GrayWolf IAQ SURVEYOR PRO User manual

AIR QUALITY MEASUREMENT
operating instructions
SURVEYOR INSTRUMENTS
Issue: 11 2004
IAQ SURVEYOR PRO / IAQ SURVEYOR II

2
MAN IAQ Surveyor Issue 11 2004

3
MAN IAQ Surveyor Issue 11 2004
Every effort has been made to ensure the accuracy in the
contents of our documents, however, GratWolf Sensing
Solutions, LTD can assume no responsibility for any errors or
omissions in our documents or their consequences.
GrayWolf would greatly appreciate being informed of any
errors or omissions that may be found in the contents of any
of our documents and to this end we include the following
form for you to photocopy, complete and return to us so that
we may take the appropriate action.
HELP US TO HELP YOU

4
MAN IAQ Surveyor Issue 11 2004
Marked up copies attached (as appropriate): Yes / No
Please inform me of the outcome of this change: Yes / No
For Marketing Communications, GrayWolf:
Actioned By: Date:
Response: Date:
I suggest the following corrections/changes be made to Section ...........
To: Marketing Communications,
GrayWolf Sensing Solutions
12 Cambridge Dr.
Trumbull, CT 06611
Tel : 203-416-0005
Fax : 203-416-0002
From :
Address :
Tel :
Fax :
email :
HELP US TO HELP YOU

5
MAN IAQ Surveyor Issue 11 2004
COPYRIGHT
This publication contains information partly derived from
proprietary data of GrayWolf. The express purpose of this
information is to assist in the operation and maintenance of the
instrument described herein. The publication of this information
does not convey any right to reproduce or use the information for
any purpose other than in the operation or maintenance of the
equipment described herein.
GrayWolf shall not be liable for any incidental or consequential
damages in connection with any deletions, errors or omissions in
this Manual.
PRODUCT WARNING
GrayWolf Sensing Solutions' IAQ Surveyor instruments are
NOT suitable for use in flammable or potentially explosive
environments. They are NOT certified intrinsically safe. They are
also NOT intended for use in confined spaces where operator
safety might be at risk due to exposure levels, such as reduced
oxygen, during the course of the instrument survey.
WARRANTY
GrayWolf warrants each SURVEYOR instrument to be free from
any defects in workmanship, materials and components for one
year from the date of purchase. Any SURVEYOR instrument
found to be defective by GrayWolf within the warranty period
will be repaired (or replaced at GrayWolf's discretion), providing
that the product is returned, shipping prepaid, to GrayWolf or an
authorized GrayWolf distributor.
SURVEYOR instrument failure must not have occurred as a
result of operation of the instrument other than in accordance
with the instructions furnished with the instrument.
This warranty does not apply if the instrument has been
subjected to unusual physical or electrical stress or on which the
original identification marks have been removed or altered.
COPYRIGHT AND WARRANTY

6
MAN IAQ Surveyor Issue 11 2004
This warranty will not apply if adjustments, repair or parts replacement
is required because of accident, hazard, misuse, transportation or
causes other then ordinary use; in which case
repair will be charged at normal rates. An estimate will be submitted
before repair work starts and a purchase order must be supplied
before work begins. Any out of warranty repairs are warranted 90
days from the date of invoice. The warranty does not apply if any
modifications have been made by anyone other than GrayWolf or an
authorized distributor.
The foregoing warranty is in lieu of all other warranties, conditions,
terms, undertakings and obligations implied by statute, common law,
custom, trade usage, course of dealing, or otherwise, all of which
are hereby excluded to the fullest extent permitted by law. This
warranty does not affect the statutory rights of a consumer. In such
transactions, the rights and obligations of the Buyer and Seller shall
be determined by statute.
GrayWolf shall have no liability or obligations under this warranty
other than to remedy breaches thereof by the provision of materials
and services within a reasonable time and without charge (or,
where appropriate In accordance with the above at normal rates).
If GrayWolf shall fail to comply with such obligations its liability for
such failure shall be limited to a sum equal to the price. The foregoing
states the entire liability of GrayWolf, whether in contract or in tort,
for defects in the instrument notified to it after the date of purchase
order other than liability arising where loss of damage to any property
or injury to or death of any person is caused by any negligent act or
omissions or wilful misconduct of GrayWolf, its employees, agents or
subcontractors or by any defect in the design or workmanship of the
instrument.
Save as otherwise stated, in no circumstances shall GrayWolf be
liable in contract or in tort or otherwise for any direct incidental or
consequential loss suffered by the purchaser or their agents.
GrayWolf does not give any warranty that the instrument is fit for
any particular purpose unless the purpose is specifically advised
to GrayWolf in writing and GrayWolf confirms in writing that the
instrument can fulfil that particular purpose. Representations and
warranties which are inconsistent with the terns of this warranty
are not valid unless approved in writing by an officer of GrayWolf .
Contact GrayWolf directly if there are any questions about this
warranty.
COPYRIGHT AND WARRANTY

