Instruction Manual
HI 93727
Color of Water ISM
http://www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI93727iswarrantedfor two years against defects in workman-
shipand materialswhenusedforits intendedpurposeandmaintained
accordingtoinstructions.
Thiswarranty islimitedto repairorreplacement freeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required,contactthedealerfromwhom youpurchasedthe
instrument.Ifunderwarranty, report the model number, dateofpur-
chase,serialnumberandthe nature of the failure. Iftherepairisnot
coveredby the warranty,youwillbe notified ofthechargesincurred. If
theinstrumentistobereturnedtoHannaInstruments,firstobtaina
ReturnedGoodsAuthorizationNumber from the Customer Service de-
partmentand thensenditwith shipmentcostsprepaid.Whenshipping
anyinstrument, makesureitisproperly packagedforcompleteprotec-
tion.
Tovalidate yourwarranty,filloutandreturntheenclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrightsare reserved. Reproduction inwholeor in part isprohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket, Rhode Island, 02895,USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives
EN 50081-1 and EN 50082-1.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Three Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93727 meter measures the true and apparent color
in water and wastewater in the 0 to 500 PCU (Platinum
Cobalt Units) range.
Themeter usesan exclusivepositive-locking systemto ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0 to 500 PCU (Platinum Cobalt Units)
Resolution 10 PCU
Accuracy ±10 PCU ± 5% of reading
TypicalEMC ±10 PCU
Deviation
Light Source Light Emitting Diode @ 470 nm
Method Adaptation of the Colorimetric Platinum
Cobalt method
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED ACCESSORIES
0.45 µm membrane for true color measurement.
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximumconcentrationreadable(see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high. Dilute the
sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
1.5 cm
ON
OFF
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays "- - -", it is
ready.
• Fill one cuvet up to 1.5 cm (¾")
below the rim with deionized water
and replace the cap. This is the
blank.
• Fill another cuvet up to 1.5 cm (¾")
below the rim with unfiltered sample
and replace the cap. This is the
apparent color.
• Filter 50 mL of sample through a
filter with a 0.45 µm membrane.
• Fill the third cuvet up to 1.5 cm
(¾") below the rim with the filtered
sample and replace the cap. This the
true color.
• Placetheblank intothe holder
and ensure that the notch on
the cap is positioned securely
into the groove.
• Press ZERO and "SIP" will appear on the display.
• Wait for a few seconds and the dis-
playwillshow "-0.0-".Nowthe meter
is zeroed and ready for measure-
ment.
1.5 cm
1.5 cm
ZERO
• Remove the blank, insert the
apparent color cuvet into the
instrument and ensure that
the notch on the cap is posi-
tioned securely into the
groove.
• Press READ and "SIP" will appear on the display.
• The meter directly displays the value of apparent color.
• Remove the cuvet, insert the
true color cuvet into the in-
strument and ensure that
the notch on the cap is posi-
tioned securely into the
groove.
• Press READ and "SIP" will appear on the display.
• Themeter directly displays concentrationin color units of
true color on the Liquid Crystal Display.
INTERFERENCES
There are no known interferences for this method.
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shaking the cuvet can generate bubbles in the sample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C(86°F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Replace the battery and the cover.
9V
SLID
E
OFF
RecommendationsforUsers
Before using these products, make sure that they are entirely suitable for the environ-
mentinwhichtheyareused.
Operationoftheseinstrumentsinresidentialareacouldcauseunacceptableinterferences
toradioandTV equipments, requiring the operatortotakeallnecessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments'EMCperformance.
Toavoiddamagesorburns,donotperformanymeasurementinmicrowaveovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
OTHER ACCESSORIES
HI710009 Blue rubber boot
HI710010 Orange rubber boot
HI721310 9V battery (10 pcs)
HI731318 Tissue for wiping cuvets (4 pcs)
HI731321 Glass cuvets (4 pcs)
HI731325 Caps for cuvets (4 pcs)
HI93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93727R2 01/00
PRINTED IN PORTUGAL

