Hangzhi PSM Series User manual

PORTABLE AC/DC STANDARD METER
USER MANUAL
PSM SERIES
SHENZHEN HANGZHI PRECISION ELECTRONICS CO. LTD
www.hangzhicn.com

1
PORTABLE AC/DC STANDARD METER
USER MANUAL
V2.2
Founded in 2017, Shenzhen Hangzhi Precision Electronics Co., Ltd. is a
technology-leading enterprise dedicated to the development, production, sales and
customization of high-precision current transducers and measuring instruments. We will
strive to build a well-known brand of precision current transducers and precision
instruments in the DC field, and become a leading international leader in precision
electronics in the field of DC systems.
Based on multi-faceted technology integration and innovation, Shenzhen Hangzhi
Precision Electronics Co., Ltd. has developed the industry's first high-precision digital
current transducer and an analog current transducer featuring high precision, low costs, low
zero drift and low temperature drift. This series of products reduces industry costs,
improves industry efficiency, enhances user experience, and creates value for customers.
The company's products have won many achievements in the national innovation and
entrepreneurial competition, and won wide attention and support from all walks of life.
As a company with strong sense of responsibility and mission, we adhere to multi-point
zero-flux technology-led approach, with client-oriented service and customized products,
and improve the operating quality by successfully capital financing. We are making our
efforts to build an innovative sharing enterprise.

2
Table of content
1Preface.........................................................................................................................................4
1.1 Packing checklist .............................................................................................................4
1.2 Accessories.......................................................................................................................5
1.3 About safety......................................................................................................................5
1.4 About label........................................................................................................................6
1.5 About measurement level...............................................................................................7
1.6 Precautions for use..........................................................................................................8
1.6.1 Inspection before use ...........................................................................................8
1.6.2 Placement environment........................................................................................8
1.6.3 Placement method................................................................................................9
1.6.4 Use of the instrument............................................................................................9
1.6.5 Before connecting the power cord......................................................................9
1.6.6 Before connecting the test cable ......................................................................10
1.6.7 Before turning on the power..............................................................................10
1.6.8 Before measurement..........................................................................................11
1.6.9 Before connecting the communication cable..................................................11
2Summary....................................................................................................................................12
2.1 Product summary...........................................................................................................12
2.2 Product characteristics..................................................................................................12
2.3 Product composition......................................................................................................13
3Product selection guide and technical parameters.............................................................15
3.1 Product selection ...........................................................................................................15
3.2 Technical parameters....................................................................................................15
3.3 Phase and frequency....................................................................................................16
3.4 Harmonic measurement ...............................................................................................17
3.5 Other technical parameters..........................................................................................17
4Instructions for use...................................................................................................................18
4.1 Steps................................................................................................................................18
4.2 Instructions of interface.................................................................................................18
4.2.1 Boot interface.......................................................................................................18
4.2.2 Main interface.......................................................................................................18
4.2.3 AC measurement interface................................................................................19
4.2.4 DC measurement interface................................................................................20
4.2.5 Setting interface...................................................................................................22
4.2.6 Power energy error .............................................................................................22
4.2.7 Daily timing error interface.................................................................................23
5Connector information.............................................................................................................24
5.1DB9 terminal definition (DB9 male).............................................................................24
5.2 Aviation terminal definition............................................................................................24
6Dimensions................................................................................................................................25
7Maintenance and service........................................................................................................26

3
7.1 Calibration and repair....................................................................................................26
7.2 Instrument transportation..............................................................................................26
7.3 Replacement of parts and life......................................................................................26
7.4 Cleaning..........................................................................................................................26
7.5 Frequently asked questions.........................................................................................26
Attachment 1 Communication agreement...................................................................................28
Attachment 2 PSM600 pulse constant comparison table.........................................................29
Attachment 3 PSM1000 pulse constant comparison table.......................................................31
Attachment 4 PSM1500 pulse constant comparison table.......................................................33

