Hanna Instruments Checker HI 716 User manual

Bromine Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 716 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• Six powder packets of Bromine Reagent
• 1 x 1.5V AAA Battery
• Instruction Manual
HI 716
Technical specifications:
Range 0.0 to 8.0 ppm
Resolution 0.1 ppm
Accuracy ±0.1 ppm ±5% of reading @ 25 °C / 77 °F
Light Source Light Emitting Diode @ 525 nm
Light Detector Silicon Photocell
Method Adaptation of the
Standard Methods for the Examination
of Water and Wastewater, 18th edition
, DPD method.
The reaction between bromine and the reagent causes a
pink tint in the sample.
Environment 0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type 1 x 1.5V AAA
Auto-Shut off After 10 minutes of non-use
Dimensions 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”)
Weight 64 g (2.25 oz.)
Functional description:
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement. Please
check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking “0.0” indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you
use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
an over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.
1. Dust cover.
2.Cuvette with cap.
3.Cuvette holder.
4.Liquid Crystal Display.
5.Button.
For more details about spare parts and accessories
see “Accessories”.
www.hannainst.com
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at [email protected] or
view our worldwide contact list at www.hannainst.com.
[email protected]
www.ponpe.com
•Turn the meter on by pressing the button. All
segments will be displayed. When the display
shows “Add”, “C1” with “Press” blinking,
the meter is ready.
•Fill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the cuvette
into the meter and close the meter’s cap.
•Press the button. When the display shows
“Add”, “C.2” with “Press” blinking the meter
is zeroed.
•Remove the cuvette, from the meter and
unscrew the cap. Add the content of one packet
of HI 716-0 reagent. Replace the cap and
shake gently for 20 seconds to dissolve most
of the reagent. Place the cuvette back into
the meter.
•Press and hold the button until the timer is
displayed on the LCD (the display will show
the countdown prior to the measurement)
or, alternatively, wait for 2 minutes and
30 seconds and press the button.
•The instrument directly displays the
concentration of bromine in ppm. The meter
automatically turns off after 10 minutes.
10 mL
REAGENT SETS
HI 716-25 Reagents for 25 Bromine tests
OTHER ACCESSORIES
HI 716-11 Bromine Certified Standard Kit
HI 731318 Cloth for wiping cuvettes (4 pcs.)
HI 731321 Glass cuvettes (4 pcs.)
HI 731225 Cuvette black cap for checker HC (4 pcs.)
HI 731353 Cuvette seal cap for checker HC (4 pcs.)
HI 740028 1.5V AAA batteries (4 pcs.)
HI 93703-50 Cuvettes cleaning solution (230 mL)
Recommendations for Users
Before using this product, make sure that it is entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipments,
thus requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For the safety of you and the
instrument do not use or store the instrument in hazardous environments.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find a Hanna Office in your area,
visit our web site:
www.hannainst.com
Hanna Instruments reserves the right to modify the design, construction or
appearance of its products without advance notice.
Battery management
IST716 03/13
Measurement procedure: Accessories:Tips for an accurate measurement
To save the battery, the instrument shuts down after 10 minutes of non-use.
One fresh battery lasts for a minimum of 5000 measurements. When the battery is dead the
instrument will display “bAd” then ”bAt” for 1 second and then turns off.
To restart the instrument, the battery must be replaced with a new one.
To replace the instrument’s battery:
•Turn the instrument off by holding the button until the meter shuts off.
•Turn the instrument upside down and remove the battery cover with a screwdriver.
•Remove the battery from its location and replace it with a new one, inserting the
negative end first.
•Insert the battery cover and replace the screw with a screwdriver.
•It is important that the sample does not contain any debris.
•Whenever the cuvette is placed into the measurement cell, it must be dry outside,
and completely free of fingerprints, oil and dirt. Wipe it thoroughly with
HI 731318 or a lint-free cloth prior to insertion.
•Shaking the cuvette can generate bubbles, causing lower readings. To obtain
accurate measurements, remove bubbles by swirling or by gently tapping the
cuvette.
•Do not let the reacted sample stand for too long after reagent is added, as accuracy
will be affected.
•After the reading it is important to immediately discard the sample, otherwise the
glass might become permanently stained.
[email protected]
www.ponpe.com

