Gio Pelle GP-PL-70 Instruction manual

Gio Peel
Installation, Operation & Maintenance Manual
Gio Peel Microdermabrader Model #GP-PL-70
7740 Records Street • Indianapolis, IN 46226
(317) 545-6196 • 800-428-1610 • Fax (317) 543-3289
www.giopelle.com
LGPMANUALVER02-11-3-11
MICRODERMABRADER

I
BENEFITS
• Treats all skin types
• Little or no downtime
• Progressive
• Minimum pain or discomfort
• Patient satisfaction typically 100% following treatment
• Stimulation of broblasts for collagen reproduction
• Hyperpigmentation reduction, possibly in combination with skin
lighteners
• Overall rejuvenation of skin health
• Enhanced penetration of carefully chosen topical actives after
treatment
• Elimination of potential for adverse chemical irritation
• May be performed during pregnancy
• May be performed shortly after surgery using appropriate
adjustment of vacuum and/or crystal ow
• May assist in healing due to increased blood ow
• Hyperpigmentation risks are minimal
• May be performed in 20-30 minute sessions

II
TREATMENTS
•SunDamagedSkin
•AgingSkin/FineLines
•Hyperpigmentation
•Acne/AcneScaring
•Scars
•StretchMarks
•IngrownHairs
•Histological evaluation reveals supercial to medium
depth,uniformabrasionoftheepidermis.
•Microdermabrasionisabuzzwordintheeldofskincare
introducedtotheUnitedStatesin1997.Europeanesthe-
ticiansbeganusingmicrodermabrasioninthe1980’s.
•Microdermabrasionexpandsthenumberoftherapyoptions
availableinyourofce,whichreinforcesyourreputation
inthelocalmarketandexpandsyourpatientbase.

III
CONTRAINDICATIONS
•Rosacea
•RaisedWarts
•Moles
•SkinTags
•Acne*
•Accutane(mustdiscontinueuse6-8weekspriortotreatment)
•EczemaandSeborrheicDermatitis
•OralBloodThinners
•Active,BrittleDiabetes
•SkinCancer
•AutoImmuneDisease
•Vitiligo
•Telangiectasias**
* Cystic acne should be treated medically and brought
undercontrolpriortomicrodermabrasion
** Treating telangiectasias should be determined by the
severityofthecondition.Wedonotbelievethiscondi-
tioncanbecorrected with microdermabrasion, there-
forevasodilatationdisordersshouldbetreated before
attemptingmicrodermabrasiontreatments.

IV
CongratulationsandthankyouforyourpurchaseoftheGioPeelMicrodermabrader.Thisprecisiondesignedpieceof
equipment,engineeredwithyouinmind,willprovidemanyyearsoftroublefreeoperation.Itwillsaveyoutimeand
increaseyourabilitytoserviceyourclientinwaysyouneverbelievedpossible.
Followtheinstructionsexactlyasdescribedinthismanualandintheorderpresented.Thismethodgivesyouthe
benetofourexperienceandisthebestwaywehavefoundtoeliminatestart-upproblems.Readcarefullyandunder-
standeachsectiononsafetybeforeproceeding.Besuretothoroughlyfamiliarizeyourselfwiththeoperationofequip-
mentpriortousingitonanypatient.
Weappreciateyoubusinessandbelievetheperformance,valueandreliabilityofourequipmentwillearnyourtrust.
Sincerely,
JoanneBianchini
President
GioPelle,Inc.reservestherighttoincorporateanymodicationsorimprovementsinmachinesandmachinespecicationswhichitconsidersnecessaryanddoes
notassumeanyobligationtomakethesesaidchangesinmachinesandequipmentpreviouslysold.Theinformationinthisdocumentissubjectochangewithout
noticeanddoesnotrepresentacommitmentonthepartofGioPelle,Inc.Nopartofthisdocumentmaybereproducedortransmittedinanyformorbyanymeans,
electronicormechanical,includingphotocopying,foranypurposewithouttheexpresswrittenpermissionofGioPelle,Inc.
INTRODUCTION
Customer Survey........................................................1
Safety and Other Considerations..............................2
Introduction & Suggestions ......................................3
Inspection Upon Arrival .............................................4
Gio Peel Unit Parts .....................................................5
Installation & Set-up...................................................6
Operation Procedures................................................7
Crystals .......................................................................8
Nozzles & Tubing........................................................9
Prior to Microdermabrasion ....................................10
Performing the Procedure ....................................... 11
Completing the Procedure.......................................12
Frequently Asked Questions...................................13
Clinical Applicator ..............................................14-16
Crystal-Free Tip Protocol.........................................17
Maintenance..............................................................18
Troubleshooting Procedures...................................19
Technical Service & Parts........................................20
Order Form...........................................................21-22
Additional Products .................................................23
Material Safety Data Sheet.......................................24
Warranty & Training Registration ...........................25
Warranty ....................................................................26
Table of Contents