4
1 Preface
Thank you for choosing HANGZHI "PSM Series Portable AC/DC Standard Meter". In order
to make full and lasting use of this product, please keep the manual properly. HANGZHI
PSM series portable AC/DC standard meter is referred as "this instrument" below.
1.1Packing checklist
When this instrument is delivered to you, please check if any abnormalities or damages
occur during transportation before using it. In particular, please pay attention to accessories,
panel, keys and other items. In case of damage or failure to work, please contact the agent
or HANGZHI service center.
Please keep the packaging material for delivery properly for future transportation.
Please make sure that the contents of the packing are correct.
Packing checklist
Item
Product photo
Specifications
Description
□ This instrument
See Part 3
Portable AC/DC standard
meter
□ User manual (This
manual)
Soft copy or hard
copy
To describe the
operational method,
specifications, etc.
□ Power line
1.5 m/3*0.75 mm2
Rated voltage: 250V
Rated current: 10A
For power supply
□ Voltage test line
1 m/2.5mm2
Rated voltage: CATIII
1000V/CATIV 600V
Rated current: 10A
To measure the voltage
input signal
□ Small current test
line
1 m/2.5 mm2
Rated current: 10A
To test small current input
signal
□ Power pulse line
1 m/5*0.3mm2
To connect power pulse
input/output interface,
clock pulse input interface
Note:

5
1) This instrument has been programmed when it was manufactured, and the latest
version can be downloaded from the homepage of our company
2) Instructions for use in other languages are available at our website:
http://www.hangzhicn.com/
1.2Accessories
This instrument has the following options (to be sold separately). Please contact the agent
or sales center if you need purchase.
Option list
Item
Product photo
Specifications
Description
□ USB to RS232
connection line
1.8 m/USB2.0/RS232
It can be used to transfer
PC interface from USB2.0
to RS232.
□ RS232
connection line
2 m/3*0.3mm2
DB9 Female to
female/23 connection
line
It can be used to connect
between RS232 and
communication interface of
this device.
□ USB to RS485
connection line
1.5 m/USB2.0/RS485
It can be used to transfer
PC interface from USB2.0
to RS485.
□ RS485
connection line
0.1 m/2*0.3mm2
DB9 Female to female
It can be used to connect
between RS485 and
communication interface of
this device.
□ USB extension
line
2.0 m/USB2.0/ Male to
male
It can be used for LCD
screen program upgrade.
□ AC adapter
For overseas usage
Power adapter for different
countries.
1.3About safety
The instrument is designed and tested in accordance with IEC61010 safety specifications,
and is shipped in a safe state. In addition, failure to comply with the instructions may
damage the functions provided by the instrument to ensure safety. Before using this
instrument, please read carefully the following safety-related matters.

6
Danger
If wrong method is used, it may lead to personal accident and instrument failure. Read the
instructions carefully and operate after fully understanding the contents.
It includes electrical hazards such as electric shock, heating, fire and arc discharge caused
by short circuit. Personnel who first use electrical measuring instruments should use them
under the supervision of senior electrical measuring personnel.
1.4About label
This manual classifies and marks the severity and risk levels of risks as follows.
Danger
A dangerous situation that is highly likely to cause death or
serious injury to the operator is described.
Warning
Situations that are likely to result in death or serious injury to
the operator are described.
Note
Conditions that may result in minor injury to the operator or
expected damage or malfunction of the instrument are
described.
Important
matters
Information or content regarding operations and
maintenance work that must be known in advance are
described.
An indication of high voltage hazard is used to warn the risk
of shock, burns and even death from electric shock due to
neglection in safety confirmation or misuse.
Prohibited behavior is indicated.
The "mandatory" matter which must be performed is
indicated.
Symbols on the instrument
To indicated caution or danger. When the symbol
is displayed on the instrument, please refer to the
corresponding position in the instruction manual.
To indicate the ground terminal.
To denote direct current (DC).
To denote alternate current (AC).
To denote power “ON”.
To denote power “OFF”.
Warning

7
Symbols related to standard
Marking of regulations on the
abandonment of electrical and
electronic equipment (WEEE
Directive) in EU countries.
Consistent with the restrictions
shown in the EC Ministerial Council
Directive (EC Directive).
1.5About measurement level
In order to use the measuring instrument safely, IEC61010 classifies the measurement into
three safety levels of CAT II to CAT IV according to the places of use.
Danger
◼Using a measuring instrument with a small
classification level in a large numerical level may cause
major accidents, so please avoid this situation absolutely
◼Using a measuring instrument without a classification
mark to measure CAT II to CAT IV may lead to major
accidents, so please avoid this situation absolutely
This instrument is suitable for CAT III 1000 V.
CAT Ⅱ:The primary side circuit of an instrument (movable tool, household appliance, etc.)
with a power cord that connects to the outlet, when the socket is directly
measured.
CAT Ⅲ:Measuring the primary side circuit of an instrument (fixed device) that is directly
powered from the switchboard, and the circuit from the switchboard to the outlet.
T
CAT IV
Electric
meter
Switchboard
Inspection entrance
Usage entrance
Internal wiring
CAT III
CATII
Socket

