FIS OralChroma CHM-2 User manual

OralChroma
HALITOSIS MEASURING DEVICE
CHM-2
Instruction Manual
Nissha FIS, Inc.

Introduction
Thank you for purchasing an OralChroma (Model: CHM-2), halitosis measuring device.
Before using this unit, please read this instruction manual carefully for correct usage, and keep
it near the product for easy accessibility..
* The contents of this manual are subject to change without prior notice.
* The contents of this manual are protected by copyright law.
* Unauthorized copying/reprinting of this manual is prohibited.
COPYRIGHT (C) Nissha FIS, Inc. All rights reserved.

Index
Safety Instructions............................................................................................................. 1
Precautionary Statements ................................................................................................. 2
Overview ........................................................................................................................... 4
Confirm supplied items ...................................................................................................... 6
Names and Functions of Parts .......................................................................................... 7
Setting up the unit ............................................................................................................. 9
Setting up the PC ............................................................................................................ 10
Installing and starting up the OralChroma DataManager................................................. 12
Importing data from the old version ................................................................................. 15
Initial set up of OralChroma DataManager ...................................................................... 19
Preparing for measurement............................................................................................. 27
Measurement .................................................................................................................. 36
Ending measurement ...................................................................................................... 45
Viewing measurement history ......................................................................................... 46
Measurement data window.............................................................................................. 58
Judgment window............................................................................................................ 68
System settings window .................................................................................................. 72
Maintenance.................................................................................................................... 81
Troubleshooting............................................................................................................... 83
Services........................................................................................................................... 84
Specifications .................................................................................................................. 85

1
Safety Instructions
Shown below are the descriptions and explanations of the warnings and precautions in this
instruction manual for the OralChroma.
DANGER:Failure to follow this instruction may result in loss of life or serious
injury.
WARNING:A warning identifies a potential hazard, which, if ignored, may result in
injury from electric shock, fire (burns), etc.
CAUTION:A caution identifies a potential hazard, which, if ignored, may result in
damage to the unit. The caution instructs you to follow the correct
operating procedure.
WARNING
Improper use of the unit may cause physical injury or unit malfunction.
For your safety, carefully read and understand this Instruction Manual
before you use the OralChroma. Please note that we are not liable for
any accident unless the cause of the accident is clearly attributable to
failure of the product.

2
Precautionary Statements
The following precautionary statements must be observed in order to use the unit safely and to
make the most of its functions:
WARNING
Avoid damaging or abusing the power cord. A short circuit may cause variations in
current, resulting in overheating and fire.
Make sure that the unit is grounded and that the plug is fully inserted into the power
supply. Ignoring this warning may cause overheating and fire.
Avoid inserting/disconnecting the plug into/from the power supply with wet hands.
Ignoring this warning may cause serious disaster due to electric shock.
Do not modify the unit in any way. It may cause an accident or a breakdown.
Do not insert the syringe for collecting oral gas samples deeply into the oral cavity.
Inserting the syringe deeply into the oral cavity may cause injury or respiratory
problems.
If you should recognize something unusual while using the machine, immediately stop
operation and contact the sales agent from whom you purchased this unit for inspection
and repair.
If the unit is used in a manner not specified in this manual, the unit’s protection may be
impaired.
CAUTION
Avoid installing, using, or storing the unit under the following conditions, which may
cause inaccurate measurements or damage of the unit, as well as significantly decrease
product life.
・Ambient temperature is outside the range of 10 to 30 °C.
・Ambient relative humidity is above 80%.
・Temperature and humidity change drastically and condensation occurs.
・Presence of odor or gas (including hydrogen, methane, odor of alcohol, formalin
and other chemicals, cigarettes, or aromatic oils, etc.)
・Vibration or impact directly affects the unit.
・Exposure to water, oil, chemical, or steam.
・Presence of dirt or dust.
・There is a serious induction problem where static electricity, magnetism, and noise
tend to be generated.
・Exposure to direct sunlight.
・The airflow from an air conditioner blows directly on the unit.
・Rooms in which gas- or kerosene-fired heaters are used without ventilation to the
outside

