GORMAN-RUPP SOTERA 820 User manual

Installation and Operation Manual
820, 825, 850 Digital Meter

sotera.com 2
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
Table of Contents
MODEL#
SERIAL#
PURCHASE DATE:
Thank You!
Thank you for your loyalty to the Sotera®brand of chemical, lubricant,
and mixed hydrocarbon transfer pumps. Your safety is important, so
please read and thoroughly understand the procedures set forth in this
manual. Protect yourself as well as those around you by observing all
safety instructions and adhering to all danger, warning, and caution
symbols. Please save these instructions for future reference and
record the model, serial number, and purchase date of your Sotera
digital meter. Please register your Sotera product via
info.fillrite.com/sotera_product_registration.
IMPORTANT RETURN POLICY
Please do not return this product to the store. For all warranty
and product questions, please contact Technical Support at
1 (800) 720-5192 or via email at SoteraTech@fillrite.com
(M-F, 8 AM – 5 PM ET).
About This Manual ............................................................................. 3
Symbols and Definitions..................................................................... 3
General Description............................................................................ 3
Safety Instructions............................................................................. 3
Technical Information......................................................................... 4
Fluid Compatibility............................................................................. 4
Options............................................................................................... 5
Installation......................................................................................... 5
Operational Functions........................................................................ 6
Use..................................................................................................... 6
Calibration Using the CAL Factor ....................................................... 6
Calibration Procedure Using the CAL Factor....................................... 7
Batteries ............................................................................................ 8
Repair ................................................................................................ 8
Maintenance ...................................................................................... 8
Troubleshooting Guide........................................................................ 9
Meter Parts List................................................................................ 10
Appendix A ....................................................................................... 11
Appendix B – Water Calibration ....................................................... 12
Appendix C – Fluid Calibration (Other Than Water).......................... 13
Appendix D – Flow Performance....................................................... 14
CE Certification Information............................................................. 14

sotera.com 3
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
About This Manual
Symbols and Definitions
General Description
Emphasizes an area in which personal injury or even death may result from failure to follow instructions properly. Mechanical damage may also occur.
These boxes contain information that illustrates a point that may save time, or be key to proper operation, or clarifies a step.
Failure to observe a “Caution” may cause damage to the equipment.
WARNING
CAUTION
NOTICE
DANGER Indicates a hazardous situation which, if not avoided, will result in death or serious injury.
From initial concept and design through final production, your Sotera product is built to provide years of trouble-free use. To ensure the safety
of yourself and those around you, it is critical that this manual is read in its entirety prior to attempting to install or operate your new purchase.
We strongly urge that any installer and operator become familiar with the terms, diagrams, and technical data in this manual and pay close attention
to any DANGER, WARNING, CAUTION, or NOTICE information. At Sotera, your satisfaction with our products is paramount to us. If you have questions
or need assistance with your product, please contact us at 1 (800) 720-5192 or via email at SoteraTech@fillrite.com (M-F, 8 AM – 5 PM ET).
The Sotera 825 and 850 Meters are nutating disk, positive displacement meters that use magnetic coupling to convert fluid flow into digital
display information. The meter can store and display the current flow amount (current total),or cumulative flow amount (totalizer) in any of five
user specified units (ounces, pints, quarts, liters, and gallons) or special units (e.g. per acre volume). The meter can be calibrated without
dispensing fluid simply by selecting a calibration factor from the 20 stored settings. Power is supplied by two AA field replaceable batteries.
Pulser models have the added ability to connect to fluid management systems for additional control and monitoring of fluid being dispensed.
Safety Instructions
To ensure safe and efficient operation,it is essential to read and follow each of the following warnings and precautions.
1. Agricultural herbicides flowing through the meter may be harmful to your health. Use and disposal of these products is controlled by
federal, state, or local laws and procedures.
2. Conform to fluid manufacturer’s recommended handling procedures when using product and when cleaning meter.
3. Do not exceed an internal meter pressure of 120 PSI/8.2 Bars.
4. Improper use or installation of this product can cause serious bodily injury or death.
5. The 825 & 850 Digital Meter is not for use with flammable fluids. DO NOT use with fluids with a flashpoint below 100°F (such as gasoline
and alcohol).
6. DO NOT REMOVE PRINTED CIRCUIT BOARD! Damage to LCD could occur, and warranty is void.
WARNING

