Your meter comes with one 1.5V AAA size alkaline battery.If the low battery symbol“ ”
appears on the screen,this indicates that the power is low and it is time to change the
battery.
• Replacing the battery does not affect the test results stored in memory.
• As with all batteries,this battery should be kept away from children.If swallowed,
promptly seek medical assistance.
• Batteries may leak chemicals if unused for a long time.Remove the battery if you are
not going to use the meter for an extended period (i.e.,3 months or more).
• Properly dispose of the battery according to your local environmental regulations.
NOTE
STEP 3. Exit the memory mode
After displaying the last test result in memory,press the Magain.The meter displays“End”
and then turns off.
• If you wish to exit memory mode before the last result is displayed,press the Mbutton
for 3 seconds.
•When the memory is full,the oldest test result will be replaced by the newest test result.
•The results of the ketone control test and β-Ketone test results will be stored in the
memory.
NOTE
2. Connecting to a personal computer
3. Data transmission
To transmit data,follow the instructions provided with the software.Results will be
transmitted with the date and time.Remove the cable and your meter will automatically
switch off.
Replacing the Battery
To replace the battery,make sure that your meter is turned off.
1.Press the edge of the battery cover and lift it up to remove.
2.Remove the old battery and replace with one 1.5V AAA size alkaline battery.
3.Close the battery cover.If the battery is inserted correctly,you will hear a“beep”
afterwards.
The“ ”appears with E-b,Error:
The power is not enough to do a test.Please change the battery
immediately.
While the meter is connecting to your PC,it is unable to perform a β-Ketone test.
Warning:
Model No.:TD-4235E
Dimensions &Weight:93 (L) x 36 (W) x 21.3 (H) mm,45.4 g (including battery)
Power Source:One 1.5V AAA alkaline battery
Display:LCD
Memory:450 measurement results with respective date and time
External output:USB
Auto electrode insertion detection
Auto reaction time count-down
Auto switch-off after 3 minutes without action
TemperatureWarning
Operating Condition:5°C to 40°C (41°F to 104°F),below 85% R.H.(non-condensing)
Meter Storage/Transportation Conditions:-20°C to 60°C (-4°F to 140°F),below 95% R.H.
Strip Storage/Transportation Conditions:2°C to 30°C (35.6°F to 86°F),below 85% R.H.
Measurement Unit:mmol/L fixed
Measurement Range:0.1 mmol/L to 8.0 mmol/L
Expected Service Life:5 years
This device has been tested to meet the electrical and safety requirements of: IEC/EN
61010-1,IEC/EN 61010-2-101,EN 61326-1,IEC/EN 61326-2-6.
SPECIFICATIONS
Caring for Your Meter
Cleaning
1.To clean the meter exterior,wipe it with a cloth moistened with tap water or a mild
cleaning agent,then dry with a soft dry cloth.Do NOT rinse with water.
2.Do NOT use organic solvents to clean the meter.
Meter Storage
• Storage conditions:-20°C to 60°C (-4°F to 140°F),below 95% relative humidity.
• Always store or transport the meter in its original storage case.
• Avoid dropping and heavy impact.
• Avoid direct sunlight and high humidity.
Meter Disposal
The used meter should be treated as contaminated that may carry a risk of infection during
measurement.The battery in this used meter should be removed and the meter disposed
in accordance with local regulations.
The meter falls outside the scope of the European Directive 2002/96/EC-Directive on waste
electrical and electronic equipment (WEEE).
MAINTENANCE
The following is a summary of display messages.If your meter displays an error message,
please follow the actions described in the table below.If the problem persists,please
contact GlucoRx Customer care for help.
DISPLAY MESSAGES AND PROBLEM-SOLVING GUIDE
BATTERY
You can use your meter with a USB cable and the GlucoRx Healthcare Software System to
view test results on your personal computer.To learn more about this Software System or
to obtain a USB cable separately,please contact GlucoRx Customer care for assistance.
Connect the cable to a USB port on your
computer.With the meter switched off,
connect the other end of the USB cable to
the meter data port." USb " will appear on
your meter display,indicating that the
meter is in communication mode.
1. Obtaining the required cable and installing the software
To download the Healthcare Software System,please visit www.glucorx.co.uk.
DOWNLOADING RESULTS ONTO A COMPUTER
TaiDoc Technology Corporation
B1-7F, No. 127, Wugong 2nd Rd.,
Wugu Dist., 24888 New Taipei City, Taiwan
www.taidoc.com
MESSAGE WHAT IT MEANS ACTION
Appears when the battery is too
low. Replace the battery immediately.
