  9. Hanna Instruments HI 93727 User manual

Hanna Instruments HI 93727 User manual

Instruction Manual
HI 93727
Color of Water ISM
http://www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI93727iswarrantedfor two years against defects in workman-
shipand materialswhenusedforits intendedpurposeandmaintained
accordingtoinstructions.
Thiswarranty islimitedto repairorreplacement freeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required,contactthedealerfromwhom youpurchasedthe
instrument.Ifunderwarranty, report the model number, dateofpur-
chase,serialnumberandthe nature of the failure. Iftherepairisnot
coveredby the warranty,youwillbe notified ofthechargesincurred. If
theinstrumentistobereturnedtoHannaInstruments,firstobtaina
ReturnedGoodsAuthorizationNumber from the Customer Service de-
partmentand thensenditwith shipmentcostsprepaid.Whenshipping
anyinstrument, makesureitisproperly packagedforcompleteprotec-
tion.
Tovalidate yourwarranty,filloutandreturntheenclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrightsare reserved. Reproduction inwholeor in part isprohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket, Rhode Island, 02895,USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Three Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93727 meter measures the true and apparent color
in water and wastewater in the 0 to 500 PCU (Platinum
Cobalt Units) range.
Themeter usesan exclusivepositive-locking systemto ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0 to 500 PCU (Platinum Cobalt Units)
Resolution 10 PCU
Accuracy ±10 PCU ± 5% of reading
TypicalEMC ±10 PCU
Deviation
Light Source Light Emitting Diode @ 470 nm
Method Adaptation of the Colorimetric Platinum
Cobalt method
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED ACCESSORIES
0.45 µm membrane for true color measurement.
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximumconcentrationreadable(see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
1.5 cm
ON
OFF
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays "- - -", it is
ready.
• Fill one cuvet up to 1.5 cm (¾")
below the rim with deionized water
and replace the cap. This is the
blank.
• Fill another cuvet up to 1.5 cm (¾")
below the rim with unfiltered sample
and replace the cap. This is the
apparent color.
• Filter 50 mL of sample through a
filter with a 0.45 µm membrane.
• Fill the third cuvet up to 1.5 cm
(¾") below the rim with the filtered
sample and replace the cap. This the
true color.
• Placetheblank intothe holder
and ensure that the notch on
the cap is positioned securely
into the groove.
• Press ZERO and "SIP" will appear on the display.
• Wait for a few seconds and the dis-
playwillshow "-0.0-".Nowthe meter
is zeroed and ready for measure-
ment.
1.5 cm
1.5 cm
ZERO
• Remove the blank, insert the
apparent color cuvet into the
instrument and ensure that
the notch on the cap is posi-
tioned securely into the
groove.
• Press READ and "SIP" will appear on the display.
• The meter directly displays the value of apparent color.
• Remove the cuvet, insert the
true color cuvet into the in-
strument and ensure that
the notch on the cap is posi-
tioned securely into the
groove.
• Press READ and "SIP" will appear on the display.
• Themeter directly displays concentrationin color units of
true color on the Liquid Crystal Display.
INTERFERENCES
There are no known interferences for this method.
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C(86°F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLID
E
OFF
RecommendationsforUsers
Before using these products, make sure that they are entirely suitable for the environ-
mentinwhichtheyareused.
Operationoftheseinstrumentsinresidentialareacouldcauseunacceptableinterferences
toradioandTV equipments, requiring the operatortotakeallnecessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments'EMCperformance.
Toavoiddamagesorburns,donotperformanymeasurementinmicrowaveovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
OTHER ACCESSORIES
HI710009 Blue rubber boot
HI710010 Orange rubber boot
HI721310 9V battery (10 pcs)
HI731318 Tissue for wiping cuvets (4 pcs)
HI731321 Glass cuvets (4 pcs)
HI731325 Caps for cuvets (4 pcs)
HI93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93727R2 01/00
PRINTED IN PORTUGAL