7
MAN IAQ Surveyor Issue 11 2004
The following may be of help as a quick reference:
How do I? You should: Or
(then press MENU/METER to Refer To
return to the Measurement screen) Section:
1 Change the units Press MENU/METER; 2.4.3.1
of measurement for select Channel Setup;
one of the sensors. select sensor to change;
(not available for select Select Units;
all sensors). select new units;
. press ACCEPT.
2 Find out when a Press MENU/METER; 2.4.3.5
sensor was last select Channel Setup;
calibrated. select sensor;
select Channel Info.
3 Calibrate a sensor. Press MENU/METER; 6.1
(Not available for select Channel Setup;
all sensors) select sensor;
select Calibrate Sensor.
Select Do Standard Low Cal;
calibrate; press ACCEPT.
Select Do Standard High Cal;
calibrate; press ACCEPT.
4 Revert to the factory Press MENU/METER; 6.3.3
calibration settings select Channel Setup;
for a sensor. select sensor;
select Calibrate Sensor;
select Clear Cal;
press ACCEPT.
5 Display the Press MENU/METER; 2.5.1
min/max/average press ACCEPT;
values. press left/right arrows to show all
` available data.
TWENTY SHORT-CUTS

8
MAN IAQ Surveyor Issue 11 2004
6 Reset the Press MENU/METER; 2.5.3
min/max/average select Clear Min/Max/Avg;
calculations. press ACCEPT.
7 Change the name Press MENU/METER; 4.2.2
of a Site/Location. select Data Logging;
select Rename Site/Loc;
use alphanumeric screen to
rename;
press ACCEPT.
8 Change the default Press MENU/METER; 4.2.4
Site/Location. select Data Logging;
press ACCEPT;
select new Site/Location;
press ACCEPT.
9 Change the default Press MENU/METER; 4.3.5
logging interval. select Data Logging;
select Log Timing;
change timing;
press ACCEPT.
10 Set the instrument Press MENU/METER; 4.3.4
to log overnight. select Data Logging;
select Log Timing;
change Auto St. to ON;
set start time/date;
set finish time/date;
press ACCEPT
TWENTY SHORT-CUTS
How do I? You should: Or
(then press MENU/METER to Refer To
return to the Measurement screen) Section:

9
MAN IAQ Surveyor Issue 11 2004
11 Find out how much Press MENU/METER; 2.2.3
log memory remains. select Instrument Info;
press ACCEPT.
12 View the log data Press MENU/METER; 4.4.2
that is stored in the select Data Logging;
instrument. select View Log Data;
select Site/Location;
select session to view;
limit data to be viewed;
press ACCEPT.
13 Print the log data Press MENU/METER; 4.4.4
that is stored in the select Data Logging;
instrument. select Print Logged Data;
select Site/Location;
press ACCEPT;
select session;
limit data to be printed;
press ACCEPT.
14 Display the battery Press MENU/METER; 2.2.3
remaining life select Instrument Info,
percentage press ACCEPT
15 Lock the keypad. Press MENU/METER; 2.3.3.4
select Instrument Setup;
select Lock Keypad;
press ACCEPT.
16 Unlock the keypad. Press the cursor left and right 2.3.3.4
arrows at the same time.
TWENTY SHORT-CUTS
How do I? You should: Or
(then press MENU/METER to Refer To
return to the Measurement screen) Section:

10
MAN IAQ Surveyor Issue 11 2004
How do I? You should: Or
(then press MENU/METER to Refer To
return to the Measurement screen) Section:
TWENTY SHORT-CUTS
17 Switch on/off the Press MENU/METER; 2.3.3.1
alarm or keypad select Instrument Setup;
bleep. press ACCEPT;
press right arrow;
press up for On or down for Off;
press ACCEPT.
18 Change the alarm Press MENU/METER; 2.4.4
levels for one of the select Channel Setup;
sensors. select sensor to change;
select Set Alarms;
change alarm level;
press ACCEPT.
19 Display the serial Press MENU/METER; 2.2.3
number of a probe. select Instrument Info;
press ACCEPT
20 Switch off one of the Press MENU/METER; 2.4.3.4
sensors. select Channel Setup;
select sensor to switch off;
select Disable Channel;
press ACCEPT.

