DR/2400 Lamp Module Replacement, continued
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:
Call 800-227-4224
Contact the HACH office or distributor serving you.
In the U.S.A. –
Outside the U.S.A. –
On the Worldwide Web – ; E-mail –
toll-free
Components
Description Cat. No.
Lamp Module Replacement
Serial Numbers less than 030900002600...................................................................................................... 59456-00
Serial Numbers 030900002600 and higher.................................................................................................. 59648-01
Instruction Sheet............................................................................................................................................. 59456-88
T10 L-Key......................................................................................................................................................... 59117-00
Torx Screws...................................................................................................................................................... 49440-00
Figure 3 Removing the Lamp Module 59456-00
(serial numbers less than 030900002600)
Note: For serial numbers 030900002600 and higher using
Cat. No. 59648-01, proceed to step 8.
4. Grasp the tab and pull the module out of the
recess.
5. Insert the replacement Lamp Module
(Cat. No. 59456-00). Push where indicated by the
arrows to fully seat the module. Ensure that the
lamp module is fully inserted before installing the
screws.
6. Reassemble the battery compartment, and replace
the batteries.
7. Return the instrument to the upright position and
restore power and other cable connections. No
calibration is necessary for the lamp module.
Figure 4 Removing Lamp Module 59648-01
(serial numbers 030900002600 and higher)
8. Grasp the tab and pull the module out of the
recess.
9. Insert the replacement Lamp Module
(Cat. No. 59648-01). Push where indicated by the
arrows to fully seat the module. Ensure that the
lamp module is fully inserted before installing the
screws.
10. Reassemble the battery compartment, and replace
the batteries.
11. Return the instrument to the upright position and
restore power and other cable connections. No
calibration is necessary for the lamp module.
© Hach Company, 2004. All rights reserved. Printed in the U.S.A. te/dk 05/04 2ed