manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Fluke
  6. •
  7. Measuring Instrument
  8. •
  9. Fluke CPS-40M-HC40 User manual

Fluke CPS-40M-HC40 User manual

PN 5007263
July 2018
©2018 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies. Specifications are subject to change without notice.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
ООО «Флюк СИАЙЭС»
125167, г. Москва,
Ленинградский проспект
дом 37,
корпус 9, подъезд 4, 1
этаж
1
CPS-40M-HC40
Contamination Prevention System
CPS-100M
High-Pressure Contamination Prevention System
Instructions
Introduction
The Fluke CPS-40M-HC40 (CPS-40M) Contamination Prevention System and CPS-100M High-Pressure
Contamination Prevention System (Product) are accessories that protect pressure controllers from
contamination by the Device Under Test (DUT).
• CPS-40M-HC40 (CPS-40M) has a maximum-working pressure of 44 MPa (6400 psi).
• CPS-100M has a maximum-working pressure of 110 MPa (16 000 psi).
When the CPS is enabled, the pressure controller controls the CPS valves to decrease the test system
pressure, purge contaminants, and completely vent the test system.
The Product is connected to the pressure controller by a high pressure hose and a control cable. There is an
optional vacuum pump connection (CPS-40M only). All upward pressure changes are accomplished by the
pressure controller. Rough downward changes in pressure are accomplished through the Product's vent
valve. This results in most of the gas being exhausted out of the Product instead of going back to the
controller. Contaminants from the DUT are filtered and liquids drop into a sump. The sump is automatically
purged into a waste bottle when the system is vented. The product includes both mesh and coalescing filters
that are replaceable.
CPS-40M-HC40/CPS-100M
Instructions
2
How to Contact Fluke Calibration
To contact Fluke Calibration, call one of the following telephone numbers:
• Technical Support USA: 1-877-355-3225
• Calibration/Repair USA: 1-877-355-3225
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31-40-2675-200
• Japan: +81-3-6714-3114
• Singapore: +65-6799-5566
• China: +86-400-810-3435
• Brazil: +55-11-3759-7600
• Anywhere in the world: +1-425-446-6110
To see product information and download the latest manual supplements, visit Fluke Calibration’s website at
www.flukecal.com.
To register your product, visit http://flukecal.com/register-product.
Safety Information
A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies
conditions and procedures that can cause damage to the Product or the equipment under test.
WWarning
To prevent personal injury:
• Read all safety information before you use the Product.
• Wear eye protection.
• Do not use with flammable or combustible liquids.
• Use the Product only as specified, or the protection supplied by the Product can
be compromised.
• Carefully read all instructions.
• Do not use the Product if it operates incorrectly.
• Do not use the product if it is altered or damaged.
• Disable the Product if it is damaged.
• Do not attempt to operate the Product above its rated pressure.
• Do not disconnect tubing when the system is pressurized.
W Caution
To avoid damage to the Product, carefully follow the cleaning and decontamination
instructions in the manual. Do not use unapproved solvents or cleaners on the Product.
Contamination Prevention System
High-Pressure Contamination Prevention System
Symbols
3
Symbols
The symbols shown in Table 1 are found in these instructions or on the Product.
Table 1. Symbols
Box Contents
The Product is either of these kits:
CPS-40M:
• CPS-40M-HC40 Test Station
• Pressure Hose, JIC #4 (AN4) fittings
• Pressure Hose Adapter Fittings, JIC #4 (AN4):
1/4 NPT; 1/4 BSP; 7/16-20 (x2)
• Vacuum Hose, 3/8 inch polyethylene
• Vacuum Hose Adapter Fittings, 3/8 tube:
1/4 NPT; 1/4 BSP; 7/16-20 (x2)
• PK-ADPTR-HC40 Adapter Set
• CPS Cable
• Instructions
CPS-100M:
• CPS-100M Test System
• High-Pressure Hose, DH500 Fittings (cone
and threaded connection compatible with
Autoclave F250C and HIP HF4)
• CPS Cable
• Instructions
Symbol Definition
W WARNING, RISK OF DANGER.
 Consult user documentation.
) Certified by CSA Group to North American safety standards.
P Conforms to European Union directives.
 Conforms to relevant Australian Safety and EMC standards.