8
1.6Precautions for use
In order to use the instrument safely and make full use of its functions, please observe the
following precautions.
1.6.1 Inspection before use
Warning
•If the test cable or the instrument is damaged, it may cause electric
shock. Be sure to do the following checks before using it.
•Before using it, please confirm that there are no problems caused
by storage and transportation, and use it after checking and
confirming the operation. If it is confirmed to be faulty, please
contact the agent or the company after-sales center.
•The outer surface of the power cord damage or exposure may
cause an electric shock or short circuit accident. Please do not use,
and contact your dealer or company after-sales center
•The outer skin of the cable damage or metal exposure may cause
short circuit or electric shock. Please replace with a device that is
not damaged.
•Check whether the instrument is damaged. If it is damaged, please
send it for repair.
•When the power is turned on and the start button is lit red, the
power cord may be broken or a malfunction occur inside the
instrument. Please send it for repair.
•After the end of the test (displaying the company LOGO), if the
main measurement function screen is not displayed, a malfunction
may occur inside the instrument. Please send it for repair.
1.6.2 Placement environment
Warning
Please do not place the instrument in the following places,
otherwise it will cause malfunction or accident of the instrument.
•Direct sunlight or high temperature places
•Locations where corrosive gases and explosive gases are
generated
•Places where strong electromagnetic waves are generated
or near charged objects
•Close to induction heating device (high frequency
induction heating device, IH induction cooker, etc.)
•Locations where mechanical vibrations are frequent
•Locations affected by water, oil, chemicals and solvents
•Wet, dew condensation
•A place with a lot of dust

9
1.6.3 Placement method
Warning
Please do not place on unstable pedestals or in inclined places.
Otherwise, personal injury or malfunction of the main unit may occur
due to falling or tipping over.
•Place the bottom side down.
•In order to prevent the temperature of the instrument from rising,
please be sure to keep a specified distance from the surroundings
when placing it.
Above 50mm Above 50mm Above 150mm
The means to cut off the power supply of this instrument is to unplug the power cord. In
case of emergency, the power cord can be unplugged to cut off the power supply
immediately, so please make sure that there is enough space for operation.
1.6.4 Use of the instrument
Danger
To prevent an electric shock, never remove the main unit casing, since
there are high voltage and high temperature parts inside.
Note
In order to prevent damage to the instrument, please avoid vibration
and collision during handling and use, and pay attention to collisions
caused by falling.
1.6.5 Before connecting the power cord
Warning
To avoid electric shock and to ensure the safety of this instrument,
please connect the supplied power cord to a three-phase outlet.

10
1.6.6 Before connecting the test cable
Danger
Be sure to connect the test cable to the secondary side of the circuit
breaker. Even if a short circuit occurs on the secondary side of the
circuit breaker, the short circuit current is cut by the circuit breaker. The
current capacity on the primary side is very large, and in the event of a
short circuit accident, damage to the instrument or equipment may
occur.
Warning
To avoid electric shock and short circuit accidents, please use the
specified test cable.
1.6.7 Before turning on the power
Warning
Before turning on the power, please confirm whether the power
voltage listed on the power connection of the instrument and the one
you are using are consistent.
Using the power supply voltage outside the specified range may cause
instrument damaged or an electrical accident.
Note
Do not use UPS and DC-AC inverters with square wave or
approximate sine wave output to drive this instrument to avoid damage
to this instrument.

11
1.6.8 Before measurement
When measuring voltage
Danger
•The maximum in-phase voltage of the voltage measurement terminal
is as follows.
CAT II: AC/DC 300 V
Without measurement classification:AC/DC 800 V
Exceedance of this voltage may cause damage to the instrument or
cause personal injury.
•The maximum input voltage of the voltage measurement terminal is
DC 1000 V and 1100 V peak.
When the voltage exceeds 800V, it can be measured only when the
object to be tested is insulated from the ground. Exceedance of this
voltage may cause damage to the instrument or personal injury.
•To prevent an electric shock, do not use the test cable tip to avoid
short circuit in the voltage-applied circuit.
1.6.9 Before connecting the communication cable
Note
When connecting or removing the communication cable, please be
sure to turn off the power of the instrument and the connected device.
Failure to do so may result in false action or malfunction.