3
・Places where gas tends to accumulate, such as a low position near a floor.
This is a single-purpose oral breath-measuring unit. Never inject any material other than
breath samples (E.g., liquid substances including chemicals, water, and saliva, or any
kind of gas). Ignoring this warning may cause mechanical failure or performance
degradation.
To obtain a stable measurement result, leave the power on during office hours. If you
have not powered on the machine for more than one week, the measured value may be
lower than the actual value immediately after power on due to the characteristics of the
sensor.
If you are not using the unit for a long period of time, disconnect the plug from the power
supply. Ignoring this warning may cause fire.
Do not wipe the surface with benzine, thinner, alcohol, etc. Use a clean, dry cloth when
cleaning. Ignoring this warning may cause inaccurate measurements, degradation of the
unit, or fire.
Do not wipe the syringe by rubbing it with alcohol. Do not touch the syringe with a bare
or gloved hand that has been exposed to an antiseptic solution.
Use only designated syringes.
Allow ten minutes before measuring if the subject has used a product containing
alcohol, such as mouthwash or toothpaste. It may affect the measurement results, or
influence measurement value.
Do not place anything on the machine. Ignoring this warning may cause overheating or
damage.
Do not disassemble the unit. We do not guarantee the accuracy of data or operation of
the unit if the unit has been disassembled.

4
Overview
The OralChroma measures and classifies volatile sulfur compounds in human breath samples
into three causal components (hydrogen sulfide, methyl mercaptan, and dimethyl sulfide), then
displays each gas concentration on your PC. It is easy to use and requires only four minutes to
analyze the sample. It provides more opportunities for people to check their oral breath.
■ Features
・Measures volatile sulfur compounds in the oral breath sample; classifies the sample
into three causal components of halitosis: hydrogen sulfide, methyl mercaptan, and
dimethyl sulfide; and displays each gas concentration on your PC in units of ng/10 ml
and ppb.
・Measures gas concentrations with a high degree of accuracy by using a supersensitive
semiconductor gas sensor and a column developed only for the purpose of detecting
halitosis.
・Does not require a carrier gas cylinder although the gas chromatograph method has
been adopted. Thus, the unit can be considered portable. It is very easy to operate and
does not require any specialized knowledge. Just connect it to the power supply and
use immediately.
・Requires only 4 minutes to measure the concentration of all three components.
・Small and light, the OralChroma is easy to carry.
・Real-time graphical representation of the measured data is produced by connecting the
machine with a PC and using specially designed software.

5
■ Mechanism of measurement
The following constitutes the procedure for measuring oral breath concentrations:
1. Collect oral breath samples and inject the specified volume of gas into the column.
2. Separate volatile sulfur compounds from the gas by using the column and detect the
gases with the semiconductor gas sensor.
3. Calculate gas concentrations from the peak patterns obtained.
■ Components and configuration
Shown below are the components and configuration for the OralChroma.
Pump Filter Flow rate sensor
Sensor
Control unit
Column Oral gas
Ambient clean air
Exhaust
Inlet
Temperature controller
Column unit
Power
supply
PC
(Data)
Supplied syringe

6
Confirm supplied items
When you receive the OralChroma, immediately check for any damage or abnormal conditions
caused during delivery. Check thoroughly to ensure that all accessories are included. If there
should be any missing components, damage, or malfunctions, i.e., the unit does not operate
according to specifications, contact the agent from whom you purchased the unit.
The unit and all accessories are included in the box. The box and all packing should be kept,
as they might be needed later if you send the unit back to us for servicing or warranty repair.
■ List of the contents
Components included Quantity
Checkbox
OralChroma unit 1
Power cable 1
USB cable 1
Application CD including:
・OralChroma DataManager software
・Instruction manual
・USB driver software
1
Syringe for collecting oral gas samples 100
Warranty 1
Notice of operation 1
Confirm that all items listed above are included in the package and place a mark in the
checkbox.

7
Names and Functions of Parts
■ Top
■ Rear
POWER lamp
ERROR lamp
READY lamp
Power switch Lid
Gas inlet
USB port Main switch Fuse holder AC power jack

8
■ Functions
Place Name Functions
Top
Lid
Manually open/close
Usually closed to prevent dust from coming
into gas inlet.
Gas inlet
Inject oral gas sample into this inlet with a
supplied syringe.
Power
switch
Power on/off the unit.
Keep pushing until buzzer sounds.
POWER
lamp
Orange light comes on when power is turned
on.
ERROR
lamp
Red light comes on when error occurs.
READY
lamp
Green light comes on when the unit is ready
for measurement.
It goes off when the unit is in Waiting Status*
or under measurement.
Rear
Main switch
Turning this switch on (“|”) enables you to
turn Power switch on.
Turn this switch off if not used for a long
period.
AC power
jack
Connect a power cable to this jack.
Fuse holder
(Power
supply fuse)
Check the fuse in the holder when
abnormality occurs.
Rating:3.15A, 250V
(Mfr: Littelfuse, Model: 213_3.15)
USB port Connect the supplied USB cable to this port.
*Waiting Status: The status under which internal sensor and column are being stabilized after
power-on or measurement.