sotera.com 4
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
Technical Information
Flow ports 1" NPT inlet / outlet ports, female threads
(3/4" BSPP also available)
Flow Range 2 to 20 U.S GPM / 7.6 to 75.7 LPM
Pressure 120 PSI / 8.2 Bar maximum @ 70°F / 21°C
50 psi / 3.4 Bars maximum @ 130°F / 54°C
Temperature Min. operating temperature = 0°F / 17°C
Max. operating temperature = 130°F / 54°C
Meter can be stored at lower temperatures but
display may not work below 0°F.
Accuracy ± 0.5%
Units of Measure Ounces, pints, quarts, liters, gallons;
special calibration option also available.
Range 9999 current total;
10,000,000 accumulated total
Technical Information
Materials of Construction
Body Polypropylene
Chamber Polyphenylene sulfide (PPS) and
303 Stainless Steel
Wetted Seals Fluorocarbon (EPDM optional)
Weather &
Dust Body Seal
BUNA-N
Display LCD (Liquid Crystal Display)
Power Two CR2032 Lithium batteries and external
5 - 24 VDC (see page 8)
The 825 & 850 Digital Meters will handle most pesticides,
automotive fluids (except gasoline), and mild acids.
It is also compatible with the following fluids**:
Aatrex 4L®
Abate 4E®
Apron®
Agrotain®
Assure II®
Atrazine 4L
Banvel®
Banvel SFG®
Bicep®
Blazer®
Treflan®
Dual®
**Camix TM
Caustic Soda (50%)
Clarity®
Command®
Conclude®
Conclude®xtra
Contour
Diesel Fuel
Doubleplay®
Dual II®
Ethylene Glycol
Flexstar®
Fultime®
Fusion®
Gramoxone lnteon™
3.38EC Pursuit®
Rezult®
Roundup®
Squadron®
Surpass®
Superboll®
Touchdown®
Treflan™ HFP
** Requires EPDM Seals.
Aatrex®, Aatrex® 4L., Bicep®, Bicep 11®, Dual®, and Dual 11® are registered trademarks
of Syngenta Corporation. Broadstrike and Treflan are registered trademarks of Dow AgroSciences.
Banvef®, BanvelSGF®, Blazer®, camix, Clarity®, Conclude®, Galaxy®, Guardsman®, Lumax,
Manifest “Marksman® Poast®, Poast HC®, Poast Plus®. Rezult®, and Storm® are registered
trademarks of BASF. DoublePlay®, Eptam® 7E, FulTime, Fusion®, Gramaxone® Extra, Karate®,
ReHex®, Surpass®. TopNotch, and Touchdown® are registered trademarks of Syngenta.
Contour®.Detail, Pursuit®, Prowl®, and Squadron® are registered trademarks of American
Cyanimld. Harness® Xtra, Roundup® are registered trademarks of Monsanto Company.
Command®, and Furadan® are registered trademarks of FMC.Agrotain® Isa registered
trademark of IMCAgrico. Superboll®is a registered trademark of Griffin.
The 825P & 850P Digital Meters are NOT compatible with very strong
acids or if fluid flash point is below 100°F (38°C). If in doubt about
compatibility of a specific fluid, contact supplier of fluid to check for
any adverse reactions to the following wetted materials:
Guardsman®
Harness xtra®
Karate®
Laddock S-12®
Lasso Micro Tech®
*Lumax TM
Manifest™
Marksman®
Maxim®
Methyl Parathion Broadstrike®+
Motor Oil Broadstrike®+
Nufos®
Phosphoric Acid
Poast®
Poast HC®
3ME Poast Plus®
Princep 4L®
Prowl®
Detail™
Oil.Adend®
Dual®
Eptam 7E®
Fallowmaster®
Frontier®
Furadan®
Gramoxone Extra®
Prowl®
Reflex®
Ridomil Gold®
Sodium Hydroxide (50%)
Storm®
100 Surpass® EC
Topnotch®
Treflan™
Water
Fluid Compatibility

