GTE KMG-2000-G User manual

EN
Operating manual
KMG-2000-G

KMG-2000-G – Operating manual
Document revision: 305-2310-001 EN34
– Translation –
Manufacturer:
GTE Industrieelektronik GmbH
Helmholtzstr. 21, 38-40
41747 Viersen
Germany
Support-Hotline: +49 2162 3703-0
E-mail: [email protected]
© 2022 GTE Industrieelektronik GmbH – This document and all figures contained may not be copied,
changed, or distributed without explicit approval by the manufacturer!
Subject to technical changes!

I
305-2310-001 EN34
BRIEF DESCRIPTION
Brief Description
The force measuring device KMG-2000-G measures dynamic and static forces at door and gate
closing edges as well as the duration of force effects. It complies with all requirements of EN 12453,
which guarantees the complete testing and documentation of the force limitation capability of a system
according to EN 12453, as well as the standard-compliant quality assurance for the production of
doors according to EN 13849-1. DIN EN 16005 for automatic doors is also completely fulfilled.
The compact and handy device shows the entire force-time curve in the integrated graphic display
immediately after the five-second measuring period has elapsed and, in the background, combines
three consecutive individual measurements into one measuring point average value in accordance
with the standard specifications. The generous data memory of the KMG-2000-G offers space for
500 complete individual measurements or 166 complete standard measurements.
The following four measured values are determined for each measurement in order to assess the
closing force in accordance with the standard.
Permitted Peak Force Opening Widths from 50 to
500 mm
Opening Widths > 500 mm
Horizontally-moved gate 400 N 1400 N
Gate rotating on an axis per-
pendicular to the ground
400 N 1400 N
Vertically-moved gate 400 N 400 N
Gate rotating on an axis pa-
rallel to the ground
400 N 400 N
Barriers 400 N 400 N
dynamic static
Time
Fd
Fs
Fe
TdTs
Fdmax. dynamic force
Fsmax. static force
Feend force
Td
Time during where the measured
force exceeds 150 N
Ts
Time during where the measured
force exceeds 25 N

TABLE OF CONTENTS
II 305-2310-001 EN34
Inhaltsverzeichnis
1 About this Manual .......................................... 1
1.1 Explanation of Symbols .................................. 1
1.2 Storing this Manual ..................................... 2
2 For your Safety ............................................ 2
2.1 Intended Use ......................................... 2
2.2 Standards and Regulations ................................ 3
2.3 Qualification of Personnel................................. 3
2.4 Modifications......................................... 3
2.5 Abbreviations......................................... 3
3 Scope of Delivery .......................................... 4
3.1 Accessories and Spare Parts ............................... 4
4 Transport and Storage Conditions ................................ 5
5 Setup and Function.......................................... 5
5.1 Overview KMG-2000-G ................................. 5
5.2 Function ............................................ 6
6 Menu Options............................................. 6
6.1 General Notes ........................................ 6
6.2 “Options“ Menu ....................................... 6
6.2.1 Option 1: Activation Mode............................. 6
6.2.2 Option 2: Delete Mode ............................... 7
6.2.3 Option 3: Check on Measured Values Held in Memory .......... 7
6.2.4 Option 4: Lightning Adjustment .......................... 8
6.2.5 Option 5: Buzzer ................................... 8
6.2.6 Option 6: Switch off Time.............................. 8
6.3 Language Selection..................................... 9
7 Measurement Preparation ..................................... 9
7.1 Turn on the KMG-2000-G ................................ 9
7.2 Performing Measurement ................................. 9
8 Carry out Measuring Operation.................................10
9 Measurement Analysis........................................10
9.1 Display Measured Values .................................10
9.2 Further Measurements ...................................11
9.3 Storage of the Measurement Data and Graph via PC ..............11

III
305-2310-001 EN34
TABLE OF CONTENTS
10 Warnings and Error Massages ..................................11
10.1 Battery Monitoring .....................................11
10.2 Memory Space .......................................12
10.3 Memory Space Empty ...................................13
10.4 Device Still Loaded . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
11 Calibration ...............................................13
12 Disposal.................................................13
13 Technical Data ............................................14

ABOUT THIS MANUAL
1305-2310-001 EN34
1 About this Manual
This manual describes the correct use of the KMG-2000-G. It is aimed exclusively at knowled-
geable specialist personnel.
1.1 Explanation of Symbols
This manual follows a certain structure to make it easy to work with and understand. The follo-
wing designations are used throughout.
Operational objectives
rational objectives specify the result to be achieved by following the subsequent instructions.
Operational objectives are shown in bolt print.
Marking Explanation
ZThis is followed by an instruction.
1 ►
Step-by-step instructions. Instructions for action are numbered in
the order of their appearance.
Results of action steps.
Lists without a defined sequence.
Warnings
This manual uses the following information types.
DANGER!
This combination of symbol and signal word indicates an immediately dangerous
situation which could lead to death or severe injuries if it is not avoided.
WARNING!
This combination of symbol and signal word indicates a possibly dangerous
situation which could lead to death or severe injuries if it is not avoided.