1
CUSTOMER SURVEY
WeatGioPelle,Inc.areproudofthequalityworkmanshipweputintoeachpieceofequipmentweofferourcustomers.
Although,webelieveourequipmentisthehighestquality,wearecurioustohearwhatyouthink.Pleasetakeafewmin-
utesgiveusyourhonestcommentsandwewillconsiderusingyourideastohelpmakeourproductsandservicesbetter.
Whenyouarenished,sealthisforminsideanenvelopeandmailittoGioPelle,7740RecordsStreet,Indianapolis,IN
46226.Thankyouforyourtime.
1.Didourliteratureaccuratelyconveythebenetsandfeaturesofourproduct? YES NO
2.Whatfactorsinuencedyourdecisiontobuythisequipment(circleallthatapply)?
Features Price Delivery Functions Simplicity Reputation Other (explain)
3.Howcouldweimprovetheliterature?
4.Giveusyouroverallimpressionoftheclarity,understandingandeffectivenessofthismanual?
5.Pickout2itemsinthemanualandtellushowwecouldimprovethem.
1.
2.
6.Didtheproductfunctioncorrectlyuponstart-up? YES NO
7.Iftherewasaproblem,whatwasthecauseandwhatdidyoudotoresolveit?
8. Whatfeatureorbenetimpressedyoumost?
9. Whatfeatureorbenetwouldyouliketoseeaddedtothisproduct?
10. Basedonthisexperience,wouldyoubeinclinedtobuyotherGioPelleproducts? YES NO
Ifno,whynot?
11. Othercommentorproblemsyouneedsolutionsfor:

2
SAFETY & OTHER CONSIDERATIONS
GioPelle,Inc.iscommittedtodesigningsafe,reliable,easy-to-operateproductsthatpromotegreaterproductivitywhile
reducingcosts.Thesesystemswillprovidemodernefcientmethodsofmeetingindustryneeds.Topreventinjury,only
personneltrainedinthismachinesproperoperationshouldusethisequipment.Alwaysobservethesafetyprecautions
listedinthismanualandallotherwarningorsafetylabelsonthemachine,wheninstalling,operatingandservicingthis
equipment.
Inpreparingtheseinstructions,wehaveattemptedtorecommendthemosteffectivemethodsandcautionstowarnagainst
actionsthatcouldcausepersonalinjuryormaketheequipmentunsafe.ItisimportanttounderstandthatGioPelle,
Inc.cannotanticipateorlistallconceivablemethodsandwarnofallpossiblehazards.Readthismanualthoroughlyto
understandallsafetyprecautionsbeforeproceedingtoinstall,operateorservicethisequipment.Theoperator’ssetting
ofthemachineandproperuseofthewandsolelydeterminestheseverityandsuccessofthetreatment.Alwaystrythe
operationoftheunitonasamplepatchofyourownskinrst.Thisnotonlyhelpsyoutodeterminethecapabilities
butalsogivesyouaclearerunderstandingofwhatthepatientfeels.Lettheirresponsesguideyouinprovidingasafe,
satisfyingapplication.
Occupational Exposure to Bloodborne Pathogens
Duringatreatment,theabrasivepowderbecomescontaminatedwithparticlesofskin,tissue,oilandpossiblybody
uids.OnDecember6,1991,theOccupationalSafetyandHealthAdministration(OSHA)promulgatedtheOccupational
ExposuretoBloodbornePathogensStandard29CFR1910.1030.ContactOSHA,localandstateauthoritiestodetermine
ifthecollectedwastepowderisconsideredbiologicalsubstancethatmayhaveparticularrequirementsfordisposal.
Breathing & Eye Protection, Clothing & Other Apparel
Duringatreatment,itismandatoryfortheoperatortowearrubbergloves,facemaskandeyeprotection.ReadtheMaterial
SafetyDataSheetinthebackofthemanualtounderstandthenatureandproperuseoftheabrasivemedia.Duetothe
machine’squietness,hearingprotectionisnotrequired.
Additional Safety Information
• American National Standards Institute (ANSI)
1430Broadway
NewYork,NY10018
• Federal Drug Administration (FDA)
Rockville,MD20857
1-888-463-6332
http://www.fda.gov
• National Institute for Occupational Safety & Health (NIOSH)
http://www.cdc.gov/niosh/homepage.html
• National Safety Council
444N.MichiganAve.
Chicago,IL60611
• U.S. Department of Labor
Occupational Safety & Health Administration (OSHA)
200ConstitutionAve.
Washington,D.C.20210
(202)693-1999
http://www.osha.gov

