06 07 08 09 10
Maintenance
CLEANING: Clean the exterior with a damp
cloth. Do not use detergents or solvents.
BATTERIES: The 9V battery must be replaced
when the low battery icon is displayed.
Pull off the rubber boot, unscrew the battery
cover, and replace the old battery.
FIELD CALIBRATION: Calibration should be
performed regularly forthehighestaccuracy.
The calibration is offsetting the temperature
forthat jack forthat particularthermocouple.
1 Stabilize distilled ice water.
2 Immerse the bead of the T1 thermocouple.
3 Select T1.
4 Use the T1 calibration pot to dial in 32.0°F.
5 Repeat for T2.
Specifications
Specifications good in ambient conditions of 73°F ±9°F
(23°C ±5°C), <75% relative humidity
Display: 3.5 digit LCD, 2000 count
Over range: "OL" or "-OL" is displayed
Range: -58 to 2000°F (-50 to 1300°C)
Resolution: 0.1°
Accuracy:
±4°F @ -58 to 32°F
±(0.3%rdg +2°F) @ 32 to 1100°F
±(0.5%rdg +2°F) @ 1100 to 2000°F;
±2°C @ -50 to 0°C
±(0.3%rdg +1°C) @ 0 to 600°C
±(0.5%rdg +1°C) @ 600 to 1300°C
Accuracy after field calibration:
±1°F @ 30 to 120°F, ±6°F @ 0 to 50°F
Temperature coefficient:
0.1 x (specified accuracy) per °C (0 to 18°C, 28 to 50°C)
Measurement rate: 2.5 readings/second
Battery type: 9V, NEDA 1604, JIS 006P, IEC 6F22
Battery life: 200 hours typical alkaline.
is displayed when the battery needs to be replaced.
Operating environment: 32°F to 122°F (0°C to 50°C) at <75% RH
Storage temperature: -4°F to 140°F (-20°C to 60°C), 0 to 80% RH
(with battery removed)
Limited Warranty
This meter is warranted against defects in material or workman-
ship for one year from date of purchase from an authorized Fieldpiece
dealer. Fieldpiece will replace or repair the defective unit, at its option,
subject to verification of the defect.
This warranty does not apply to defects resulting from abuse,
neglect, accident, unauthorized repair, alteration, or unreasonable use
of the instrument.
Any implied warranties arising from the sale of a Fieldpiece prod-
uct, including but not limited to implied warranties of merchantability
and fitness for a particular purpose, are limited to the above. Fieldpiece
shall not be liable for loss of use of the instrument or other incidental or
consequential damages, expenses, or economic loss, or for any claim of
such damage, expenses, or economic loss.
State laws vary. The above limitations or exclusions may not apply
to you.
Obtaining Service
Visit www.fieldpiece.com/rma for the latest information on
how to obtain service.
www.fieldpiece.com
© Fieldpiece Instruments, Inc 2017; v10
More Fieldpiece
There's always something new coming from
Fieldpiece Instruments. Find the latest news
and updates at www. fieldpiece.com
SC260 True RMS
Compact Clamp
Meter
SC660 True RMS Wireless
Swivel Clamp Meter
SMAN460 4 Port Wireless
Manifold and Micron Gauge
COM
CAT III
600V
Fieldpiece
Select
M/m
SC260
TRUE RMS
VAC
NCV
AAC
°F°C VDC
MFD
TEMP
AUTO OFF
TYPE K30V
MAX