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 96813 User manual

Hanna Instruments

Hanna Instruments HI 96813 User manual

Hanna Instruments HI 93707 User manual

Hanna Instruments

Hanna Instruments HI 93707 User manual

Hanna Instruments GROLINE MONITOR HI981420 User manual

Hanna Instruments

Hanna Instruments GROLINE MONITOR HI981420 User manual

Hanna Instruments HI 93749 User manual

Hanna Instruments

Hanna Instruments HI 93749 User manual

Hanna Instruments HI97723 User manual

Hanna Instruments

Hanna Instruments HI97723 User manual

Hanna Instruments HI 97500 User manual

Hanna Instruments

Hanna Instruments HI 97500 User manual

Hanna Instruments HI98194 User manual

Hanna Instruments

Hanna Instruments HI98194 User manual

Hanna Instruments HI99111 User manual

Hanna Instruments

Hanna Instruments HI99111 User manual

Hanna Instruments HALO 2 HI9810402 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810402 User manual

Hanna Instruments HI 96804 User manual

Hanna Instruments

Hanna Instruments HI 96804 User manual

Hanna Instruments HI981954 User manual

Hanna Instruments

Hanna Instruments HI981954 User manual

Hanna Instruments Combo Grochek HI 991404 User manual

Hanna Instruments

Hanna Instruments Combo Grochek HI 991404 User manual

Hanna Instruments HI 9147 User manual

Hanna Instruments

Hanna Instruments HI 9147 User manual

Hanna Instruments HI 4111 User manual

Hanna Instruments

Hanna Instruments HI 4111 User manual

Hanna Instruments HI 98401 User manual

Hanna Instruments

Hanna Instruments HI 98401 User manual

Hanna Instruments DiST HI 98302 User manual

Hanna Instruments

Hanna Instruments DiST HI 98302 User manual

Hanna Instruments HI 9835 User manual

Hanna Instruments

Hanna Instruments HI 9835 User manual

Hanna Instruments Silica ISM 93705 User manual

Hanna Instruments

Hanna Instruments Silica ISM 93705 User manual

Hanna Instruments HI 99141 User manual

Hanna Instruments

Hanna Instruments HI 99141 User manual

Hanna Instruments HI97738 User manual

Hanna Instruments

Hanna Instruments HI97738 User manual

Hanna Instruments HI 9033 User manual

Hanna Instruments

Hanna Instruments HI 9033 User manual

Hanna Instruments HI 705 User manual

Hanna Instruments

Hanna Instruments HI 705 User manual

Hanna Instruments HI 8314 User manual

Hanna Instruments

Hanna Instruments HI 8314 User manual

Hanna Instruments HI83303 User manual

Hanna Instruments

Hanna Instruments HI83303 User manual

Popular Measuring Instrument manuals by other brands

AEMC MN353 user manual

AEMC

AEMC MN353 user manual

Eickemeyer LIFEVET O2 user manual

Eickemeyer

Eickemeyer LIFEVET O2 user manual

Giropes GI620 COMPACT user manual

Giropes

Giropes GI620 COMPACT user manual

Megger DCM340 user guide

Megger

Megger DCM340 user guide

TINRS Edgecutter manual

TINRS

TINRS Edgecutter manual

Bender PEM353 quick start guide

Bender

Bender PEM353 quick start guide

Knick SensoGate WA 131 H user manual

Knick

Knick SensoGate WA 131 H user manual

3M The Edge quick start guide

3M

3M The Edge quick start guide

Campbell LI200R instruction manual

Campbell

Campbell LI200R instruction manual

GE SimpliNano Original instructions

GE

GE SimpliNano Original instructions

Balmac 205ce Operation manual

Balmac

Balmac 205ce Operation manual

Shanghai Grandway Telecom Tech. Co., Ltd. FHP12 user guide

Shanghai Grandway Telecom Tech. Co., Ltd.

Shanghai Grandway Telecom Tech. Co., Ltd. FHP12 user guide

Nikon Nivo Series instruction manual

Nikon

Nikon Nivo Series instruction manual

Elma Instruments RT606 A user manual

Elma Instruments

Elma Instruments RT606 A user manual

Minolta SPOTMETER F - IR REMOTE CONTRO LRC-3 manual

Minolta

Minolta SPOTMETER F - IR REMOTE CONTRO LRC-3 manual

Megger DLRO-10 user guide

Megger

Megger DLRO-10 user guide

iSTEK K100-TB instruction manual

iSTEK

iSTEK K100-TB instruction manual

AITE ATC1006A Operation manual

AITE

AITE ATC1006A Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.