This product complies with WEEE Directive marking requirements. The affixed label indicates that
you must not discard this electrical/electronic product in domestic household waste. Product
Category: With reference to the equipment types in the WEEE directive Annex I, this product is
classed as category 9 “Monitoring and Control Instrumentation” product. Do not dispose of this
product as unsorted municipal waste.
CPS-40M-HC40/CPS-100M
Instructions
4
Features
The CPS-40M features are shown in Table 2. The CPS-100M features are shown in Table 3.
Table 2. CPS-40M
5
3
2
1
6
10
987
4
ifc01.eps
Number Description
 Stand
 Pressure Hose
 Tee
D Vacuum Hose
E TEST port connector
F LED Indicator
G Drain Tube
H Filter Cover
I Filter Housing
J Body
Contamination Prevention System
High-Pressure Contamination Prevention System
Features
5
Table 3. CPS-100M
4
5
8
7
6
3
1
2
ifc02.eps
Number Description
 Stand
 High-pressure Hose
 DUT Port
D LED Indicator
E Body
F Drain Tube
G Filter Cover
H Filter Housing
CPS-40M-HC40/CPS-100M
Instructions
6
Set Up
This section details Product set up.
W Warning
To prevent personal injury, use thread sealing tape only on NPT fittings to avoid
incorrect sealing. Do not use on JIC #4 (AN4), BSP, SAE or DH500 fittings.
CPS-40M-HC40
For atmospheric calibration:
1. Ensure the O-ring is on the SAE side of a SAE to JIC #4 (AN4) adapter.
2. Install the above adapter to the CONTROLLER port of the CPS.
3. Install the applicable adapter (JIC #4 (AN4) to NPT, BSP or SAE) to the TEST port of the pressure
controller.
4. Connect the pressure hose to the two adapters and tighten with a wrench.
5. Attach the CPS cable to the connector on the rear panel of the CPS, plug the other end into the solenoid
driver connector on the rear-panel of the pressure controller.
6. Install the free end of the drain tube into a suitable container to collect fluid contaminants.
For sub-atmospheric calibration:
1. Ensure O-ring is on the SAE side of a SAE to 3/8 tube adapter.
2. Install the above adapter to the VAC port of the CPS and tighten with a wrench.
3. Install the applicable adapter (3/8 tube to NPT, BSP or SAE) to the EXHAUST port of the pressure
controller and tighten with a wrench.
4. Place the CPS in desired location noting that test gauges will likely be mounted on top of it.
5. Plan the cuts on the 3 needed vacuum tubing sections; vacuum pump to tee, tee to CPS, and tee to
pressure controller. See Table 2. Cut the vacuum tubing into these three pieces.
6. Connect the vacuum tubing to the tee and adapters (to vacuum pump, VAC port on CPS, and
EXHAUST port on pressure controller) by normal nut and ferrule tube connection methods.
CPS-100M
To set up the CPS-100M:
1. Slide the gland nut over the threaded center of the hose end fitting as shown in Figure 1.
2. Screw the collar onto the left-hand thread hose end fitting, counter-clockwise.
3. Repeat steps 1 and 2 for the other hose end.
4. Install the High-Pressure Hose from the CONTROLLER port on the CPS to the TEST port on the
pressure controller. Torque to 15 N · m (11 lbf · ft) with a wrench. Excessive torque will lead to
premature leaks that will require replacement of the fitting.
5. Attach the CPS cable to the connector on the rear panel of the CPS, plug the other end into the solenoid
driver connector on the rear panel of the pressure controller.
6. Install the free end of the drain tube into a suitable container to collect fluid contaminants.
Contamination Prevention System
High-Pressure Contamination Prevention System
Operation - DUT Connection
7
A
C
B
ife004,eos
Figure 1.
W Caution
To prevent Product damage, before you use the CPS-100M, enable the CPS on the
pressure controller (SETUP>Instrument Setup External Values>CPS) – refer to
controller operators manual. The CPS must be enabled or the controller could be
