Gehaka G610i User manual

User manual
Moisture meter G610i
Version: 1.00.000
Date: 11/2016


Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation of G610i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functions diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1. Measure - measuring moisture . . . . . . . . . . . . . . . . . .
2 Battery test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3 Set-up system . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1 Set date and time . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Set number of media elements . . . . . . . . . . . . . . . . .
3.3 Set number of decimals . . . . . . . . . . . . . . . . . . . . . .
3.4 Set printing mode . . . . . . . . . . . . . . . . . . . . . . . . . .
3.5 Set number of printed copies . . . . . . . . . . . . . . . . . .
3.6 Set contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.7 Load Internet equation . . . . . . . . . . . . . . . . . . . . . .
3.8. Load standard equation . . . . . . . . . . . . . . . . . . . . . .
3.9 Select curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.10 Auto report . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.11 Select language . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.12 Auto shuto . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.13 Auto oset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Serial Port RS232 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Change the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Extreme conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Technical specications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4
6
8
9
10
13
15
15
16
17
18
19
20
21
23
24
25
26
27
28
29
30
31
33
35
37
39
40

4I Moisture meter G610i
The Moisture Meter G610i gathers all the characteristics to provide the most accurate
reading on grains moisture.
With a single operation and without depending of the operator, the model G610i provides
a moisture reading withing few seconds, for a wide range of products.
With a simple keyboard of only 5 keys, it is possible to operate the instrument and
quickly setup it without requiring a long user training.
G610i enables the moisture analysis using the FLOW THRU technology. It is not necessary
to provide the sample external weighting, the temperature corrections, ad weights are
automatically performed, without using another resources, or additional tables.
The calibration curves for each cereal were obtained at the Gehaka laboratories, using
samples collected from several planting areas.
Using the standard stove method as reference, the measurement scales for each
product were created. These curves were transferred to the G610i memory and are
easilyidentiedbytheproductnameanditsversion.
If it is necessary to develop new calibration curves, there is a function in the web site
www.moisturetester.com.br that allows the creation, edition and installation in your
moisture meter G610i.
The G610i has three instruments into one, managed by a state-of-the-art internal
microprocessor. They are these three instruments: an electronic scale, which measures
the sample weight; a built-in digital thermometer, which measures the sample
temperatureinsidethechamberandtheG610itemperature,andnallyaCapacitance
meter, which indicated the percentage of moisture to
suer the correction from the other parameters further
up.
Its program runs all the required calculations, resulting in
extremely reliable and repetitive readings. It still performs:
automatic setting and checking of all the electronic circuits
for the good operation.
Introduction

Moisture meter G610i I 5
The FLOW THRU instrument drawing allows that quick readings are accomplished, that
is, in less than 40 seconds.
Simply select the product, spill the sample from the G610i Weighting Vessel, until it
BEEPS, load the sample,, the measurements will be taken and the moisture content will
beautomaticallyprovided.ThisallisdisplayedinanalphanumericdigitalLCDdisplay
(LiquidCristalDisplay), of easyreadingwith16characters and 2lines,inPortuguese
language.
TheG610ialsohasabidirectionalcommunicationporttypeserialRS232C.Therefor,it
is possible to connect the G610i to a printer so that the media can be printed in a label,
makingthecounterproofidenticationeasier,oritcanevenconnectthemetertoaPC
and transfer the information to a management system.
The G610i is run with an external power source type FULL RANGE, that is, the instrument
operationisassuredtomainsthatvaryfrom90to240VAC,protectingtheinstrument
against the mains oscillation.
It is agreed that the indication in the G610i display means:
ON Switches on the G610i
-> Right arrow
<- Left arrow
YES Operationconrmation
OFF Quit,Switcho

6I Moisture meter G610i
Description
1. SAMPLE LOAD HOPPER
The location where the sample of the cereal will
be spilled in order to determine the moisture.
2. DISCHARGE BUTTON
Through this Button, we discharge the sample
fromtheMeasurementChamber.Itwillbeused
at the end of the Moisture measurements.
3. LCD DISPLAY
Displays the results and settings with
alphanumeric characters.
4. ON
Switches on the G610i.
5. YES
ConrmstheSelectionofafunctionorsetting.
6. RIGHT ARROW
Displays the next function or increments a
value of a division.
7. LEFT ARROW
Displays the previous function or decrements a
value of a division.
8. OFF (ESCAPE)
Switches o the G610i, quick a function or
setting.
9. DRAWER
Receives the cereals sample from the chamber,
after the measurement was carried out.