3
INTRODUCTION & S U G G E S T I O N S
If items are damaged or if you have questions about the operation or maintenance of the unit
STOP and contact us immediately!
ThisoperationmanualprovidesnecessaryinformationforthesafeuseoftheGioPeelMicrodermabraderhowever,itis
meanttosupplement,notreplaceaformaltrainingprogramintheuseofthisdevice.
General Description
TheGioPeelMicrodermabrasionsystemisdesignedtoexfoliatebyabradingtheskinwithhigh-puritygradeAluminumOxide
Crystalormayalsobeusedforcrystal-freeoperation.Themajorcomponentsofthesysteminclude:
• Powerfuldiaphragmvacuumpump
• Onesimplevacuumowcontrolvalve
• Plasticcrystalvolumeowcontrolscrewnut(noadjustingneededafterinitialadjustment)
• Prelledandsealed“CrystalCapsule™”containingenoughabradingmediafor7to9completetreatments
• CrystalCapsule™alsocontainsasealedusedcrystalwastereceptaclelocatedonthebottomhalfofthedisposablecapsule
• Disposableplastichandpiecewithconnectingtubing
• DisposableCrystal-FreeKit
NotonlyisyourGioPeellightweightandeasytomovefromroomtoroom,thehandpieceisveryeasytomanipulateanditssizeand
shapehasbeenengineeredtoprovideprecisionandaccuracyforthemicrodermabrasionprocedure.Thetipisdesignedtoprovide
theproperangletoabradetheepidermissafelyandgradually.Thetubingforthehandpieceisconnectedtotwopostslocatedinthe
centerfronthousingoftheCrystalCapsule.Thetubingconnectsthesystem’scrystalsupplytothehandpiecethenisvacuumedup
anddepositedinthewastesectionofthesamecapsule!Onlywhenthetipisincontactwiththeskinisthevacuumcreatedbyclos-
ingtheloopafterconnectingthedisposableCrystalCapsulethevacuumportpostonthefrontoftheGioPeel.Thevacuumgauge
indicatesthevacuumlevelthatisregulatedbytheowcontrolvalve.
How To Get The Most From This Manual
Properly follow the instructions in the order listed below, to avoid injuries or jeopardize any warranty coverage.
• Usethismanualasacheckofflist.Checkmarkeachtaskwhencompleted.
• Concentrateonthematerial,pre-readthemanualrstbeforedoinganythingelse.
• Readinanisolatedareaawayfromdistractions.
• Doitourwayrst.We’vespentalotoftimedevelopingthemostefcientmethodsforyou.
• Beprepared.Haveeverythingreadybeforestarting.
• Takenotestomakethesubjectclearerortoenhancethematerial.
• Learnatyourownpace.Don’tgiveup.Itwillcometoyouaspracticemakesperfect.
• Learnfromyourmistakes.Correcttheerrorandtryagain.
• Don’ttakechances.Onlyperformastepwhenitisclearexactlywhattodo.
Hints, Tips & Precautions
• AlwayswearaN95/N99NIOSHapprovedrespirator,glovesandANSIZ87.1approvedeyeprotection.
• Themediarecommendedforusewiththisdeviceisaluminumfree.
• Onlyoperateunitintemperaturerangesof60°Fto85°F.
• Never,underanycircumstance,operatewithwetordamphandsorwhilestandinginliquid.
• Onlyusea3wiregroundedelectricaloutletcapableofdelivering115vacat10amps.
• NeveroperatetheGioPeelwhilepluggedintoa2wireextensioncord.
• NEVER LOAD OR UNLOAD THE GIO PEEL WHILE THE UNIT IS OPERATING.
• OnlyoperatethisunitwithdrymediarecommendedbyGioPelle,Inc.
• UnplugtheGioPeelfromthewallelectricaloutletwhennotinuse.
• Inspecttheequipmentfordamageandweareachtimebeforeusing.
• VerifytheGioPeelmediasizeisproperlysuitedforthepatientorapplication.
• NeveroperatetheGioPeelunattended.
• Onlyuseaccordingtotheinstructionsgiveninthismanual.
• Neveroperatewithliquidmediaorliquidcleaners.
• Donotattempttoalterormodifytheunitoroperateitwithanysortofattachmentsordevicesunlessspecicallyprovidedor
recommendedbythemanufacturer.
• Donotoperatethisdevicewithadamagedcordorplugoraftertheunitmalfunctionsorhasbeendamaged.Ifdamaged,return
theunitforexamination,repairoradjustment.

4
INSPECTION UPON ARRIVAL
Remove the Gio Peel Microdermabrader from the shipping container and inspect for damage.
Contact us immediately if there is any type of problem.
Please take the time to familiarize yourself with these items.
Contact us immediately at 800-428-1610 if there are any items missing or
there is any indication of a problem.
(#1) Gio Peel Microdermabrader
(#2) Crystal Capsules
(#3) Filter
(#4) Clear Filter Cover
(#5) Vacuum Port Cover
Figure 1
12
3
4
5
6
Inspection Checklist
Verifytheitemsshowningure1areincludedwithyourGioPeelMicrodermabrader.
7
(#6) Twin Line Disposable Tubing
(#7) Reusable Nozzles (blue)
(#8) Disposable Crystal-Free Tips
(#9) Electrical power cord
9
8