exposed to contamination. When enabled, the LED illuminates.
Operation - DUT Connection
W Caution
To prevent Product damage or damage to a gauge:
• To prevent incorrect sealing that creates an unsafe condition, do not use thread-
sealing tape or any other sealing method on the DUT or the adapters. The Gauge
Adapter sealing system can be hand-tight sealed up to 44 MPa (6400 psi).
Wrenches or similar tools are not required. Over tightening can cause damage to
threads or sealing faces.
• Before connection, ensure that there are O-rings fitted to the test port and gauge
adapter.
• Check that the sealing face of the device to be fitted is clean and undamaged, as
scratches or dents can form leak-paths.
• Any DUT that is known to be full of liquid, or grossly contaminated, should be
cleaned before using with a CPS and pressure controller. Also use the CPS Purge
and CPS Cleanout Tasks to perform a final clean of the DUT before calibration.
CPS-40M-HC40/CPS-100M
Instructions
8
CPS-40M-HC40
1. Select the appropriate adapter from the adapter set to match the thread on the DUT.
2. Screw the adapter fully onto the DUT so that the bottom face of the DUT seals on the O-ring inside the
adapter. Hand-tight is sufficient.
3. Turn the adapter collar counter-clockwise until the adapter seals on the O-ring in the mounting post to
mount the DUT/adapter assembly to the test port. Hand-tight is sufficient.
4. To adjust the position of the DUT, hold the adapter, and turn the collar clockwise 1/4-turn.
5. Position the DUT to face the desired direction, and turn the collar counter-clockwise to re-seal.
CPS-100M
The DUT mounts directly to the test port on the CPS-100M, with a metal-to-metal DH500 connection.
Torque to 15 N · m (11 lbf · ft) with a wrench.
1. To adjust the position of the DUT, unscrew the gland nut counter-clockwise 1/4-turn.
2. Position the DUT to face the desired direction, and tighten the gland nut clockwise re-seal.
Operation - CPS Tasks
Tasks are accessed by the Setup>Tasks menu path on the controller’s display screen.
Purge
Purge is used to void the test system connected to the CPS of liquid or particulate contaminants by
pressurizing and venting the test system plumbing through the CPS (same as normal operation). Fluke
Calibration recommends to do this before a calibration is performed if gross contaminants might be present.
To purge the system, the Product pressurizes to the user-defined purge pressure and then vents to the drain
tube. The controller repeats the process for the number of cycles that are set.
To run a purge process:
1. Select the Purge Task to enter the menu.
2. Set the Pressure, 350 kPa (50 psi) for example, but not higher than the rating of the DUT.
3. Set the number of cycles.
4. Select Run to start the purge process.
Figures 2 and 3 show the normal operaton and the Purge function of both models in a cross-section.
Contamination Prevention System
High-Pressure Contamination Prevention System
Purge
9
Ifc06.eps
Figure 2. Cutaway of the CPS-40M-HC40 during a Normal Downward Pressure Excursion, or Purge Task
ifc15.eps
Figure 3. Cutaway of the CPS-100M Purge Task during a Normal Downward Pressure Excursion, or Purge Task
CPS-40M-HC40/CPS-100M
Instructions
10
CPS Cleanout
CPS Cleanout is a function that pushes contaminants from the sump of the CPS out the drain tube. A low-
pressure setpoint is automatically set to establish a flow to clean out the CPS for the specified amount of
time. This same function occurs each time the CPS is controlling downward or venting with the pressure
less than 170 kPa (25 psi).
Disconnect the Product
To disconnect the Product, ensure that the system is vented and then disable the CPS in the controller
settings. Make sure that the CPS is disabled in the firmware if it is physically disconnected from the
