Process Control January 2023 • Rev. 0 • Standard Devices User Manual 19
Ramping Options
Ramping options control when ramping occurs. This can be
when the setpoint changes, when the device powers on, or
when setting a zero setpoint.
• Ramp Up can toggle between on and o. When o,
the ramp rate will not be honored when increasing
flow or pressure to reach a given setpoint.
• Ramp Down can toggle between on and o. When o,
the ramp rate will not be honored when decreasing
flow or pressure to reach a given setpoint.
• Power Up toggles between Allow Ramp and No
Ramp. If set to No Ramp, the device ignores the ramp
rate just after powering on, otherwise it honors the
ramp rate starting at a zero setpoint. This functions
the same as power-up setpoint with ramping
(page 18). Changing one overwrites the other.
• 0 Setpt determines if the controller ramps when
a zero setpoint is set. If this setting is set to No
Ramp when given a zero setpoint, the controller
immediately moves to the zero setpoint. Otherwise,
the controller ramps at the selected rate.
✓ Note: Setpoint ramping can be used with flow
or pressure setpoints, depending on the control
loop selected. Ramping for pressure control
limits how quickly pressure changes before
reaching the setpoint. To limit flow rates directly
while controlling pressure, see page 21.
Batch Dispensing Flow Controllers
BATCH or MENU →CONTROL →Batch 1 or Batch 2
Batch dispensing flows a set volume of gas. Once that volume
of gas flows through the controller, the valve closes and flow
stops. You can repeat batches with a single button press.
To utilize batching, at least one totalizer must be enabled.
The BATCH button can be found on the totalizer screen
(page 14). When selecting from the
CONTROL
menu,
Batch
1is for totalizer 1 and Batch 2 is for totalizer 2. Only set one
batch at a time to avoid premature stoppage in flow.
Start a Batch
. Choose the total quantity to be dispensed in each batch.
Press SET to accept the new batch size.
2.
Once a batch size has been set, give the controller a
setpoint (page 17). Flow begins as soon as you press
SET.
✓ Note: Batch dispensing requires an active batch size
and a non-zero setpoint. If your controller already
has a non-zero setpoint, batching begins as soon
as you press SET from the batch size screen.
While a batch is being dispensed, the remaining quantity of
to be dispensed is displayed about the RESET and BATCH
buttons. When the batch is complete, –DONE– displays just
above the BATCH button and flow stops automatically. The
setpoint is not cleared and remains the same.
The batch size can be changed while a batch is in progress.
If the new batch size is larger than the current totalized flow,
then flow continues until the new value is reached. If the new
batch size is smaller than the current totalized flow, then the
flow stops immediately. Press RESET to start the new batch.
Setpoint Ramping
Setting a maximum ramp rate.
Configuring delta / time ramping.
Setpoint ramping menu.
Selecting a batch size of 1000 SL.
Totalizer displaying a mass
flow average without a batch,
with a batch in progress,
and with a finished batch.