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 991000 User manual

Hanna Instruments

Hanna Instruments HI 991000 User manual

Hanna Instruments HI 721 User manual

Hanna Instruments

Hanna Instruments HI 721 User manual

Hanna Instruments Checker HI772 User manual

Hanna Instruments

Hanna Instruments Checker HI772 User manual

Hanna Instruments HI 60503 User manual

Hanna Instruments

Hanna Instruments HI 60503 User manual

Hanna Instruments HI 8424NEW User manual

Hanna Instruments

Hanna Instruments HI 8424NEW User manual

Hanna Instruments HI97779 User manual

Hanna Instruments

Hanna Instruments HI97779 User manual

Hanna Instruments HI6421P User manual

Hanna Instruments

Hanna Instruments HI6421P User manual

Hanna Instruments HI 2210 User manual

Hanna Instruments

Hanna Instruments HI 2210 User manual

Hanna Instruments HI 771 User manual

Hanna Instruments

Hanna Instruments HI 771 User manual

Hanna Instruments HI753 User manual

Hanna Instruments

Hanna Instruments HI753 User manual

Hanna Instruments HI96101C User manual

Hanna Instruments

Hanna Instruments HI96101C User manual

Hanna Instruments HI 991001 User manual

Hanna Instruments

Hanna Instruments HI 991001 User manual

Hanna Instruments BL 7917 User manual

Hanna Instruments

Hanna Instruments BL 7917 User manual

Hanna Instruments HI84531 User manual

Hanna Instruments

Hanna Instruments HI84531 User manual

Hanna Instruments HI 98140 User manual

Hanna Instruments

Hanna Instruments HI 98140 User manual

Hanna Instruments HI96729 User manual

Hanna Instruments

Hanna Instruments HI96729 User manual

Hanna Instruments PCA 310 User manual

Hanna Instruments

Hanna Instruments PCA 310 User manual

Hanna Instruments HI 22091 User manual

Hanna Instruments

Hanna Instruments HI 22091 User manual

Hanna Instruments HI5321 User manual

Hanna Instruments

Hanna Instruments HI5321 User manual

Hanna Instruments HI 84533 User manual

Hanna Instruments

Hanna Instruments HI 84533 User manual

Hanna Instruments HI96725C User manual

Hanna Instruments

Hanna Instruments HI96725C User manual

Hanna Instruments HI 9835 User manual

Hanna Instruments

Hanna Instruments HI 9835 User manual

Hanna Instruments DiST HI 98302 User manual

Hanna Instruments

Hanna Instruments DiST HI 98302 User manual

Hanna Instruments HI774 User manual

Hanna Instruments

Hanna Instruments HI774 User manual

Popular Measuring Instrument manuals by other brands

Biogal CombCam user guide

Biogal

Biogal CombCam user guide

Megger ATEC 210415 Series manual

Megger

Megger ATEC 210415 Series manual

ACS A1270 EMAT Operation manual

ACS

ACS A1270 EMAT Operation manual

Vanguard Instruments ATRT-01 S3 user manual

Vanguard Instruments

Vanguard Instruments ATRT-01 S3 user manual

Altronics Redback A 2067 instructions

Altronics

Altronics Redback A 2067 instructions

GQ Electronics EMF-360V2 quick start guide

GQ Electronics

GQ Electronics EMF-360V2 quick start guide

Schonstedt Instrument TraceMaster instruction manual

Schonstedt Instrument

Schonstedt Instrument TraceMaster instruction manual

Primes Cube M operating manual

Primes

Primes Cube M operating manual

Otodata TM6030 Installation procedures

Otodata

Otodata TM6030 Installation procedures

VOLTCRAFT IR-SCAN-350RH/2 operating instructions

VOLTCRAFT

VOLTCRAFT IR-SCAN-350RH/2 operating instructions

HT Italia HT52 user manual

HT Italia

HT Italia HT52 user manual

Thermo Scientific LevelPRO user manual

Thermo Scientific

Thermo Scientific LevelPRO user manual

Tecnosystemi Galaxy Proair GAZ100001 manual

Tecnosystemi

Tecnosystemi Galaxy Proair GAZ100001 manual

Irwin Geiger Counter instructions

Irwin

Irwin Geiger Counter instructions

Applent Instruments AT682 user manual

Applent Instruments

Applent Instruments AT682 user manual

Marantz professional PMD580 user guide

Marantz professional

Marantz professional PMD580 user guide

Keysight Technologies U4205A user guide

Keysight Technologies

Keysight Technologies U4205A user guide

HP 4154A Operation and service manual

HP

HP 4154A Operation and service manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.