Appears when a used ketone
test strip is inserted. Repeat with a new test strip.
Appears when ambient
temperature is above or below
system operation range.
System operation range is 5°C to 40°C
(41°F to 104°F).Repeat the test after
the meter and test strip are in the
above temperature range.
Appears when the wrong ketone
code strip is inserted or other
coding errors.
Make sure to insert the correct code
strip. If the problem persists,please
contact GlucoRx Customer care for
help.
Problem with the meter.
Repeat the test with a new test strip.
If the meter still does not work,
please contact GlucoRx Customer
care for assistance.
Appears when ketone test strip is
removed while counting down. Review the instructions and repeat
test with a new strip.
If the problem persists,please contact
GlucoRx Customer care for help.
STEP 2. Recall test results
If you continue to press M,you can then review the last 450 results in the memory.
MESSAGE WHAT IT MEANS
< 0.1 mmol/L
0.1 to 8.0 mmol/L
> 8.0 mmol/L
SYMBOL INFORMATION
Symbol Referent
Consult instructions for use
Temperature limitation
Use by
Humidity limitation
Caution,consult accompanying
documents Collection for electrical and
electronic equipment
Batch code
CE mark
Manufacturer
Serial number
Symbol Referent
STEP 1. Enter the memory mode
With the meter turned off,press Mtwice.The first test
result will appear,indicating that you are in the memory
mode.
When using the meter for the first time or if the results
have been deleted,“---”will appear,indicating that there
are no test results in the memory.
Viewing Results
Your Meter stores the 450 most recent β-Ketone test results with date and time in its
memory.You can review the test results with these easy steps.
VIEWING THE METER MEMORY
STEP 4. Get result
The result of your cow‘s β-Ketone test will be stored in the meter memory automatically.
If you do not apply a blood sample to the test strip within 3 minutes,the meter will
automatically turn off.You must remove and reinsert the test strip to start a new test.
NOTE
1.Please do not change your treatment on cows based on the ketone result without
first consulting your veterinarian.
2.Turn the meter off by removing the ketone test strip.Discard the used test strip and
syringe carefully according to your local regulations.
Warning:
If the confirmation window is not filled completely before your meter begins to count
down,do not add more blood to the test strip.Discard the test strip and start again.If you
have trouble filling the test strip,please contact your veterinarian or GlucoRx Customer
care for assistance.
Gently apply a drop of blood to the absorbent hole of the ketone test strip at a tilted angle.
The confirmation window should be completely filled if enough blood sample has been
applied.Do NOT remove the cow‘s root-tail until you hear a beep sound.
STEP 3. Apply blood sample
Draw blood sample from the cow's root-tail.Obtain blood drop volume of at least 0.7 µL.
STEP 1. Insert the test strip to turn on your meter
Wait for the meter to display code no.and then a flashing “ ”.
STEP 2. Select the appropriate measuring mode by pressing M
For selecting the measurement mode,please refer to the“TWO MEASURING MODES”
PLEASE
AFFIX
STAMP
GlucoRx Ltd
4 Wintonlea
Monument Way West
Woking
Surrey
GU21 5EN
United Kingdom
TERMS & CONDITIONS OF WARRANTY
1. GlucoRx warrants this product to be free of defects in workmanship and materials within the said warranty period
on the Warranty Certificate.
2. During the warranty period if this product is found to be defective, you may send it with the Warranty Certificate to
our office for warranty service. GlucoRx will then repair or replace defective parts, or exchange the whole product as
GlucoRx may choose, with no additional charge to the original owner. After such repair, replacement or exchange,
the product will be warranted for the remaining time of the warranty period.
3. This warranty is valid only if the Warranty Certificate and Warranty Registration Card are duly completed with date
of obtaining product, serial number and if the Warranty Registration Card is sent to our office (or you register on
our website www.glucorx.co.uk) no later than 6 months from the date obtained.
4. This warranty is void if this product has been repaired or serviced by an unauthorised person. This warranty does
not cover defects caused by misuse, abuse, accident, tampering, lack of reasonable care, fire or any other acts
beyond human control.
5. Except as stated in the above paragraphs, GlucoRx disclaim all other warranties, including those of merchantability
of fitness for a particular purpose with respect to the use of this product. GlucoRx shall not be liable for any direct,
consequential or incidental damages arising out of the use or inability to use this product.
Distributed by
GlucoRx Ltd
4 Wintonlea, Monument Way West
Woking, Surrey, GU21 5EN, UK
Website: www.glucorx.co.uk
Customer careline: 01483 755133