controller. Failure to do so could result in the controller not being able to vent pressure. Once disabled in the
firmware, the CPS can be disconnected both electrically and pneumatically. Do not leave the CPS physically
connected to the controller if it has been disabled in the firmware.
Maintenance
Inspect the drain tube and waste bottle/container after use. If the drain tube or liquid from the tube shows
signs of contamination (particulates or oily substances), the coalescing filter and mesh filter should be
inspected. The frequency of inspection will vary depending on the condition of Devices Under Test (DUT)
that have been calibrated. With time and experience, you can determine if a regular inspection interval is
more appropriate, or if the filters should be inspected when the waste bottle/container is full.
Push on the plastic ring on fitting  to disconnect drain tube , and pull gently on the drain tube. See
Figure 2.
1
2
3
ifc03.eps
Figure 2.
Contamination Prevention System
High-Pressure Contamination Prevention System
Maintenance
11
1. Unscrew the filter cover  counter-clockwise to release the coalescing filter D and O-ring (see Table 4)
E. See Figure 3.
3
4
5
ifc05.eps
Figure 3.
2. Unscrew filter housing F counter-clockwise to release mesh filter H, spring I and O-ring (see Table 4)
G. See Figure 4.
CPS-40M-HC40/CPS-100M
Instructions
12
6
7
8
9
ifc04.eps
Figure 4.
The coalescing filter D cannot be cleaned, and should be replaced when heavily discolored or showing
signs of deterioration (see Table 4).
3. Clean the mesh filter H with suitable solvent and replace if necessary (see Table 4).
4. Clean interior chambers of filter housing F to remove any contaminant particles or residue.
5. Fit the O-ring, mesh filter, and spring to housing and screw clockwise into the main body.
6. Fit coalescing filter and the O-ring to the filter cover  and screw clockwise into the filter housing.
Replacement Parts
Replacement parts are listed in Table 4.
Table 4. Replacement Parts
Number
(See Figures 3 and 4) Description Part Number
D Coalescing Filter 4578779
E O-ring (above coalescing filter) 4840807
G O-ring (below mesh filter) 3922019
H Mesh Filter 4976723
Contamination Prevention System
High-Pressure Contamination Prevention System
Specifications
13
Specifications
Operating Temperature ............... -20 °C to +50 °C
Storage Temperature ................... -20 °C to +70 °C
Operating Humidity ...................... 5 % to 95 % relative humidity, non-condensing
Wetted Materials .......................... Stainless Steel, Aluminum, Nitrile, PTFE, borosilicate glass
CPS-40M-HC40
Pressure Range ........................ -0.1 MPa to 44 MPa (-15 to 6400 psi)
Weight ...................................... 3900 g (8.0 lb)
Dimensions ............................... 175 mm x 185 mm x 300 mm (6.9 in x 7.3 in x 11.8 in)
CPS-100M
Pressure Range ........................ 0 MPa to 110 MPa (16 000 psi)
Weight ...................................... 3630 g (8.6 lb)
Dimensions ............................... 175 mm x 185 mm x 270 mm (6.9 in x 7.3 in x 10.7 in)
Safety
General ..................................... IEC 61010-1; Pollution Degree 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This
warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration,
contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other
warranty on Fluke’s behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service
center to obtain return authorization information, then send the product to that Service Center with a description of the
problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL
OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or
countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this
limitation of liability may not apply to you.
11/99