12
2 Summary
2.1Product summary
PSM series portable AC/DC standard meter is a new generation of standard meter
produced by our company. The product adopts a new software and hardware design, which
can simultaneously measure single-phase AC and DC voltage, current, frequency, phase,
active power, reactive power, inspecting power, power factor, active energy, and reactive
energy etc., at the same time, high-order harmonic distortion (up to 63 harmonics) can be
measured. It can be widely used in metrology research institutes, government inspection
agencies and other metrology fields for the metrological verification of single-phase AC and
DC voltmeters, ammeters, electric energy meters and power meters.
2.2Product characteristics
➢To measure single-phase AC and DC voltage, current, frequency, phase, active
power, reactive power, inspecting power, power factor, active energy, and reactive
energy etc.
➢To detect power error and daily timing error
➢To perform electric energy word test
➢Ripple test can be performed to detect AC ripple below 1 kHz
➢To measure the distortion of voltage and current, perform 2~63 harmonic analysis,
display real-time harmonic components, harmonic content, harmonic distortion, etc.
➢Equipped with energy pulse input interface which can be used for real-time
verification of various active energy meters, reactive energy meters, etc.
➢Equipped with standard energy pulse output interface for automatic setting of pulse
constant
➢Equipped with RS232, RS485 and optional CAN communication interfaces, and it
can communicate directly with PC
➢Multi-range can be switched automatically for voltage and current, and 110% of
measuring range limit can be reached
➢Equipped with product program online upgrade function

13
2.3Product composition
Front
1
Company logo
Company logo
2
Product model
The model and name of the
corresponding product
3
Current measuring
hole
Perforation in the specified direction
when measuring large current
4
Small current
measuring terminal
Terminals that are directly inserted
during small current measurement, with
red terminal connected to current input,
and black terminal connected to current
output.
5
Display area
For user interaction, to display
measurement data, parameter settings,
etc.
Back
6
Vents
For body cooling
1
2
3
4
5
6

14
Left
7
DB9 interface
Used for RS232, RS485, CAN communication
8
USB interface
Used for display program upgrade
9
Aviation interface
For pulse input and output
10
Product serial number
The unique serial number of the product
11
Manufacturing nameplate
For management purpose, please don’t peel off.
12
Ground terminal
Used to connect to the ground when using this instrument
Right
13
Voltage measurement
terminal
Terminals that are directly inserted during voltage
measurement, with red terminal connected to
positive voltage, and black terminal connected to
negative voltage.
14
Power input and main power
switch
To connect power and for ON/OFF of the main
power
15
Fan hole
For body cooling
7
8
9
10
11
12
13
15
14

15
3 Product selection guide and technical parameters
3.1Product selection
PSM Series Product Selection
PSM600-E
PSM1000-E
PSM1500-E
DC Voltage Measurement (DCV)
20V~1000V
20V~1000V
20V~1000V
DC Current Measurement (DCI)
500mA~600A
1A~1000A
75A~1500A
AC Voltage Measurement (ACV)
30V~707V
30V~707V
30V~707V
AC Current Measurement (ACI)
500mA~424A
1A~707A
50A~1000A
3.2Technical parameters
PSM600-E
PSM1000-E
DC Current
Measuring Limit
600A
1000A
Measuring Range
(0~110%)RG
(0~110%)RG
Accuracy
±0.02%RD(500mA≤I≤600A)
±0.02%RD(1A≤I≤1000A)
Resolution
0.002%RD(500mA≤I≤600A)
0.002%RD(1A≤I≤1000A)
DC Voltage
Measuring Limit
1000V
1000V
Measuring Range
(0~110%)RG
(0~110%)RG
Accuracy
±0.02%RD(20V≤U≤1000V)
±0.02%RD(20V≤U≤1000V)
Resolution
0.002%RD(20V≤U≤1000V)
0.002%RD(20V≤U≤1000V)
AC Current
Measuring Limit
424A
707A
Measuring Range
(0~110%)RG
(0~110%)RG
Accuracy
±0.05%RD(500mA≤I≤424A)
±0.05%RD(1A≤I≤707A)
Resolution
0.005%RD(500mA≤I≤424A)
0.005%RD(1A≤I≤707A)
AC Voltage
Measuring Limit
707V
707V
Measuring Range
(0~110%)RG
(0~110%)RG
Accuracy
±0.05%RD(30V≤U≤707V)
±0.05%RD(30V≤U≤707V)
Resolution
0.005%RD(30V≤U≤707V)
0.005%RD(30V≤U≤707V)
DC power/
Electric
Energy
Accuracy
±0.02%RD
(20V≤U≤1000V, 500mA≤I≤600A)
±0.02%RD
(20V≤U≤1000V, 1A≤I≤1000A)
AC power/
Electric
Energy
Accuracy
±0.05%RD
(30V≤U≤707V, 500mA≤I≤424A)
±0.05%RD
(30V≤U≤707V, 1A≤I≤707A)
Note:RD denotes reading value,RG denotes measuring range value (The same below).

