Halma Ocean Optics DH-3PLUS-CAL User manual

Deuterium-Halogen Calibration
Light Source
Installation and Operation Manual
For Products: DH-3PLUS-CAL, DH-3PLUS-BAL-CAL
Document: 355-00000-003-01
Version: 1.0

ii 355-00000-003-01
AMERICAS & WORLD HEADQUARTERS
Phone: +1 727-733-2447
Fax: +1 727-733-3962
Sales: info@oceanoptics.com
Orders: orders@oceanoptics.com
Support: techsupport@oceanoptics.com
EUROPE, MIDDLE EAST & AFRICA
Phone: +31 26-319-0500
Fax: +31 26-319-0505
Email: info@oceanoptics.eu
Germany : +49 711-341696-0
UK : +44 1865-811118
France : +33 442-386-588
ASIA
Phone: +86 21-6295-6600
Fax: +86 21-6295-6708
Email: asiasales@oceanoptics.com
Japan & Korea: +82 10-8514-3797
www.oceanoptics.com
Copyright © 2017 Ocean Optics, Inc.
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, by any means,
electronic, mechanical, photocopying, recording, or otherwise, without written permission from Ocean Optics, Inc.
Trademarks
All products and services herein are the trademarks, service marks, registered trademarks or registered service marks of their
respective owners.
Limit of Liability
Every effort has been made to make this manual as complete and as accurate as possible, but no warranty or fitness is implied. The
information provided is on an “as is” basis. Ocean Optics, Inc. shall have neither liability nor responsibility to any person or entity
with respect to any loss or damages arising from the information contained in this manual.
Ocean Optics, Inc.
830 Douglas Ave.
Dunedin, FL 34698
USA
Manufacturing & Logistics
4301 Metric Dr.
Winter Park, FL 32792
USA
Ocean Optics Asia
666 Gubei Road
Kirin Tower Suite 601B
Changning District
Shanghai
PRC, 200336
Sales & Support
Geograaf 24
6921 EW Duiven
The Netherlands
Manufacturing & Logistics
Maybachstrasse 11
73760 Ostfildern
Germany

355-00000-003-01 iii
Table of Contents
Table of Contents.....................................................................................................................iii
About This Manual..................................................................................................................v
Document Purpose and Intended Audience ........................................................................v
What’s New in this Document..................................................................................................v
Document Summary............................................................................................................v
Product-Related Documentation..........................................................................................v
Upgrades ...............................................................................................................................vi
Important Safety Notices.........................................................................................................vi
Warranty ..................................................................................................................................vi
Certifications and Compliance...............................................................................................vii
Chapter 1.................................................................................................................................. 1
Setup......................................................................................................................................... 1
Overview................................................................................................................................ 1
Models Available.................................................................................................................... 2
Unpacking the DH-3PLUS Series Light Source...................................................................... 2
Contents ................................................................................................................................ 2
Connecting a Spectrometer to the DH-3PLUS ....................................................................... 3
Chapter 2.................................................................................................................................. 4
DH-3PLUS Specifications........................................................................................................ 4
Specifications......................................................................................................................... 4
Operating Environment .......................................................................................................... 5
DH-3PLUS Components........................................................................................................ 6
Front Panel......................................................................................................................... 6
Rear Panel......................................................................................................................... 7
Calibration Label ................................................................................................................ 8
Pinout Information.................................................................................................................. 9
Pinout Diagram .................................................................................................................10
Parts List...............................................................................................................................10
Chapter 3.................................................................................................................................11

Table of Contents
iv 355-00000-003-01
Operating Instructions............................................................................................................11
Operating the Lamp ..............................................................................................................11
Starting Main Power..........................................................................................................11
Manual Operation of Lamp....................................................................................................11
Starting the Lamp..............................................................................................................11
Turning the Lamp Off ........................................................................................................12
Warming Up the Lamp.............................................................................................................12
Manual Operation of the Shutter ....................................................................................................12
Remote Operation of the LAMP and Shutter.........................................................................13
Chapter 4.................................................................................................................................14
Troubleshooting......................................................................................................................14
Bulb Replacement.................................................................................................................15
Appendix A..............................................................................................................................16
Calibration Basics...................................................................................................................16
Overview...............................................................................................................................16
Calibration for Absolute Irradiance ........................................................................................16
Uncertainty Level Calculation for DH-3PLUS-BAL-CAL.........................................................16
Appendix B..............................................................................................................................19
DH-3PLUS Use Log.................................................................................................................19
DH-3PLUS Series Use Log...................................................................................................19
Index........................................................................................................................................21