2
305-2310-001 EN34
FOR YOUR SAFETY
CAUTION!
This combination of symbol and signal word indicates a possibly dangerous
situation which could lead to minor injuries if it is not avoided.
NOTICE!
This combination of symbol and signal word indicates a possibly dangerous
situation which could lead to property damage if it is not avoided.
Tips and Recommendations
This type of information provides information that is of direct importance for the
further operation of the device.
1.2 Storing this Manual
Keep this documentation at hand so that you can look it up if necessary.
2 For your Safety
In order to use the KMG-2000-G safely, it is essential to read, understand and follow these
instructions and the safety instructions contained therein.
2.1 Intended Use
The force measuring device KMG-2000-G is used for routine checking of closing forces on
gates and doors, after new installation or for the prescribed annual inspection of existing gates
with entry in the inspection book belonging to each gate. Via the built-in USB interface, the
measured values can be transferred to a laptop or PC, which allows an exact analysis of the
measured values and possible target deviations. The necessary software can be purchased
additionally.
Intended use includes observance of these instructions and compliance with all applicable
country-specific regulations. Furthermore, please observe the permissible peak forces of the
applicable standards.

FOR YOUR SAFETY
3305-2310-001 EN34
2.2 Standards and Regulations
When testing power operated doors and gates, the safety and accident prevention regulations
applicable to the specific application must be observed.
The following standards and directives are of particular importance when handling the KMG-
2000-G:
Regulation Description
EN 12453 Industrial, commercial and garage doors and gates – Safety in use
of power operated doors – Requirements and test methods
EN 13849 Safety of machinery – Safety-related parts of control systems – Part 1:
General principles for design
EN 16005 Power operated pedestrian doorsets - Safety in use - Requirements
and test methods
EN 16361 Power operated doors - Product standard, performance characte-
ristics - Door systems other than swing doors, intended for power
operation
ASR 1.7 Technical rule for workplaces – doors and gates
TS011 and TS012 DHF Code of practice for powered gates & barriers
2.3 Qualication of Personnel
All work with the KMG-2000-G may only be performed by qualified personnel. Qualified
persons are those who can perform work on electrical systems due to their technical training,
their knowledge and experience as well as knowledge of the relevant regulations and who
can recognize possible dangers.
2.4 Modications
WARNING!
Property damage due to unauthorized changes
Any form of unauthorized change can lead to property damage.
−Modication of the KMG-2000-G is explicitly prohibited!
2.5 Abbreviations
Abb. Meaning
StM Standard measurement
SiM Single measurement

4
305-2310-001 EN34
SCOPE OF DELIVERY
3 Scope of Delivery
No. Description Quantity
1KMG-2000-G 1
2USB-cable 1
3Charging adapter 2
3.1 Accessories and Spare Parts
NOTICE!
Property damage caused by using the wrong components
The use of parts other than the original spare parts and the original accessories
of the manufacturer can lead to property damage.
−Only original spare part and original accessories of the
manufacturer may be used!
The following accessories are available for the KMG-2000-G:
Product
Distance Set
Software KMG-VD 2005
1
23

TRANSPORT AND STORAGE CONDITIONS
5305-2310-001 EN34
4 Transport and Storage Conditions
NOTICE
Material damage due to incorrect storage conditions
Mechanical loads caused by improper transport will result in material damage to
the KMG-2000-G.
−No mechanical stress may occur during transport.
5 Setup and Function
5.1 Overview KMG-2000-G
12 3 45 6 1. Measuring surface
2. Enclosure
3. Display
4. Softkeys
5. Handle
6. Battery compartment