5
Line
+
ON
OFF
Regulator
1515
20
25
10
5
30
Vacuum Meter
Front of Unit
GIO PEEL MICRODERMABRADER UNIT PARTS
Please take the time to familiarize yourself with the different parts
of the Gio Peel Microdermabrader.
1
23
Figure 2
**Gio Pelle reserves the right to substitute motor or accessory specications with an alternate without notice
4
Side & Back of Unit
Figure 3
04.Vacuum Post—ConnectsCrystalCapsule™to
machine.
05.Crystal Capsule™ Stabilizer—Helps keep
CrystalCapsule™securedinuprightposition.
06.External Filter & Cover—Prevents crystal
fromgettingintomotorportscausingdamage.
WARNING: Never attempt to lift the machine
by the lter assemblies.
07.Main Power Switch—Turnsuniton.
08.Power Receptacle—Connects power cord to
machine.
09.Carrying Handle—Foreasyportability.
10.Cooling Fans—Located on back of the unit.
Keepsthemotorrunningcool.Do Not Block!
1.Vacuum Gauge—Continuously displays the
operatingvacuumlevel.
2.Vacuum Control Valve—Regulatestheoperating
vacuum level. Turning the knob clockwise in-
creasesthevacuum,counterclockwiseforless.
3.Power Switch—Vacuumowon.
56
7
8
9
10

6
D
e
c
r
e
a
s
e
I
n
c
r
e
a
s
e
+
ALLUMINI OXIDE CAPSULE
PRODUCT NO: IM-CAP-003
LOT NO: 20050219
Before You Begin
• Readalltheseinstructionspriortooperation.
• Carefullyremovethesystemfromtheshippingboxandfoampacking.
• Removetheplasticcoveringandplacemachineonahardatsurface.
• UnpacktheCrystalCapsule™,handpieceandtubing.
• Removethe2rubbercapsfromtheportposts.Donotdiscardrubbercaps.
Youwillneedtherubbercapstoplugtheportpostswhendisposingofused
capsule(g.4).
• Holdcapsuleupright.Removethelabelcoveringvacuumhole(g.5).
Attaching the Tubing & Nozzle
• Connectthepropersizetubingtotheappropriatepostonthecapsule.Push
tubingrmlyontotheposts(g.6).
• Connectthepropersizetubingtotheatendofthenozzle.Pushtubingrmly
intothenozzle(g.7).
Attaching the Media Canister
• SlidetheCrystalCapsule’s™openingontothemalevacuumportpoststicking
outofthefrontofyourGioPeel.Vacuumportpostislocatedbetweenthe
CrystalCapsule™stabilizer(g.8).
• Wipe post clean priortoinserting.Besureyouslidestraight ontothepost.
DO NOT TWISToryouwilldamagethesealing o-rings.
• Insertcapsulefullyontothepostwithbotho-ringsdowninsidetheopening
inthecapsule.
Turning the Unit On
• PlacethemasterpowerswitchonthesideoftheunitintheOFFposition.
• Plug the POWER CORD into the connector located on the back of the
microdermabrader.
• AttachtheotherendofthePOWER CORD toaproperlygroundedoutlet
supplying115vac.
• DepresstheIportionofthemasterpowerswitchlocatedonthebackofthe
GioPeeltoturntheuniton.Thefanwillstart.
• DepresstheIportionofthepowerswitchlocatedonthefrontoftheGioPeel
toturnthesuctionon.
Adjusting Crystal Flow
• Unscrewingthenutcounter-clockwiseadjustscrystalow.Youshouldnotice
somemovement,orbubbling,ofthenewabradingmediainsidetheCrystal
Capsule™whenpreformingprocedures(g.9).
• Iflesscrystalowisdesired,unscrewthenutuntilyouseeaslightdecreaseof
thecrystalbubblinginsidethecapsule,ornomovementatallwhichindicates
minimumvolumeofcrystalow(g.9).
Vacuum Control Valve
• TurntheVacuumControlValveclockwisetoincreasethevacuumorcounter-
clockwisetodecreasevacuum(g.10).
INSTALLATION & SET UP
Contact us if the unit fails to operate during these steps.
Figure 4 Figure 5
Figure 6 Figure 7
Figure 8
Figure 9
Figure 10
Vacuum
Port
&
Sealing
O-Rings
Insert
Capsule
STRAIGHT
onto
Vacuum
Post
Rubber Caps Vacuum Hole
Large
Tube Small
Tube
Large Tube
Small
Tube
Nozzle
Capsule
Nut
Crystal Capsule™
Stabilizer

