Forte Zephair Installation guide

1
Pain Management System Maintenance &
Care Instruction

DISCLAIMER
This product user manual is provided as a guide for the safe operation of the Forté Zephair Mattress.
Whilst all care has been taken to comprehensively identify any risk associated with the operation of
the Zephair Mattress, Forté Healthcare Solutions cannot be held liable for circumstances outside of
it’s control including inappropriate device usage. We strongly recommend that for any clarification
required by care sta or indeed anyone using the Zephair Mattress that appropriate consultation is
made before using the device. Forté Healthcare cannot be held liable for patient outcomes due to
clinical, circumstantial and care factors outside of its control.
TABLE OF CONTENT
OPERATION 3
PATIENT SAFETY AND CARE 4
TROUBLE SHOOTING 5
MAINTENANCE 6
CLEANING - Cover 7
CLEANING - Foam core 7
MANUAL HANDLING 7
2

1 The Zephair Mattress is designed to be used with the pump at all times
2 Position the Zephair Mattress on the bed in a manner that prevents trapping and/or damage.
Pay careful attention to bed rails (if fitted) and ensure they will not rub excessively against the
Zephair. Mattress damage as a result of incorrect fitment is not covered by warranty.
3 It is preferable to mount the pump to the foot-end of the bed. The pump may be placed on the
floor if needed. If the pump is placed on the floor it is important it is suciently protected from
damage.
4 Connect the pump to a power outlet. The Pump power lead may be located within the power
cord loops on the side of the Zephair Mattress. Never use the Zepahir Mattress if the power cord
is in any way damaged.
5 Visually inspect the pump before operation. If there is any evidence of damage or is defective,
contact an authorized representative before use.
6 To activate, turn the pump on by the top switch
7 Use the lower switch to set either Alternating or Static Therapy. Alternating Therapy will provide
the greatest level of pressure injury protection.
8 Use the weight gauge on the right hand side of the pump face to adjust the pump pressure to
ensure the patient is suciently supported by the Zephair System.
9 Ensure Mattress Zipper remains closed at all times, only open for necessary inspection or
maintenance
10 Ensure foam core is not exposed for extended periods of time to UV light
11 Ensure the pump is not used within a wet environment or where there is risk of the pump coming
in contact with water or liquid
12 Ensure the outer covering is protected against damage by penetration of sharp objects. Damage
sustained to the cover as a result of puncture by sharp objects is not covered under warranty.
13 Regular observations must be made of the Zephair system whilst in operation to ensure that
there is no device malfunction or damage that will impact the patient in a negative manner.
14 Do not use the Zephair near fire sources or excessive heat. Product damage caused by heat
damage e.g. cigarette burns will not be covered by warranty.
OPERATION
3

Pain Management, High Level Palliative Care Mattress
PATIENT SAFETY & CARE
Maximum Patient Weight Capacity:180kg
Therapeutic Features
- Alternating 1:2 Cell System for Pressure Redistribution
- Narrow Cell Design Minimizing Patient Disturbance
- Quiet pump operation minimizing patient discomfort
- Dual interface immersion foams provide comfort, reduce ‘alternating sensation/motion
discomfort’, stabilize temperature and provide envelopment
- Sloped heel region for added pressure relief for susceptible heels
- Firm strengthened sides for lateral stability and support during repositioning and patient transfer
- Adjustable patient weight gauge for comfort and appropriate care
1 The Premiflex Ultra outer covering is waterproof yet vapour permeable. The outer covering
provides a hygiene barrier for the inner air and foam system. The outer covering alone is not
designed as an ongoing continence barrier nor is it a substitute for continence care products
such as pads.
2 Patients must be positioned correctly to avoid risk of suocation.
3 Avoid using excessive layers of bedding and/or covers as this can negatively impact the vapour
permeability and pressure redistribution/relief properties of the Zephair system.
4 Continual routine monitoring is required for patients on the Zephair system. Clinical judgement
must be used to ensure sucient frequency of intervals.
5 The Zepahir is not intended to be used in substitution of necessary patient care practices such
as adequate nutrition, repositioning, sucient continence management, nursing care as required
and wound/skin dressings, ointments and creams as deemed clinically appropriate.
6 Ensure the Zephair is cleaned regularly for patient hygiene and infection control. Refer to
cleaning guidelines for recommended cleaning procedure and cleaning agents.
4

