
845A
b. Turn the recorder on.
c. Proceed as outlined in paragraph 2-13 or 2~15, as
desired.
d. Adjust the ISOLATED OUTPUT LEVEL control for
the desired output to the recorder. This control
has alog taper so that smooth control is possible
at both high and low settings.
and forth will dissipate this charge, eliminating the
problem, Witii ahigh source impedance, the response
of the instiTiment is unavoidably slow due to the low pass
filter used to suppress superimposed noise. However,
the design of the low pass filter is such that common
mode rejection is extremely high while the response
time for the normally encountered low source imped-
ances is very fast.
2-31. OVERLOAD VOLTAGES
The ISOLATED OUTPUT current capability
is 100 microan^eres with a5kilohm source
impedance.
2-24. OPERATING NOTES
2-25. SPURIOUS VOLTAGES AND CURRENTS
2-26, Voltage measurements at the microvolt level
involve the persistant problems of thermoelectric ef-
fects. These effects may be compensated for by tempo-
rarily disconnecting the voltage from the circuit under
measurement and noting the meter deflection of the
Model 845 on the desired range. This reading must
then be subtracted from all subsequent voltage measure-
ments. Athorough understanding of these effects can
lead to reducing or eliminating them completely.
2-27. THERMOELECTRIC VOLTAGES
2-28. Ha circuit is composed of two dissimilar metals,
anet voltage will result if the two dissimilar junctions
are maintained at different temperatures. These ther-
moelectric voltages, also known as thermals, thermo-
couple voltages, or Seebeck voltages, can be reduced
by using metals having low thermoelectric potentials and
keeping all junctions at the same temperature. The
terminals of the Model 845 are made of pure copper,
gold-flashed to prevent tarnish. For lowest thermal
voltages, all connections to the Model 845 should be
made with pure copper wire. Silver plated copper or
solder coated copper also produce satisfactory results.
Tinned copper is less satisfactory than silver plated or
solder coated copper .Nickel and nickel- based alloys
are not suitable for connections to the instrximent. Ex-
cellent results can be obtained using ordinary TV twin
lead, or even lamp cord if high insulation resistance
is not required, if shielding is necessary, use alength
of flat braid over the cable.
2-29. HIGH SOURCE IMPEDANCE
2-30. Due to the very high input resistance and ex-
treme sensitivity of the Model 845, it is charge sensi-
tive. Thus, aperson’s body potential, an electrostatic
voltage, can cause charge redistribution at the input to
the instrument and result in meter needle deflection as
a hand approaches the input terminals. Careful shielding
will eliminate this problem. Also, due to charges that
may be deposited on the input terminals when the OPR
switch is set to ZERO, an appreciable transient will
result when the switch is set to OPR if nothing is con-
nected to the input terminals. Turning the switch back
2-32. The instrument is designed to withstand up to
HOO volts dc or 1100 volts peak ac continuously applied
between any two of the three input terminals, or between
cabinet ground and any of the three input terminals,
regardless of the setting of the RANGE or OPR switch.
However, repeated or continuous overloads above 200
volts in the ranges below 3millivolts will result in ex-
cessive dissipation in the protective, low-pass -filter
resistor RllO. This will result in thermal voltages
which may take several minutes to subside after the
overload is removed.
2-33. GUARDING
2-34. The instrument has an inner chassis connected
to the GUARD terminal on the front panel. Ordinarily,
this GUARD terminal is striped to the COMMON termi-
nal. When connected in this way, the inner chassis
serves as ashield. This greatly improves the leakage
resistance to ground and the common mode rejection.
However, since the inner chassis is available at the
GUARD terminal, it may be driven at the same voltage
as the COMMON terminal. This further increases the
leakage resistance and common mode rejection by about
ten times. The voltage used to drive the GUARD termi-
nal should be obtained from aseparate source or by
means of avoltage divider connected directly across
the source so that the leakage currents do not cause
voltage drops across impedances in the circuit under
measurement.
2-35. INCREASING INPUT RESISTANCE
2-36. In the 1microvolt to imillivolt ranges, a10
megohm resistor is connected directly across the input
of the instrument. The input resistance may be in-
creased on these ranges by disconnecting the 10 megohm
resistor where it attaches to the RANGE switch. How-
ever, the input resistance will no longer be well defined.
Typical input resistances with the 10 megohm resistor
removed, are as follows:
RANGE INPUT RESISTANCE
1microvolt
3microvolt
10 microvolt
30 microvolt to 1millivolt
300 megohms
1000 megohms
3000 megohms
10, 000 megohms
2-5