16
PSM1500
DC Current
Measuring Limit
1500A
Measuring Range
(0~110%)RG
Accuracy
±0.02%RD(75A≤I≤1500A)
Resolution
0.002%RD(75A≤I≤1500A)
DC Voltage
Measuring Limit
1000V
Measuring Range
(0~110%)RG
Accuracy
±0.02%RD(20V≤U≤1000V)
Resolution
0.002%RD(20V≤U≤1000V)
AC Current
Measuring Limit
1000A
Measuring Range
(0~110%)RG
Accuracy
±0.05%RD(50A≤I≤1000A)
Resolution
0.005%RD(50A≤I≤1000A)
AC Voltage
Measuring Limit
707V
Measuring Range
(0~110%)RG
Accuracy
±0.05%RD(30V≤U≤707V)
Resolution
0.005%RD(30V≤U≤707V)
DC power/Electric Energy
Accuracy
±0.02%RD
(20V≤U≤1000V, 75A≤I≤1500A)
AC power/Electric Energy
Accuracy
±0.05%RD
(30V≤U≤707V, 50A≤I≤1000A)
3.3Phase and frequency
Technical parameters of phase and frequency
Phase
measurement
Measuring range
0.000°~359.999°
Accuracy
±0.02°
Resolution
0.001°
Frequency
measurement
Measuring range
40Hz~70Hz
Accuracy
±0.01Hz
Resolution
0.001Hz

17
3.4Harmonic measurement
Allowable error of harmonic measurement
Grade
To be measured
Condition
Allowable error
A
Voltage
Current
Note: UNdenotes nominal voltage, Ubdenotes harmonic voltage, INdenotes rated
current, Ibdenotes harmonic current.
3.5Other technical parameters
Other technical parameters
Power pulse output
Power pulse output parameter (r/kwh) can be set from 1 to
2,000,000,000,and the maximum pulse frequency is 160kHz.
The power pulse is a TTL/CMOS compatible level output with a
load capacity greater than 20 mA.
The default pulse output frequency is 0.2Hz~160KHz.
Power pulse input
The calibration of the energy meter pulse constant (r/kWh) is set
from 1 to 2,000,000,000.
The setting range of the check circle is from 1 to 999,999,999.
The maximum received pulse frequency is 50 kHz
Daily timing pulse input
Clock frequency (Hz) setting range is from 0.01 to 50,000.00.
The setting range of the check circle is from 1 to 999,999,999.
The maximum received pulse frequency is 50kHz.
Working power voltage range
AC85V~265V,50/60Hz
Power consumption
<30VA
Preheat time
≤30 minutes
Working temperature
10℃~35℃
Relative humidity
≤85%,Non-corrosive gas
Dimensions
Around 373mm×211.5mm×131.5mm (Length×Width×Depth) (No
protrusions)
Weight
5.5kg

18
4 Instructions for use
4.1Steps
1) Place the instrument
2) Check before measurement
3) Connect the power cord
4) Connect the test cable
5) Turn on the power
6) Start measuring
7) Record data
8) Measurement completed
4.2Instructions of interface (For reference only, subject to delivery)
4.2.1 Boot interface
The boot interface is displayed within 1-2 seconds after the power is turned on, and the
boot interface is as shown below.
4.2.2 Main interface
After the boot screen, the main screen as shown below will appear. The main interface has
a total of 5 buttons, which are AC, DC, settings, electrical energy error and daily timing
error.

19
4.2.3 AC measurement interface
After clicking the “AC” button on the main interface, the AC measurement interface as
shown below will appear. The AC interface can display voltage, current, frequency, phase,
and active power.
Click “Heavy current”or “10A”to select corresponding gear while “Heavy current”can be
used in direct perforation measurement, "10A" can be used when accessing 10A small
current.
Click the "Word test" to enter the AC word test interface.
Other manuals for PSM Series
1
This manual suits for next models
3
Table of contents
Other Hangzhi Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

PBI
PBI SAS SUPER IAQ instruction manual

Elster
Elster EnCal 3000 Quad Hardware manual

Siemens
Siemens SOMATOM Sensation Cardiac Version A60 Application guide

ITRON
ITRON Bluetooth RF Master user guide

Rice Lake
Rice Lake 1280 Enterprise Series Technical manual

Eaton
Eaton Power Xpert PXM 4000 User and installation manual