sotera.com 5
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
Use Teflon tape or thread compound on all threaded joints.
1. Determine direction for fluid flow and point arrow on meter body in
that direction.
2. Thread hose or pipe into ports until snug. Be careful not to cross
thread when starting threads.
Installation Hint
To prevent cross threading, turn the pipe / hose backwards
(counterclockwise) until you feel it engage threads, then tighten.
Changing Meter Readout Position
If it is necessary to change position of the meter readout,
follow these steps (Refer to exploded view).
1. Unscrew meter cap (item 1). Use a strap type oil
filter wrench if too tight to unscrew by hand.
2. Insert a wide, flat-head screwdriver into the upper slot and
gently pry up electronics module (see Figure 1).
3. Gently rotate electronics module to desired location.
4. Press electronics module down into meter cover in the correct
orientation.
5. Thread on meter cap until hand tight. To check tightness, there
should be approximately 1/16" gap between the cap and ridge
on outlet port (see Figure 3).
UPPER PRY SLOT
FIGURE 1
O-RING
FIGURE 3
1/16" GAP
FIGURE 2
InstallationOptions
• EPDM Seals
DANGER Not for use with fluids that have a flash point below
100°F (37.8°C, ie: gasoline, alcohol). Refer to NFPA
325M (Fire Hazard Properties of Flammable Liquids,
Gases, and Volatile Solids) for flash points of
common liquids. Static electricity buildup and
discharge could result in arc and explosion!
CAP
RIDGE ON
OUTLET PORT

sotera.com 6
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
• Turns meter on when off.
• Displays accumulated total as long as it is pressed.
If accumulated total is larger than 9999, the
numbers will scroll across the screen.
• When held for 3 seconds, it resets current total
to zero. Also resets to normal operating mode
when in CAL or FLSH mode.
• When held for 3 seconds, it allows changes to
the calibration factor displayed in the bottom left
corner. Repeated pressing will step the number
up to 19 and back to zero. When desired number
is displayed, press button 2to lock in the new
number and return to normal operation.
• When held for 3 seconds, flow rate is displayed.
Fluid dispensed is still added to the accumulated
total and current total. Press button 4to return
to normal operation.
1. Press button to turn meter on. Current total, unit of measure, and
calibration factor are displayed. The meter also turns on automatically
and begins recording when fluid starts flowing through it.
2. Hold button 2for one second to reset current total to “0.00.”
3. Begin dispensing.
NOTE: Meter display automatically goes blank after 60 seconds of
inactivity and automatically comes back on when flow resumes.
No data is lost during periods of inactivity.
The THINNER the fluid, the LOWER the CAL number.
The THICKER the fluid, the HIGHER the CAL number.
• CAL 4 is set for thin fluids like water.
• CAL 19, the highest number, is set for very thick fluids like
cold molasses.
• Each number changes the meter accuracy by about 1%.
Table #1: Suggested CAL Factor Settings for Common Fluids
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
Water
Kerosene
GRAMOXONE INTEON™ @ 50°F
Antifreeze @ 70° F
ROUNDUP® @ 50°F
TREFLAN™ @ 50°F
ATRAZINE 4L @ 70°F
10W Oil @ 70°F
DUAL® @ 70°F
BICEP® @ 70°F
Molasses @ 32°F
NOTE: The suggested CAL factors are for REFERENCE ONLY.
Wear proper safety equipment when handling
hazardous fluids.
CAUTION
Operational Functions Use
Calibration Using the CAL Factor
CAL Factor
Meter will count air if you dispense air. Before initial
operation or when air has entered the system, prime
the meter by dispensing fluid until all trapped air has
been removed. Meter is now ready to operate.
CAUTION