This manual suits for next models

1

Other Fluke Measuring Instrument manuals

Fluke 323 Use and care manual

Fluke

Fluke 323 Use and care manual

Fluke 1630-2 User manual

Fluke

Fluke 1630-2 User manual

Fluke i800 User manual

Fluke

Fluke i800 User manual

Fluke RUSKA 2465A-754 User manual

Fluke

Fluke RUSKA 2465A-754 User manual

Fluke 43B User guide

Fluke

Fluke 43B User guide

Fluke OptiView Series User manual

Fluke

Fluke OptiView Series User manual

Fluke 7 Series User manual

Fluke

Fluke 7 Series User manual

Fluke SONOCHEK User manual

Fluke

Fluke SONOCHEK User manual

Fluke 6050H User manual

Fluke

Fluke 6050H User manual

Fluke i430-Flexi-TF-II User manual

Fluke

Fluke i430-Flexi-TF-II User manual

Fluke endurance series User manual

Fluke

Fluke endurance series User manual

Fluke ScopeMeter 123B Operating instructions

Fluke

Fluke ScopeMeter 123B Operating instructions

Fluke 831 User manual

Fluke

Fluke 831 User manual

Fluke VT305 User manual

Fluke

Fluke VT305 User manual

Fluke 635 User manual

Fluke

Fluke 635 User manual

Fluke i800 User manual

Fluke

Fluke i800 User manual

Fluke i5s-PR 1750 5A User manual

Fluke

Fluke i5s-PR 1750 5A User manual

Fluke 945 User manual

Fluke

Fluke 945 User manual

Fluke 345 Instruction manual

Fluke

Fluke 345 Instruction manual

Fluke 5606 User manual

Fluke

Fluke 5606 User manual

Fluke 3210-PR-TF-II User manual

Fluke

Fluke 3210-PR-TF-II User manual

Fluke PQ Window User manual

Fluke

Fluke PQ Window User manual

Fluke LS-1310 Manual

Fluke

Fluke LS-1310 Manual

Fluke 481 User manual

Fluke

Fluke 481 User manual

Popular Measuring Instrument manuals by other brands

BOREAL LASER GasFinderMC-2 Operation manual

BOREAL LASER

BOREAL LASER GasFinderMC-2 Operation manual

Omni Instruments bmcm USB-AD12f manual

Omni Instruments

Omni Instruments bmcm USB-AD12f manual

CLA-VAL X144D e-FlowMeter Quick Start Installation & Removal Instructions

CLA-VAL

CLA-VAL X144D e-FlowMeter Quick Start Installation & Removal Instructions

catec HG600 user manual

catec

catec HG600 user manual

Qiagen Q-Base user manual

Qiagen

Qiagen Q-Base user manual

JENCO 3321 Operation manual

JENCO

JENCO 3321 Operation manual

Lorex LW2760 Series quick start guide

Lorex

Lorex LW2760 Series quick start guide

geo-FENNEL FL 510HV-G TRACKING user manual

geo-FENNEL

geo-FENNEL FL 510HV-G TRACKING user manual

Cortalk RMU5 Generic Installation and configuration guide

Cortalk

Cortalk RMU5 Generic Installation and configuration guide

Bosch gll3-80 P Original instructions

Bosch

Bosch gll3-80 P Original instructions

GS Instruments Protek 7830 user manual

GS Instruments

GS Instruments Protek 7830 user manual

netvox R718N3 E Series user manual

netvox

netvox R718N3 E Series user manual

Servomex NanoTrace DF-745SG Operator's manual

Servomex

Servomex NanoTrace DF-745SG Operator's manual

Blue-White F-300 quick start guide

Blue-White

Blue-White F-300 quick start guide

Pepperl+Fuchs IMUF99 B16 Series manual

Pepperl+Fuchs

Pepperl+Fuchs IMUF99 B16 Series manual

Meter-Master 100EL operating instructions

Meter-Master

Meter-Master 100EL operating instructions

POSTBERG+Co measuringSYSTEM MA-Di operating instructions

POSTBERG+Co

POSTBERG+Co measuringSYSTEM MA-Di operating instructions

Nellcor OXIMAX N-560 Service manual

Nellcor

Nellcor OXIMAX N-560 Service manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.