7
Operation Procedure
• Operatorsareadvisedtoalwayswearamaskandgloveswhileperformingmicrodermabrasion.
• Besureanewnozzleisproperlyattachedtothetubing.Newtubingisrecommendedifanymedicalskinproblemisnoted.
• Cleanthenozzlewithamilddisinfectant(youmayalsoautoclavethe“blue”reusablenozzles).
• Plugthetipopeningwithyourthumbandturnthevacuumcontrolknobtoreacha10”Hgsetting.
• StartwiththeCrystalFlowControlNutscreweddownfully(maximumow).
• PlacethetipopeningagainsttheskinandcrystalwillbegintoagitateintheCrystalCapsule™(asdescribedonpage7)andbegin
owingthroughthenozzle.
• Maintainnozzleopeninginfullimmediatecontactwiththeskintoachievedsmoothresults.
• Increaseordecreasevacuumleveldependingonpatient’stoleranceandthemanufacturersrecommendedsettings.
• Increaseordecreasecrystalow,dependingonresultsandpatienttolerance,byturningthreadnut.Iflesscrystalowisdesired,
unscrewthenutuntilyouseeaslightdecreaseofthecrystalbubblinginsidethecapsule,ornomovementatall,whichindicates
minimumvolumeofcrystalow.
• Oncesatisfactoryresultsareobtained,notethevacuumsettingsforthatpatientforfuturetreatments.
• Uponcompletionofthemicrodermabrasiontreatment,turnthemachineoffandcleanaccordingtorecommendedcleaningpro-
tocol.
• Cleanoutsideofmachinewithamilddisinfectantonly.DO NOT USE ALCOHOL,whichmaydamageunitandltercover.
• Proceedwithanafter-treatmentregimentasrecommendedbyyourskincarespecialist.
Safety
• Whenthenozzleopeningisinfullimmediatecontactwiththeskin,avacuumiscreatedtherebyactivatingthepressuresystem.
Thevacuum maybe visuallyconrmed bywatching theclean crystal,in thetop halfof the pre-lledsealed capsuleagitate
throughout.
• Theairowfromthesystembeginsatthevacuumportpost,thenpassesthroughtheCrystalCapsule™andcarriestheMicro-Crystal
throughthetubingtothenozzle.Thespeciallyengineerednozzleallowsforuniform,steadyandeffectiveabradingaction.Once
theabradingmediaispassedovertheskin,itisdrawnawayandsuctionedbackintothebottomhalfofthesealedwastereceptacle
portionoftheCrystalCapsule™.
• Crystalemissionshouldbeactivatedonlywhenthenozzleisinfullimmediatecontactwiththetreatmentarea.Immediatelybegin
continuouslymovingthenozzleoverthetreatmentarea.
Warning
• Holding the nozzle in STATIONARY CONTACT with the skin can cause SERIOUS DEEP ABRASION,
BLEEDING AND SCARRING!
•TheGioPeelMicrodermabrasionmachineisaclosedlooptypesystem,whichpreventscross-contamination.Thenozzlemaybe
disinfected.Disposablenozzleseliminatetheneedforsterilization,butcouldbeusedacoupleoftimesifyouhaveanultraviolet
orsteamsterilizer.
• Insertcapsulefullyontothepostwithbotho-ringsdowninsidetheopeninginthecapsule(g.8).
Important
• Neverreusecrystalfromthewastereceptaclehalfofthecapsule.DisposeoftheentireCrystalCapsule™intoatrashreceptacle,
whichyoucansealtopreventaccidentalinhalation.Reusetheyellowandorangecapsandtapeoverthevacuumpostsopening
tohelpsealupthecapsulebeforedisposingforaddedprotectionagainstspillsorinhalation.
• Glovesandmasksarerecommendedfortheoperatorduringuseandcleaning.
• Alwaysprotectthepatient’seyesfromcrystals.Thepatientshouldremovecontactlensesandeyeglasses.Eyes must be protected
with cotton balls, goggles, tanning eyecups or a combination of these.
OPERATION PROCEDURES, SAFETY & WARNINGS
Do not operate your Gio Peel Microdermabrasion System until you have successfully completed the proceeding
installation and set-up procedures listed on page 6, have read and understand the complete operating instruc-
tions, and have signed the training certication form at the back of this manual.
The Gio Peel System must be set-up, regulated, and used only in accordance with the
operating instructions set forth in this manual.
NEVER ATTEMPT TO OPEN THE GIO PEEL ENCLOSURE.
This will void your warranty. There are no user serviceable components inside.

8
AluminumOxide(Al2O3)soldtobeusedwiththisequipmentisthemanmadeequivalentofamineralcalled
Corundum.Thiscrystallizedpowderistheperfectabrasivebecauseitissurpassedinhardnessonlybyadia-
mondandisablockshaped,sharpedgedcrystal.Whenfoundinnature,thepure,transparentortranslucent
corundumisprizedasagem.Othermorefamiliarvarietiesofthesamemineralfamilyarenamedaccording
totheircolorsuchassapphire(blue)ruby(red)amethyst(purple)oremerald(green).Thismineralcom-
poundshouldnotbeconfusedwithAluminum(Al)orproductsmadefromAluminum(Al)orthepotential
healthrisksthatmebeassociatedwithAluminum(Al)duringtopicalskintreatmentsorbyingestion.This
productcontainsnoAluminum.
Weofferavarietyofthreecrystalsizesthatallowyoutocustomizeyourtreatmentfortheneedsofyour
patient.
Normal #12
Developedforallskintypes.Greatforgeneralpolishingoftheskin.Alsousedforsensitiveskinandolder,thinskinned
patientsthatbruiseeasily.
Aggressive #10
Usedonmostskintypesexceptforsensitiveskin.YoucanstartthepatientwithNORMAL #12andincreasetoAG-
GRESSIVE #10asneeded.ProvidesaslightlydeeperabrasionthantheNormal#12andisthemostcommonlyused
ofthethreecrystals.
Fast Cut #9
Themostabrasiveofthethreecrystalsoffered.Usedonthick,oilyskintypes,acnescarring,stretchmarks,scarrevi-
sionandotherareasrequiringdeeperabrasion.Priortotheprocedure,andevaluationoftheskinwilldeterminewhich
crystalsizeismorebenecial.Whenyouareunsurehowthepatientwilltoleratethetreatment,alwaysstartwiththe
NORMAL #12crystals.
CRYSTALS