sotera.com 7
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
Changing the CAL Factor
• Hold button 3until the display only shows CAL and number.
• Press 3repeatedly until you reach the desired number.
NOTE: Number will step up to 19, then back to zero.
• Press 2to return to normal operating mode.
Calibration Procedure
A) Set CAL factor to 10.
Hold button 3for three seconds. Press 3again until the number 10
shows below “CAL” (NOTE: If you go past 10 keep pressing 3because
the number will return to 0 after passing 19).
Press 2to get back to normal operating mode.
NOTE: If your fluid is listed on Table 1, use that number in step A above
in place of 10.
B) Prime pump and meter by dispensing 2 - 3 gallons of fluid back
into the bulk tank.
With the outlet valve closed and the pump still running, reset the
meter to 0.00.
C) Fill the Proving Can exactly to the 5.0 gallon line.
Focus on the Proving Can, do not look at the meter at this point. Keep the
hose end nozzle wide open as long as possible for best accuracy.
D) Adjust Meter CAL Factor.
• If meter reads high, increase the CAL factor. Each CAL # changes the
accuracy by about 1%. For a 5 gallon proving can, 1% = 0.05 gallons.
• If 5.10 is displayed, this is 2% over 5.00; the CAL factor should be
changed to CAL 12.
• If the meter reads low, lower the CAL factor. For example, if 4.90 is
displayed, it is 2% less than 5.00, so the CAL factor should be set to 8.
• When finished with the CAL factor procedure, press 2to return
to normal mode and to reset the meter to 0.00. The meter is now
calibrated and ready to use.
Calibration Procedure Using the CAL Factor

sotera.com 8
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
NOTE: Low battery icon will appear when batteries require replacement.
Meter still functions properly for several days after the icon appears.
Neither calibration, current total or totalizer quantities will be lost when
you replace batteries.
To Replace Batteries (refer to exploded view).
1. Unscrew meter cap (item 1). Use a strap type oil filter wrench or
large 5" jaw pipe wrench if needed.
2. Insert a flat-head screw driver into the top slot (see Figure 4) and
gently pry up electronics module.
3. Remove old batteries and insert new batteries, making sure battery
polarity is correct, or meter damage could occur.
4. As noted in Figure 5, reinstall O-ring on electronics module.
Align sensor receptacle in proper location. Press module gently
down into meter cover.
5. Thread on meter cap until hand tight. To check tightness, there
should be approximately 1/16" gap between cap and ridge on
outlet port. (See Figure 6).
If any meter components are damaged, they should be replaced. See
meter kits drawing on page 12 for correct replacement part information
before ordering.
Meter should be flushed between uses with water to prevent chemicals
from drying and plugging meter.
Thorough Cleaning (refer to exploded view)
If meter is plugged due to hardened chemical or debris, do the following:
1. Drain all fluid from meter.
2. Unscrew meter cap (item 1). Use a strap type oil filter wrench or
large 5" jaw pipe wrench if necessary.
3. Insert a flat-head screwdriver in the lower slot (see Figure 7) and
turn to pry up meter cover (item 6).
4. The meter chamber (item 8) can now be removed.
5. Rinse all meter components with flushing fluid. DO NOT submerge
display assembly. Be careful not to get any fluid or dirt in the
electronics module.
6. Reassemble meter.
CALIBRATION NOTE: Over time, the chamber inside the meter will wear,
requiring the meter to be recalibrated with water. When this should be done
depends on the amount and type of fluid dispensed. In most crop protection
fluid uses (less than 1000 gallons of a clean fluid per year), the meter will
remain accurate for many years without recalibration. On the other hand,
dispensing an abrasive fluid may require more frequent recalibration.
The 825 / 850 meter is designed to be calibrated with clean water for
safe handling. See “Water Calibration” section in Appendix B.
Storage
Store in a cool, dry place. Drain out all fluid that could freeze in the meter.
LOWER PRY SLOT
FIGURE 7
DO NOT get fluid or dirt in electronics area.
CAUTION
Follow fluid manufacturer’s recommended procedures
for handling and disposing of metered fluids.
CAUTION
Batteries
Maintenance
Repair
UPPER PRY SLOT
FIGURE 4
O-RING
FIGURE 5
FIGURE 6
1/16" GAP
CAP
RIDGE ON
OUTLET PORT