11
MAN IAQ Surveyor Issue 11 2004
CONTENTS
Section Page
Twenty Short-cuts 7
1. INTRODUCTION 17
1.1 Introduction 17
1.2 Main Features 18
1.2.1 IAQ Surveyor Pro 18
1.2.2 IAQ Surveyor II 18
1.3 Power Supplies 19
1.4 Battery Life 20
1.5 Probe Connections 20
1.6 Connecting Auxiliary Equipment 21
1.7 Keypad Functions 22
1.8 Modes of Operation 24
1.8.1 General 24
1.8.2 Meter Mode 25
1.8.3 Menu Mode 26
2. USING THE INSTRUMENTS 27
2.1 Getting Started 27
2.2 Using the Instrument Information Facility 28
2.2.1 General 28
2.2.2 Getting to the Instrument Info Screen 28
2.2.3 Details on the Instrument Info screen 29
2.3 Using the Instrument Setup Facility 31
2.3.1 General 31
2.3.2 Getting to the Instrument Setup screen 32
2.3.3 Using the Instrument Setup Options 32

12
MAN IAQ Surveyor Issue 11 2004
CONTENTS
Section Page
2.4 Using the Channel Setup Facility 36
2.4.1 General 36
2.4.2 Getting to a Channel Setup Screen 36
2.4.3 Using the Channel Setup Options 37
2.4.4 Changing the Alarm Limits 39
2.4.5 Displayed Channel Information 40
2.5 Using the Min/Max/Average Facility 41
2.5.1 General 41
2.5.2 Viewing the Mm/Max/Average Values 42
2.5.3 Resetting the Mm/Max/Average Values 43
3. MICROMANOMETER AND THERMOCOUPLE 45
3.1 General 45
3.2 Thermocouple 45
3.3 Micromanometer 45
3.3.1 General 45
3.3.2 Pressure Measurements 46
3.3.3 Autozero Facility 47
3.3.4 Pressure Connections 47
3.3.5 Volume Flow Rate 48
3.3.6 Duct Averaging - Storing Readings 50
3.3.7 Duct Averaging - Viewing Readings 50
4. DATALOGGING FACILITIES 53
4.1 General 53
4.2 Configuring the Site/Location References 53
4.2.1 General 53
4.2.2 Changing an Existing Site or Location 53
Reference
4.2.3 Adding a New Site or Location Reference 54
4.2.4 Changing the Default Site/Location Reference 55
4.2.5 Deleting a Site or Location 55
4.2.6 Navigating to a Site and Location 56
4.2.7 Using the Alphanumeric Selection Screen 57

13
MAN IAQ Surveyor Issue 11 2004
CONTENTS
Section Page
4.3 Logging the Environmental Conditions 58
4.3.1 General 58
4.3.2 Storing Individual Probe Measurements 59
4.3.3 Logging Continuously 59
4.3.4 Using the Data Logging Autostart Facility 60
4.3.5 Setting the Default Logging Interval 63
4.4 Accessing the Logged Data 64
4.4.1 General 64
4.4.2 Viewing Logged Data 65
4.4.3 Deleting Logged Data 66
4.4.4 Printing Logged Data 68
5. PC SOFTWARE 71
5.1 General 71
6. CALIBRATION PROCEDURES 73
6.1 Introduction 73
6.2 Using the Quick Cal Facility 73
6.3 Changing the Calibration Levels 74
6.3.1 General 74
6.3.2 Setting New Calibration Levels 75
6.3.3 Reverting to the Factory Settings 76
7. MAINTENANCE PROCEDURES 77
7.1 Introduction 77
7.2 Preventive Maintenance 77
7.3 Battery Maintenance 77
7.3.1 Dry Cell Batteries 77
7.3.2 Rechargeable Batteries 78
7.4 Troubleshooting 78
7.5 Service 78

14
MAN IAQ Surveyor Issue 11 2004
Section Page
8. SPECIFICATION 85
8.1 Dimensions and Weight 85
8.2 Display 85
8.3 Serial Interface (Printer) 85
8.4 Housing 85
8.5 Battery Lifetime at 20oC 85
8.6 Environmental 86
8.7 Electromagnetic Compatibility (EMC) 86
8.8 Measurement Ranges 87
9. PROBES AND ACCESSORIES 89
9.1 GrayWolf Probes Available for use with the 89
Surveyor instruments
9.2 GrayWolf Accessories Available for use with the 90
Surveyor instruments
10. PITOT STATIC TUBES AND ACCESSORIES 91
10.1 GrayWolf Pitot Static Tubes 91
10.2 Specifications 92
10.2.1 Nose and Head Section 92
10.2.2 Flexible Tubes Types 50-4, 229MS-EXT, 92
10.2.3 Rigid Tubes Types 229MS, 229MM, 299ML 92
10.2.4 Dimensions and Weight (Type 50-4) 92
CONTENTS