6
305-2310-001 EN34
MENU OPTIONS
5.2 Function
The KMG-2000-G force-measuring device is used for the routine, on-the-spot checking of clo-
sing forces on gates, e.g. following the installation of a new gate or for the prescribed annual
inspection of existing gates, with the result being recorded in the inspection log book relating
to each gate. The measured values can be transferred to a laptop or PC via the built-in USB
interface. By using the KMG-VD 2005 software, an exact analysis of the measured values
and possible target deviations can be carried out. The required software can be ordered as
an accessories. The centerpiece of the KMG-2000-G contains an LCD display, on which all
the measured values, the force curves, the measured-value memory management function, and
user guidance are displayed.
6 Menu Options
6.1 General Notes
The function keys F1, F2 and F3 are used for calling up menu functions and for operating the
device. The function of the keys is given by the symbols F1, F2 and F3 on the display.
If keys are not represented by these F symbols on the display, then these keys have no function
or all keys have the same function. For approx. 5 seconds after switching on the KMG a menu
is displayed allowing “Options” and “Language“ to be selected.
F1: Options Call up menu for settings
F2: New NM Initiate new normal measurement
F3:Activate Activate measuring
6.2 “Options“ Menu
6.2.1 Option 1: Activation Mode
Exit:
Options:
next function:
Activ.-mode:
1/6
F1
F2
F3
The various options can be selected one after the
other via button F1. F2 activates the option displayed.
The number of the current option of a total of 6 is
displayed top right.
In the activation mode the type of “ready to measure”
mode can be set. Button F1 switches the activation
mode between “Automatic” and “Manual” and swit-
ches back to the main menu. The illustration shows that
the “Manual” activation mode has been selected. But-
ton F3 cancels the activation menu.
Exit:
Man (Auto)
Active Mode:
F1
F3

MENU OPTIONS
7305-2310-001 EN34
Activation Mode “Auto”: After every measurement the device is automatically switched to
“ready to measure”. As soon as a force > 20N acts on the test surface, a new force/time
graph is plotted and then the current graphic is displayed and the measured values stored.
This mode is used to record several measurements one after another which should be triggered
automatically by hitting the test plate.
Activation Mode „Manual“: Prior to each new measurement the display must be switched
from showing the graph to showing the measured values using any button as desired. A new
measurement is then activated using button “F3”. At this setting the measurement of the force
cannot be triggered inadvertently by hitting the test plate.
6.2.2 Option 2: Delete Mode
After actuating one of the selection buttons relating to the type of measurement values, you are
asked once more in a further menu whether the measured values selected should be deleted.
F1 = Yes
F3 = No
By actuating F1 the data selected will be deleted.
6.2.3 Option 3: Check on Measured Values Held in Memory
In order to prevent accidental deletion, button F2
must be pressed for longer than 3 seconds.
The following delete functions can be selected on
the delete menu:
F1: entire memory is deleted
F2: last individual measurement is deleted
F3: last standard measurement along with its
3 individual measurements is deleted.
By actuating button F2 the control display relating
to those measured values deposited in the memory
is called up.
The item stored „normal measurement / single mea-
surement no.“ is displayed top right, e.g. in this
case: 2 individual measurements within the 28th
standard measurement
F1 previous measurement
F2 next measurement
F3 display of the relevant force/time graph
Exit:
Options:
Next function:
Delete results:
2/6
F1
F2
F3
last SiM:
all results: Delete results: F1
F3
last StM: F2
Exit:
Options:
Next function:
Show results:
3/6
F1
F2
F3
Td:
Fd:
Fs:
Fe:
325N
51N
0N
40ms
28/2
<F1
>F2
F3

8
305-2310-001 EN34
MENU OPTIONS
6.2.4 Option 4: Lightning Adjustment
6.2.5 Option 5: Buzzer
6.2.6 Option 6: Switch off Time
Switch off the KMG-2000-G
The device is switched on by pressing any button as desired and automatically
switches off after a set time.
After browsing all individual measurements the rele-
vant mean values appear. These are identified by
an “m” on the display.
Tdm:
Fdm:
Fsm:
Fem:
325N
51N
0N
40ms
28/3
<F1
>F2
F3
On this menu the lighting of the display can be
switched on or off using button F1.
Exit:
Options:
Next function:
Illumination:
4/6
F1
F2
F3
On this menu the internal buzzer can be
switched on or off using button F1.
The buzzer gives an audible signal once measuring
has been completed. Exit:
Options:
Next function:
Beeper:
5/6
F1
F2
F3
The device does not have an on/off switch. After
the device has been switched on by pressing any
button as desired, it switches off automatically after
a period of time that can be selected once no more
buttons have been actuated or there is no commu-
nication.
Using button “F2” the time is increased
Using button “F1” the time is decreased
Different times can be set ranging from 2 - 30 mi-
nutes
Exit:
Options:
First function:
Turn-off time:
6/6
F1
F2
F3
Exit:
Turn-off time:
6 minutes: < F1
> F2
F3