Moisture meter G610i I 7
10. AC ADAPTER
Converter, switched source, it operates from
90 to 240VAC, if removed, is starts operating
with internal 9V battery.
11. SERIAL PORT CONNECTOR RS232
Data output to the printer or microcomputer.
12. WEIGHTING VESSEL
Sample receiver for weighting and used to spill
the sample from the Weighting Vessel.

8I Moisture meter G610i
Checkthepackagecarefully,iftherearemarksanddamagesinthecardboardbox,it
indicates that there were problems during the transportation.
Open the package, check the state of the G610i. Remove the equipment and check if all
the accessories are included.
Place the G610i on a stable table, so that it does not move during the operation. Remove
the adhesive tapes that hold the Drawer and check if it slides smoothly.
Connect the AC Adapter in the connector in the real panel and t it securely, taking
care to see if the plug is totally inserted in the connector. It is not necessary to check
the power supply voltage, the source is “Full Range”, and the mains selection is not
necessary. It is also recommended to use a voltage stabilizer.
When pressing the ON key, the G610i will start the operation, running a self-diagnosis
to check if its components are working properly. In the case the instrument veries
that any of is components is working incorrectly, it will display an error message in the
display regarding the failure.
The G610i can work connected in the outlet, or using an internal 9V battery. This
switching is automatically selected. It is important to not that, when working with the
9Vbattery,theDisplayBacklightwillbeswitchedoinordertosavethebatterycharge.
When the G610i is started, a percentage indicating the Battery status is displayed during
the self-diagnosis, note:
30% a 100% Normal operation
0% a 30% Request the battery change or charge
0% Indication of low battery, the G610i no longer works
The function “2. Battery Test” indicates the Battery voltage beyond the statue in the
percentage format.
The G610i shall be operated inside a location free from excessive dust and with a
temperaturewithin0ºto45ºC.
Installation of G610i

Moisture meter G610i I 9
Below, there is a diagram that indicates all the G610i functions and will help understanding
the operation.
Functions diagram

10 I Moisture meter G610i
1. Select“Measure”inthemainmenu,pressYEStoconrm;
2. Choicethetypeofcerealtomeasurethemoisture.UsetheRight/LeftARROWS
keysuntilyoundthedesiredproduct.Theindicationsbelowtheproductsnames
are the minimum and maximum moisture limits that G610i is able to read. Press
YEStoconrmyourchoice.
3. Get a product sample of at least 200g to which you want to know the moisture
and, using a plastic cup, slowly pour the cereal sample in the Weighting Vessel,
until the indication of 100% is displayed in the display, a beep will sound indicating
that the sample weight is correct. In the case that weight is exceeded, the display
will indicate a value higher than 100%, and the sample excess shall be removed
from the Weighting Vessel.
1. Measure - measuring moisture
WHEAT
Moisture: 13.09%
100.0% |
W a it ...
S pill
T h e S a m ple...
W a it ...

Moisture meter G610i I 11
4. Pour the weighted sample in the G610i Load Hopper,
using the Weighting Vessel edge as reference.
Observe the drawing at the side, the Weighting Vessel
shall pour the sample from backward to forward the
G610i, the sample shall be poured quickly, do not
pour it slowly because it decreases the instrument
repeatability.
5. After few seconds, the G610i will indicate the value for the product moisture in the
display. If the Right/Left keys are pressed, the G610i will display all the measures
it has performed, that is:
When you select the Right/Left Keys, we will have:
Thermometer
Measuring...
Te m p.: 2 4.8 oC
Date :11/16/16 >
Wheat
Moisture: 13.09% >
Capacitance meter
Measuring...
Curve: 001765
Simple: 9 >

12 I Moisture meter G610i
6. After successfully concluding the
measurement, the G610i will send all
the measurement data to the printer or
computer through the serial output. An
example of the report is shown below.
NOTE: If there is a large dierence in
temperature between the product
and the G610i Chamber, where
the Temperature Sensor is locate,
it will wait until there is a thermal
balance between both. It can cause
the temperature measurement
to take a longer time to be carried
out. Thus, we have improved the
instrument accuracy, measuring
the sample real temperature.
7. After carrying out the reading discharge
theChamberusingtheDischargeButton.
8. When the G610i is indicating the
measurement results on the display and
a new sample is poured in the Weighting
Vessel, the G610i will carry on a new
measurement using the same settings as
the previous measurement. If the OFF key
is pressed, the function MEASURE will be
quit, returning to the main menu.
NOTE: If the moisture content is higher
than 22%, the G610i will request
the sample to be measured again,
repeating it three times to assure a
higher accuracy in the reading.

