Hanna Instruments HI 93753 User manual

Instruction Manual
HI 93753
Chloride ISM
www.hannainst.com ThisInstrumentisin
Compliancewith the CE Directives
HannaInstrumentsreservestherighttomodifythedesign,construction
andappearanceofitsproductswithoutadvancenotice.
WARRANTYWARRANTY
WARRANTYWARRANTY
WARRANTY
HI93753iswarrantedfortwoyears against defects in workman-
shipand materialswhenusedforits intendedpurposeandmaintained
accordingtoinstructions.
Thiswarrantyislimitedtorepairorreplacementfreeofcharge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenanceare not covered.
Ifserviceis required, contactthedealerfromwhomyou purchased the
instrument.Ifunderwarranty,reportthe model number, date of pur-
chase,serialnumberandthenatureof the failure. If the repair is not
coveredbythe warranty,youwillbenotified ofthechargesincurred.If
theinstrumentistobereturnedtoHannaInstruments,first obtain a
ReturnedGoodsAuthorizationNumberfrom the Customer Service de-
partmentand thensenditwithshipment costsprepaid.Whenshipping
anyinstrument, makesureitisproperly packagedforcompleteprotec-
tion.
Tovalidate yourwarranty,filloutandreturnthe enclosedwarrantycard
within14 daysfromthedateofpurchase.
Allrights are reserved. Reproductionin whole or inpart is prohibited
without the written consent of the copyright owner, Hanna Instru-
mentsInc.,Woonsocket,Rhode Island, 02895 , USA.
DearCustomer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V battery
• Two sample cuvettes and caps
• One transport cap
Note: Keep all packing material until the instrument has
beenobserved tofunction correctly.Anydefective item
must be returned in its original packing.
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATIONPRELIMINARY EXAMINATION
PRELIMINARY EXAMINATION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTIONGENERAL DESCRIPTION
GENERAL DESCRIPTION
The HI 93753 meter measures the chloride (Cl–) content in
water and wastewater in the 0 to 20 mg/L (ppm) range.
Themeter usesan exclusivepositive-locking systemtoensure
that the cuvet is in the same position every time it is placed
into the measurement cell.
The reagents are in liquid form and are supplied in bottles.
The amount of reagent is precisely dosed to ensure maxi-
mum repeatability.
Display codes aid the user in routine operations.
The meter has an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SPECIFICATIONS
Range 0.0 to 20.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.5 mg/L±6% of reading
TypicalEMC ±0.1 mg/L
Deviation
Light Source Light Emitting Diode @ 470 nm
Method Adaptation of the mercury(II)
thiocyanate method. The chloride ion
displaces thiocyanate ion from
mercury(II).The iron(III) present forms
with thiocyanate an orange colored
complex. The intensity of color is
proportional to the chloride ion
concentration.
Light Detector SiliconPhotocell
Environment 0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off After 10' of non-use
Dimensions 80 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight 290 g (10 oz.)
REQUIRED REAGENTS
Code Description Quantity
HI 93753A-0 Displacing Reagent 1.0 mL
HI93753B-0 Complexing Reagent 1.0 mL
REAGENT SETS
HI93753-01 Reagents for 100tests and two 1mL syringes
HI93753-03 Reagents for 300 tests and six 1 mL syringes
This indicates that the meter is in a
ready state and zeroing can be per-
formed.
Sampling in Progress. This prompt ap-
pears each time the meter is performing
a measurement.
This indicates that the meter is in a
zeroed state and measurement can be
performed.
A zero reading was not taken. Insert a
sample before adding reagent and press
ZERO.
Under range. A blinking "0.00" indicates
that the sample absorbs less light than
the zero reference. Check the procedure
and make sure you use the same cuvet
for reference (zero) and measurement.
Over range. A flashing value higher than
the maximum range (see specifications)
indicates that the sample absorbs too
much light and that the concentration is
too high. Dilute the sample.
Light over range. The cuvet is not in-
serted correctly and an excess ambient
light is reaching the detector. If the cover
is properly installed, then contact your
dealer or the nearest Hanna Customer
Service Center.
Light under range. The zero sample is
too dark for proper zeroing. If this is not
the case, contact your dealer or the
nearest Hanna Customer Service Center.
The "V" indicates that the battery volt-
age is getting low and the battery needs
to be replaced.
This indicates that the battery is dead
and must be replaced.
Note: once this indication is displayed,
the meter will lockup. Change the bat-
tery to restart.
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDEDISPLAY CODE GUIDE
DISPLAY CODE GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing
ON/OFF.