355-00000-003-01 v
About This Manual
Document Purpose and Intended Audience
This document provides you with an installation section to get your system up and running,
basic information about the calibration light source, and a log to save the time of the light
source’s use.
What’s New in this Document
This version of the Deuterium-Halogen Calibration Light Source Installation and Operation
Manual describes the re-engineered DH-3plus family of light sources. To verify that you have a
re-engineered unit, look for the phrase “Assembled in USA” on the product label.
Document Summary
Chapter
Description
Chapter 1: Setup
Contains a list of package contents and unpacking
instructions. Also contains procedures for connecting to a
spectrometer.
Chapter 2: DH-3PLUS Specifications
Contains operating environment specifications, as well as
other physical details of the product.
Chapter 3: Operating Instructions
Provides instructions for operating the DH-3PLUS lamp and
the TTL shutter.
Appendix A: Calibration Basics
Provides an overview of the physics for radiometrically
calibrated systems.
Appendix B: DH-3PLUS Use Log
Provides a sample log for recording lamp use.
Product-Related Documentation
You can access documentation for Ocean Optics products by visiting our website at
http://www.oceanoptics.com. Select Support → Technical Documents, then choose the
appropriate document form the available drop-down lists.
Ocean Optics offers a Glossary of spectroscopy terms to help you further understand your
state-of-the-art products and how they function, located at: http://oceanoptics.com/glossary/.
Detailed instructions for OceanView Spectrometer Operating Software is located at:
http://oceanoptics.com/wp-content/uploads/OceanViewIO.pdf.

About This Manual
vi 355-00000-003-01
Upgrades
Occasionally, you may find that you need Ocean Optics to make a change or an upgrade to
your system. To facilitate these changes, you must first contact Customer Support and obtain a
Return Merchandise Authorization (RMA) number. Please contact Ocean Optics Application
Scientist for specific instructions when returning a product.
Important Safety Notices
1. Do not remove or modify any installed safety device on this equipment. Doing so will
void your warranty and create an unsafe operating environment.
2. Dangerous voltages are present in this device. There are NO user serviceable parts
inside.
3. Only allow qualified personnel to operate this unit.
4. Do not use the unit if it is damaged in any way. Contact your dealer for repair or
replacement information.
5. Always screw in the fiber optic cables before starting the instrument.
Protective eyewear must be worn when using this equipment. Intense
ultraviolet radiation is present.
Never look directly into the light beam, as this can cause eye damage.
WARNING
Warranty
Light Source products are covered by Ocean Optics Exclusive Three Year Warranty. For
details, please visit the following webpage:
https://oceanoptics.com/wp-content/uploads/Warranty-Sheet.pdf
There are no warranties for the bulbs/modules.
This instrument should not be used for any Clinical or Diagnostic purposes. Data
generated in these areas is not warranted in any way by Ocean Optics, Inc.

About This Manual
000-10000-191-02-201601 vii
Certifications and Compliance
ISO CERTIFICATION
Ocean Optics, the industry leader in miniature photonics, has been certified for ISO 9001:2008
applicable to the design and manufacture of electro-optical equipment since 2009.
WEEE COMPLIANCE
The WEEE symbol on the product indicates that the product must not be
disposed of with normal household waste. Instead, such marked waste
equipment must be disposed of by arranging to return to a designated
collection point for the recycling of waste electrical and electronic
equipment. Separating and recycling this waste equipment at the time of
disposal will help to conserve natural resources and ensure that the
equipment is recycled in a manner that protects human health and the
environment.
This device has been tested and complies with the following standards:
EN 61326-1:2013
EN 61000-4-2:2009 / IEC 61000-4-2:2008
EN 61000-4-3:2006 / IEC 61000-4-3:2006
EN 61000-4-5:2006 / IEC 61000-4-5:2005
EN 61000-4-11:2004 / IEC 61000-4-11:2004
EN 55011: 2009/A1:2010 Group 1 Class A
ANSI C63.4:2003

About This Manual
viii 355-00000-003-01
FCC COMPLIANCE
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates uses and can radiate radio
frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which the user
will be required to correct the interference at his own expense.