Moisture meter G610i I 13
The Battery function is used to perform the measurement and indicate the battery
status at the moment. Its main function is to prevent it from running out of power during
an external measurement. Execute the procedures:
1. Press YES to enter in the “Battery” function. A screen will be displayed indicating
the Battery voltage and the charge percentage still remaining until it is totally
discharged. It is possible that a new battery indicates a percentage higher than
100% and it will be considered as discharged when it displays 0%. It is important
to note that the systems only switched to the Battery when the power Adapter it
is removed from the mains.
2. There are two errors that the Battery measurement system displays, they are
respectively:
This error indicates that an improper Adapter is being used to power-up
the G610i, there is a high risk of damages to the equipment electronics;
please, immediately remove the source that is being used and acquire a
new power Adapter at Gehaka.
This error shows that the battery is completely discharged; please provide
the battery change as shown later in this manual.
3. If the RIGHT ARROW is pressed, a report with the result will be printed, as shown
at the side:
9.1 Volts
100 %
Error 16:
Source > 13 Volts
Error 17:
Discharge Battery
2. Battery test

14 I Moisture meter G610i
4. Gehakastilloersanoptional,whichisthe
Rechargeable Battery for the G610i. This
Battery has the same mechanical aspect
as the disposable Battery; however it is
rechargeableNickelCadmium.Onlyinthis
case the power Adapter will e another with
voltage of 12V, enough to recharge the
Battery in 12 hours of loading. To verify if
the battery was totally charged, observe
the loading percentage in this function,
it must achieve approximately 120%,
and at this moment the Adapter can be
unplugged from the mains. The expected
autonomy for this Battery is of 10 hours
of continuous operation. There can have
variations depending on the model of
battery used.
Tooptimizethebatteryduration,letinONinthefunction“3.12AutoShuto”.
5. ToconcludethisfunctionpressOFF(ESCAPE),thereisasummaryoftheactions
of keys in this function below:
YESKEY: Noeectinthisfunction;
LEFT ARROW: Sends report by Serial;
RIGHT ARROW: Sends report by Serial;
PRESS OFF: Quit the function;

Moisture meter G610i I 15
3. Set-up system
Themode“3.Setup”allowsperformingsettingsorcongurationsintheG610i.Wewill
describe each of these functions below.
They are:
3.1. Set Date and Time
3.2. Set Number of Media Elements
3.3. Set Number of Decimals
3.4. Set Printing Mode
3.5. SetNumberofPrintedCopies
3.6. SetContrast
3.7. Load Internet Equation
3.8. Load Standard Equation
3.9. SelectCurves
3.10. Self-Diagnosis Report
3.11. Select Language
3.12. AutoShuto
3.13. SetOset
Observe that the number displayed at the beginning of each function corresponds to
the same number displayed in the G610i Menu; therefore, it is simpler to identify where
the explanation of each function is in the manual.
We will describe each of them below.
ThisfunctionperformsDateandTimesettingforthetealtimeclock(RTC).Thisvalueof
date and time will be sent to the reports by the Serial Port RS232 after the measurement.
The G610i is already supplied with the clock set at the factory, but it can eventually be
readjusted for the daylight saving time.
Internally, there is a battery that keeps the clock working even when the G610i is switched
ofromtheoutlet.Thisbatteryistolastmorethan5years,andtotestit,justswitch
otheG610iandobserveiftheclockstartsindicatingthetimeas00:00:80,ifitoccurs,
3.1. Set date and time