• Whenthe LCD displays "- - -",itis
ready.
• Fill one cuvette (#1) with 10 mL
ofdistilled water (up to the mark).
• Fill another cuvette (#2) with 10
mL of sample (up to the mark).
Note: for low chloride ions
concentration, rinse the cuvette a couple of times with
sample before filling it with 10 mL of sample (up to the
mark).
Note: for most accurate results, use two graduated
pipettes to deliver exactly 10 mL of distilled water and
10 mL of sample to the cuvets.
• Using the 1 mL syringe, add 0.5 mL of HI 93753A-0
Displacing Reagent to each cuvet.
Note: to dose exactly 0.5 mL of reagent with the
syringe, push the plunger completely into the syringe
and insert the tip into the HI 93753A-0 Displacing
Reagent bottle. Pull the plunger out until the lower
edge of the seal is on the 0.0 mL mark of the syringe.
Then hold the syringe above the cuvet and push the
plunger down to the 0.5 ml mark.
• Replace the caps and mix each
cuvetby turningthe cuvetteupside
downforapproximately30 seconds.
• Remove the caps and use the second 1 mL syringe to
add 0.5 mL of HI 93753B-0 Complexing Reagent to
each cuvet.
ON
OFF
• Replacethe caps and mix each
cuvet by turning the cuvette
upsidedown for approximately
30seconds.
• Place the cuvet with the reacted
distilled water (#1) into the
instrument.
• PressREADTIMEDandthedisplaywillshowthecountdown
prior to zeroing the blank or, alternatively, wait for 2
minutes and press ZERO. In both cases "SIP" will
appear during measurement.
• Waitfora few seconds and the
displaywill show"-0.0-”. Now
the meter is zeroed and ready
formeasurement.
• Remove the cuvette.
• Place the other cuvet with the
reacted sample (#2) into the
instrument.
• Press READ DIRECT and "SIP" will blink on the LCD
during measurement.
• Theinstrumentdirectlydisplaysconcentrationinmg/Lof
chloride on the Liquid Crystal Display.
• To convert the reading to mg/L of Sodium Chloride
(NaCl), multiply the reading by a factor of 1.65.
The instructions below should be carefully followed during
testing to ensure best accuracy.
• Do not touch the cuvette walls with your fingers.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvette to prevent any contamination.
• Do not let the test sample stand too long after adding
the reagents or accuracy will be compromised.
• Whenever the cuvette is placed into the measurement
cell,itmustbe completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvette is used for zeroing
and measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
• Shakingthe cuvette can generate bubbles in thesample,
causing higher readings. To obtain accurate measure-
ments, remove such bubbles by swirling or by gently
tapping the vial.
• All the reaction times reported in this manual are
referred to 20°C (68°F). As a general rule of thumb,
they should be doubled at 10°C (50°F) and halved at
30°C(86°F).
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATETIPS FOR AN ACCURATE
TIPS FOR AN ACCURATE
MEASUREMENTMEASUREMENT
MEASUREMENTMEASUREMENT
MEASUREMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENTBATTERY REPLACEMENT
BATTERY REPLACEMENT
Battery replacement must only take place in a non-hazard-
ous area using a 9V alkaline battery.
Simply slide off the battery cover on
the back of the meter. Detach the
battery from the terminals and at-
tach a fresh 9V battery while paying
attention to the correct polarity. Re-
place the battery and the cover.
9V
SLID
E
OFF
Recommendations for
Users
Beforeusingtheseproducts,
make sure that they are
entirely suitable for the
environment in which they
areused.
Operation of these
instruments in residential
area could cause
unacceptable interferences
to radio and TV
equipments, requiring the
operator to take all
necessary steps to correct
interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments'EMCperformance.
Toavoiddamagesorburns,donotperformanymeasurementinmicrowaveovens.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
REAGENT SETS
HI93753-01 Reagents for 100 tests
HI93753-03 Reagents for 300 tests
OTHER ACCESSORIES
HI710009 Blue rubber boot
HI710010 Orange rubber boot
HI721310 9V battery (10 pcs)
HI731318 Tissue for wiping cuvets (4 pcs)
HI731321 Glass cuvettes (4 pcs)
HI731325 Caps for cuvettes (4 pcs)
HI93703-50 Cuvettes cleaning solution (230 mL).
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
IST93753R3 09/03
READ
TIMED
10 mL
#1
10 mL
#2
#1 #2
#1
ZERO
#2
INTERFERENCES
The pH of sample after addition of reagents should be about
2. For alkaline samples, neutralize before adding reagents.
Intensely colored samples will cause interference, therefore
they should be adequately treated before performing the
test. Suspended matter in large amounts should be re-
moved by prior filtration.
ACCESSORIES
• 2 syringes (1 mL), included in the HI 93753 reagents
set.
#1 #2