355-00000-003-01 1
Chapter 1
Setup
Overview
The DH-3PLUS Series of Light Sources for the UV-VIS-NIR (210nm-2400nm, depending upon
configuration) is a deuterium and halogen light source that provides known absolute intensity
values at several wavelengths, expressed in µW/cm²/nm. Since the spectral intensity of the DH-
3PLUS Series can be traced to an intensity standard traceable to the National Institute of
Standards and Technology (NIST), it is specifically designed for calibrating the absolute spectral
response of your system.
Calibrated sources are guaranteed for the lesser of 50 hours of operation or 1 year from factory
calibration. Due to low usage for guaranteed calibration, they are not intended to be used as an
illuminant in day-to-day measurements.
Note
If you have a spectrometer setup that is highly sensitive, you may not be able to
use the DH-3PLUS Series as a calibration source.
DH-3 plus Calibrated Light Source

1: Setup
2355-00000-003-01
Models Available
The following models are available in the DH-3PLUS Series:
Model
Default
Calibration
Wavelength
(nm)
Optional Calibration Wavelength (nm)
Balanced
Spectral
Output
DH-CAL *
DH-CAL-BF**
DH-EXT-CAL*
DH-EXT-CAL-BF**
DH-3P-CAL
210-1100
250-1100
210-2400
250-2400
No
DH-3P-BAL-CAL
230-1100
250-1100
230-2400
250-2400
Yes
*Cosine Corrected
**Bare Fiber
The following sections provide instructions on unpacking and setting up your DH-3PLUS Series
Light Source.
Unpacking the DH-3PLUS Series Light Source
►Procedure
1. Unpack your lamp assembly and power supply carefully. Although the lamp is rigidly
mounted, dropping this instrument can cause permanent damage.
2. Inspect the outside of the instrument and make sure that there is no damage. Do not use
the instrument if damage is present.
3. Use this instrument in a clean laboratory environment (see

1: Setup
000-10000-191-02-201601 3
4. Operating Environment).
Contents
Your DH-3PLUS Series Light Source package should contain the following:
DH-3PLUS light source unit
Power cord
UV safety goggles
One IC-DB15-2 interface cable for shutter operation
Lamp Calibration Report for using the DH-3PLUS light source with a CC-3-UV-S cosine-
corrected irradiance probe
USB containing lamp calibration files
Uncertainty report for unit
Calibration certificate for unit
Allen wrench for adjusting the cosine corrector
Connecting a Spectrometer to the DH-3PLUS
Before using a DH-3PLUS Series light source for the first time, check for transport damage. Be
sure to adhere to all warnings on the unit and in this operational manual. Make sure you have
your DH-3PLUS light source, your spectrometer, a CC-3 cosine-corrected irradiance probe, and
Ocean Optics spectrometer operating software.
►Procedure
Use the following procedure to connect your DH-3PLUS light source to a spectrometer:
1. Loosen the set screw on the DH-3PLUS's SMA connector.
2. Screw the CC-3 cosine corrector onto the end of the fiber. The connection should be
tight. Insert the CC-3 all the way into the DH-3PLUS's SMA connector.
3. Tighten the setscrew on the SMA connector of the DH-3PLUS with an Allen wrench.
Connect the other end of the fiber into the SMA connector of the spectrometer.
4. Plug in the power cord.
5. Turn the unit on using the ON/OFF switch on the back of the light source.
6. Turn on the deuterium light and the halogen light on the front of the light source.
7. Allow the lamp to warm up for at least 15 minutes before using it. Place the DH-3PLUS
horizontally.
8. Insert the calibration USB drive that came with your light source into your computer. This
disk contains one ASCII file with the same information as the Lamp Calibration Reports
that came with your DH-3PLUS. The file name contains the lamp's serial number,
followed by CC.LMP. Copy these files to the desired directory on your computer.
9. Start your spectrometer operating software. See your spectrometer manual for
instructions on calibrating the spectrometer with the DH-3PLUS.

1: Setup
4355-00000-003-01
Tips
When you perform measurements with the calibrated spectrometer system,
record a new dark spectrum before each measurement. This minimizes the
influence of a shifted dark spectrum.
Use a log to monitor the operation time of the light source. Recalibration of the
light source after 50 hours of operation is recommended. A sample log is
provided in Appendix B: DH-3PLUS Use Log.