Pain Management, High Level Palliative Care Mattress
TROUBLE SHOOTING
Low Pressure Alarm
1 Should the Low Pressure Alarm sound and/or Low Pressure Indicator be illuminated on the Pump
Control Face, perform the following inspection;
a. Check Air Supply Leads from Mattress are Connected Correctly to Pump
b. Check CPR port and ensure it is correctly fastened closed
c. Check Air Supply Leads for any damage
Should the fault persist, unzip the outer cover and lower compartment of inner cover and
perform the following inspection;
d. Check Air-Cells for any puncture by a sharp object. A quick check can be to inspect the
underside of the top cover as this will show any stain caused by fluid penetration through
the cover if a sharp object has penetrated
e. Check Air Cell Connections that they are secure and are not leaking
f. Should faults persist, contact your distributor or Forte Healthcare for further assistance
2 Cell/s not inflating;
a. Unzip the outer cover and base compartment of inner cover
b. Identify the cell/cells
c. Inspect the system for any air tube connectors that may have become loose or may be
leaking
d. Reconnect any air cells or connectors that require it
e. Replacement connectors and parts are available as spare parts if required
5

MAINTENANCE
Regularly assess for soiling. Refer to the below cleaning instructions
- If the product is used directly upon or in contact with a bed surface, periodically and ongoing
assess the bed surface for any sharp objects that may cause damage to the mattress cover
- Support Surfaces should not be lifted by the cover only. Support the foam core as well as the
cover.
- Always disconnect the pump from electrical power before beginning maintenance or repair
work. Never attempt to work on a pump that has been in contact with water/liquid without
consulting an approved technician.
- Use of unauthorized spare parts and/or accessories may void warranty
6

CLEANING & HANDLING
Cover Cleaning
Wipe o cover with an approved cleaning chemical (refer to approved chemical listing). Excrement,
blood spots, spills of any kind must be removed immediately with an approved cleaning agent or soapy
warm/cold water. Cleaning chemical usage instructions must be followed.
Usage of high concentrations of Phenol‐based and Chlorine chemicals and abrasive cleaning agents
can have a negative impact on the integrity of the mattress fabric coatings. . Forte Healthcare stipulates
a maximum Chlorine concentration of 1000ppm for all standard fabrics if it must be used for cleaning/
disinfection. All residue of cleaning agents should be completely dried from fabrics shortly after
cleaning.
Sterilisation might have a negative eect on polyurethane coatings and is therefore not recommended.
If one does decide to do so, a reduced lifetime of the product should be taken into account. This is
inevitable for all thermoplastic polyurethanes, due to their chemical nature.
Foam Core Cleaning
The outer covering protects the foam core, which means cleaning of the core seldom is necessary.
Should the foam core need to be cleaned, clean aected area with a mild cleaning chemical and water
(eg spray on). Gently press out excess water. Allow the foam core to dry completely in a ventilated area
avoiding direct exposure to sunlight as UV light can aect foam performance. Ensure that the foam
core is completely dried before using mattress.
Manual Handling
Ensure mattress is carried centrally and with 2 hands.
It is recommended for transport of Support Surfaces King Single Size and above that 2 people are
involved moving the mattress due to the mattress bulk and weight.
7
Table of contents
Popular Personal Care Product manuals by other brands

Permobil
Permobil ROHO LTV owner's manual

Laica
Laica NE1001 Instructions and warranty

Vulpes Electronics
Vulpes Electronics Smart heated kidney belt Welcome guide

Lookah
Lookah Seahorse Max user manual

Hans Grohe
Hans Grohe Logis Universal 73570 Series Instructions for use/assembly instructions

PROZIS
PROZIS Satin user manual

Dräger
Dräger PAS Micro Series Instructions for use

mitoredlight
mitoredlight MitoADAPT quick start guide

ATMOTUBE
ATMOTUBE Plus quick start guide

Orliman
Orliman M710 Use and maintenance instructions

Indoor biotechnologies
Indoor biotechnologies Ventia Instructions for use

Dicarre
Dicarre Noble ER 90 User Guide Application Instruction