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 9033 User manual

Hanna Instruments

Hanna Instruments HI 9033 User manual

Hanna Instruments HUALIX HI 715 User manual

Hanna Instruments

Hanna Instruments HUALIX HI 715 User manual

Hanna Instruments HI 96707C User manual

Hanna Instruments

Hanna Instruments HI 96707C User manual

Hanna Instruments HI 95762C User manual

Hanna Instruments

Hanna Instruments HI 95762C User manual

Hanna Instruments HI708 User manual

Hanna Instruments

Hanna Instruments HI708 User manual

Hanna Instruments HI96709C User manual

Hanna Instruments

Hanna Instruments HI96709C User manual

Hanna Instruments IC-HI98165 User manual

Hanna Instruments

Hanna Instruments IC-HI98165 User manual

Hanna Instruments HI97732 User manual

Hanna Instruments

Hanna Instruments HI97732 User manual

Hanna Instruments HI 96771C User manual

Hanna Instruments

Hanna Instruments HI 96771C User manual

Hanna Instruments HI 83740 User manual

Hanna Instruments

Hanna Instruments HI 83740 User manual

Hanna Instruments HI 96822 User manual

Hanna Instruments

Hanna Instruments HI 96822 User manual

Hanna Instruments HI 83208 User manual

Hanna Instruments

Hanna Instruments HI 83208 User manual

Hanna Instruments HI 726 User manual

Hanna Instruments

Hanna Instruments HI 726 User manual

Hanna Instruments HI 991002 User manual

Hanna Instruments

Hanna Instruments HI 991002 User manual

Hanna Instruments HI83325 User manual

Hanna Instruments

Hanna Instruments HI83325 User manual

Hanna Instruments HI 9033 User manual

Hanna Instruments

Hanna Instruments HI 9033 User manual

Hanna Instruments HI 991301 User manual

Hanna Instruments

Hanna Instruments HI 991301 User manual

Hanna Instruments HI 993310 User manual

Hanna Instruments

Hanna Instruments HI 993310 User manual

Hanna Instruments HI747 User manual

Hanna Instruments

Hanna Instruments HI747 User manual

Hanna Instruments HI708 User manual

Hanna Instruments

Hanna Instruments HI708 User manual

Hanna Instruments HI781 User manual

Hanna Instruments

Hanna Instruments HI781 User manual

Hanna Instruments HI 4111 User manual

Hanna Instruments

Hanna Instruments HI 4111 User manual

Hanna Instruments Silica ISM 93705 User manual

Hanna Instruments

Hanna Instruments Silica ISM 93705 User manual

Hanna Instruments HI 2210 User manual

Hanna Instruments

Hanna Instruments HI 2210 User manual

Popular Measuring Instrument manuals by other brands

Haefely WA 2293 quick start guide

Haefely

Haefely WA 2293 quick start guide

DigiCore Technology Ctrack iS100 Series II installation manual

DigiCore Technology

DigiCore Technology Ctrack iS100 Series II installation manual

Sierra 780S Series Flat-Trak instruction manual

Sierra

Sierra 780S Series Flat-Trak instruction manual

Mastech MS2002 Operator's instruction manual

Mastech

Mastech MS2002 Operator's instruction manual

Siemens SMART 7KT manual

Siemens

Siemens SMART 7KT manual

Sealey AK9621EV instructions

Sealey

Sealey AK9621EV instructions

Ecolab OGM PLUS Short operating instructions

Ecolab

Ecolab OGM PLUS Short operating instructions

GOK GWG Assembly and operating manual

GOK

GOK GWG Assembly and operating manual

CS Instruments DS 300-P Operation manual

CS Instruments

CS Instruments DS 300-P Operation manual

Textron GREENLEE CMGRT-100A-C instruction manual

Textron

Textron GREENLEE CMGRT-100A-C instruction manual

Teac TD-700T Instructions for use

Teac

Teac TD-700T Instructions for use

MREL MicroTrap Operation manual

MREL

MREL MicroTrap Operation manual

WeatherHawk MyMET manual

WeatherHawk

WeatherHawk MyMET manual

Omicron Lab SPECTANO 100 quick start guide

Omicron Lab

Omicron Lab SPECTANO 100 quick start guide

FTS Bubbler Installation and quick start guide

FTS

FTS Bubbler Installation and quick start guide

IDEAL TightSight 61-757 Operation and safety manual

IDEAL

IDEAL TightSight 61-757 Operation and safety manual

ProCharger 10037598 Instructions for use

ProCharger

ProCharger 10037598 Instructions for use

Xylem WTW SensoLyt 900 operating manual

Xylem

Xylem WTW SensoLyt 900 operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.