Introduction ACS 900
Page 2
Standby position = Position 5 of turn table is in front. Bracket of basket holder
is in front. An arrow on the display is moving from left to right.
The machine is ready to start.
Caution:
Do not place anything on top of the instrument that could block the ventilation.
The ventilator fan that evacuates the fumes during the cleaning process from the machine is
located on the top. When process is finished, ventilator stops automatically after 5 min.
Preparation of the baths
Pos. 1 = cleaning bath (with or without ultra-sound)
Pos. 2 = cleaning bath or 1st rinsing bath
Pos. 3 = rinsing bath
Pos. 4 = cleaning or lubrication bath
Pos. 5 = remains empty (drying position)
Take care not to exceed the maximum level of fluids indicated on the jar and ultrasonic
tank.
(quantity 0,55 l).
Unpack the ACS-900 and install it on the working place.
Open the front doors and remove all packing material inside the instrument.
1. Main switch off.
2. Connect the ACS 900 with the delivered power-
cable to the power socket.
Please observe the voltage (230 or 115 Volts)
marked on the label.
3. Switch on the main power switch.
The basket-holder moves upwards, the turn-table
turns to Pos. 5 in front and the bracket of basket
holder turns to front.
The instrument is now in stand-by position.
4. Press the „LOAD“ key on the keyboard to install
the ultra-sound and the glass jars.