MEASUREMENT PREPARATION
9305-2310-001 EN34
6.3 Language Selection
Language selection is only available after the device
is switched on. As the device does not have an off
switch, you must wait until it switches off automatical-
ly and then switch the device on again.
You have access to the menu for selecting national
languages from the loaded menu after switching
the device on by pressing button F2.
Actuating the relevant button selects the desired
national language and switches the device over to
“ready to measure” mode.
With F3 you navigate through the individual pages
of the language selection.
7 Measurement Preparation
7.1 Turn on the KMG-2000-G
►Press any key
KMG-2000-G is switched on.
After switching on, the “start” display appears for
5 seconds:
During this time the language selection or the options
menu can be selected.
After 5 seconds the device is ready to carry out force
measurements. The current status of the measured-
value memory is displayed top left, e.g.:
2. Normal measurement
3. Individual measurement
7.2 Performing Measurement
A measuring operation is activated by pressing
button F3. The device is now ready to take mea-
surements. The measuring operation is started if a
force > 20 N acts on the test surface. The force/
time measuring operation then runs (for a measuring
period of 5 seconds).
StM: 2 SiM: 3
F1
Activate:
F2
F3
New StM:
Options:
KMG-2000-G
F3
Language:
Exit:
F1
F2
Options:
StM: 2 SiM: 3
F1
Start:
F2
F>20N
New StM:
Options:
Exit:
KMG-2000-G
Options:
Language:
F1
F2
F3
English:
German: F1
F2
> F3

10
305-2310-001 EN34
CARRY OUT MEASURING OPERATION
8 Carry out Measuring Operation
Hold the test area centrically (inner circle) to the mea-
suring points according to the standards. The measu-
ring points are specifically prescribed in relation to
the relevant type of gate construction.
Once the trigger threshold of 20 N has been pas-
sed, the force is measured over a period of 5 se-
conds. During this time the message “Measurement
- Please wait” is displayed.
Once measuring has been completed, the force/
time graph is analyzed automatically and the value
relevant to the standard are determined.
9 Measurement Analysis
9.1 Display Measured Values
Following evaluation, the force-time graph over the
measurement period (5secs) is initially displayed and
the dynamic force Fd (peak force) is displayed as
a value.
Button F1: To switch between “normal display“
(5 sec. measuring range) and „zoom display“
of the dynamic range (1sec measuring range)
Button F2: To switch to display of 4 measured values:
Max. dynamic force: Fd in N
Static force: Fs in N
End force after 5 seconds: Fe in N
Duration of dynamic force range: td in ms
Button F3: Cancellation of display mode
Once the third measurement has been taken, the
totals of 3 individual measurements required for the
standard measurement have been made. If button F3
is now actuated, the mean values of the 3 standard
measurements are displayed. The mean values are
Measurement
Please wait
Measurement
is calculated
Please wait
Td:
Fd:
Fs:
Fe:
326N
82N
17N
128ms
3/1
<F1
>F2
F3
Tdm:
Fdm:
Fsm:
Fem:
237N
58N
12N
86ms
3/3
<F1
>F2
F3
362N
362N

WARNINGS AND ERROR MASSAGES
11 305-2310-001 EN34
identified by the addition of an “m”. Buttons F1 and F2 do not have any function as the graphic
representation of the mean values doesn‘t make sense. If the mean values are displayed, the
display mode can be cancelled using F3.
9.2 Further Measurements
9.3 Storage of the Measurement Data and Graph via PC
The KMG-2000-G can be connected to any PC via a data link (USB). To do this, the cable
supplied is connected to the USB jack on the KMG-2000 G and to a USB interface on the
PC. Before data can be read from the KMG-2000-G the device must be switched on and the
„KMG-VD 2005“ program on the computer has to be started. The software tests the connecti-
on and alerts you in case of mistakes in connection or malfunctions.
The KMG-2000-G can store a total of 500 graphs (equivalent to 166 standard measurements
with 3 individual measurements). The data is preserved, even after the device has been swit-
ched off, with the result that the force graphs can also be read off the KMG-2000-G during
a subsequent period.
The individual functions can be called up via the program menu of the software. You can find
further explanations in the online help.
10 Warnings and Error Massages
10.1 Battery Monitoring
Activate:
StM: 2
Options:
New StM:
SiM: 3
F1
F2
F3
You can return to the main menu by actuating
button “F3” and
to the “Start F>20N” display by actuating
button “F3” once again Start:
StM: 2
Options:
New StM:
SiM: 3
F1
F2
F>20N
Whilst in operation the state of the rechargeable
battery is checked. If the battery charge is low, the
following message appears:
With the present level of charge further measure-
ments can still be made.
Exit:
Charge
batteries
F3