9
Disposable Crystal-Free Tips
Designed for single client use. Tips are available in six styles including massage
(GP-DMT-0),coarse(GP-DCFT-1),medium(GP-DCFT-3),ne(GP-DCFT-6),
x-smallcoarse(GP-DCFT-XS-1)andx-smallne(GP-DCFT-XS-6).
Tipsarecolorcodedforeasyidentication.
Requires GP-DCFT-HP-D Handpiece and GP-RTP5 Tubing.
GP-DCFT-HP-D Angle Reusable Crystal-Free Handpiece
GP-DCFT-HPS Straight Reusable Crystal-Free Handpiece
Disposable Crystal-Free Tips slide onto handle. Tubing connects to bottom of handle.
Handlecanbeautoclaved.
Requires Disposable Crystal-Free Tips and GP-RTTP5 Tubing.
GP-RTP5 Disposable Crystal-Free Tubing
Reusabletubingisattachedtoeitherthedeposablenozzleorthereusablenozzle.Sinceno
partofthetubingcomesintocontactwiththepatient,tubingmaybeusedmultipletimes
beforechanging.
Requires Disposable Crystal-Free Tips and GP-DCFT-HP Handpiece.
GP-IM-CAN-2 Disposable Waste Canister
Forusewithdisposableandresusablecrystal-freetips.Wasteissuctionedinotself-contained
canisterandcanbeusedsealedwithrubberlidfordisposal.
Requires Crystal-Free Tips and GP-DCFT-HP-D or GP-DCFT-HPS Handpiece
and GP-RTP5 Disposable Crystal-Free Tubing.
FOR USE WITH CRYSTAL-FREE TIPS, HANDPIECES & TUBING ONLY.
GP-AHP-1B Reusable Nozzle for Crystals
Designedformultiplepatientuse.Thisnozzlecanbeautoclavedandreusedonmorethan
onepatient.Requires GP-RT5 Reusable Tubing.
GP-RT5 Reusable Tubing for Crystals
Reusabletubingisattachedtoeitherthedeposablenozzleorthereusablenozzle.Sinceno
partofthetubingcomesintocontactwiththepatient,tubingmaybeusedmultipletimes
beforechanging.
Requires GP-AHP-1B Reusable Nozzle.
NOZZLES & TUBING

10
1. Conductconsultationincludingpatienthistoryandexpectationspriortostartingaseriesoftreatments.
2. Patient should not have used any exfoliation productsor solutions 24 hours prior to treatment
(Glycolic,Retin-A,etc.)
3. Theoperatorshouldalwaysweargloves(eyeprotectionandmaskareoptional).
4. PatientMUSTremovecontactlenses.
5. Drapepatient’shairandclothing.
6. Evaluatetheskintodeterminetheappropriatecrystalsizeforthetreatment.Ifyouareunsurehowthe
patientwilltoleratetreatment,alwaysstartwithNORMAL#12.Youcanalwaysprogresstoamore
aggressivecrystal.
7. StartbycleansingtheskinwithChamomileCleanser.
8. Theskinmustbecompletelydrybeforetheprocedureisconducted.
9. Withgloveson,testthepressureofthemachinewith yournger.Thegaugeshouldreadabout15
(outsidenumbers)tobegin.
PRIOR TO BEGINNING MICRODERMABRASION
Notes:

11
1. Thetreatmentiseasieriftheprocedureiscompletedononesideofthefaceandthenrepeatedonthe
otherside.
2. Holdtheskintautandbeginthetreatmentinthemiddleoftheforeheadjustabovethenose.Runthe
handpiecealongtheforeheadandabovetheeyebrow,stoppingatthetemple.Liftthehandpieceand
returntotheareawhereyoustartedtheprocedure(askthepatientiftheyarecomfortable.Eithercon-
tinuethetreatmentoradjustthesuctionlevelaccordingly).Repeatthisactionuntilyouhavecovered
thathalfoftheforehead.Thenrepeatthisactioninthreedifferentdirections.
3. Continuefromthetemplesdownthesideofthefaceandcheekarea.Again,repeatthismotioninthree
differentdirections.
4. Sweepacrossthecheekstothenose(avoidingtheundereyearea).Thenosecanbeadifcultareato
treat,thismaytakepractice.Repeatinthreedifferentdirections.
5. Next,continuewiththeupperliparea,thechinandanyremainingareasonthesideoftheface.Repeat
inthreedifferentdirections.
6. Repeatabovestepsontheoppositesideoftheface.
7. Lowerthesuctiontoaboutlevel10andperformthetreatmentontheneck.Runthehandpieceinan
upwardmotion,thenpullandliftoutwardtowardtheears.Twopassesaresufcienttothisarea.
8. Repeatundertheeye,includingouterareaandcrows’feet.Thisshouldbeperformedinshort,quick
strokestoinsurepatientcomfort.
9. Thelipscannowbetreatedinthesamemanner.Alwaysholdthelipstautanduseshort,quickstokes.
BesuretoworktheVermilionBorder,reducingnelinesandleavingthelipsrefreshedandfull.
10. Removethetubefromthecleanmediacanister.Nowyouareworkingwithsuctiononly,nocrystals.
Repeattheproceduretocleanupresidualcrystals.Thiswillalsofurtherstimulatebloodow,leaving
thepatientwithahealthy,beautifulglow.
PERFORMING THE PROCEDURE
Notes:

