait for 60 seconds and the display will revert to the normal
state.
Note!
Press key formorethan5secondsanaddionaldefrostwillbeacvated(dEF).
3. Additional defrost (manual defrost)
• Press the buon for 5 seconds, and the screen will display the text “SET”. Keep the
buon pressed down and the text ”dEF” will appear on the screen. The extra thawing will
start. The addional thaw funcon cannot be switched o using the buon. Thawing will
end automacally.
Note!
Addionaldefrostdoesnotturnonifthereisnoneedfordefrosng.
4. To lock and unlock the keyboard
• Keep pressed together for more than 3 s the and keys. The “POF” message
will be displayed and the keyboard will be locked. In lock mode, only minimum and
maximum temperatures can be checked. Other funcons are locked.
• Keep pressed together for more than 3 s the and keys ll the “Pon” message
will be displayed.
5. Control of the minimum and maximum temperatures recorded (P4 additional probe)
• Press buon once. ”Hit” will appear on the display for a moment followed by the
maximum temperature logged.
MED SAVE XW757K
14

ENGLISH
• Press buon once. ”Lot” will appear on the display for a moment followed by the
minimum temperature logged.
• wait for 5 seconds and the display will revert to the normal state.
6. Reset the the minimum and maximum temperatures recorded
• Press buon ”Hi” or ”Lo” once and the recorded temperature will appear on
the display.
Immediately press buon more than 3 s and the ”rst” will blink on the display.
Recorded temperature will be reset during blinking. The equipment will revert to the
normal operaon automacally.
Reset both minimum ”Lo” and maximum ”Hi” recorded temperature separately.
Settheclockme
Set the clock me by pressing 3 s and message will be displayed
Press and message will be displayed
Conrm by pressing and select HOURS (for example. 9) by buons och .
Conrm by pressing and go through all the me sengs:
Hur = hour
Min = minutes
Udy = day of the week (Sun,Mon,tuE,Ued,thE,Fri,SAt)
dAy = day of the month
Mon = month
yEa = year
Wait for 60 seconds and the display will revert to the normal state.
MED SAVE XW757K
15

ENGLISH
Storageoftemperatureandalarmdata
Press the buon for 3 seconds. Use the arrow keys to select or parameter
“YES” and conrm the funcon with the buon , so the recording starts and the
indicator light is displayed . The device stores the informaon in its memory.
When you want to stop recording, select “no” according to the previous instrucon, and
the recording stops and the indicator light the display turns o.
HowtoexportdataandalarmstoUSB
If you want to copy the data stored on the device, connect the external USB memory
device to the USB connector of the device.
Press the buon for 3 seconds, in which case the stored data is copied to an external
USB memory. During copying, texts dAt, Lod, ALr, and Lod are displayed.
When the data is copied, the word “End” is displayed.
If data copying fails, “Err” will be displayed.
Clearingtemperatureandalarminformaonfromthedevices
memory
Press simultaneously + for 3 seconds, whereby text “Pr1” will be displayed.
Conrm the selecon with and select with arrows and “rSd” or stored
temperature data.
Press the buon and use the arrow keys to select and ”YES”. Conrm the
selecon with the buon , whereby the temperature data is erased from the memory.
MED SAVE XW757K
16

ENGLISH
MED SAVE XW757K
17
When the temperature data is erased from the memory, “rSA” i.e., the stored alarms is
displayed.
Press the buon and use the arrow buon to select ”YES”. Conrm the selecon
with , whereby the alarm informaon is erased from the memory.
Wait 60 seconds aer which the unit will return to normal mode.
Alarms
HA I High temperature alarm (internal temperature probe 1, TR3)
The temperature inside the cabinet has exceeded the set alarm temperature. The alarm
can be insgated by sudden icing of the evaporator. Iniate addional manual defrost.
See instrucons on the previous page. Ensure the cabinet has not been loaded with warm
products. Check that a door has not been le open. Check that any product is not blocking
air circulaon within the cabinet. Temperature alarm will be cancelled automacally when
the inside temperature has returned to set limits. The cabinet is meant to keep ready
chilled products cold.
LA I Low temperature alarm (internal temperature probe 1, TR3)
The temperature inside the cabinet has fallen below the set value. Ensure that the products
put into the cabinet are not too cold.
The temperature alarm will reset automacally aer the interior of the cabinet returns to
the set temperature range. If this does not happen, move the products to another cabinet
and contact your service provider.
PFI Internal temperature probe fail (internal temperature probe 1, TR3)
PF2 Evaporator probe fail (probe 2)
HA3 Condenser high temperature alarm (condenser probe 3, TL1)
Cabinet alarms of dirty condenser / condenser lter. Check that ambient temperature is
not more than +32°C and clean condenser / lter. If alarm recurs call to service.
LA3 Condenser low temperature alarm (condenser probe 3, TL1)
PF3 Condenser probe fail (condenser probe 3, TL1)
HA4 Addional probe high temperature alarm (addional probe 1, TL2)
LA4 Addional probe low temperature alarm (addional probe 1, TL2)
dA Door open alarm
Close the door carefully and the alarm will reset automacally.