12
305-2310-001 EN34
WARNINGS AND ERROR MASSAGES
Changing batteries is not necessary as the device contains 2 rechargeable batteries. Changing
batteries is only required if the batteries are damaged. As the batteries are monitored by an
integrated charging regulator whilst charging and discharging, exchanging the batteries is only
to be expected at the end of their useful life (which is several years).
Charging of the batteries is essentially carried out via the USB jack of the KMG. Three char-
ging modes are provided for in this connection:
• The included power supply is equipped with a USB jack.
• The measuring device is connected via the USB cable to an adaptor used for car cigarette
lighters.
• The KMG is charged whilst the USB is connected to a computer via the USB interface.
The shortest charging time is achieved using the power supply unit. If the batteries are empty,
the charging time is approx. 3 - 4 hours.
Operating display:
a) without KMG connection: LED green
b) charging process: LED red
c) charging process completed, holding charge: LED green for 19 secs and red for 1 sec.
The average power consumption in case of an unilluminated display is approximately 20
mA, and in case of an illuminated display approximately 40 mA. This gives an operating life
of approximately 80 hours in normal display mode and approx. 40 hours in the case of an
illuminated display.
10.2 Memory Space
This means:
a) 166 standard measurements à three individual measurements
b) 500 standard measurements as single measurement
c) other combinations of standard/individual measurements
Whenever the KMG switches to the “ready to mea-
sure” mode while the memory is full, the following
message is displayed:
After actuating F1 branching off to the options menu
Notice: The device has a measured-value memory
capable of storing a total of 500 graphs.
Memory full
Options F1
If the “batteries low” message appears, the batteries
are depleted and must be charged up.
Exit:
Batteries
low
F3

CALIBRATION
13 305-2310-001 EN34
10.3 Memory Space Empty
10.4 Device Still Loaded
By pressing the F3 key, the measurement can be continued despite the basic force. If this
warning is indicated without a load on the measurement area, the force gauge should be sent
for maintenance to your service department.
11 Calibration
As a demand of EN 12453 and EN 16005, each force-measuring device has to be sent to
for calibration at least once a year.
12 Disposal
Send the KMG-2000-G back to the manufacturer at the
end of its useful life. This ensures environmentally friendly
disposal of all components.
If the KMG-2000-G is loaded with a force of more
than 50 N when a measurement is started manual-
ly, further measurement is possible (F3 key), but the
measurement accuracy is no longer guaranteed. The
following message is displayed:
Device still
loaded
Exit: F3
If no data is found while performing menu in 6.2.3,
„Check on Measured Values Held in Memory“, the
following message appears:
Memory empty
Exit: F3

14
305-2310-001 EN34
TECHNICAL DATA
13 Technical Data
Power supply: 2 x 1,2 V NiMH-battery
Power consumption: 20 mA
Interface: USB
Memory: 500 Individual measurements, corresponds
to 166 Standard measurements with 3 indi-
vidual measurements each
Temperature range -10 ... +60 °C
Relative humidity 20 ... 90 % rF(non-condensing)
Test surface dimensions: 80 mm Ø, height 50 mm
Dimensions: 310 mm x 80 mm x 50 mm (l x b x h)
Weight: 1400 g
Measuring range: 25 N ... 2000 N
Measuring accuracy: Typ. < 10 N
Measurement error: 25 N ... 200 N ±10 N
200 N ... 2000 N ± 5 % v. measured
value
Spring constant: 500 N/mm
Rise/fall time: ≤5 ms

Other manuals for KMG-2000-G
1
Table of contents
Other GTE Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

PCB Piezotronics
PCB Piezotronics T356A30 Installation and operating manual

horiba
horiba HR-200RT instruction manual

Neptune
Neptune E-CODER Quick install guide

Sylvac
Sylvac SCAN 52 Instruction Manual Quick Start Version

Monitor Sensors
Monitor Sensors Power + SWR instruction manual

Romus
Romus HC-3 Manual instruction

ESD-PROTECT
ESD-PROTECT EFM 822 manual

IOTRONIC
IOTRONIC AQUACON PH-CL operating manual

C-LOGIC
C-LOGIC 3400 instruction manual

Hanna Instruments
Hanna Instruments HI 93748 instruction manual

TESTO
TESTO Hygrotest 6337 9742 instruction manual

Spectrum Technologies
Spectrum Technologies Field Scout TDR 150 product manual