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 9142 User manual

Hanna Instruments

Hanna Instruments HI 9142 User manual

Hanna Instruments HI 8424NEW User manual

Hanna Instruments

Hanna Instruments HI 8424NEW User manual

Hanna Instruments HI 96715C User manual

Hanna Instruments

Hanna Instruments HI 96715C User manual

Hanna Instruments HI 8734 User manual

Hanna Instruments

Hanna Instruments HI 8734 User manual

Hanna Instruments HI 93749 User manual

Hanna Instruments

Hanna Instruments HI 93749 User manual

Hanna Instruments HI 8043 User manual

Hanna Instruments

Hanna Instruments HI 8043 User manual

Hanna Instruments HI 83746 User manual

Hanna Instruments

Hanna Instruments HI 83746 User manual

Hanna Instruments HI8417 User manual

Hanna Instruments

Hanna Instruments HI8417 User manual

Hanna Instruments HI97732 User manual

Hanna Instruments

Hanna Instruments HI97732 User manual

Hanna Instruments HI 99121 User manual

Hanna Instruments

Hanna Instruments HI 99121 User manual

Hanna Instruments HI6421 User manual

Hanna Instruments

Hanna Instruments HI6421 User manual

Hanna Instruments Checker HI 711 User manual

Hanna Instruments

Hanna Instruments Checker HI 711 User manual

Hanna Instruments HI 38018 User manual

Hanna Instruments

Hanna Instruments HI 38018 User manual

Hanna Instruments HYGROCHECK HI 98601 User manual

Hanna Instruments

Hanna Instruments HYGROCHECK HI 98601 User manual

Hanna Instruments HI96700C User manual

Hanna Instruments

Hanna Instruments HI96700C User manual

Hanna Instruments HI 991301 User manual

Hanna Instruments

Hanna Instruments HI 991301 User manual

Hanna Instruments HI 5421 User manual

Hanna Instruments

Hanna Instruments HI 5421 User manual

Hanna Instruments HI 95724C User manual

Hanna Instruments

Hanna Instruments HI 95724C User manual

Hanna Instruments HI 84100 User manual

Hanna Instruments

Hanna Instruments HI 84100 User manual

Hanna Instruments PCA 300 Series User manual

Hanna Instruments

Hanna Instruments PCA 300 Series User manual

Hanna Instruments PICCOLO User manual

Hanna Instruments

Hanna Instruments PICCOLO User manual

Hanna Instruments HI 83414 User manual

Hanna Instruments

Hanna Instruments HI 83414 User manual

Hanna Instruments HI 98140 User manual

Hanna Instruments

Hanna Instruments HI 98140 User manual

Hanna Instruments HI 83141V User manual

Hanna Instruments

Hanna Instruments HI 83141V User manual

Popular Measuring Instrument manuals by other brands

QUARTA RADEX ONE user manual

QUARTA

QUARTA RADEX ONE user manual

Katronic KATflow 170 Operating and safety instructions

Katronic

Katronic KATflow 170 Operating and safety instructions

Lumel N30P user manual

Lumel

Lumel N30P user manual

Easy-Laser E980 manual

Easy-Laser

Easy-Laser E980 manual

Elcontrol STAR 3 user manual

Elcontrol

Elcontrol STAR 3 user manual

Agilent Technologies NFA Series Setup guide

Agilent Technologies

Agilent Technologies NFA Series Setup guide

Endress+Hauser nivotester FTC 420 Quick installation guide

Endress+Hauser

Endress+Hauser nivotester FTC 420 Quick installation guide

Flybox Omnia57-80 Series manual

Flybox

Flybox Omnia57-80 Series manual

bmcm AMS42-LAN16fx manual

bmcm

bmcm AMS42-LAN16fx manual

Schäfer RMS 442 quick start guide

Schäfer

Schäfer RMS 442 quick start guide

PCB Piezotronics 352A21/NC Installation and operating manual

PCB Piezotronics

PCB Piezotronics 352A21/NC Installation and operating manual

Schaller humimeter BMA user manual

Schaller

Schaller humimeter BMA user manual

THEIS TKL-7 operating instructions

THEIS

THEIS TKL-7 operating instructions

HYDROMETER IZAR CENTER user manual

HYDROMETER

HYDROMETER IZAR CENTER user manual

Omega HHF143 user guide

Omega

Omega HHF143 user guide

amc VIBRO AVM 1002D user manual

amc VIBRO

amc VIBRO AVM 1002D user manual

Mollet LF20 Series Instruction

Mollet

Mollet LF20 Series Instruction

DayTronic 3000PLUS instruction manual

DayTronic

DayTronic 3000PLUS instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.