9
Setting up the unit
■ How to set the unit before measurement.
1. Place the unit on a flat and stable place.
CAUTION
Position the OralChroma unit so that power cord and USB cable can be easily
connected to or disconnected from the unit.
2. Connect power cord to the unit, then to receptacle of AC supply.
3. Do not connect the unit to PC with the supplied USB cable until the USB driver soft is
installed. See page 10, “Setting up the PC” for details.
CAUTION
Do not start measurement.
PC setup is required initially.

10
Setting up the PC
You need a personal computer (PC) and the supplied CD including “OralChroma
DataManager” software to start measurement.
Prepare the PC specified below by yourself.
■ System requirements
OS
Windows XP(SP3)、Windows Vista(32/64bit),
Windows 7(32/64bit), Windows 8(32/64bit)
Computer Hardware on which the above OS runs without problems.
Display XGA(1024×768) or higher
I/O USB port: 1 port
CD drive 1 unit
Note: Install USB driver software in the supplied CD before installing DataManager.
CAUTION
Only the user with administrator privilege should install/un-install OralChroma
DataManager.

11
■ Installing USB driver software
1. Power on the PC and start Windows.
2. Insert the supplied CD into the CD drive.
3. When the set up window appears, click “USB driver.”
4. Follow the directions displayed on the window to proceed with the installation.
5. When the window below appears, click OK to complete the installation.

12
Installing and starting up the
OralChroma DataManager
■ Installing OralChroma DataManager
1. Insert the supplied CD into the CD drive.
2. The setup program is automatically started. At this time, the warning window below may
appear depending on your security status. If it does, click YES.
3. When the setup window appears, click “OralChroma DataManager” to proceed the
installation.
4. When the window below appears, click “OK” to complete the installation.

13
5. The following folder where the DataManager is installed appears.
6. Double-click “CreateShortcutIcons” to create shortcut icons for DataManager and its
manual on the desktop.
7. Close the setup window. At this time, the confirmation window below may appear
depending on your security status. If it does, click “This program installed correctly.”

14
■ Starting OralChroma DataManager
1. Double-click “OralChroma” shortcut icon on the Windows desktop. When you first
activate the program after installation, a message “Select COM port.” appears. Click
“OK.”
2. The OralChroma DataManager top window appears as below. This window is referred to
as “Main window.”
■ Uninstalling OralChroma DataManager
Delete “OralChroma” folder which is usually created under “C:\.” Since the DATA folder is
initially in this “OralChroma” folder, copy the DATA folder to another place by using “Export
data folder” (See page 76.)

15
Importing data from the old version
Data files between current and old version DataManager softwares are not compatible.
If your PC has the old version, its data should be imported into the current version in advance.
■ How to import
1. Click “System settings…” on Main window.
2. Select “General” tab to click “Import data….”

16
3. Data importing window appears.
4. Select “CHM-1” for “Data of” column.
5. Click “Search the PC for CHM-1 data folder.”

17
6. When PC finds the CHM-1 data folder, it is automatically imported in the “Folder” column
and the data are shown as below. Click “Import.”
Note: If CHM-1 data folder cannot be found…
When your OS is Windows Vista, 7, or 8, the CHM-1 data folder may not be found,
depending on your security settings on the OS. In this case, check “Show hidden files,
folders, and drives” in View tab of Folder Options window. After that, click “Browse” to
specify the folder below:
C:\Users\XXX\AppData\Local\VirtualStore\Program Files\OralChroma\DATA
XXX: Your log-in name
Program Files: This may be Program Files (x86) if the OS is 64 bit.
7. The message, “Start to import” appears. Click “OK.”
Table of contents
Popular Laboratory Equipment manuals by other brands

Noztek
Noztek fusionX user manual

Velp Scientifica
Velp Scientifica MST DIGITAL F203A0450 instruction manual

Asylum Research
Asylum Research cypher user guide

Oxford Instruments
Oxford Instruments Cryojet Operator's handbook

Sutter Instrument
Sutter Instrument P-97 Operation manuals

Four E's Scientific
Four E's Scientific MI0103002 operating instructions