Instruction Manual
HI 93732N
Dissolved Oxygen ISM
www.hannainst.com This Instrument is in
Compliance with the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI 93732N is warranted for two years against defects in
workmanshipand materialswhen usedfor itsintended purposeand
maintainedaccordingtoinstructions.
Thiswarrantyislimited torepairorreplacementfree ofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenancearenot covered.
Ifserviceis required,contactthe dealerfromwhomyoupurchasedthe
instrument.Ifunderwarranty,reportthemodelnumber,dateofpurchase,
serialnumberand thenatureofthefailure.Iftherepairisnotcovered
bythewarranty,youwillbenotifiedofthechargesincurred.
Iftheinstrument istobe returnedtoHanna Instruments,firstobtain a
Returned Goods Authorization Number from the Customer Service
departmentandthensenditwithshipmentcostsprepaid.Whenshipping
any instrument, make sure it is properly packaged for complete
protection.
Tovalidateyourwarranty, filloutandreturntheenclosedwarrantycard
within14daysfrom thedateofpurchase.
Allrights are reserved.Reproduction in wholeor inpartis prohibited
withoutthewrittenconsentof thecopyrightowner, HannaInstruments
Inc.,Woonsocket,RhodeIsland,02895 ,USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
This instrument is in compliance with directives.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• Two Sample Cuvets and Caps
• One Transport Cap
• One 60 mL glass Bottle with Stopper
• 9V Battery
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing with
thesuppliedaccessories.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93732N meter measures the content of dissolved
oxygen (O2) in surface, feed, natural and waste waters in
the 0.0 to 10.0 mg/L (ppm) range.
Themeterusesan exclusive positive-locking systemtoensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid form and are supplied in bottles.
Theamountofreagent is precisely dosedtoensuremaximum
repeatability.
Display codes aid the user in routine operations. The meter
has an auto-shut off feature that will turn itself off after 10
minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0.0 to 10.0 mg/L (ppm)
Resolution 0.1 mg/L
Accuracy ±0.2 mg/L ± 3% of reading
TypicalEMC ±0.1 mg/L
Deviation
Light Source Light Emitting Diode @ 470 nm
Method Adaptation of the Standard Methods for
theExaminationofWaterand Wastewater
(18th edition), azide modified Winkler
method. The reaction between dissolved
oxygen and the reagent causes a yellow
tint in the sample.
Light Life Life of the instrument
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery 1 x 9 volt/40 hours
Type/Life
Auto-Shut off After 10' of non-use
Dimensions 180 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.).
REQUIRED REAGENTS
Code Description Quantity
HI 93732A-0 Reagent A 5 drops
HI 93732B-0 Reagent B 5 drops
HI 93732C-0 Reagent C 10drops
REAGENT SETS
HI93732-01 Reagents for 100 tests
HI93732-03 Reagents for 300 tests
This indicates that the meter is in a
readystateandzeroing canbeperformed.
Sampling in Progress. This prompt
appearseachtime the meterisperforming
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
themaximumconcentrationreadable (see
specifications) indicates that the sample
absorbs too much light, meaning that
the concentration is too high.
Lightoverrange.The cuvet is not inserted
correctly and an excess ambient light is
reaching the detector. If the cover is
properlyinstalled,then contactyourdealer
or the nearest Hanna Customer Service
Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery voltage
is getting low and the battery needs to
be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the
battery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE