Giropes GI650 User manual

V.1.4_03/05/2022
EN
USER MANUAL
GI650
WEIGHBRIDGE
El fabricante se reserva el derecho de modificar sin previo aviso las características de sus productos para introducir mejoras técnicas o cumplir con nuevas
regulaciones oficiales./Le constructeur se réserve le droit de modifier les caractéristiques de ses produits en vue d’y apporter des améliorations techniques
ou de respecter de nouvelles réglamentations./The manufacturer reserves the right to modify the specifications of its products in order to make technical
improvements or comply with new regulations.
Pol. Empordà Internacional C/Molló 3
17469 VILAMALLA - (Girona) SPAIN
T. (34) 972 527 212

2
USER MANUAL GI650 ENEN

3
1. INTRODUCTION 4
1.1 CHARACTERISTICS 4
CONNECTION OF THE LOAD CELL 4
OPERATION INTERFACE 4
SERIAL COMMUNICATION 4
POWER SUPPLY 4
OPERATING CONDITIONS AND MECHANICAL DATA 5
1.2 GENERAL FUNCTIONAL CHARACTERISTICS 5
1.3 DIMENSIONS 6
1.4 DISPLAY 7
1.5 KEYS AND FUNCTIONS 8
MAIN-DISPLAY 8
KEYBOARD-DISPLAY 9
NUMERIC-DISPLAY 9
1.6 FUNCTIONS WHEN STARTING THE INDICATOR 10
2. WEIGHING INSTRUCTIONS 10
ZERO 10
TARE 10
TARE MANUAL 11
DEACTIVATE TARE 11
MEMORIZED TARE 11
CHECKWEIGHER 11
NET WEIGHT USING 2 WEIGHINGS 12
NET WEIGHT USING 1 WEIGHING 13
3. MENU STRUCTURE 15
3.1 MENU 16
COMMUNICATIONS 16
SECURITY 17
DATA BASE 17
GENERAL 17
FUNCTIONALITY 18
PRINTERS 18
ADDITIONAL 19
WEIGHING 19
CALIBRATION AND METROLOGY 21
DIAGNOSTIC 23
ABOUT THE DEVICE 23
3.2 DATA BASE 23
4. ERROR MESSAGES 28
5. CONFIGURATION OF PERSONALIZED TICKETS 28
6. CONNECTIONS 34
7. DSD-MEMORY 35
8. REMOVAL OF ELECTRONIC EQUIPMENT 35
9. WARRANTY 35
10. ANNEX 36
EN USER MANUAL GI650 ENEN

ENENUSER MANUAL GI650
4
1. INTRODUCTION
Dear User,
the performance that the equipment you have acquired will offer you, will be all the greater the more attention you devote to the instructions
for use included in this manual. Please break our traditional apathy towards the manuals and do not wait for “everything goes wrong” for a
consultant, so you can enjoy optimal benefits from the execution of the first weight.
This equipment is designed to work on one hand as basic equipment to carry out weighing with or without tare, assigning these weights to their
corresponding code. Up to a maximum of 9200 weighings can be memorized.
1.1 CHARACTERISTICS
FEATURES OF THE LOAD CELL
Precision Class III or IIII
Fraction factor p’I = 0,5
Minimum input voltage per VSI > 0.3 uV
Excitation voltage 5 V DC
Min. transducer resistance 29 Ω
Max. transducer resistance 1200 kΩ
Maximum length of cable 5714 m/mm2
Temperature range -10°C à +40°C
Type of Range Mono-range; Multi-range; Multi-intervall
OPERATION INTERFACE
Display LCD tactil 800x600
COMMUNICATION
Port Tx/Rx: (Channel 1) RS-232 bi-directional
Port Tx/Rx: (Channel 2) RS-232 bi-directional
Port Tx/Rx: (Optional) RS-485 bi-directional
Port Ethernet
Port USB Compatible with USB 2.0
Transmission Speed 300 - 115200
Bits 7 and 8 bits
Parity No; Even and Odd
POWER SUPPLY
Power Source 110 - 230VAC, 50-60 Hz

5
ENENUSER MANUAL GI650
OPERATING CONDITIONS AND MECHANICAL DATA
Operation Temperature Range from -10ºC till 40ºC
Max. Temperature Range from -25ºC till 70ºC
Size without Support (mm) 242 x 212 x 84
Weight without Support ~ 2.8Kg
Support Desktop and Wall
1.2 GENERAL FUNCTIONAL CHARACTERISTICS
Display
The GI650 has a Touchscreen Display with integrated back-light.
Password
Password protection for metrological and security Parameters
Theoric Calibration
If the reference weight is not available, you can perform a theoretical
calibration by using the capacity and sensitivity values (mV/V) of the load
cells used.
Import / Export Data
It is possible to export the Code-files and weighings via USB to a PC to
manage / modify the Categories.
Manage Files
It is possible to store and manage up to 1900 code-files. These files are
used in 5 different types: Product (1000), Customer (400), Provider
(100), Target (100) and Plates (300). It is also possible to program a
specific name for each code.
Manage Weighings
It is possible to store and manage up to 9200 weighings in 2 different
types: open weights and Weights (4600 each).
Manage memorized Tares
It is possible to store and manage Tara-values in the Plates-register.
Manage the Printer
It is possible to work with 2 printer/labeler on the GI650. You can also free
configure the printouts and on which printer, which information is printed.
Software
GI Manager
Repeater available
It is possible to work with a Repeater through the serial RS-232 Port or the
optional RS-485 Port
Alibi-Memory
Alibi-Memory (DSD) is an available Optional.
External Keyboard
It is possible to plug an external Keyboard through the serial USB-Port

ENENUSER MANUAL GI650
6
1.3 DIMENSIONS
BENCH TOP
1
1
2
2
3
3
4
4
5
5
6
6
A A
B B
C C
D D
Diseñado: Revisado: Fecha: Unidades del plano: Número SAP:
Tamaño del papel:
Escala:
Página:
Propiedad:
Giropès S.L.
Pol. Empordà Internacional.
Calle Molló. Parcela 15-16.
17469 Vilamalla (Girona).
AMS NT 13/05/2019 mm
A3
GI450T8
Vista general Pedido nº:
Validado:
Nº plano: IDD-VIGI-GI650-C2DD-0003-00
126
242
226
84
164
212
242
272
ON WALL
1
1
2
2
3
3
4
4
5
5
6
6
A A
B B
C C
D D
Diseñado: Revisado: Fecha: Unidades del plano: Número SAP:
Tamaño del papel:
Escala:
Página:
Propiedad:
Giropès S.L.
Pol. Empordà Internacional.
Calle Molló. Parcela 15-16.
17469 Vilamalla (Girona).
AMS NT 13/05/2019 mm
A3
GI450T8
Vista general Pedido nº:
Validado:
Nº plano: IDD-VIGI-GI650-C2DD-0003-00
126
242
226
84
164
212
242
272

7
ENENUSER MANUAL GI650
1.4 DISPLAY
1
11
12
2
8
9
4
5
6
7
10
3
13
Number Function
1Platform-number and Configuration
2Date
3Time
4Charging-indicator
5Zero-indicator
6Stable-indicator
7Weight-range
8Gross-weight
9Net-weight
10 Tare-weight
11 Truck - status
12 Categories
13 Command-Keys

ENENUSER MANUAL GI650
8
1.5 KEYS AND FUNCTIONS
It is possible to choose between 3 different Keyboards:
QWERTZ, QWERTY and AZERTY
To change the Keyboard you have to access to the Main-menu by press
Choose the Menu point “General” and change the settings.
MAIN-DISPLAY:
Number Function
1Zero
2Tare (blocked by default)
3Deactivate Tare
4Store the weight
5Manual Tare
6Open PIT - List
7Reprint last Ticket
8Fastkey to access Database
9Activate / Deactivate the Checkweigher
10 Start a new Weight without Zero-Step
11 Access to Main-menu
123
4567 8 9 10 11

9
ENENUSER MANUAL GI650
KEYBOARD-DISPLAY:
Key Function
1Exit the Keyboard
2Clear all letters
3Delete the last letter
4Save the changes
5Change to the special characters
6Arrow Keys
NUMERIC-KEYBOARD:
Key Function
1Exit the Keyboard
2Clear all numbers
3Delete the last number
4Save the change
1 2 34

ENENUSER MANUAL GI650
10
1.6 FUNCTIONS WHEN STARTING THE INDICATOR
To turn on the indicator press the key
The Display turns on, showing the Firmware-, Bootloader- and Weight model-version, also the CRC-value and the number
of Calibrations during the initialization.
After finishing the initialization, the indicator is performing the initial zero.
To do this, the indicator must be connected to the Load cell and following conditions must be made:
•The weight must be stable (indicator for stability is on )
•The weight must be inside the range of initial Zero (15% in positive and 5% in negative)
During this operation, the Display will show “0 init” till the weight is in range.
2. WEIGHING INSTRUCTIONS
PLEASE NOTE:
This equipment comes with a standard factory configuration. It is strongly recommended to program the databases and the
different codes as required.
Before do any weighing you have to insert the Plate.
Without doing this, it is not possible to store any weight.
You can put the Data directly into the respective Field or select from Database (see point 3.2)
ZERO
When you press the key ,this assumes the metrological function of weight zeroing.
To do this, the following conditions must be made:
•The weight must be stable (indicator for stability is on )
•Weight value not less than -2% of the Max. in relation to the zero of the calibration
•Weight value not higher than 2% of the Max. in relation to the zero of the calibration
The sign indicates that the function has been done.

11
ENENUSER MANUAL GI650
TARE
There are different types of tare that are described in the following sections.
TARE MANUAL
Pressing the Key will open the Numeric-Keyboard.
to enter the Tare-value. After saving the value with the OK button
the value will be shown in negative in the indicator.
DEACTIVATE TARE
To deactivate the tare, press the button
MEMORIZED TARE
It is possible to store a Tare-value in the Plates (will be explained in point 3.2)
CHECKWEIGHER
The GI650 is equipped with a Checkweigher function. The different areas are displayed in different Colours.
Low High AlarmOK
You can set the value in the menu. At first you have to access to the main-menu by pressing . Choose the Menu-point “Functiona-
lity” and you will see the settings of Check Weight Low, Check Weight High and Check Weight Alarm (the “OK-range” automatically is
set between the Low and High) Press the showed value you want to modify (by default it is set on 0) and the Numeric-keyboard will appear.
Input all 3 values, when you finish press the key to save the changes The Checkweigher is NOT activating automatically.
To activate the Checkweigher you have to press in the weigh-window.
PLEASE NOTE :
It is ONLY possible to store weights that are between the High and Low Limits (indicated in Green)

ENENUSER MANUAL GI650
12
NET WEIGHT USING 2 WEIGHINGS
Before you can do the 1st weighing, you always have to manually input a Plate (and also all the other information if needed).
After this you can place the Truck on the Weighbridge, please note that the weight have to be higher than the minimum-weight.
Wait till the stability-indicator turn on and press to store the weight.
If you want to perform the 2nd weighing, you have to perform a Zero-step (Weighbridge empty) first. Also you can press the Button
to directly start a new weighing without perform the Zero-step. You also have to input the Plate. You can do this in 2 different
ways, input the Plate manually, like you did it in the 1st weighing, or choose from the list of Plates In Transit.
For choose from the List of Plates In Transit, you have to do following steps:
Press the column Plate for open the List of stored Plates.
Now choose the Plates In Transit in the down right corner.

13
ENENUSER MANUAL GI650
Select the Plate In Transit for which you want to perform the 2nd weight and validate.
In the weight-window the Plate is loading and you can perform the second weight pressing .
NET WEIGHT USING 1 WEIGHING
Whenever you want to perform a weighing with Tara-function it is an Unique Weight.
You can do this in 2 different Ways:
Loading a stored plate from the Database.
(How to input and load files from the Database will be explained in point 3.2)
The other option is to input the Information manually. For this you have to first input the information of the Plate (and also all the other
information if needed).

ENENUSER MANUAL GI650
14
Now press the Button , input the Tare-value and validate.
The weight is changed to Unique Weight now, also the Net-indicator turn on and the Tare-value will be displayed, now you can place the
truck and perform the weiging pressing .

15
ENENUSER MANUAL GI650
3. MENU STRUCTURE
Communications
COM 1
COM 1
COM 2
COM 2
ETHERNET
RS-485
General
Security
Database initialization
Database configuration
Database tables
Import / Export
Date
Hour
Language
Keyboard Type
Tare block
Check Weight Low
Check Weight High
Check Weight Alarm
Print mode
Trigger type
Baudrate
Number of bits
Parity
Stop bits
Protocol
Trigger type
Interval time
Baudrate
Number of bits
Parity
Stop bits
Protocol
Trigger type
Interval time
Protocol
Trigger type
Interval time
Baudrate
Number of bits
Parity
Stop bits
Protocol
Trigger type
Interval time
Address
Database
General
Functionality
Printers
Additional
Weighing
Diagnostic
About the device
Ticket ID
Scales configuration
DSD
Channel1
Printer
Configuration
Custom tickets
Printer
Configuration
Custom tickets
Parameters
Manual calibration
Weight parameters
Theoric calibration
Legally relevant data
Restore all to default
MENU
User password
Default parameters

