Setting-Up the Nebulizer
1. Securely attach the nebulizer inlet to the port on the front panel and the nebulizer outlet to the
patient circuit
2. On the ventilator front panel, press the Extended button.
3. Tap the Nebulizer control button. The control button turns orange, and a pop-up list displays the two unit
options: ON and OFF.
4. To activate the nebulizer, tap the control button to select ON; to deactivate the nebulizer, tap the control
button to select OFF. When the nebulizer is activated a countdown timer will appear on the upper right
corner indicating the amount of time left.
5. Press the OK (Enter) button to confirm your selection.
Your selection (ON or OFF) is displayed on the control button.
The nebulizer period controller is configured when setting up the ventilator.
Configuring the nebulizer period controller
1. On the ventilator front panel, press the Technical button.
2. Tap the Nebulizer Period control button. The control button turns orange.
3. Adjust the nebulizer period using the +/- button until you reach the desired value.
4. Press the OK (Enter) button to confirm your selection.
Your selection is displayed on the control button.