16 I Moisture meter G610i
contact the Gehaka Technical Assistance to change the battery.
To perform the Date and Time setting:
1. Search the function “Set-up” and press YES;
2. Search the function “3.1 Set Date and Time”, press YES;
3. Now, use the Right/Left ARROWS to Increase/Decrease respectively the value
forTIME.Whenthevalueiscorrect,pressYES,G610iwillskiptothenexteldto
be adjusted;
4. Repeat this procedure to perform the adjustment of MINUTES, DAY,
MONTH and YEAR.
NOTE: An underline will be displayed below the value that is being adjusted, after
pressing YES it will advance to the next item to be set.
3:00:00 PM
11/27/2013
3.2 Set number of media elements
This function performs the setting of the number of elements that will compose the
reading media used to calculate the moisture percentage nal value. We can select
between 1 to 5 elements.
To perform the setting for the number of Media Elements:
1. Search the function “Set-up” and press YES;
2. Search the function “3.2 Set Number of Media Elements”, press YES;
3. Use the Right/Left ARROWS to Increase/Decrease the value for the Number of
Media Elements. This setting is limited between 1 to 5 elements;
Measurements = 1
- < YES > +

Moisture meter G610i I 17
This function sets the number of decimal places shown in the display and in the reports
for the moisture. If one decimal place is used, the rounding of the second place will be
automatically performed. The standard from the factory is one decimal place.
To perform the setting for the number of decimal places proceed:
1. Search the function “Set-up” and press YES;
2. Search the function “3.3 Set Number of Decimals”, press YES;
3. Use the Right/Left ARROWS to increase/decrease the value for Decimals. This
setting is limited between 1 to 2 decimals;
4. Press YES to conclude the setting.
Places = 1
- < YES > +
4. Press YES to conclude the setting.
NOTE: It is important to note that independently from the setting performed in
this function when the moisture content exceeds 22%, the reading will be
ALWAYS performed with the average of three elements.
3.3 Ajusta número de decimais

18 I Moisture meter G610i
3.4. Set printing mode
This function directs all the data resulting from the measurement to the Printer or
Computer.
A. Printer:
In this mode, all the measurement data will be sent in the Report format as indicated
below by the Serial Port RS232.
If the G610i has a built-in Thermal Printer, this report will be simultaneously printed and
sent by the Serial Port RS232.
B. Computer:
TheComputerModeonlysendsdata,disregardingtheformating.Thedataisseparated
by “;” to make the separation easier. This report can be easily captured and interpreted
by a computerized system that is connected to the Serial Port RS232.
AnexampleofStringsentisprovidedbelow.TheconclusioniswithCRandLF.
15; 12.35; 141.7; 0.0; 26.9; 27.4; 66.4; Wheat; 20111101; G610i;
1.00.000;1.00;16:51;18/01/13;<CR><LF>
Descriptionofthedataeldssubmitted:
Sa m ple;
Moisture;
Weight;
Density;
Sample Temperature;
Instrument Temperature;
Scale _ A (Capacitance);
Product Name;
Equation Version;
Instrument Model;
Firmware Version;
Hardware Version, Time;
Date;
<CR><LF>

Moisture meter G610i I 19
To perform the selection of the printing mode, proceed:
1. Search the function “Set-up” and press YES;
2. Search the function “3.4 Set Printing Mode”, press YES;
3. Use the Right/Left ARROWS to select between two printing modes, Printer or
Computer;
4. Press YES to conclude the setting.
> Printer
Computer
This function allows the selection of the quantity of copies that will be generated in the
printed report.
It is possible to select from 0 to 5 copies, where zero copies means do not print the
report.Foreachsentcopy,ascreenwillbedisplayedtotearothepaperandpressYES
so that the next copy can be printed.
To perform the selection of the number of copies, proceed:
1. Search the function “Set-up” and press YES;
2. Searchthefunction“3.5SetNo.ofPrintedCopies”,pressYES;
3. Use the Right/Left ARROWS to select the number of copies desired;
4. Press YES to conclude the setting.
Tear it off and
press YES to Print...
Copies= 1
- < YES > +
3.5. Set number of printed copies

20 I Moisture meter G610i
ThisfunctionallowssettingtheLCDDisplaycontrast.
This value can be adjusted between 55% and 100%, being 100% the maximum contrast.
The setting will be stored in the G610i memory.
ToperformthesettingfortheContrast,proceed:
1. Search the function “Set-up” and press YES;
2. Searchthefunction“3.5SetContrast”,pressYES;
3. UsetheRight/LeftARROWStoincrease/decreasethevalueforContrast.These
setting advances in increments of 5%;
4. Press YES to conclude the setting.
Contrast= 90%
- < Setting > +
3.6. Set contrast
Other manuals for G610i
1
Table of contents
Other Gehaka Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Lake Shore
Lake Shore F71 quick start guide