ENENUSER MANUAL GI650
16
3.1 MENU
COM 1
RS-232
COM 2
RS-232
ETHERNET
RS-485
300; 600; 1200; 2400; 4800; 9600; 14400;
19200; 28800; 38400; 56000; 57600; 115200
7; 8
None; Even; Odd
1; 2
None; Giropes; Sipi 2; Alibi; Sensocar; Printer; GiConf2; Smart; Bilancai;
A+; Bizerba; Epel; Eric; Mettler; SB400; Soehnle; PT6S3
No trigger; Weight trigger; Continuous
Freely Configurable (ms)
300; 600; 1200; 2400; 4800; 9600; 14400;
19200; 28800; 38400; 56000; 57600; 115200
7; 8
None; Even; Odd
1; 2
None; Giropes; Sipi 2; Alibi; Sensocar; Printer; GiConf2; Smart;
Bilancai; A+; Bizerba; Epel; Eric; Mettler; SB400; Soehnle; PT6S3
No trigger; Weight trigger; Continuous
Freely Configurable (ms)
None; Giropes; Sipi 2; Alibi; Sensocar; GiConf2; Smart; Bilancai;
A+; Bizerba; Epel; Eric; Mettler; SB400; Soehnle; PT6S3
No trigger; Weight trigger; Continuous
Freely Configurable (ms)
300; 600; 1200; 2400; 4800; 9600; 14400;
19200; 28800; 38400; 56000; 57600; 115200
7; 8
None; Even; Odd
1; 2
None; Giropes; Sipi 2; Alibi; Sensocar; GiConf2; Smart; Bilancai;
A+; Bizerba; Epel; Eric; Mettler; SB400; Soehnle; PT6S3
No trigger; Weight trigger; Continuous
Freely Configurable (ms)
Freely Configurable
COMMUNICATIONS
Baudrate
Baudrate
Protocol
Baudrate
Number of bits
Parity
Stop bits
Number of bits
Parity
Stop bits
Number of bits
Parity
Stop bits
Protocol
Protocol
Protocol
Sending mode
Interval time
Sending mode
Interval time
Sending mode
Interval time
Sending mode
Interval time
Adress

17
ENENUSER MANUAL GI650
SECURITY
Inside the menu are 2 sub-menus: User Password and Default parameters
• User Password: Inside here you can change the Password for. The saved menus.
• Default parameters: It is restoring to Factory settings, except the Database, Ticket-number and the weighing parameters.
PLEASE NOTE:
For access to the menu you need to insert a Password, by default 0000
DATABASE
In the Database you can modify and manage the different Code-files and weighings.(More in point 3.2)
GENERAL
When access to this menu, you can modify the Date and Time as well as the Language and the Keyboard Type.
The Date and Time are freely modifiable.
• Languages available: Spanish; English; French; Italian; Portuguese and German
• Keyboards available: QWERTZ ; QWERTY and AZERTY

ENENUSER MANUAL GI650
18
FUNCTIONALITY
In the functionality menu you have access to following working configurations: Tare-setting and the Checkweigher-settings. You can decide
to work with a blocked or unblocked Tare (By default this option is unblocked)
•You can modify the Checkweigher-settings (Explained in point 2)
PRINTERS
Here you have access to the general ticket configuration and also to the detailed configuration of each Com Port.
• General tickets configuration
Here you can set the Ticket-ID (ticket number) and the Header / Footer lines.
In the different Port Com are several sub menus, them are even on both Com.
• Printer configuration
Here you can set the Printer type (Custom+, AscII, ESC-POS or Lx350) You can configure which Format is used for each Printout. The
number of ticket-copies and Advanced lines. Also if this Com-port is used for print lists and if a cutter is used or not.

19
ENENUSER MANUAL GI650
• Custon Tickets
It is possible to customize the ticket formats, explained in point 9.
ADDITIONAL
Inside this menu you can check and delete the stored Alibi-weighings. (when Alibi-Memory is plugged)
PLEASE NOTE:
Erase the Alibi-Memory is only working with Cal_open
WEIGHING
PLEASE NOTE:
For access to the menu you need to insert a Password, by default 0000
When you access to this Menu, you have only access to the non-metrological. (Cal_close)
To have full access to All metrological Parts, you need to press the Cal_switch. (Cal_ouvert)

ENENUSER MANUAL GI650
20
• Cal_close:
You only have access to the Sub-menu Legally relevant data. Inside this Menu are different Software-information about the device.
• Cal_open:
You have access to all Sub-menus included the Calibration
Scale configuration:
Inside this menu, you will find all the calibration settings and metrological settings. (explained in the next point)
Legally relevant data:
Inside this menu you can find all legally relevant data, such as WM-version, Events, Errors and Updates.
Restore all default parameters:
with this setting you restore All parameters back to fabric, included the Calibration. (except the Database and Ticket-number)
Other manuals for GI650
1
Table of contents
Other Giropes Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Endress+Hauser
Endress+Hauser dosimass operating instructions

Myron L
Myron L ULTRAPEN PT3 quick start guide

ATD Tools
ATD Tools 5597 operating instructions

Agilent Technologies
Agilent Technologies FRG-700 Short operating instructions

Armstrong
Armstrong AMF Installation & operation manual

Swann
Swann AMU-II Oxytrace Operator's manual