12
Patient Care
1. CleansethefaceandneckwithChamomileCleanser.
2. SpraywithAlvrinSpray.
3. MassageafewdropsofHydrasilkintothefaceandneck.
4. MassageasmallamountofIntensiveMoisturizerPlusintothefaceandneck(ifpatientisacne
prone,skipthisstep).
5. ApplySunblockSPF25evenlyoverentirefaceandunderthechin.
Machine Care
1. IfusingGP-AHP-1BReusableNozzle,removenozzleandautoclaveit.Replacewithaneworclean
GP-AHP-1BReusableNozzle.Neverreuseanozzleondifferentpatients.IfusingoneoftheDisposable
Crystal-FreeNozzles,pleasedisposeofaccordingtolocal,stateandfederalguidelines.
2. IfusingGP-RT5ReusableTubing,youmayreuseitwithoutcleaningunlessithasbeencontaminated
orhascomeindirectcontactwiththepatient.Tubingshouldbereplacedwhenwornout.
3. Wipeanydebrisorcrystalsoffofmachinewithawet,dampcloth.
NOTE: Dispose of crystals, nozzles and tubing and other contaminated materials as re-
quired by local, state and federal laws.
COMPLETING THE PROCEDURE
Notes:

13
Q.How many treatments are needed?
A. Atypicalpatientreceives6treatmentsin2weekintervalstocompleteaseries.
Q.How long does a patient typically continue receiving treatments?
A. Aftertheinitial 6treatments,atypicalpatientreturns onamonthlybasisformaintenance.
Somemayreturneveryfewmonthsasneeded.Thisisusuallydeterminedbythephysicianor
technician.
Q.Can the Gio Peel be used on all skin types?
A. Absolutely.Unlikeacidpeelsthatcancausehyperpigmentationondarkerskin,microdermabra-
sionissafeforallskintypes.
Q.Can the Gio Peel be used on pregnant women?
A. Yes.Microdermabrasionissafeforpregnantorlactatingwomen.
Q.Can the Gio Peel be used on acne patients?
A. Yes.Aseriesoftreatmentswillactuallybenetacnemorethantheuseoftopicalproducts.
Q.Can you treat cystic acne with the Gio Peel?
A. Cysticacneshouldbetreatedwithanoralortopicalantibioticpriortobeginningmicrodermabra-
siontreatments.Thephysicianshoulddeterminewhentreatmentcanbegin.
Q.Should a patient with Rosacea receive treatments?
A. ActiveRosaceashouldneverbetreated.Iftheconditioniscalm,thepatientcanbetreatedat
therequestofthephysician.Normal#12crystalsshouldalwaysbeusedonpatientswithRo-
sacea.
Q.Can ingrown hairs be treated?
A. Yes.Thisprocedureisexcellentforpatientswithingrownhairsandforpatientsundergoing
hairremovaltreatments.TheGioPeelkeepsthehairfollicleuncloggedbetweentreatments.
Q.Should a patient be exposed to sun immediately following a treatment?
A. No.Athreedaywaitingperiodisimportanttopreventsundamagetotheskin.Also,always
applyasunblockofSPF25ormore.
FREQUENTLY ASKED QUESTIONS
Notes:

14
Operator Training
Waiver of Liability
GioPelle,Inc.expresslydisclaimsanyandallliabilitywhatsoevereitherincidentalorconsequentialforanyandalldamagesand/
orinjuriesofanykindtoanypersonsorpropertyresultingeitherdirectlyorindirectlyfromuseofthisproduct.UseoftheGio
PeelSystemisatthesoleriskoftheoperatorandisintendedforandlimitedtouseonlybyorundertheclosedirectsupervisionof
atrainedcertiedmedicalprofessional.GioPelle,Inc.furtherexpresslydisclaimsanyandallliability,includingbutnotlimited
to,anyandalldamagesorinjuriesresultingeitherdirectlyorindirectlyfromtheuser’serrors,omissions,ormistakeswhetherac-
cidentalorotherwise.
Depth of Abrasion
• Thedepthofabrasionoftheepidermislayerscanvarydependingonthepatient.
•WARNING: An esthetic procedure must always remain at the epidermis level and NEVER induce bleeding or bruising.
CLINICAL APPLICATION
Indications for Use
AlloftheGioPellemodelsimproveskintexturethroughprogressiveoperatorcontrolledmechanicalexfoliationoftheskin.
Microdermabrasion is recommended for skin with:
• Sundamage
• Finelinesandwrinkles
• Enlargedpores
• Minoracnescars
• Blackheadsandcomedones
• Agespots/supercialpigmentation
• Adjuncttherapytocosmeticsurgery
• Blendingpostlaseredskin
Treatment
Pre-treatment Recommendations
The quality of results of any dermatological procedure depends on a thorough understanding of the skin history and previous
treatments.Assessmentoftheproblemandcurrentskinconditionshouldbefollowedbyatreatmentplan.Afterdeterminingthat
microdermabrasionisthetreatmentofchoice,atestareashouldbedoneonthebackofthehandtodemonstratethesensationprior
toapplyingtreatmenttothefaceorotherareas.
Contradictions
Thefollowingisapartiallistofskinconditionsthatshouldnotbetreatedbymicrodermabrasion,butcontraindicationsarenot
limitedtothislist:
• Lesions
• Inammatoryacneorinfectedpustules
• Dermatitisincludingpsoriasis
• HerpesSimplex
• Thickandevolutionisterythematouswithill-denedborders
• Sunburns
• Ultrasensitiveskin
• PeoplewithDiabetes,AIDSorusersofanticoagulantsorAccutane®withthelastsixweeks
AccutaneisaregisteredtrademarkofRocheLaboratories,Inc.
Clinical Application Instructions continued on page 10...
Depth of Abrasion Skin Reaction Operator Level
Epidermis Coenocytes Layers Redness; No Bleeding Esthetic Level
Malpighian Layers Suffusion; Moist Medical Level
Basal Layers Blood Specs Medical Level
Epidermal/Dermal Layers Bleeding Medical Level

15
Treatment Recommendations
GioPeelMicrodermabrasionproceduresarebestperformedoveraperiodofseveralweekswithamaintenancetreatmentscheduled
everymonthortwothereafter.Treatmentsrangefrom4weeklytreatmentsfornormalskintoasmanyas12ormoreforscars.The
dermaenhancingtreatmentisperformedconsecutively7-10daysbetweentreatments.
Amajorbenetofamechanicalabrasiontreatmentisthatitisfast,predictable,painlessandeffective.Someadvantagestothe
patientarehavingnodowntimefromworkorsocialactivitieswitharapidrecoverytimeinaslittleas6to8hoursdependingon
depthoftreatment.
Treatment Procedure
NOTE: Operator should now have on gloves.
Preparetheskinbycleaningtheareatobetreated.Nooilsormake-upshouldbeleftontheskin.Useanon-oilycleanserandrinse
untilclean(preferablyChamomileCleanser).
Important:
• Dry the skin completely. Any remaining moisture may clog system and reduce treatment efcacy.
• Use only ordinary gauze or sponge for cleaning. Do not use cotton! Residual bers can be sucked into the tip and clog the
system.
Exfoliation Rate
Therateatwhichtissueisremovedisdirectlyrelatedtothespeedandvolumeatwhichthecrystalpassesthroughthenozzle.The
settingofthesetwocontrolsincreasesordecreasestherateofexfoliation:
• Vacuum Level:TurningtheVacuumControlValvewillincreaseordecreasethevacuumlevel.
• Crystal Flow:TurningtheCrystalFlowControlNutwillincreaseordecreasethecrystalow
(clockwise-tight=maximumcrystalow,counter-clockwise=lesscrystalow).
Eye, Ear & Hair Protection
Thepatient’shairshouldbecoveredwithadisposablecapthatcoverstheearstoprotectthemfromcrystals.Thepatientshould
removecontactlensesandeyeglasses.The eyes MUST be protected with eye patches, cotton balls, goggles, tanning eyecups or
a combination of these.GioPellerecommendsourEyeShields.ContactGioPelleforthesemusthaveitems.
Double Check the Gio Peel and the Nozzle Connections
CheckthattheGioPeelisinproperworkingorderandhasanewcleannozzlethathasbeenproperlyinstalledontothetubing.
NOTE: Operator should now have on gloves & a mask.
Treatment Tips
NOTE: Some people have more sensitive skin than others. Tolerance to treatment will vary. People with more
sensitive skin may require lower vacuum levels with longer, lighter treatments and more treatment sessions.
• Oncethepropervacuumsettingischosen,theoperatorshouldusethehandthatisnotholdingthenozzletostretch the skinunder
thetiptocreateaneffectivesuction.
NOTE: When the nozzle opening is not in full immediate contact with the skin, external air will be pulled
into the system and temporarily interrupts system vacuum, operation and performance.
• Manydifferenttechniquescanbeusedtoensureauniformexfoliation.Beginthetreatmentbygentlymovingthenozzleoverthe
treatmentareainanupanddownpattern.Covertheareacompletelyinonedirectionthengooverthesameareaintheopposite
direction.
CLINICAL APPLICATION
(Continued from page 16)
Clinical Application Instructions continued on page 11...