355-00000-003-01 5
Chapter 2
DH-3PLUS Specifications
This section provides information on the operating environment and specifications of the DH-
3PLUS Series light sources.
Specifications
Specification
Value
Spectral Range (calibrated):
Cosine corrected
DH-3PLUS-BAL-CAL
DH-3PLUS-BAL-CAL
DH-3PLUS-CAL
DH-3PLUS-CAL
Bare Fiber
DH-3PLUS-BAL-CAL
DH-3PLUS-BAL-CAL
DH-3PLUS-CAL
DH-3PLUS-CAL
230-1100 standard
230-2400 extended
210-1100 standard
210-2400 extended
250-1100 standard
250-2400 extended
250-1100 standard
250-2400 extended
Dimensions (LWH):
13.8 cm x 6.3 cm x 6.1 cm; 5.4" x 2.5" x 2.4"
Power Requirements:
85-264 V, 50/60 Hz
Power Consumption:
Approximately 78VA
Time to Stabilized Output
40 minutes –deuterium
20 minutes –tungsten halogen
Connector
CC-3-UV-S or 6.35 mm barrel for cosine corrector

2: DH-3PLUS Specifications
6355-00000-003-01
Operating Environment
The following table provides information on optimizing the operating environment of your DH-
3PLUS calibration light source.
Moisture
5-95% without condensation at 40°C
Ventilation
Unit should be situated so that its location or position does not interfere with
proper ventilation.
Temperature
5°C –35°C. Unit should be situated away from any device that emits excessive
heat.
Object and Liquid
Entry
Unit should be positioned so that objects do not fall on top of the unit.
Additionally, ensure that no liquids are spilled into the enclosure through
openings.

2: DH-3PLUS Specifications
355-00000-003-01 7
DH-3PLUS Components
The following sections describe the components located on the front and rear of the DH-3plus
unit.
Front Panel
Component
Description
Deuterium On/Off
Press the DEUTERIUM button to turn the Deuterium lamp on or off.
The bulb requires a warm-up time of 20 seconds before the
Deuterium lamp is illuminated. You must allow the bulb to warm up to
receive accurate data from the lamp.
LED to the left of the button lights green upon successful illumination or
red to indicate lamp malfunction.
Power LED
Indicates the power state of the DH-3PLUS.
Mechanical Protection –
SMA Connector
Covered to protect users from unintentionallylooking directlyat the
beam of light. Connect the fiber cable to the DH-2000 BEFORE
turning the lamp on to avoid unnecessaryexposure to UV radiation.
Always wear proper eye protection when using the DH-2000 lamp.
Halogen On/Off
Press the HALOGEN button to turn the Halogen lamp on or off.
LED to the left button lights green upon successful illumination or red to
indicate lamp malfunction.
Shutter Mode Switch
Sets the operational mode of the shutter. Open indicates that
the shutter is constantly open; Closed indicates that the
shutter is constantly closed; TTL indicates that the shutter is
operated via the rear panel connector (see Rear Panel).
Power LED
Halogen On/Off Switch
Halogen Status LED
Shutter Switch
Mechanical Protection
–SMA Connector
Deuterium Status LED
Deuterium On/Off
Switch

2: DH-3PLUS Specifications
8355-00000-003-01
Rear Panel
Component
Description
Main Power Switch
Turn on to supply power to the unit. The Power LED lights when
this switch is in the ON position.
Fuse Compartment
Contains the fuse to protect the unit against overload:
2.0 Amp, 250 V slow blow. Littlefuse part # 0218002
Input Terminal for Power
Cord
Plug power cord into this terminal.
Note: Only connect the power cable to the lamp when the Main
Power Switch is in the OFF position.
THE INSTRUMENT MUST BE CONNECTED TO A GROUNDED
(EARTHED) OUTLET
TTL Input
D-SUB 15-pin connector for automatic shutter control and remote
lamp enable
Unit Label
- Model DH-3PLUS-CAL Deuterium-Halogen Lightsource
-or-
DH-3PLUS-BAL-CAL Deuterium-Halogen Lightsource
- Version Europe / USA
- Main connection 110 / 240V 50-60Hz
- Power consumption 80 Watt
- Max. Ambient Temperature 35°C
- Serial-No. xxxxxx
Cooling Fan
Cools the interior of the DH-3PLUS. Do not obstruct.
Unit Label
Cooling Fan
TTL Input
Input Terminal for
Power Cord
Main Power Switch
Fuse Compartment

