
5
be dis layed rom t function, arameter values and error codes in
interactive.
(2) TDS indicator light. When the light is lit in the measuring state,means
the measured value is TDS; When the light is lit in the setting state,
means the in ut arameter is TDS.
(3) H alarm light,Under the conditions of setting high alarm, when the
measured H value of the solution is higher than the value of High
alarm, the H alarm light will be turned on and the high relay (N/O)
close; When the measured H value of the solution is lower than the
value of High alarm, furthermore lower than the hysteresis, the H
alarm light will be turned off and the high relay (N/O) o en.
(4) L alarm light,Under the conditions of setting low alarm, when the
measured H value of the solution is lower than the value of low alarm,
the L alarm light will be turned on and the low relay (N/O) close; When
the measured H value of the solution is higher than the value of low
alarm, furthermore higher than the hysteresis, the L alarm light will be
turned off and the low relay (N/O) o en.
(5) SET indicator light, the light will be turned on when entering the
setting state.
(6) uS/cm indicator light. When the light is lit in the measurement state,
means the measured value is conductivity, and unit is uS/cm.
(7) mS/cm indicator light. When the light is lit in the measurement state,
means the measured value is conductivity, and unit is mS/cm.
(8) MΩ
ΩΩ
Ω·
··
·cm indicator light. When the light is lit in the measurement state,
means the measured value is resistivity, and unit is MΩ·cm.
(9) ℃
℃℃
℃ indicator light. When the light is lit in the measurement state,
means the measured value is Tem erature; When the light is lit in the
setting state, means the in ut arameter is Tem erature.