sotera.com 9
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
Problem Possible Cause Solution Notes
Meter won’t turn on • Dead batteries
• Damaged or contaminated
electronics module.
• Replace batteries.
• Replace electronics module & gaskets.
Seal to electronic chamber
is broken if display label is
removed or punctured.
Flashing decimal Current total has rolled over. Reset display to zero by pressing button 2.Meter will continue to
operate normally.
Flashing or dim display Low batteries. Replace batteries. Use CR2032 batteries.
Leaking fluid at
inlet/outlet port
• Need thread sealant.
• Cross-threaded port.
• Add Teflon pipe tape to joint.
• Replace body.
Fluid flows;
meter won’t count
• Meter disk sticking.
• Damaged driver or magnet.
• Meter failure.
• Clean out meter chamber.
• Repair or replace chamber assembly.
• Repair or replace meter.
Meter reads high • Air in system.
• Wrong calibration factor.
• Prime system, fix suction leak at pump.
• Use a higher calibration factor.
See 1-Step procedure.
• See “Check Meter” in Appendix A.
Meter will count air.
Chemical formulations
sometimes change.
Meter reads low by 10%
or less
Wrong calibration factor. • Use a lower calibration factor.
• See 1-Step procedure on page 7page 7.
Chemical formulations
sometimes change.
Temperature also
affects accuracy.
Meter reads low by more
than 10%
• Meter chamber is worn.
• Chamber is partially plugged.
• Damaged or very worn
chamber.
• Recalibrate meter with water.
See Appendix B.
See “Check Meter” in Appendix A.
• Clean chamber.
• Replace chamber and recalibrate meter.
Meter is not consistent • Air in system.
• Particulates in fluid.
• Worn or damaged meter
chamber.
• Prime system, fix suction leak at pump.
• Put screen in front of meter.
• Clean chamber.
• Replace chamber.
40 mesh minimum.
Err0 • Calibration error.
• Damaged chamber.
• Recalibrate meter with a more
accurate container.
• Replace chamber.
Indicates fluid calibration is
out of acceptable window.
Volumetric container may be
off, there may be air in the
system, or the meter chamber
may be damaged.
Err1 • Damaged electronics.
• Software fault.
• Repair or replace electronics.
• Press 2then recalibrate meter
Contact factory.
Err2 Bad eeprom. Replace electronics. Meter still functions, but all
data will be lost if batteries
are removed.
Troubleshooting Guide

sotera.com 10
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
825 / 850 Meter Parts List
Item # Part # Description
1 KITPOLYRING Replacement Ring Kit
2 825G8559/850G8569 Replacement Electronics Kit 825/850
3 825KTF1582 Meter Chamber Kit and Seals
When ordering repair parts, be
sure to give the replacement part
number, the date of manufacture,
and the meter series number.
This will ensure the correct
replacement part is supplied.
Technical Support:
(800) 720-5192