ENGLISH
Pr1 user level parameters for temperature and door alarms and
theirdelaymesaswellassensortemperaturevalues
1. Pr1 user level parameters for temperature and door alarms
Press simultaneously + buons for about 3 seconds whereby text “Pr1” will be
displayed.
Press the buon and use the arrow keys to search and muuteava parameter
AlC, AlU, AlL, Ald, A4U, A4L, A4d or did to be changed.
Press the buon at the parameter to be changed.
Change the value with the arrow buon and . Conrm your selecon with
buon.
By default, the alarm can be accessed by holding the door ajar (door switch releases) for
one minute, aer which the door-open alarm releases (checking for potenal-free alarm
relay).
AlC = alarm mode
AlU = upper limit for alarm P1
AlL =lowerlimitforalarmP1
AlL = alarm delay P1
A4U= upper limit for alarm P4
A4L=lowerlimitforalarmP4
A4d = alarm delay P4
did = door alarm delay
MED SAVE XW757K
18

ENGLISH
2. P1 probe Alarms
AlCalarmmode(rELorAbS)
rEL (selected as factory default):
• the temperature alarm limit changes when the temperature changes.
For example, with the seng Set = 4 and AlU = 4 and AlL = 2, alarms come at + 8C
and + 2 ° C aer Ald delay
AlUmaximumtemperaturealarm
• seng value 4 for C and F models
AlL minimum temperature alarm
• seng value 2 for C models and 3 for F models
AbS(selectable):
• xed alarm limit, whereby AlU and AlL do not change when the temperature is changed
• AIU and AIL are set to preferred alarm threshold
3. P4 bole sensor alarms
P4 sensor has always set (AbS) alarm thresholds, which are set with parameters A4L and A4U:
Factoryvalue: MEDchiller MEDfreezer
A4L A4U A4L A4U
1 19 -30 -13
4. Delays
• Ald temperature alarm delay (seng value 20 min) P1
• A4d temperature alarm delay (seng value 10 min) P4
• did Door alarm delay alarm (Setpoint 1 = 1 minute, 0 = immediately)
5. db1-db4 temperature values
db1=controlsensorP1
db2=defrostsensor(vaporsensor)P2
db3=sensorforcondenser/condenserlterP3
db4=addionalsensorfortemperaturemonitoring(so-calledP4monitoringsensor)P4
Press the buon for 3 seconds and use the arrow keys and to move to the
desired sensor and press , whereby the temperature value of the selected sensor is
displayed.
Press the buon , to access the next parameter. If the sensor is not in use, text ”nPr” is
displayed.
Keep browsing the menu by buon. The unit will automacally return to normal mode
within 60 seconds.
MED SAVE XW757K
19

ENGLISH 20
P4 control sensor
The P4 control sensor is installed in the liquid bole on the
rear wall of the cabinet. Loosen the bole cap to remove it
from the rack.
Fill the bole almost full with liquid and push the sensor
through the cap into the bole.
Check that the sensor is in the middle of the liquid so it
does not catch the edges of the bole.
Baerybackup(standardonBBmodels,oponalonMED
models) A
1. Connect the cabinet to the mains and turn the key (A) 90° to the right,
whereby the baery backup is switched on.
Note!
Do not leave the key in the lock, but store the key in an agreed locaon.
2. Baery backup can be turned o during transport or transfer of the unit. Turn
the key 90° to the le, so that baery backup is switched o.
This manual suits for next models