2: DH-3PLUS Specifications
355-00000-003-01 9
Calibration Label
A calibration label is attached to both the shipping box and the light source. The label identifies
the calibration performed on the unit (shown in red oval, below) and the date of the calibration,
as well as the serial number of the light source.
The calibration acronyms are defined in the following table:
Calibration Acronym
Description
CC
Cosign Corrector - Standard Range
CC EXT
Cosign Corrector –Extended Range
BF
Bare Fiber - Standard Range
BF EXT
Bare Fiber - Extended Range
CC & BF
Cosign Corrector and Bare Fiber
CC& BF EXT
Cosign Corrector and Bare Fiber Extended Range

2: DH-3PLUS Specifications
10 355-00000-003-01
Pinout Information
The following table contains pinout information for the DH-3PLUS Light Sources:
Pin
Description
1
Must be pulled up by use to 2.6V to 5 Volts
2
Deuterium Lamp –Turns the Deuterium Lamp on/off when 2.6V to 5V is applied to
the pin (Rising Edge ON, Rising Edge OFF). *
3
Halogen Lamp - Turns the Halogen Lamp on/off when 2.6V to 5V is applied to the
pin (Rising Edge ON, Rising Edge OFF). *
4
Light Source Ground
5
HAL_BAD: In the event the Halogen lamp is bad, this will be active low, otherwise it
will be hi. This equates to the Red Halogen LED on the front. Note this LED usually
flashes briefly at power up
6
NA
7
NA
8
D_BAD: In the event the Deuterium lamp is bad, this will be active low, otherwise it
will be hi. This equates to the Red Deuterium LED on the front
9
D_GOOD: This is active low when the Deuterium is on. It will toggle at 1 Hz during
warm up, just like the Green Deuterium LED on the front
10
Shutter Ground
11
NA
12
NA
13
Lamp Enable Signal (Rising Edge ON, Rising Edge OFF, 2.6V to 5V). *
14
HAL_GOOD: This is active low when the Halogen is on. Just like the Green
Halogen LED on the front.
15
SHTR_STAT: Should be active HI if the shutter is OPEN
15
SHTR_STAT: Should be active HI if the shutter is OPEN
*This pull up is only required to use the OUTPUT signals on pins 5, 8, 9, 14, and 15. The input control signals
on pins 2, 3, and 13 do not require this pull up.
A label on the rear of the unit also provides pin information.

2: DH-3PLUS Specifications
355-00000-003-01 11
Pinout Diagram
Parts List
Part Number
Description
DH-3P-CAL
Deuterium Tungsten-Halogen Light Source, Calibrated, 210-
1100nm
DH-3P-BAL-CAL
Deuterium Tungsten-Halogen Light Source, Balanced and
Calibrated, 230-1100nm
Littlefuse 0218002
Fuse
DH-CAL
Calibration service for DH light sources
DH-3P-CAL: 210-1100nm
DH-3P-BAL-CAL: 230-1100nm
DH-CAL-BF
Bare fiber calibration service for DH sources
DH-3P-CAL: 250-1100nm
DH-3P-BAL-CAL: 250-1100nm
DH-EXT-CAL
Extended range calibration service for DH sources
DH-3P-CAL: 210-2400nm
DH-3P-BAL-CAL: 230-2400nm
DH-EXT-CAL-BF
Extended range bare fiber calibration service for DH sources
DH-3P-CAL: 250-2400nm
DH-3P-BAL-CAL: 250-2400nm

355-00000-003-01 12
Chapter 3
Operating Instructions
Operating the Lamp
The following sections provide instructions on operating the deuterium and halogen lamps in the
DH-3PLUS Light Source. The DH-3PLUS unit must be in a horizontal position for it to work.
Starting Main Power
To apply main power to the unit,
1. Ensure that the Main Power Switch is OFF. Then, connect the power cord to the
Terminal Input on the rear panel of the unit.
2. Turn the Main Power Switch ON. The Power LED lights, indicating that the unit is
receiving power. The Status LED blinks momentarily after the main power is switched
on. The deuterium and halogen bulbs remain off after power-on; they must be turned on
separately.
Manual Operation of Lamp
Starting the Lamp
Press the DEUTERIUM or HALOGEN On/Off switch down to ignite the desired lamp.
The LED blinks green until the lamp illuminates (approximately 20 seconds for the
deuterium lamp, 1-2 seconds for the halogen lamp.)
Deuterium Status LED
Halogen Status LED
Halogen On/Off Switch
Deuterium On/Off
Switch
This manual suits for next models
1
Table of contents
Other Halma Industrial Equipment manuals