sotera.com 11
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
To Change Units of Measure
The units of measure can be changed to ounces (OZ), pints (PT),
quarts (QT), gallons (GAL), or liters (LITER) without recalibrating
the meter. If special units are desired, see note below.
1. Hold buttons 2and 4for three seconds. Display will show
current software revision level (i.e: r1.02).
2. Press button 3. Display will show the current units.
3. To change units, press the button repeatedly until the desired
unit is displayed.
4. Press button 2. Display will show firmware revision.
5. To return to normal operating mode press button 2again.
6. The units selected will be displayed. Current or accumulated total
will change to reflect the new units.
Special Units
To use special units, you need to know how many ounces are in your
special unit. Here is an example: You want to use “acres” as your
“special” unit. The fluid is to be applied at 18 ounces per acre.
These are the additional steps to set the meter to “special” units
(ignore steps 4 & 5 above):
1. After selecting “special” in step #3 above, press button 2.
2. Enter the number of ounces in a special unit by pressing button 4
to increment the digit, and the button to move the flashing
digit to the right. If you make a mistake, press button 2to start
back at the left most digit. Per our example, we would enter 018.0.
3. Press the button again. Current firmware will display.
4. To get back to the normal operating mode, press button 2.
Reset Accumulated Total
1. Press 2and 4simultaneously and hold for 3 seconds.
Display will read the version of the software loaded in the meter
(example: “r1.02”).
2. Press to display accumulated total. If over 9999,
display will scroll across the screen.
3. Hold buttons and 4for 5 seconds to reset
accumulated total.
4. Press button 2twice to get back to normal operating mode.
Meter Check
You can check the calibration in your meter.
1. Set to CAL 4 (See “to change the calibration factor”).
2. Hold button 3and 4for 3 seconds.
Meter will display “FLSH”.
3. Hold buttons & 3together. A number will display that indicates
the pulses per unit used to calculate flow (ie: pulses per gallon).
When new, this number is between 120.0-127.0 pulses per gallon.
If you find a number higher than 127, recalibrate with water
(see Appendix B). If this number is lower than 120, the meter
chamber may need to be replaced.
4. Press 2to get back to normal operating mode.
FLSH (Flush) Mode
The 825/850 Meter can be flushed without adding to the totalizer.
Turn meter on by pressing the button.
Press 3and 4simultaneously and hold for 3 seconds. Display will
show FLSH. Flush meter with suitable fluid (water is suitable for most
herbicides). When completed, press 2to leave FLSH mode and return
to normal operation. Quantity of fluid flushed will not be added to total.
APPENDIX A

sotera.com 12
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
The 825 / 850 meter is designed to be recalibrated with water for
safe handling. Over time, the chamber inside the meter will wear.
Recalibrating the meter with clean water will insure that Table #1
(calibration table page 6) is most accurate.
You will need a container of known volume, at least 5 gallons or larger.
Do not exceed a 60 gallon container.
1. Press the 2& 4buttons simultaneously and hold for 3 seconds.
The display will read the version of the software loaded in the meter
(example: “r1.02”).
2. Press button 3to enter calibration mode.
The unit of measure will be displayed.
3. Press button 3.
The display will read “FILL”.
4. Now dispense water into your container. “FILL” will flash on the display.
5. After dispensing, press the button.
The left digit of the display will blink.
6. Press the 4button to increment the digit to the amount of fluid
dispensed (example: 05.00). Press the button to move to the right.
If you make a mistake, press button 2to start back at the
left-most digit.
7. After number is loaded, press the button again to accept.
8. Display now shows .
Since you are calibrating with water, accept this by pressing
the button again. If calibrating with a fluid other than water,
see Appendix C. Display will again show “r1.02”.
NOTE: If the value entered is out of an acceptable range, the display
will read “Err0” and the meter will revert to the previous settings. See
Troubleshooting Guide for more information.
9. Press 2to get back to the normal operating mode.
APPENDIX B – Water Calibration