15
MAN IAQ Surveyor Issue 11 2004
FIGURES
Figures Page
1 Surveyor Instruments 17
TABLES
Table Page
1 Instrument Display Ranges 49
2 Basic Problems and Suggested Remedies 80
3 Summary of the Warning Messages 82
APPENDICES
A Glossary of Terms and Abbreviations 95
B Useful Conversions 99
CONTENTS

16
MAN IAQ Surveyor Issue 11 2004

17
MAN IAQ Surveyor Issue 11 2004
1. INTRODUCTION
1.1 INTRODUCTION
The IAQ Surveyor Pro and IAQ Surveyor II (Figure 1) are two
instruments capable of measuring a variety of air parameters.
These parameters include airspeed, temperature, relative
humidity, CO, CO2, etc depending on the type of probe
connected to the instrument. It should be noted that certain
probes will only operate with the IAQ Surveyor Pro. Up to
four measured parameters are continuously displayed on
the instrument's large graphics display and these readings
may be stored internally. A maximum of 16 parameters are
available depending upon the type of probes fitted to the
instrument.
The instruments can be used for hand held walk through
surveys or set up to periodically log readings which can later
be downloaded to a PC for further analysis.
This manual covers both instruments and therefore some
features listed may not be available on the Surveyor II.
Liquid Crystal
Display
Accessories
Connecting Point.
For PC, Printer or an
external ac supply.
Probe Mounting.
Showing one of the
probes available.
Probe
Connections
Simple to
use Keypad
Figure 1 Surveyor Instruments
Pressure Ports
(Surveyor Pro only)
+
-
Battery
Compartment
Start/Stop Average
Datalog.
Thermocouple
Socket
(Surveyor Pro
only)

18
MAN IAQ Surveyor Issue 11 2004
1.2 MAIN FEATURES
1.2.1 IAQ Surveyor Pro
a. Auto data logging with log download to PC.
b. Two smart probe sockets.
c. Supports all types of smart probe.
d. Supplied with 1260GSS IAQ Multi-parameter probe
as standard.
e. Built-in micromanometer and Type k thermocouple
socket.
f. Duct Averaging facility.
1.2.2 IAQ Surveyor II
a. Auto data logging with log download to PC.
b. Two smart probe sockets.
c. Supports all types of smart probe.
d. Supplied with 1260GSS IAQ Multi-parameter probe
as standard.
1. INTRODUCTION

19
MAN IAQ Surveyor Issue 11 2004
1.3 POWER SUPPLIES
Power supplies for the instruments may be:
a. Three alkaline dry cells.
Duracell MN1400 LR 14 types are recommended.
b. Three rechargeable cells.
Nickel Cadmium, Nickel Metal Hydride (NiMH) or
Lithium Ion (Li Ion).
Note: Alkaline dry cell batteries are used for battery life
indications these indications are not valid with any
other type of
battery.
Insert the three
batteries, positive
terminal first, into the
battery compartment
and securely tighten the battery end cap.
Rechargeable batteries must be removed from the instrument
and recharged externally using a proprietary battery charger.
Alternatively the instrument may be operated from an
external GrayWolf AC Adaptor operating from a 110/240 V
ac supply.
Note: When the adaptor is plugged into the instrument the
internal batteries are automatically disconnected.
There is therefore no need to remove the batteries
from the instrument when an ac adaptor is being
1. INTRODUCTION

20
MAN IAQ Surveyor Issue 11 2004
used.
1.4 BATTERY LIFE
Approximately 30 hours: Using a1260GSS IAQ Multi-
parameter probe
Approximately 170 hours: Using a 2300AHTS Multi-
parameter probe (assumes zero
air speed)
1.5 PROBE CONNECTIONS
Both versions of the instrument may operate two smart
probes simultaneously. Probes are connected to sockets
located at the bottom of the instrument as shown below:
When connecting a probe ensure that the locating peg on
the probe plug is aligned with the corresponding slot in the
socket. Push the plug firmly into the socket and carefully
screw the plug collar into the security ring.
With the exception of the IAQ Surveyor II, an additional
external temperature measurement can be made using a
suitable thermocouple connected to the k type socket on the
top of the instrument.
Two ports on the top of the instrument allow the connection
of tubing for the measurement of differential pressure
(micromanometry).
Location of the Probe Connectors
Two 8 way connectors
suitable for all probes
1. INTRODUCTION
This manual suits for next models
1
Table of contents
Other GrayWolf Measuring Instrument manuals