sotera.com 13
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
You will need a container of known volume, at least 5 gallons or larger.
Do not exceed a 60 gallon container.
1. Press the 2& 4buttons at the same time and hold for 3 seconds.
The display will read the version of the software loaded in the meter
(example: “r1.02”).
2. Press button 3to enter calibration mode.
The unit of measure will be displayed.
3. Press the button to change unit of measure, if required. This is
ONLY necessary if calibrating a different unit of measure.
4. Press button 3.
The display will read “FILL”.
5. Now dispense fluid into your container. “FILL” will flash on the display.
For best results, dispense fluid at the same flow rate that will be used
in actual use.
6. After dispensing, press the button.
The left digit of the display will blink.
7. Press the 4button to increment the digit to the amount of fluid
dispensed (example: 05.00).
Press the button to move to the right.
If you make a mistake, press button 2to start back at the left-most digit.
8. After number is loaded, press button again to accept.
9. Display now shows .
This is the default for water. Check Table 1 for your fluid. Press 3to
change the Cal #. Press to accept. Display will again Show “r1.02”.
NOTE: IF YOU ACCEPT CAL 4, USE THE METER ON CAL 4 WHEN
DISPENSING THIS FLUID.
NOTE: If the value entered is out of an acceptable range, the display
will read “Err0” and the meter will revert to the previous settings.
See Troubleshooting Guide for more information.
10. Press 2to get back to the normal operating mode.
DO NOT perform this calibration unless you fully understand how CAL factors work.
CAUTION
APPENDIX C - Fluid Calibration (Other Than Water)
Calibrating with a fluid other than water voids
Table #1 (calibration table page 6).
After calibration, set the meter to CAL 4, and use
the meter on CAL 4 (unless you input a different
number during step 9 below).
CAUTION

sotera.com 14
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
NOTE: Graphs are accurate with original factory calibration, or a water calibration.
METER CALIBRATION FACTOR SELECTION BASED ON FLUID VISCOSITY
VISCOSITY (CPS)
VISCOSITY (CPS)
VISCOSITY (CPS)
CALIBRATION FACTOR
Low Viscosity (thin) Fluids
CALIBRATION FACTOR
High Viscosity (very thick) Fluids
CALIBRATION FACTOR
Medium Viscosity Fluids
PERFORMANCE CURVE
APPENDIX D – Flow Performance
Pressure Drop (PSI)
Flow Rate (GPM)
The 820, 825 and 850 meters bearing the CE mark have been certified to the following European directives:
2011/65/EU: Restriction of the use of certain hazardous substances in electrical and electronic equipment.
2004/108/EC: Electromagnetic Compatibility
The following standards were used to test and show compliance:
Emissions
EN 55011:2009/A1:2010 Group 1, Class B, Industrial, Scientific, and Medical (ISM) Equipment
Immunity
EN 61326-1:2006, Electrical Equipment for Measurement, Control and Laboratory Use
IEC 61000-4-2: ESD
IEC 61000-4-3: Radiated Immunity
IEC 61000-4-8: Magnetic Field
CE Certification Information

sotera.com 15
| 820,825, 850 DIGITAL PULSE OUTPUT METERS INSTALLATION AND OPERATION MANUAL
This page has been intentionally left blank.

DC000578-000 Rev.7
Fill-Rite Company
8825 Aviation Drive
Fort Wayne, Indiana 46809 USA
T1 (800) 720-5192
1 (260) 747-7524
F1 (800) 866-4681
fillrite.com | sotera.com | gormanrupp.com
This manual suits for next models
2
Table of contents
Other GORMAN-RUPP Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Sony
Sony PZW-4000 Service manual

HydroPoint
HydroPoint WaterCompass FM300 manual

Dwyer Instruments
Dwyer Instruments EMMA user manual

Sigma Metalytics
Sigma Metalytics Precious Metal Verifier quick start guide

Renishaw
Renishaw OMM-S SPRINT quick start guide

Landis+Gyr
Landis+Gyr ULTRAHEAT T230 Operating and installation instruction